1From the Department of Ophthalmology, Université Pierre et Marie Curie-Paris 6, Paris, France; 2Department of Ophthalmology, Centre Hospitalier National dOphtalmologie des Quinze-Vingts, Paris, France; 3Department of Ophthalmology, Fondation Ophtalmologique Rothschild, Paris, France; and 4Laboratory of Cellular and Molecular Physiopathology of the Retina, National Institute for Health and Medical Research (INSERM U592), Paris, France.
PURPOSE. To analyze high-resolution color photographs of the mouse fundus.
METHODS. A contact fundus camera based on topical endoscopy fundus imaging (TEFI) was built. Fundus photographs of C57 and Balb/c mice obtained by TEFI were qualitatively analyzed.
RESULTS. High-resolution digital imaging of the fundus, including the ciliary body, was routinely obtained. The reflectance and contrast of retinal vessels varied significantly with the amount of incident and reflected light and, thus, with the degree of fundus pigmentation. The combination of chromatic and spherical aberration favored blue light imaging, in term of both field and contrast.
CONCLUSIONS. TEFI is a small, low-cost system that allows high-resolution color fundus imaging and fluorescein angiography in conscious mice. Panretinal imaging is facilitated by the presence of the large rounded lens. TEFI significantly improves the quality of in vivo photography of retina and ciliary process of mice. Resolution is, however, affected by chromatic aberration, and should be improved by monochromatic imaging.
This article has been cited by other articles:
![]() |
J.-L. Guyomard, S. G. Rosolen, M. Paques, M.-N. Delyfer, M. Simonutti, Y. Tessier, J. A. Sahel, J.-F. Legargasson, and S. Picaud A Low-Cost and Simple Imaging Technique of the Anterior and Posterior Segments: Eye Fundus, Ciliary Bodies, Iridocorneal Angle Invest. Ophthalmol. Vis. Sci., November 1, 2008; 49(11): 5168 - 5174. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |