IOVS Journal of Neurophysiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2007;48:5494-5498.)
© 2007 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.06-1479

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Romppainen, T.
Right arrow Articles by Thiel, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Romppainen, T.
Right arrow Articles by Thiel, M. A.

Effect of Riboflavin-UVA–Induced Collagen Cross-linking on Intraocular Pressure Measurement

Timo Romppainen,1 Lucas M. Bachmann,2 Claude Kaufmann,1 Christoph Kniestedt,1 Michael Mrochen,3 and Michael A. Thiel1,4

1From the Department of Ophthalmology and the 2Horten Centre for Patient-Oriented Research, University of Zurich, Zurich, Switzerland; the 3IROC Institute for Refractive and Ophthalmic Surgery, Zurich, Switzerland; and the 4Department of Ophthalmology, Cantonal Hospital Lucerne, Lucerne, Switzerland.

PURPOSE. Corneal collagen cross-linking (CCL) with riboflavin and ultraviolet A irradiation has recently been introduced for treatment of corneal ectasia. Yet a CCL-induced increase in corneal rigidity may interfere with intraocular pressure (IOP) measurements. In an investigation of the effect of CCL on the accuracy of IOP measurements, IOP readings before and after CCL were compared.

METHODS. Ten human eye bank corneas were de-epithelialized and mounted on an artificial anterior chamber. The hydrostatically controlled reference pressure in the chamber was adjusted from 10 to 40 mm Hg in 5-mm Hg steps. IOP was measured by Goldmann applanation tonometry (GAT; Haag Streit, Könitz, Switzerland), dynamic contour tonometry (DCT; Pascal tonometer; Ziemer Ophthalmics, Port, Switzerland), and the TonoPen XL (TP; Tono-Pen XL, Medtronic, Jacksonville, FL) before and after CCL, which was performed with a 0.1% riboflavin solution and 30 minutes of UVA irradiation.

RESULTS. Before CCL, GAT, and DCT readings showed an excellent concordance with the manometric reference pressure, whereas TP overestimated the true IOP. After CCL, the reliability of IOP readings decreased with all three tonometers. This decrease resulted in a slight overestimation of mean IOP, but there were also some potentially dangerous underestimations in some individual corneas. The mean (±SD) difference between IOP readings after and before CCL was +1.8 (3.5) mm Hg for DCT, +2.9 (6.1) mm Hg for GAT, and +3.1 (8.3) mm Hg for TP (P ≤ 0.002 for DCT versus GAT or TP).

CONCLUSIONS. In this in vitro model on human corneas, CCL resulted in an overestimation of true IOP by all the tested tonometers. Although the magnitude of this effect was small, care should be taken when measuring IOP with GAT after CCL, as it results in less accurate, much more variable IOP readings.