From the Department of Ophthalmology, University Clinic of Navarra, Pamplona, Navarra, Spain.
PURPOSE. To quantitatively assess retinal thickness by optical coherence tomography (OCT) in normal subjects and patients with diabetes. This study was intended to determine which retinal thickness value measured with OCT best discriminates between diabetic eyes, with and without macular edema.
METHODS. OCT retinal thickness was measured by a manual technique in a total of 26 healthy volunteers (44 control eyes) and 85 patients with diabetes (148 eyes) with the clinical diagnosis of no diabetic retinopathy (45 eyes), nonproliferative diabetic retinopathy without clinically significant macular edema (CSME; 54 eyes), proliferative diabetic retinopathy without CSME (21 eyes), and 28 eyes with diabetic retinopathy with CSME. Independent predictors of the presence of CSME were quantified by using univariate and multivariate logistic regression analyses. Receiver operating characteristic (ROC) curves were generated to evaluate and compare the predictor variables. The correlation of retinal thickness measurements and visual acuity was calculated.
RESULTS. There were statistically significant differences in foveal thickness between control eyes and all the other eye groups (P = 0.001). Diabetic eyes with CSME had a statistically significant greater thickness in each of the areas compared with the other groups. In a multivariate logistic regression model, foveal thickness was a strong and independent predictor of CSME (odds ratio [OR], 1.037; 95% confidence interval [CI] 1.021.05). The area under the ROC curve of this predictor variable was 0.94 (P = 0.001). For a cutoff point of 180 µm, the sensitivity was 93%, and specificity was 75%. Foveal thickness correlated with visual acuity in a log minimum angle of resolution (logMAR) scale (Spearmans
= 0.9, P = 0.001).
CONCLUSIONS. These results suggest that foveal thickening over 180 µm measured by OCT may be useful for the early detection of macular thickening and may be an indicator for a closer follow-up of the patient with diabetes.
This article has been cited by other articles:
![]() |
G. Virgili, F. Menchini, A. F. Dimastrogiovanni, E. Rapizzi, U. Menchini, F. Bandello, and R. G. Chiodini Optical Coherence Tomography versus Stereoscopic Fundus Photography or Biomicroscopy for Diagnosing Diabetic Macular Edema: A Systematic Review Invest. Ophthalmol. Vis. Sci., November 1, 2007; 48(11): 4963 - 4973. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Biallosterski, M. E J van Velthoven, R. P J Michels, R. O Schlingemann, J H. DeVries, and F. D Verbraak Decreased optical coherence tomography-measured pericentral retinal thickness in patients with diabetes mellitus type 1 with minimal diabetic retinopathy Br. J. Ophthalmol., September 1, 2007; 91(9): 1135 - 1138. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Campbell, S. G. Coupland, R. R. Buhrmann, and P. J. Kertes Optimal Optical Coherence Tomography-Based Measures in the Diagnosis of Clinically Significant Macular Edema: Retinal Volume vs Foveal Thickness Arch Ophthalmol, May 1, 2007; 125(5): 619 - 623. [Abstract] [Full Text] [PDF] |
||||
![]() |
K A Goatman A reference standard for the measurement of macular oedema. Br. J. Ophthalmol., September 1, 2006; 90(9): 1197 - 1202. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Vujosevic, E. Midena, E. Pilotto, P. P. Radin, L. Chiesa, and F. Cavarzeran Diabetic Macular Edema: Correlation between Microperimetry and Optical Coherence Tomography Findings. Invest. Ophthalmol. Vis. Sci., July 1, 2006; 47(7): 3044 - 3051. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Garcia-Layana, A Salinas-Alaman, M J Maldonado, C Sainz-Gomez, and A Fernandez-Hortelano Optical coherence tomography to monitor photodynamic therapy in pathological myopia Br. J. Ophthalmol., May 1, 2006; 90(5): 555 - 558. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Chan, J. S. Duker, T. H. Ko, J. G. Fujimoto, and J. S. Schuman Normal macular thickness measurements in healthy eyes using stratus optical coherence tomography. Arch Ophthalmol, February 1, 2006; 124(2): 193 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Moreno-Montanes, A. Alvarez, and M. J. Maldonado Objective Quantification of Posterior Capsule Opacification after Cataract Surgery, with Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., November 1, 2005; 46(11): 3999 - 4006. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Chan and J. S. Duker A Standardized Method for Reporting Changes in Macular Thickening Using Optical Coherence Tomography Arch Ophthalmol, July 1, 2005; 123(7): 939 - 943. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Costa, D. Calucci, M. Skaf, J. A. Cardillo, J. C. Castro, L. A. Melo Jr, M. C. Martins, and P. K. Kaiser Optical Coherence Tomography 3: Automatic Delineation of the Outer Neural Retinal Boundary and Its Influence on Retinal Thickness Measurements Invest. Ophthalmol. Vis. Sci., July 1, 2004; 45(7): 2399 - 2406. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Brown, S. D. Solomon, S. B. Bressler, A. P. Schachat, C. DiBernardo, and N. M. Bressler Detection of Diabetic Foveal Edema: Contact Lens Biomicroscopy Compared With Optical Coherence Tomography Arch Ophthalmol, March 1, 2004; 122(3): 330 - 335. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Terasaki, T. Kojima, H. Niwa, C.-H. Piao, S. Ueno, M. Kondo, Y. Ito, and Y. Miyake Changes in Focal Macular Electroretinograms and Foveal Thickness after Vitrectomy for Diabetic Macular Edema Invest. Ophthalmol. Vis. Sci., October 1, 2003; 44(10): 4465 - 4472. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. Neubauer, U. C. Welge-Lussen, M. J. Thiel, C. Alge, S. G. Priglinger, C. Hirneiss, M. W. Ulbig, and A. Kampik Tele-Screening for Diabetic Retinopathy With the Retinal Thickness Analyzer Diabetes Care, October 1, 2003; 26(10): 2890 - 2897. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Drexler, H. Sattmann, B. Hermann, T. H. Ko, M. Stur, A. Unterhuber, C. Scholda, O. Findl, M. Wirtitsch, J. G. Fujimoto, et al. Enhanced Visualization of Macular Pathology With the Use of Ultrahigh-Resolution Optical Coherence Tomography Arch Ophthalmol, May 1, 2003; 121(5): 695 - 706. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |