1 From the Medical Imaging Unit, Department of Physics, Kings College; and the 2 Department of Ophthalmology, St. Thomas Hospital, London, United Kingdom.
PURPOSE. To describe a software program developed to provide an objective assessment of the amount of posterior capsular opacification (PCO) in high-resolution digital images of the posterior capsule after cataract surgery.
METHODS. Images are analyzed by a set protocol of defining the area of the posterior capsule, removing the Purkinje light reflexes by intensity segmentation, contrast enhancement, filtering to enhance low-density PCO, and variance analysis using a co-occurrence matrix to assess texture. The accuracy of the system was tested for validity and repeatability.
RESULTS. The software developed has been demonstrated to be an objective method of quantifying PCO. In validation tests, the image analysisderived measure of PCO showed good agreement with clinically derived measures of PCO. Clinicians assessed PCO on a computer screen image and also under slit lamp examination (Pearson correlation coefficient for both methods >0.92). The entire acquisition and analysis system was demonstrated to have a confidence limit for 2 SDs of 9.8% for group data.
CONCLUSIONS. This system is capable of producing an accurate and reproducible measure of PCO that is relevant to assessing techniques of PCO prevention.
This article has been cited by other articles:
![]() |
N Lois, R Dawson, J Townend, A D McKinnon, G C Smith, R v. Hof, N Van Rooijen, and J V Forrester Effect of short-term macrophage depletion in the development of posterior capsule opacification in rodents Br. J. Ophthalmol., November 1, 2008; 92(11): 1528 - 1533. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Grewal, R. Jain, G. S. Brar, and S. P. S. Grewal Pentacam Tomograms: A Novel Method for Quantification of Posterior Capsule Opacification Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 2004 - 2008. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Moreno-Montanes, A. Alvarez, and M. J. Maldonado Objective Quantification of Posterior Capsule Opacification after Cataract Surgery, with Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., November 1, 2005; 46(11): 3999 - 4006. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. R. Meacock, D. J. Spalton, J. Boyce, and J. Marshall The Effect of Posterior Capsule Opacification on Visual Function Invest. Ophthalmol. Vis. Sci., November 1, 2003; 44(11): 4665 - 4669. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Findl, W. Buehl, H. Siegl, and A. Pinz Removal of Reflections in the Photographic Assessment of PCO by Fusion of Digital Retroillumination Images Invest. Ophthalmol. Vis. Sci., January 1, 2003; 44(1): 275 - 280. [Abstract] [Full Text] [PDF] |
||||
![]() |
T M Aslam, B Dhillon, N Werghi, A Taguri, and A Wadood Systems of analysis of posterior capsule opacification Br. J. Ophthalmol., October 1, 2002; 86(10): 1181 - 1186. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |