IOVS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2005;46:2433-2439.)
© 2005 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.04-1108

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Boden, C.
Right arrow Articles by Sample, P. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Boden, C.
Right arrow Articles by Sample, P. A.

Relationship of SITA and Full-Threshold Standard Perimetry to Frequency-Doubling Technology Perimetry in Glaucoma

Catherine Boden,1 John Pascual,1 Felipe A. Medeiros,1 Makoto Aihara,1,2 Robert N. Weinreb,1 and Pamela A. Sample1

1From the Hamilton Glaucoma Center and Visual Function Laboratory, University of California at San Diego, La Jolla, California; and the 2Department of Ophthalmology, University of Tokyo, Tokyo, Japan.

PURPOSE. To compare full-threshold (FT) and SITA algorithms for standard automated perimetry (SAP) with frequency-doubling technology perimetry (FDT) in glaucoma, to help clinicians to relate results in patients who have had two or more of these tests during follow-up.

METHODS. This study was a retrospective analysis of data from a longitudinal prospective study at the University of California, San Diego. One hundred four eyes of 104 patients with glaucomatous optic neuropathy detected by optic disc stereophotographs were included. All patients had standard perimetry (SITA and FT) and FDT within 3 months of each other. Global indices, abnormality and severity using two threshold algorithms of standard perimetry were compared with FDT.

RESULTS. More eyes had normal visual fields by SAP-FT (57 eyes) than by either SAP-SITA (42 eyes) or FDT (45 eyes), although SAP-FT agreed more closely with FDT ({kappa} = 0.54 ± 0.08) on the presence of a visual field defect than did SAP-SITA ({kappa} = 0.34 ± 0.08). Correlations of FDT to standard perimetry global indices were similar regardless of the threshold strategy used for standard perimetry, yielding r2 = 0.38–0.57 for SAP-FT with FDT, 0.36–0.54 for SAP-SITA with FDT.

CONCLUSIONS. Despite many similarities of SAP-SITA and SAP-FT, switching the standard of comparison from SAP-FT to SAP-SITA changes the relationship to FDT with regard to visual field abnormality, but not correlations of global indices. FDT detected abnormal fields in more eyes than SAP-FT. SAP-FT tended to detect a subset of eyes found abnormal by FDT. Visual field defects may be detected more often by FDT and SAP-SITA in eyes with early visual field loss, but these two tests may not agree on which eyes show field loss in patients who undergo both tests at follow-up.





This article has been cited by other articles:


Home page
Arch OphthalmolHome page
L. M. Sakata, J. DeLeon-Ortega, S. N. Arthur, B. E. Monheit, and C. A. Girkin
Detecting Visual Function Abnormalities Using the Swedish Interactive Threshold Algorithm and Matrix Perimetry in Eyes With Glaucomatous Appearance of the Optic Disc
Arch Ophthalmol, March 1, 2007; 125(3): 340 - 345.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Ophthalmol.Home page
E Convento, E Midena, M T Dorigo, V Maritan, F Cavarzeran, and I A Fregona
Peripapillary fundus perimetry in eyes with glaucoma
Br. J. Ophthalmol., November 1, 2006; 90(11): 1398 - 1403.
[Abstract] [Full Text] [PDF]