From the Tennent Institute of Ophthalmology, University of Glasgow, Glasgow, Scotland, United Kingdom.
PURPOSE. To assess the repeatability and interoperator and intersession reproducibility of central corneal thickness (CCT) measurements made by a commercially available optical coherence tomography (OCT) system.
METHODS. Intersession and interoperator reproducibility and repeatability were defined according to the guidelines of the British Standards Institution and examined in a control group of 14 normal subjects. An in-house computer program was used to evaluate central corneal thickness from these scans.
RESULTS. The coefficient of interoperator reproducibility was 0.18%, whereas that for intersession reproducibility was 1.11%. Wilcoxon analysis (5% level of significance) showed that there was no statistically significant difference between scans acquired during different sessions or by different operators. Coefficients of repeatability were all less than 3%. The average CCT was 526 ± 28 µm (SD) and the range of normal CCT between 5th and 95th percentiles was 498 to 576 µm.
CONCLUSIONS. Although the commercially available OCT scanner was designed for retinal imaging, with a few minor modifications, the system may be used to image the anterior segment. Previous studies have shown that OCT measurements correlate well with those from conventional techniques, and it has the added advantage of being a noncontact technique. This study further demonstrates that the OCT measurements show a high degree of repeatability and reproducibility. Thus, OCT is emerging as a promising tool for evaluation of CCT in the clinical setting.
This article has been cited by other articles:
![]() |
S. Haque, D. Fonn, T. Simpson, and L. Jones Epithelial Thickness Changes from the Induction of Myopia with CRTH RGP Contact Lenses Invest. Ophthalmol. Vis. Sci., August 1, 2008; 49(8): 3345 - 3350. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Lewis, A. E. Knellinger, A. M. Mahmoud, and T. F. Mauger Effect of Soft Contact Lenses on Optical Measurements of Axial Length and Keratometry for Biometry in Eyes with Corneal Irregularities Invest. Ophthalmol. Vis. Sci., August 1, 2008; 49(8): 3371 - 3378. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mohamed, G. K. Y. Lee, S. K. Rao, A. L. Wong, A. C. K. Cheng, E. Y. M. Li, S. C. C. Chi, and D. S. C. Lam Repeatability and Reproducibility of Pachymetric Mapping with Visante Anterior Segment Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., December 1, 2007; 48(12): 5499 - 5504. [Abstract] [Full Text] [PDF] |
||||
![]() |
M J Gallagher, T Yilmaz, R A Cervantes-Castaneda, and C S Foster The characteristic features of optical coherence tomography in posterior uveitis Br. J. Ophthalmol., December 1, 2007; 91(12): 1680 - 1685. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Martin, V. de Juan, G. Rodriguez, R. Cuadrado, and I. Fernandez Measurement of Corneal Swelling Variations without Removal of the Contact Lens during Extended Wear Invest. Ophthalmol. Vis. Sci., July 1, 2007; 48(7): 3043 - 3050. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, J. Aquavella, J. Palakuru, S. Chung, and C. Feng Relationships between Central Tear Film Thickness and Tear Menisci of the Upper and Lower Eyelids. Invest. Ophthalmol. Vis. Sci., October 1, 2006; 47(10): 4349 - 4355. [Abstract] [Full Text] [PDF] |
||||
![]() |
K A Goatman A reference standard for the measurement of macular oedema. Br. J. Ophthalmol., September 1, 2006; 90(9): 1197 - 1202. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, J. Aquavella, J. Palakuru, and S. Chung Repeated measurements of dynamic tear distribution on the ocular surface after instillation of artificial tears. Invest. Ophthalmol. Vis. Sci., August 1, 2006; 47(8): 3325 - 3329. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Chen, B. Cense, M. C. Pierce, N. Nassif, B. H. Park, S. H. Yun, B. R. White, B. E. Bouma, G. J. Tearney, and J. F. de Boer Spectral Domain Optical Coherence Tomography: Ultra-high Speed, Ultra-high Resolution Ophthalmic Imaging Arch Ophthalmol, December 1, 2005; 123(12): 1715 - 1720. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Moreno-Montanes, A. Alvarez, and M. J. Maldonado Objective Quantification of Posterior Capsule Opacification after Cataract Surgery, with Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., November 1, 2005; 46(11): 3999 - 4006. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Radhakrishnan, J. Goldsmith, D. Huang, V. Westphal, D. K. Dueker, A. M. Rollins, J. A. Izatt, and S. D. Smith Comparison of Optical Coherence Tomography and Ultrasound Biomicroscopy for Detection of Narrow Anterior Chamber Angles Arch Ophthalmol, August 1, 2005; 123(8): 1053 - 1059. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, J. Thomas, I. Cox, and A. Rollins Noncontact Measurements of Central Corneal Epithelial and Flap Thickness after Laser In Situ Keratomileusis Invest. Ophthalmol. Vis. Sci., June 1, 2004; 45(6): 1812 - 1816. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Miglior, E Albe, M Guareschi, G Mandelli, S Gomarasca, and N Orzalesi Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness Br. J. Ophthalmol., February 1, 2004; 88(2): 174 - 177. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, D. Fonn, T. L. Simpson, and L. Jones Precorneal and Pre- and Postlens Tear Film Thickness Measured Indirectly with Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., June 1, 2003; 44(6): 2524 - 2528. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, D. Fonn, and T. L. Simpson Topographical Thickness of the Epithelium and Total Cornea after Hydrogel and PMMA Contact Lens Wear with Eye Closure Invest. Ophthalmol. Vis. Sci., March 1, 2003; 44(3): 1070 - 1074. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |