From the Arlene R. Gordon Research Institute, Lighthouse International, New York, New York.
PURPOSE. To develop a small-format letter contrast sensitivity test with improved accuracy, suitable for portable use and near testing, and having a simple and unambiguous scoring system that can be used with reference to existing norms.
METHODS. A near test, printed on resin-coated paper and mounted on plastic was developed by using Sloan letters, with 0.04 log unit contrast decrements between each letter and a simple scoring procedure. Monte Carlo methods and a Weibull function model of visual performance were used to assess test accuracy.
RESULTS. The new test has 28% lower score standard deviations than the Pelli-Robson Contrast Sensitivity Chart, over a wide range of low misreporting rates, while maintaining test scores that differ from the Pelli-Robson by 0.01 log unit, permitting use of norms collected with the Pelli-Robson test.
CONCLUSIONS. While maintaining comparability with Pelli-Robson norms, the new test has improved accuracy in comparison with the Pelli-Robson chart and several other advantages that result from its smaller size.
This article has been cited by other articles:
![]() |
K. Thayaparan, M. D Crossland, and G. S Rubin Clinical assessment of two new contrast sensitivity charts Br. J. Ophthalmol., June 1, 2007; 91(6): 749 - 752. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Haymes, K. F. Roberts, A. F. Cruess, M. T. Nicolela, R. P. LeBlanc, M. S. Ramsey, B. C. Chauhan, and P. H. Artes The letter contrast sensitivity test: clinical evaluation of a new design. Invest. Ophthalmol. Vis. Sci., June 1, 2006; 47(6): 2739 - 2745. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |