IOVS Stem Cells
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2004;45:2682-2690.)
© 2004 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.03-1024

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow View responses
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (12)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tognetto, D.
Right arrow Articles by Ravalico, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tognetto, D.
Right arrow Articles by Ravalico, G.

Analysis of the Optical Quality of Intraocular Lenses

Daniele Tognetto,1 Giorgia Sanguinetti,1 Paolo Sirotti,2 Paolo Cecchini,1 Luana Marcucci,1 Enzo Ballone,3 and Giuseppe Ravalico1

1From the Eye Clinic and the 2Department of Electrotechnics, Electronics and Informatics, University of Trieste, Italy; and the 3Department of Medicine and Aging, University of Chieti, Chieti, Italy.

PURPOSE. To evaluate the optical quality of different intraocular lenses (IOLs).

METHODS. An optical test bench and suitable software were used to assist in analysis of the optical Fourier transform (OFT) of a test image and to determine the quality of the lens in terms of spatial frequency response. The OFT was automatically converted, by means of an optical–electronic calibration procedure, into a modulation transfer function (MTF) for each lens. The passband value calculated by computer analysis of the MTF is an objective index of the lens quality. Three randomly acquired samples of 24 different models of foldable IOLs were compared. Statistical analysis was performed with two-way and one-way ANOVA for repeated measurements and with the Ryan-Einot-Gabriel-Welsch multiple F test.

RESULTS. The method was demonstrated to be precise and accurate. A large range of passband values was found. Statistically significant differences between the mean passband values for different lenses were found. The lowest passband value (125.60 line pairs [lp]/mm) was measured for the IOL (Lenstec LH3000; Lenstec, Inc., St. Petersburg, FL) and the highest (191.48 lp/mm) for the Acrysof SA30AL (Alcon, Fort Worth, TX).

CONCLUSIONS. Different IOLs can transmit different spectra of spatial frequencies. The best frequency response was provided by acrylic IOLs, particularly those with an asymmetrically biconvex profile. This could be due to a reduction of optical degradation provided by this type of profile. A lens with a higher frequency response should determine a better quality of vision once implanted and the frequency response should therefore be considered when choosing the intraocular lens model.





This article has been cited by other articles:


Home page
Br. J. Ophthalmol.Home page
L Mitchell, A C B Molteno, T H Bevin, and G Sanderson
Star testing: a novel evaluation of intraocular lens optical quality
Br. J. Ophthalmol., May 1, 2006; 90(5): 586 - 592.
[Abstract] [Full Text] [PDF]

eLetters:

Read all eLetters

Analysis of the Optical Quality of Intraocular Lenses
Ian N. Hickling
IOVS Online, 3 Jan 2005 [Full text]
Author Response: Analysis of the Optical Quality of Intraocular Lenses
Daniele Tognetto
IOVS Online, 3 Jan 2005 [Full text]



HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS