1From the Department of Ophthalmology, Herlev Hospital, University of Copenhagen, Copenhagen, Denmark; 2Retinalyze A/S, Hørsholm, Denmark; the 3Department of Ophthalmology, Odense University Hospital, Odense, Denmark; the 4Department of Ophthalmology, Malmö University Hospital, Malmö, Sweden; the 5Department of Ophthalmology, Sahlgrenska University Hospital, Göteborg, Sweden; and the 6Department of Diabetes Research, University of Wales College of Medicine, Cardiff, Wales, United Kingdom.
PURPOSE. To compare a fundus image-analysis algorithm for automated detection of hemorrhages and microaneurysms with visual detection of retinopathy in patients with diabetes.
METHODS. Four hundred fundus photographs (35-mm color transparencies) were obtained in 200 eyes of 100 patients with diabetes who were randomly selected from the Welsh Community Diabetic Retinopathy Study. A gold standard reference was defined by classifying each patient as having or not having diabetic retinopathy based on overall visual grading of the digitized transparencies. A single-lesion visual grading was made independently, comprising meticulous outlining of all single lesions in all photographs and used to develop the automated red lesion detection system. A comparison of visual and automated single-lesion detection in replicating the overall visual grading was then performed.
RESULTS. Automated red lesion detection demonstrated a specificity of 71.4% and a resulting sensitivity of 96.7% in detecting diabetic retinopathy when applied at a tentative threshold setting for use in diabetic retinopathy screening. The accuracy of 79% could be raised to 85% by adjustment of a single user-supplied parameter determining the balance between the screening priorities, for which a considerable range of options was demonstrated by the receiver-operating characteristic (area under the curve 90.3%). The agreement of automated lesion detection with overall visual grading (0.659) was comparable to the mean agreement of six ophthalmologists (0.648).
CONCLUSIONS. Detection of diabetic retinopathy by automated detection of single fundus lesions can be achieved with a performance comparable to that of experienced ophthalmologists. The results warrant further investigation of automated fundus image analysis as a tool for diabetic retinopathy screening.
This article has been cited by other articles:
![]() |
A. D Fleming, K. A Goatman, S. Philip, G. J Williams, G. J Prescott, G. S Scotland, P. McNamee, G. P Leese, W. N Wykes, P. F Sharp, et al. The role of haemorrhage and exudate detection in automated grading of diabetic retinopathy Br J Ophthalmol, March 4, 2010; (2010) bjo.2008.149807v2. [Abstract] [Full Text] |
||||
![]() |
M. D. Abramoff, M. Niemeijer, M. S.A. Suttorp-Schulten, M. A. Viergever, S. R. Russell, and B. van Ginneken Evaluation of a System for Automatic Detection of Diabetic Retinopathy From Color Fundus Photographs in a Large Population of Patients With Diabetes Diabetes Care, February 1, 2008; 31(2): 193 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Philip, A D Fleming, K A Goatman, S Fonseca, P Mcnamee, G S Scotland, G J Prescott, P F Sharp, and J A Olson The efficacy of automated "disease/no disease" grading for diabetic retinopathy in a systematic screening programme Br J Ophthalmol, November 1, 2007; 91(11): 1512 - 1517. [Abstract] [Full Text] [PDF] |
||||
![]() |
G S Scotland, P McNamee, S Philip, A D Fleming, K A Goatman, G J Prescott, S Fonseca, P F Sharp, and J A Olson Cost-effectiveness of implementing automated grading within the national screening programme for diabetic retinopathy in Scotland Br J Ophthalmol, November 1, 2007; 91(11): 1518 - 1523. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Niemeijer, B. van Ginneken, S. R. Russell, M. S. A. Suttorp-Schulten, and M. D. Abramoff Automated Detection and Differentiation of Drusen, Exudates, and Cotton-Wool Spots in Digital Color Fundus Photographs for Diabetic Retinopathy Diagnosis Invest. Ophthalmol. Vis. Sci., May 1, 2007; 48(5): 2260 - 2267. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Larsen, J. Godt, M. Grunkin, H. Lund-Andersen, and M. Larsen Automated Detection of Diabetic Retinopathy in a Fundus Photographic Screening Population Invest. Ophthalmol. Vis. Sci., February 1, 2003; 44(2): 767 - 771. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |