IOVS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 2007;48:3178-3183.)
© 2007 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.06-1315

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (2)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Leung, C. K.-s.
Right arrow Articles by Lam, D. S. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Leung, C. K.-s.
Right arrow Articles by Lam, D. S. C.

Optic Disc Measurements in Myopia with Optical Coherence Tomography and Confocal Scanning Laser Ophthalmoscopy

Christopher Kai-shun Leung, Arthur Chak Kwan Cheng, Kelvin Kam Lung Chong, King Sai Leung, Shaheeda Mohamed, Charles Sing Lok Lau, Carol Yim Lui Cheung, Geoffrey Chin-hung Chu, Ricky Yiu Kwong Lai, Calvin Chi Pui Pang, and Dennis Shun Chiu Lam

From the Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, Peoples Republic of China.

PURPOSE. To evaluate the relationships between optic disc measurements, obtained by an optical coherence tomograph and a confocal scanning laser ophthalmoscope, and myopia.

METHODS. One hundred thirty-three eyes from 133 healthy subjects with mean spherical equivalent –6.0 ± 4.2 D (range, –13.13 to +3.25 D) were analyzed. Optic disc measurements including disc area, rim area, cup area, cup-to-disc area, and vertical and horizontal ratios were obtained with an optical coherence tomograph (StratusOCT; Carl Zeiss Meditec Inc., Dublin, CA) and a confocal scanning laser ophthalmoscope (Heidelberg Retina Tomograph, HRT 3; Heidelberg Engineering, GmbH, Dossenheim, Germany). The modified axial length method derived from prior published work was used to correct the OCT measurements for ocular magnification. Bland-Altman plots were used to evaluate the agreement for each optic disc parameter. Associations between optic disc area and axial length/spherical equivalent were evaluated by linear regression analysis.

RESULTS. Disc area increased with the axial length/negative spherical equivalent in the HRT and the corrected OCT measurements although opposite directions of associations were found when the OCT measurements were not corrected for magnification. The difference of the corrected OCT and HRT disc area (corrected OCT disc area minus HRT disc area) was correlated with the axial length (r = 0.195, P = 0.025). When the ametropia was limited to –8.0 to +4.0 D, the correlations became insignificant in the HRT. Using the corrected OCT measurements, disc area, rim area, and cup area, cup-to-disc area, and cup-to-disc horizontal and vertical ratios were significantly larger than those measured by the HRT, with a span of 95% limits of agreement at 1.99, 1.33, and 1.86 mm2 for the areas, 0.34, 0.53, and 0.58 for the ratios, respectively.

CONCLUSIONS. While optic disc area generally increased with the axial length and myopic refraction, the HRT measurements demonstrated that optic disc size was largely independent of axial length and refractive error between –8 and +4 D. OCT may overestimate optic disc size in myopic eyes and results in poor agreement between the two instruments.





This article has been cited by other articles:


Home page
Br. J. Ophthalmol.Home page
A Nagai-Kusuhara, M Nakamura, Y Tatsumi, Y Nakanishi, and A Negi
Disagreement between Heidelberg Retina Tomograph and optical coherence tomography in assessing optic nerve head configuration of eyes with band atrophy and normal eyes
Br. J. Ophthalmol., October 1, 2008; 92(10): 1382 - 1386.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Ophthalmol.Home page
S A Vernon, A P Rotchford, A Negi, S Ryatt, and C Tattersal
Peripapillary retinal nerve fibre layer thickness in highly myopic Caucasians as measured by Stratus optical coherence tomography
Br. J. Ophthalmol., August 1, 2008; 92(8): 1076 - 1080.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
M. He, B. Liu, W. Huang, J. Zhang, Q. Yin, Y. Zheng, D. Wang, and J. Ge
Heritability of Optic Disc and Cup Measured by the Heidelberg Retinal Tomography in Chinese: The Guangzhou Twin Eye Study
Invest. Ophthalmol. Vis. Sci., April 1, 2008; 49(4): 1350 - 1355.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2007 by the Association for Research in Vision and Ophthalmology