PURPOSE. To compare intraocular pressure (IOP) measurements of the Ocular Blood Flow (OBF) pneumotonometer, Goldmann applanation tonometer, and Tono-Pen in eyes with normal corneas of various thicknesses.
METHODS. IOP was measured with an OBF pneumotonometer, Tono-Pen and Goldmann applanation tonometer in random order in 181 eyes with normal corneas. Central corneal thickness (CCT) was measured using an ultrasonic pachymeter after all IOP determinations had been made. Right and left eyes were analyzed separately for statistical purposes.
RESULTS. With all instruments, IOP varied with CCT, even though the variation in IOP was large. Readings with the OBF pneumotonometer showed a mean increase in IOP with increasing CCT of 0.28 mm Hg/10 µm, an increase of 0.23 mm Hg/10 µm with the Goldmann tonometer, and of 0.10 mm Hg/10 µm with the Tono-Pen. The OBF pneumotonometer consistently recorded comparatively higher IOPs than the other two instruments.
CONCLUSIONS. The Tono-Pen is least affected by CCT when used to measure IOP in eyes with normal corneas. The OBF pneumotonometer appears to be more affected by variation in CCT than the Goldmann tonometer. This is contrary to expectations, based on the mechanism of measurement of IOP of the OBF pneumotonometer.
This article has been cited by other articles:
![]() |
S. A. Read, M. J. Collins, and D. R. Iskander Diurnal Variation of Axial Length, Intraocular Pressure, and Anterior Eye Biometrics Invest. Ophthalmol. Vis. Sci., July 1, 2008; 49(7): 2911 - 2918. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. A. Somner, D. S. Morris, K. M. Scott, I. J. C. MacCormick, P. Aspinall, and B. Dhillon What Happens to Intraocular Pressure at High Altitude? Invest. Ophthalmol. Vis. Sci., April 1, 2007; 48(4): 1622 - 1626. [Abstract] [Full Text] [PDF] |
||||
![]() |
J M Gonzalez-Meijome, J Jorge, A Queiros, P Fernandes, R Montes-Mico, J B Almeida, and M A Parafita Age differences in central and peripheral intraocular pressure using a rebound tonometer Br. J. Ophthalmol., December 1, 2006; 90(12): 1495 - 1500. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, Y. C. Robert, and M. A. Thiel Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry. Arch Ophthalmol, August 1, 2006; 124(8): 1104 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Kotecha, E T White, J M Shewry, and D F Garway-Heath The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry Br. J. Ophthalmol., December 1, 2005; 89(12): 1572 - 1575. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kniestedt, S. Lin, J. Choe, A. Bostrom, M. Nee, and R. L. Stamper Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry Arch Ophthalmol, November 1, 2005; 123(11): 1532 - 1537. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. S. Lim, S. S. Wickremasinghe, M. F. Cordeiro, C. Bunce, and P. T. Khaw Accuracy of Intraocular Pressure Measurements in New Zealand White Rabbits Invest. Ophthalmol. Vis. Sci., July 1, 2005; 46(7): 2419 - 2423. [Abstract] [Full Text] [PDF] |
||||
![]() |
P-A Tonnu, T Ho, T Newson, A El Sheikh, K Sharma, E White, C Bunce, and D Garway-Heath The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry Br. J. Ophthalmol., July 1, 2005; 89(7): 851 - 854. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. G. Pallikaris, G. D. Kymionis, H. S. Ginis, G. A. Kounis, and M. K. Tsilimbaris Ocular Rigidity in Living Human Eyes Invest. Ophthalmol. Vis. Sci., February 1, 2005; 46(2): 409 - 414. [Abstract] [Full Text] [PDF] |
||||
![]() |
A C Browning, A Bhan, A P Rotchford, S Shah, and H S Dua The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology Br. J. Ophthalmol., November 1, 2004; 88(11): 1395 - 1399. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kniestedt, M. Nee, and R. L. Stamper Dynamic Contour Tonometry: A Comparative Study on Human Cadaver Eyes Arch Ophthalmol, September 1, 2004; 122(9): 1287 - 1293. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Li, L. W. Herndon, S. G. Asrani, S. Stinnett, and R. R. Allingham Clinical Comparison of the Proview Eye Pressure Monitor With the Goldmann Applanation Tonometer and the TonoPen Arch Ophthalmol, August 1, 2004; 122(8): 1117 - 1121. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Gunvant, M Baskaran, L Vijaya, I S Joseph, R J Watkins, M Nallapothula, D C Broadway, and D J O'Leary Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer Br. J. Ophthalmol., April 1, 2004; 88(4): 518 - 522. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, and M. A. Thiel Intraocular Pressure Measurements Using Dynamic Contour Tonometry after Laser In Situ Keratomileusis Invest. Ophthalmol. Vis. Sci., September 1, 2003; 44(9): 3790 - 3794. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. J. Lehmann, S. Tuft, G. Brice, R. Smith, A. Blixt, R. Bell, B. Johansson, T. Jordan, R. A. Hitchings, P. T. Khaw, et al. Novel Anterior Segment Phenotypes Resulting from Forkhead Gene Alterations: Evidence for Cross-Species Conservation of Function Invest. Ophthalmol. Vis. Sci., June 1, 2003; 44(6): 2627 - 2633. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |