IOVS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2003;44:5043-5048.)
© 2003 by The Association for Research in Vision and Ophthalmology, Inc.
doi:10.1167/iovs.03-0058

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Anderson, A. J.
Right arrow Articles by Johnson, C. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Anderson, A. J.
Right arrow Articles by Johnson, C. A.

Anatomy of a Supergroup: Does a Criterion of Normal Perimetric Performance Generate a Supernormal Population?

Andrew John Anderson and Chris A. Johnson

From Discoveries in Sight, Devers Eye Institute, Portland, Oregon.

PURPOSE. To interpret individual results from automated perimeters, a normative database must be developed. Typically, a set of criteria determines those subjects that may be included in the database. This study examined whether a criterion of normal performance on an established perimeter generates a subgroup with supernormal perimetric performance.

METHODS. The right-eye perimetric results of 100 subjects were analyzed. Subjects had visual acuities of 6/12 or better, no history of eye disease, and normal slit lamp biomicroscopic and ophthalmoscopic examinations. Subjects performed test–retest visual field examinations on a Humphrey Field Analyzer (HFA) 24-2 test (Zeiss Humphrey Systems, Dublin, CA), and on a custom frequency-doubling (FD) perimeter with targets spaced in the same 24-2 pattern.

RESULTS. Test–retest correlation (Spearman rank correlation coefficients, rs) for mean defect (MD) and pattern SD (PSD) were 0.65 and 0.40 (HFA), and 0.82 and 0.39 (FD perimeter). Three subjects with HFA MDs in the lower 5% had similarly low MDs on retest, whereas no subject was common between the test and retest for the lower 5% of HFA PSD. Correlation between the HFA and FD test results were 0.41 (MD) and 0.05 (PSD). Based on these correlations, the bias introduced into perimetric probability limits were determined, by using Monte Carlo simulations.

CONCLUSIONS. Although a criterion of a normal MD may produce a subpopulation with supernormal perimetric performance, a criterion of a normal PSD is less likely to do so. Also, a criterion on one test type is less likely to create a supernormal group on a different test type. The bias introduced into perimetric probability limits is small.





This article has been cited by other articles:


Home page
IOVSHome page
P. H. Artes and B. C. Chauhan
Signal/Noise Analysis to Compare Tests for Measuring Visual Field Loss and Its Progression
Invest. Ophthalmol. Vis. Sci., October 1, 2009; 50(10): 4700 - 4708.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
A. J. Anderson, C. A. Johnson, M. Fingeret, J. L. Keltner, P. G. D. Spry, M. Wall, and J. S. Werner
Characteristics of the Normative Database for the Humphrey Matrix Perimeter
Invest. Ophthalmol. Vis. Sci., April 1, 2005; 46(4): 1540 - 1548.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2003 by the Association for Research in Vision and Ophthalmology