IOVS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2005;46:1280-1285.)
© 2005 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.04-1000

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (22)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bourne, R. R. A.
Right arrow Articles by Weinreb, R. N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bourne, R. R. A.
Right arrow Articles by Weinreb, R. N.

Comparability of Retinal Nerve Fiber Layer Thickness Measurements of Optical Coherence Tomography Instruments

Rupert R. A. Bourne, Felipe A. Medeiros, Christopher Bowd, Keyvan Jahanbakhsh, Linda M. Zangwill, and Robert N. Weinreb

From the Hamilton Glaucoma Center and Department of Ophthalmology, University of California, San Diego, La Jolla, California.

PURPOSE. To compare retinal nerve fiber layer (RNFL) measurements between two ocular coherence tomography (OCT) instruments (OCT 2000 and Stratus OCT; Carl Zeiss Meditec, Dublin, CA) and compare their diagnostic precision.

METHODS. One hundred thirty-nine consecutive subjects were imaged (3 x 3.4-mm diameter circular scans) on the same day with each instrument. Thirty-five patients were excluded due to poor-quality images. RNFL thicknesses measured by the two instruments were compared, and receiver operating characteristic (ROC) curves were used to determine diagnostic precision.

RESULTS. A randomly selected eye of each of 104 participants (28 with open-angle glaucoma, 40 with suspected glaucoma, and 36 healthy subjects) was analyzed. RNFL thickness measurements generally were thicker with OCT 2000 than with Stratus OCT. The difference in global RNFL thickness between instruments was within 20 µm in 66 (65%) of subjects and within 10 µm (the instrument’s limit of resolution) in 25 (25%) subjects. Application of a correction factor to OCT 2000 measurements predicted Stratus OCT RNFL thickness within 10 µm of the observed measurement in 75% of the eyes. For both instruments, highest ROC curve areas (better discrimination between glaucomatous and normal eyes) were found in the inferior sector. Discrimination using global RNFL thickness was better with Stratus OCT than OCT 2000 (P = 0.043).

CONCLUSIONS. RNFL thickness measurements measured by OCT 2000 can be approximated to measurements made by Stratus OCT using correction factors calculated by this study. However, there remains considerable variability that exceeds the limits of resolution afforded by the instruments themselves. Therefore comparisons between instruments using these approximations should be interpreted with caution.





This article has been cited by other articles:


Home page
Br. J. Ophthalmol.Home page
C. Bowd
Optical coherence tomography for clinical detection and monitoring of glaucoma?
Br. J. Ophthalmol., July 1, 2007; 91(7): 853 - 854.
[Full Text] [PDF]


Home page
IOVSHome page
M.-L. Huang, H.-Y. Chen, and J.-C. Lin
Rule Extraction for Glaucoma Detection with Summary Data from StratusOCT
Invest. Ophthalmol. Vis. Sci., January 1, 2007; 48(1): 244 - 250.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
C. K.-S. Leung, S. Mohamed, K. S. Leung, C. Y.-L. Cheung, S. L.-w. Chan, D. K.-y. Cheng, A. K.-c. Lee, G. Y.-o. Leung, S. K. Rao, and D. S. C. Lam
Retinal Nerve Fiber Layer Measurements in Myopia: An Optical Coherence Tomography Study
Invest. Ophthalmol. Vis. Sci., December 1, 2006; 47(12): 5171 - 5176.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2005 by the Association for Research in Vision and Ophthalmology