IOVS News in Physiological Sciences
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Originally published In Press as doi:10.1167/iovs.07-1229 on March 7, 2008
(Investigative Ophthalmology and Visual Science. 2008;49:2512-2517.)
© 2008 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.07-1229

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
iovs.07-1229v1
49/6/2512    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Xu, J.
Right arrow Articles by Schuman, J. S.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Xu, J.
Right arrow Articles by Schuman, J. S.

Automated Assessment of the Optic Nerve Head on Stereo Disc Photographs

Juan Xu,1 Hiroshi Ishikawa,1,2 Gadi Wollstein,1 Richard A. Bilonick,1 Kyung R. Sung,1 Larry Kagemann,1,2 Kelly A. Townsend,1 and Joel S. Schuman1,2

1From the UPMC Eye Center, Eye and Ear Institute, Ophthalmology and Visual Science Research Center, Department of Ophthalmology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania; and the 2Department of Bioengineering, Swanson School of Engineering, University of Pittsburgh, Pittsburgh, Pennsylvania.

PURPOSE. To develop automated software for optic nerve head (ONH) quantitative assessment from stereoscopic disc photographs and to evaluate its performance in comparison with human expert assessment.

METHODS. A fully automated system, including three-dimensional ONH modeling, disc margin detection, cup margin detection, and calculation of stereometric ONH parameters, was developed and tested. One eye each from 54 subjects (23 healthy, 17 suspected glaucoma, and 14 glaucoma) was enrolled. The majority opinion of three experts defined disc and cup margins on the disc photographs was used for comparison. Seven ONH parameters, disc area, rim area, rim volume, cup area, cup volume, cup-to-disc (C/D) area ratio, and vertical C/D ratio, were computed based on both machine- and expert-defined margins and compared between the methods.

RESULTS. All automated ONH measurements showed good correlation with the expert defined margins (Pearson r = 0.90, disc area; 0.56, rim area; 0.78, rim volume; 0.88, cup area; 0.93, cup volume; 0.69, C/D area ratio; and 0.67, vertical C/D ratio; all P ≤ 0.0001). No statistically significant difference was found in the glaucoma-discriminating ability of all seven ONH parameters (P ≥ 0.21). The mean or median of automatically defined disc and cup areas was significantly higher than the subjective assessment (disc area P = 0.0001, t-test; cup area P = 0.036, Wilcoxon signed ranks test), although they had high correlation coefficients. The software failed to detect the disc margin for all the disc photographs with peripapillary atrophy.

CONCLUSIONS. The automated ONH analysis method provides an objective and quantitative ONH evaluation using widely available stereo disc photographs.








HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2008 by the Association for Research in Vision and Ophthalmology