|
|
||||||||
From the Department of Ophthalmology, Rayne Institute, United Medical and Dental Schools of Guys and St. Thomas, St. Thomas Hospital, London, United Kingdom.
Abstract
PURPOSE. To determine the relationship between optical coherence tomography (OCT) images of the retina and retinal substructure in vitro and in vivo.
METHODS. In vitro, OCT images of human and bovine retina were acquired after sequential excimer laser ablation of the inner retinal layers. Measurements of bands in the OCT images were compared with measurements of retinal layers on histology of the ablated specimens. In vivo, OCT images were acquired of retinal lesions in which there was a displacement of pigmented retinal pigment epithelial (RPE) cells: retinitis pigmentosa and laser photocoagulation (eight eyes each).
RESULTS. The mean thickness of human inner OCT bands (131 µm; 95% confidence interval [CI], 122140 µm) was 7.3 times that of the retinal nerve fiber layer (RNFL). This band persisted despite ablation greater than 140 µm. The inner aspect of the outer OCT band corresponded to the apical RPE, but the mean thickness of this band in human tissue (55 µm; 95% CI, 4862 µm) was 2.6 times the thickness of the RPEchoriocapillaris complex. OCT measurement of total retinal thickness was accurate (coefficient of variance, 0.05) and precise (coefficient of correlation with light microscopy, 0.98). Hyperpigmented lesions gave rise to high signal, attenuating deeper signal; hypopigmented lesions had the opposite effect on deeper signal.
CONCLUSIONS. The inner band is not RNFL specific, partly consisting of a surface-related signal. The location, not thickness, of the outer band corresponds to RPE melanin. Given the additional effect of polarization settings, precise OCT measurement of specific retinal layers is currently precluded.
High-resolution axial measurements of ocular tissues have become important in both the medical and surgical management of disease. For example, the advent of the intraocular lens led to a demand for precise measurement of the axial length of the globe and encouraged the development of ultrasonic scanning systems.1 More recently, an increasing requirement for the measurement of retinal thickness has become apparent, particularly in relation to the management of macular disease.2 To date, estimations of retinal thickness have been achieved by stereoscopic biomicroscopy,3 and measurements have been obtained using laser-generated slits4 or scanning slit devices.5 The advent of optical coherence tomography (OCT) offers the theoretical possibility of high-resolution measurements of both retinal thickness and the dimensions of the retinal component layers.
Detailed accounts of the underlying principles of OCT have been reported elsewhere.6 7 8 In summary, however, images are generated as a result of the interaction between a partially coherent beam of optical radiation and tissue components. The most important optical phenomena in OCT signal generation are the scatter,9 reflection,10 and absorption of incident light. In current biomedical applications, near infrared radiation is used almost exclusively as a source of illumination. The resolution of spatial information within OCT images has both theoretical and practical limitations. In axial, or z-plane, measurements the resolution is 17 µm in air,6 which is inversely related to the coherence length of the superluminescent diode light source. In biologic tissue it is somewhat larger but always better than the 120 µm currently available with ultrasound.11 In the xy plane, the maximum resolution is diffraction limited by the properties of the ocular media and is approximately 25 µm. Empirically, however, the overall resolution across the image is limited by the design of the instrument used (Humphrey Instruments, San Leandro, CA), in which 100 individual z-axis scans are required to generate each image, irrespective of the field of the image.
Commercial instruments became available late in 1996, and several groups have published pseudocolored images displaying several layers. These layers have stereospatial locations that apparently correlate with the different cell layers of the retina and choroid. Therefore, the spatial separation of the colored bands in these pseudocolor images is similar to that of a histologic section. There are four pseudocolored bands in images of extrafoveal retina that, on passing from the vitreous surface toward the sclera, have the following sequence: red-white, yellow-green, black, and red-white. Previous reports have assumed a correlation between the four bands and the four cell layers of the retina seen on histology.6 7 12 However, the dimensions of the pseudocolor bands do not display the same ratios as the cell layers in histologic sections. It is therefore extremely difficult to assign the pseudocolor bands in the OCT images to specific anatomic components.
In previous reports the innermost (red-white) band was thought to correlate with the retinal nerve fiber layer (RNFL), whereas the outermost (red-white) band was thought to be some function of the retinal pigment epitheliumchoriocapillaris complex (RPE-CC)6 7 12 (Fig. 1 A). Crude measurements of the separation between the innermost borders of these bands approximate to the average thickness of the neural retina. However, given the pixelated nature of the image, it is difficult to determine precisely the location of the borders of these and the other pseudocolored bands.
|
In view of the number of recently published papers using OCT to measure changes in retinal dimensions as a function of disease,7 8 it is surprising that only one empiric attempt has been made to relate pseudocolor banding to structural components.14 In this study the histologic images were manipulated in an attempt to superimpose one on the other, using the inner limiting membrane as an inner reference location. The outer reference chosen was the RPE. An assumption was made that this layer correlated with the superficial border of the outer band. A more direct correlation could be obtained in two ways: either by empirically changing retinal dimensions in a controlled fashion or by studying pathologic conditions in which a highly visible marker cell is known to have changed its location within the retina. The former could be performed using the excimer laser, which has the ability to remove large areas of tissue with precise control of ablation depth. If the ablation rate of retina is similar to that of cornea, then at a radiant exposure of 180 mJ/cm2 each laser pulse would remove approximately 0.25 µm of tissue.15 The latter experimental approach could be investigated by examining conditions in which RPE cells migrate into known regions16 17 of the neural retina. The present article describes an experimental investigation of OCT images in which both approaches have been taken.
Methods
In the first part of this study we observed changes in OCT images of cadaveric human and bovine retina in vitro after dissection and controlled removal of superficial retinal layers. A pilot study using human cadaveric retinal tissue had already shown the lamination pattern and dimensions of OCT images of isolated retina to be the same as human retina in vivo (Chauhan DS, Marshall J, unpublished data). For experimental studies, bovine tissue was chosen because it was readily available and had a number of features of its vitreoretinal architecture similar to those of human retina. In addition, bovine retina has tapetal and nontapetal regions that differ in thickness and histologic structure, tapetal retina being thicker. The tapetum separates the CC from the medium and large vessels of the choroid, and there is no melanin pigment within the RPE of this region.18 All studies were performed in accordance with the ARVO Statement for the Use of Animals in Ophthalmic and Vision Research and the Declaration of Helsinki.
The second part of the study was based on the observation that, for light passing through the retina, the site of greatest change in optical properties is the layer of melanin within the RPE. We therefore examined lesions in which abnormal changes in the distribution of pigmented cells were apparent. We chose retinitis pigmentosa (RP) as a condition in which pigmented RPE cells migrate forward into the inner layers of the neural retina and panretinal photocoagulation as a model for the redistribution of RPE pigmentation within the plane of the RPE. These studies were in accordance with the Declaration of Helsinki recommendations.
OCT Imaging.
Images were acquired using an optical coherence tomography scanner
(Humphrey Instruments), which directed a beam of light in a horizontal
plane through the pupil. Each composite OCT image or scan consisted of
a linear array of 100 juxtaposed individual z-axis scans,
each with a depth of 2 mm. The thickness of each individual
z-axis scan was 13 µm, and the separation between them
increased with the field of the image.
The length of the OCT scan line in the xy plane could be varied, ranging from 2.5 to 19 mm in air at a working distance of 43 mm. This represented a scanning angle of between 3° and 25°. For a flat specimen, this results in an overestimate of thickness by a factor of 1.00 to 1.024, respectively, at the lateral extremes of a scan. For this reason, the shortest scans possible were used in the in vitro studies, even though posterior pole specimens maintained the concavity of the eye. For an emmetropic eye the system-corrected range for scan length in the xy plane of the retina was 1.13 to 9.08 mm. The angular orientation of the scan line could also be controlled within the xy plane of the tissue examined. The device had an internal fixation light visible to the patient but not on the operators viewing screen. A second light, generated by a HeNe laser, was seen on the viewing screen and could be moved under the control of the operator to fall on any given retinal feature. The spatial separation between this light and the ends of the scan line was stored by the computer for each scan orientation. Thus, on repeat visits, if the operator relocated the HeNe light over the original retinal feature, subsequent scans would always be at the same location and orientation. The acquisition time for each composite OCT image was 0.9 seconds, independent of scan length and orientation. The images produced were dependent on the optical properties of tissues in the z-plane and underwent processing by the commercial software into a logarithmic pseudocolor scale in which white, red, yellow, green, blue, and black represented the range of signal intensity from high to low. All scans obtained with this instrument were accompanied by a digital image of the tissue recorded with the scan line superimposed.
The power incident on the eye was constant throughout (750 µW). Imaging began with crude focusing of the image on the operators viewing screen. After this, the continuously scanning real-time OCT image was centered on the computer screen and the focusing knob adjusted by small increments until the image was at its brightest throughout the tissue. The polarization setting was then varied for each image set to optimize the signal intensity for the inner band, because the outer band varied little with polarization. This was done to maximize the contrast in signal intensity of the inner part of the image and therefore the clarity of borders. Finally, the focusing knob was readjusted, if necessary, according to the criteria described.
In vivo examinations were performed after pupillary dilation with 1% tropicamide and 10% phenylephrine. To avoid degradation of images, only scans with minimal eye movement were recorded for later analysis. Internal fixation was used as far as possible, but external fixation was occasionally necessary to scan peripheral lesions.19 Subjects were asked to blink between acquired images to minimize corneal drying, which may have caused both discomfort for the subject and a degradation of the OCT image. Because the measurements of interest in this study were all in the z-plane and taken from several individual z-axis scans in each OCT image, z-axis alignment software was not used.
In Vitro Studies
OCT Imaging.
For the purposes of in vitro examination of tissues, the OCT scanner
was modified by placing a plane mirror above and at 45° to the
objective lens, thus allowing the acquisition of images of horizontally
oriented tissues. Extensive validation studies demonstrated that the
images acquired with this modification were identical in pattern and
dimensions to those without it. When comparing the measurement of the
thickness of glass slides and signals from glassair interfaces with
and without the plane mirror, the coefficients of variation of
measurements were 0.04 and 0.03, respectively. The correlation between
measurements made with and without the mirror was very strong
(R2 = 0.97). To avoid the inherent
inversion of vertical images obtained from such an optical system,
scans were taken in the transverse horizontal axis.
Specimens.
Fresh bovine eyes were obtained from an abattoir, in accordance with
European Union statute 716/96.20
Four cadaveric human eyes
were obtained from the National Transplant Service (Bristol, UK) with a
mean postmortem time of 50 hours. Retinal tissue was examined from two
further human eyes immediately after their enucleation for choroidal
malignant melanoma. The latter and the bovine eyes were intact, whereas
the cadaveric eyes from the eye bank were supplied after corneal
excision.
Globes were hemisected by a circumferential penetrating incision 5 mm posterior to the sulcus sclerae. With the anterior globe facing downward, the posterior globe was then separated from this and the vitreous by gently lifting and rotating the optic nerve.21 In bovine eyes full-thickness specimens were obtained from both tapetal and nontapetal regions using an 8-mm trephine. Similarly, specimens were trephined from human eyes to include attached retina from the posterior pole in cadaveric eyes and from the quadrant opposite the choroidal melanoma in the tumor-affected eyes.
OCT Image Acquisition.
For OCT imaging, these full-thickness specimens were placed in a petri
dish and covered to a depth of 5 mm with fresh normal saline. Using the
modified scanner, multiple contiguous scans (1.7 mm) were performed
across the entire diameter of each sample, using the HeNe relocation
feature to position scans end to end. All scans on each specimen were
performed at the same focal plane using the same power and polarization
settings and repeated three times. All scans were acquired by a single
observer in these studies except when stated otherwise.
Specimens were then fixed in 2.5% glutaraldehyde buffered in 0.1 M sodium cacodylate containing 1 mg/ml calcium chloride with a final pH of 7.4. In this solution, the areas scanned before fixation were then rescanned at the same focal plane and using the same settings at 1, 15, 35, and 60 minutes and 24 hours after fixation.
To determine the dependence of measurements derived from OCT images on the system settings for polarization, the results of 10 independent observers were obtained. Using a single fixed specimen of human extramacular retina, each of the observers was asked to obtain the "best" image by altering the polarization setting. All other control settings remained constant. Observers were masked to each other, and the whole image sequence was performed within a 1-hour period. Only two of these observers had any previous experience in OCT image acquisition.
Excimer Laser Ablation.
Each full-thickness trephined specimen was removed from the
glutaraldehyde solution and placed on a fixative-soaked piece of filter
paper in a covered petri dish. The petri dish was then placed on a
laboratory jack under the exit port of an excimer laser (Apex Plus,
Summit Technology, Boston, MA). The petri dish was manipulated so that
the specimen was centered using the HeNe laser-aiming beams. After the
petri dish was uncovered, surface fluid was removed using the capillary
action of a Wechsel sponge, and the HeNe beams were then focused on the
surface of the tissue. With a radiant exposure of 180
mJ/cm2 per pulse and the program designed for
phototherapeutic keratectomy, concentric ablations were performed.
The initial ablation zone diameter used was 6.5 mm, and this was reduced by 0.5 mm for each successive ablation to a minimum diameter of 1 mm. This resulted in 11 steps between zones. Fifty pulses were applied per ablation zone for both the bovine and human specimens. A typical procedure took between 12 and 20 minutes and was limited by the reprogramming time for each successive ablation zone. The lid of the petri dish was removed only during the periods of excimer ablation in an attempt to minimize drying of the tissues. Immediately after the completion of the ablation process, the specimens were reimmersed in glutaraldehyde solution. The contoured sample was then processed for light and scanning electron microscopy.
Measurements on Microscopy.
Retinal features were measured by two independent observers using a
light microscope with a calibrated measuring eyepiece. In all specimens
the baseline thicknesses of individual layers of choroid and retina
were measured in an area that had not been ablated. In addition, the
total remaining thickness of retina in each ablation zone was measured,
as was the height of each step.
OCT Image Analysis.
The pseudocolored images produced were presented as raw images in the
first instance and were then postprocessed using the manufacturers
programs. The raw images and the normalized, median-smoothed, and
gaussian-smoothed images were all converted to tagged information file
format (TIFF) using a screen-capture program (JasCapture Version
Shareware 2.0; JASC, Eden Prairie, MN). These files were then imported
into an image-processing program (Photoshop, ver. 4.0, Adobe, San Jose,
CA) and magnified without altering the image pixel dimensions. The
borders of the inner and outer bands of high signal were defined as the
point at which the signals false color changed from white or red to a
color indicating lower signal intensity.
To determine the variation in measurements between the eight independent observers, four image parameters were measured on the unablated specimens. These were the distance between the inner borders of the inner and outer bands, the overall thickness of the inner band, the overall thickness of the outer band and, finally, the distance between the outer borders of the inner and outer bands (Fig. 1A) . All measurements were repeated eight times by a single experienced image analyzer for each scan.
For the ablated specimens, measurements of the thickness of the colored bands were made using the measuring facility of the image-processing program. The exact location of the borders was determined by the operator, as described. The overall thickness of the unablated retina and the overall thickness at each ablation zone were also measured. The former was taken to be the distance between the superficial borders of the inner and outer bands in nonablated tissue.12 14 These measurements were converted to micrometers using a manufacturers conversion factor of 4 µm per z-axis pixel (Humphrey Instruments). The thickness of the remaining retina in each ablation zone was expressed in relation to the thickness of the unablated retina by subtracting the sum of the heights of the steps peripheral to the zone in question. The height of each step was measured as the distance between the extrapolated superficial borders of the inner bands. The same measurements were also attempted using the pixelated calipers available on the OCT scanners analysis software. However, such measurements were abandoned because of the technical limitations of the system.
In Vivo Studies
Subjects.
The 10 independent observers each imaged the same area of extramacular
retina of a single normal subject. They again altered the polarization
setting only to obtain the "best" image.
For a single-observer study, 10 subjects with normal retinas were then selected. Each was less than 30 years of age, had uncorrected visual acuity of 6/6 or better, and had no significant medical or ophthalmic history. In this group a series of radial 3-mm linear scans was obtained, each centered on the fovea. For each of these subjects, these scans were optimized to provide the maximal signal intensity to the inner band using all available control settings. After these images were acquired for each subject, a further scan was undertaken with the polarization adjusted to provide the minimum inner band signal intensity. All other control settings were maintained in a constant position.
Four patients with RP were examined, two with Ushers syndrome (US; 42 and 73 years old) and two with autosomal dominant (AD) RP (47 and 62 years old). Four patients with panretinal laser photocoagulation for vascular disease were also examined. All had lesions of at least 3 months duration with a hyperpigmented center and pale rim. All lesions were produced with an argon laser (Novus 2000, Coherent, Palo Alto, CA) with a spot size of 200 µm, pulse duration of 0.1 second, and power of 170 to 300 mW.
OCT Image Acquisition.
Images were optimized for each scan to produce the highest intensity
and definition to the inner and outer bands by manipulating the
polarization control.
Each control subject had eight 3-mm scans taken, centered on the foveola and at 0°, 30°, 45°, 60°, 90°, 120°, 135°, and 150°. The patients with RP had scans taken in two configurations. The first was a short (1.13.0-mm) linear scan taken through the center of one or more bone spicule lesions with normally pigmented intervening retina. The second was a longer linear scan spanning across the macula and the surrounding area of hypopigmentation when present. Photocoagulation lesions had short linear scans performed passing through their centers. In each case the scans were extended to include adjacent retina of normal appearance.
OCT Image Analysis.
In the study, to derive the interobserver error, the same parameters
were measured on the in vivo images as for the in vitro specimens. In
the single-observer study of 10 normal subjects, these parameters were
measured both in the center of the foveola and the edge of the foveal
pit for the images with the polarization maximized and minimized.
There were two aspects to the analysis of OCT images of 54 areas of bone spicule pigmentation. First, any discrete areas of higher signal within the inner band were identified and their locations in the xy plane related to those of bone spicule pigmentation. The depth of each lesion was measured on OCT as the distance between the superficial border of the inner band and the center of the discrete high signal. Second, discontinuities in the outer band were identified and measured in the xy plane of the image. The location of each discontinuity was also related to that of the bone spicule pigmentation. In the scans passing through the macula and surrounding hypopigmented retina, the thickness of the whole retina and outer band in each region was also measured.
For the OCT scans of laser lesions, a number of parameters were measured in both the vertical (z) and the horizontal (xy) axes. The z-axis measurements were made using the same measuring facility described earlier. They were undertaken at the center of the lesions and on adjacent tissue that appeared normal on both sides of the lesion. The depression of the center of the lesion relative to the surrounding retina was also measured. The xy plane measurements were made after vertical borders had been superimposed on the images (Fig. 1B) . These borders were defined by the sharp xy-axis transition in the signal intensity deep to the outer band. Pixel values in the xy plane were converted to micrometers by correcting for the given length of scan line, which was taken as that on the OCT display. Because the maximum refractive error of the subjects was 2.5 D of myopia, any discrepancy between real and actual scan length was small.
Results
In Vitro Studies
The bovine eyes were well preserved with the retina attached. The
cadaveric human eyes often had retinas that had partially detached
after death. Small serous detachments were also noted in the freshly
enucleated human eyes adjacent to the melanoma.
Glutaraldehyde fixation caused no significant change in either the pattern or measured thickness of the individual OCT bands or the total retinal thickness. The signal intensity of each band increased, however, and reached a maximum within 35 minutes of fixation.
Signal intensity and dimensions in OCT images were found to vary with control settings. Using a constant power and an optimally focused image, manipulation of the polarization control resulted predominantly in variations in the inner band. The setting of the polarization control varied among individual observers in their attempts to obtain an optimized image. Statistical parameters of the variation among observers are shown in Table 1 , in which altering polarization settings can be seen to result in a coefficient of variation in measurements of up to 0.30.
|
In all human and bovine eyes, the location and thickness of the inner band of high OCT signal corresponded, on light microscopy, to the sum of RNFL, ganglion cell layer (GCL), inner plexiform layer (IPL), and part of the inner nuclear layer (INL). The mean thickness of the inner band was 7.3 times that of the RNFL (Fig. 2) . In 25% of the eyes, the outer plexiform layer (OPL) was included within this band. Similarly, the outer band of high signal corresponded to the RPE, CC, and tapetum in tapetal retina. However, in bovine nontapetal and human retina this band corresponded to the RPE, CC, and just under half of the choroid. Here, the mean thickness of the outer band was 2.6 times that thickness of the RPE-CC complex. The thickness of the dark band of low OCT signal, between the inner and outer bands, was greater than that of the inner and outer segments of the photoreceptors in tapetal retina but less in bovine nontapetal and human retina.
|
OCT Signal Changes.
With the progressive ablation of tissue, changes in the location and
thicknesses of one or both of the major OCT bands were observed in all
specimens.
Inner Band of High OCT Signal.
The number of pulses required to remove a depth of retina corresponding
to the inner band measured in unablated tissue varied from one specimen
to another but was typically between 150 and 250. This corresponded to
between 74 and 123 µm. In practice, it was not possible to remove the
inner band by ablation, because it migrated down through the tissue
with progressive laser pulses. Although the band was always present,
its width decreased linearly with ablation until the OPL had been
reached. After this point the band remained roughly constant with a
mean thickness of 32 µm (95% CI, 2836 µm) but moved deeper,
related to the tissue surface, with subsequent ablation pulses (Fig. 2) .
Outer Band of High OCT Signal.
The outer band maintained its location and thickness after ablation of
retina in all specimens except bovine tapetal retina. In the latter,
the thickness of this band increased with progressive ablation of
overlying retinal tissue. In both human and bovine specimens the inner
border of the outer band remained within 28 µm of the location of the
apical RPE, as defined by light microscopy (Fig. 2)
.
Dark Band of Low OCT Signal.
The thickness of this band was stable in all specimens until the
ablation process reached the level corresponding to the ONL on light
microscopy (Fig. 2)
. After this, a linear reduction in the thickness of
this band was observed.
In Vivo Studies
The reproducibility of OCT measurements of total retinal thickness
among the 10 observers was high (coefficient of variance, 0.05). It was
lower (coefficient of variance, 0.18) for the distance between the
outer borders of the major bands (Fig. 1A)
.
In most subjects the inner band was absent at the foveola and occupied half the total retinal thickness at the edge of the foveal pit (Table 1) . The mean total retinal thickness at the foveola was 159 µm (95% CI, 152166 µm), and that at the edge of the foveal pit was 258 µm (95% CI, 253263 µm). The respective outer band thicknesses were 81 µm (95% CI, 7587 µm) and 71 µm (95% CI, 6775 µm) and differed significantly (P = 0.05).
At the extreme settings of polarization, the overall retinal thickness and the thickness of the outer band remained constant. By contrast, there was a highly significant change in the thickness of the inner band, which varied from a mean thickness of 11 µm (95% CI, -123 µm) at the minimum setting to 128 µm (95% CI, 114142 µm) at the maximum. The reproducibility of these measurements was very low, with a coefficient of variance of 0.97.
RP.
In scans of patients with RP, although inner and outer bands could be
readily identified, the inner band was less intense. Fifty-four areas
of bone spicule hyperpigmentation were imaged in the four patients. All
scans passing through these lesions had discrete regions of very high
signal coincident with the bone spicules, and deep to the innermost
border of the inner band (Fig. 3
A). There was also a discrete reduction in signal intensity of the outer
band under such high-signal areas. Three of the patients had
high-signal regions imaged with a median depth of 48 µm, and the
other patient had similar regions at a median depth of 116 µm. The
patient with the deeper signals was 47 years old with AD RP. Two
patients (one with US and one with AD RP) had a central macular area of
normal appearance with a surrounding area of hypopigmented retina.
Scans spanning these areas had three distinctive features coincident
with the hypopigmented retina (Fig. 3B)
. First, the inner band of the
OCT image of the hypopigmented retina had an increased signal. Second,
there was a reduced distance between the superficial borders of the
inner and outer bands (i.e., total retinal thickness). Finally, the
outer band was much thicker than that of the central macula.
|
Discussion
The optical coherence tomographic imaging of tissues is dependent on their optical properties. In any illuminated tissue the only structures to generate high signal are those that return light close to or along the axis of the incident light. Tissue components that absorb light or direct it away from the OCT detector have correspondingly low signal intensities. Thus, the intensity of a recorded signal is related to the proportion of low-coherence light returned (reflectance) to the detector.
The optical phenomena involved in the generation of OCT signals in tissues are likely to be reflection, scatter, and birefringence.9 10 Theoretical and empiric studies on multilayered model systems have demonstrated that the change in refractive index at an interface is responsible for most of the signal, the next most important factor being scatter.22 In these model studies, refractive index changes gave rise to specular reflection and sharp signal peaks. Scattering properties contributed to the band thickness. It was found that the more a medium scattered light, the more discrete were its deep borders, and therefore, the thinner it appeared. This may have been caused either by the loss of coherence with multiple scattering of light,23 resulting in the detection of low interferometric signal, or by absorption and scattering of light so that less was returned to the OCT detector. Overall, therefore, the z-axis resolution of OCT within poorly scattering tissues may be limited. In the retina, change of refractive index may be responsible for the inner aspect of the inner band, and scattering could define its outer border.
On entering the OCT detector system, light that was returned from the tissue being examined was allowed to interfere with that from the reference arm. The resultant intensity of this interference was recorded at 500 points along an individual z-axis scan and was represented by a logarithmic pseudocolor scale. The arbitrary units of the signal intensity itself had a range of 0 to 1600, whereas the pseudocolor scale consists of 16 colors. This degradation of signal resolution is in contrast with other imaging systems in which attempts at improving image resolution have involved extending gray scale. The result of the pseudocolor scale is likely to be a grouping of a wide range of higher signal intensities into single-color bands of red or white. Significant variations in signal intensity within the inner and outer bands of retinal OCT images thus may not be displayed, with a consequent loss of spatial resolution in the z-axis. When imaging retina, the effect may be to compress a 13-layered system13 into an image with just three or four bands.
Postprocessing of OCT images may also limit resolution. The z-plane is always displayed in the vertical axis, and each pixel has a depth measurement of 4 µm. The spatial resolution of OCT images in the xy plane is dependent on both the optical limitations of the ocular media and the design constraints of the instrument. These limitations are apparent when attempting to image over a wide xy field. Each OCT image is constructed from 100 equally spaced individual z-axis scans, the xy dimensions of each being constant, irrespective of the area of retina scanned. Thus, the greater the xy field imaged, the greater the spacing between the individual z-axis scans, ranging from 10 to 110 µm, and the lower the resolution in the xy plane.12 Thus, for example, when the longest scan is used, the foveola may be missed, because it is only 150 µm across. It is therefore evident that, in all three dimensions, the pixelation of images gives rise to both a mismatch between theoretical and practical resolution and indistinct borders on images of tissues.
The appearance of a recorded image is also dependent on the preference of individual operators. Although the focusing of an image is readily and accurately repeatable and the power setting can be standardized, polarization is a variable that cannot be quantified on current scanners, and settings vary among observers. In this study, by examining raw images, we have shown polarization to have a significant effect on the measurement of some parameters, both among observers and observations by a single operator. The thickness of the inner band, for example, varied by up to 30% among observers and had a maximum of 18% for a single observer. The variances of inner and outer band thickness measurements were least forimages that had undergone gaussian smoothing. For measurements of the total retinal thickness, however, there was no significant difference among raw, normalized, median-smoothed, and gaussian-smoothed images.
Given the difficulties in resolving the various components of the OCT image, it is surprising that only one other study has been undertaken to correlate OCT images with histology. Toth et al.14 assumed that the innermost aspect of the inner high-signal band corresponded to the inner limiting membrane and that the innermost aspect of the outer band corresponded to the apical surface of the RPE. They then compressed the histologic images, by between 4% and 12%, so that the distance between these two components matched those of their OCT scans. They reported that the outer aspect of the inner band was coincident with the outer aspect of the RNFL. Thus their images had four high-signal bands that corresponded to the RNFL, IPL, the residual element of the fiber layer of Henle, which they erroneously described as the OPL, and the RPE.
In all the high-signal areas identified by Toth et al.14 subcellular components are present, with dimensions between 1 and 3 µm, which are potentially high scatterers of light. In the low-signal areas the nuclei have dimensions between 4 and 9 µm.14 Although their experimental findings seem to indicate a direct correlation between OCT strata and retinal cellular components, the match is only superficial. For the bands said to correspond to the RNFL and OPL, the ranges of thickness measurements were 3.3 and 7.9 times greater on OCT than light microscopy for the published images. Thus, if, for example, the furthest excursion of the inner band were taken as the true extent of this layer, it would include both the RNFL and 80% of the GCL.
Although the previous study did not state that image postprocessing programs had been used, the images would suggest that they had. It is not clear which program was used, however, and we have used the manufacturers gaussian smoothing program in our study, because this resulted in the lesser variance in measurements (coefficient of variance, 0.08; cf. 0.12 for median smoothing).
Although Toth et al.14 concluded that there was a close correlation between the OCT bands and specific retinal layers, such as the RNFL, our findings failed to support this completely. In particular, the persistence of an inner band after the deliberate destruction of inner retinal layers contradicts the notion of tissue-specific signal.
Toth et al.14 assumed that the inner border of the outer band corresponded to the apical RPE, and their histologic images were manipulated to generate a match in total retinal thickness. In our study, no image manipulation was required to generate matches in retinal thickness, because they could be directly correlated with ablated samples. Direct evidence was determined, locating the inner border of the outer band at the level of the RPE. Bone spicules are formed in RP by the migration and aggregation of pigmented RPE cells surrounding retinal capillaries. Although their precise location cannot be determined on biomicroscopy for direct comparison with OCT, their abnormal retinal distribution is well established.17 24 The high signal corresponding to the location of these pigmented RPE cells was strongly suggestive of a role for this cell in OCT signal generation. In contrast to this situation of high melanin concentration, the RPE cells at the pale rim of a retinal laser lesion differ from RPE cells in normal nearby retina, in that they have little or no melanin within them. The corresponding loss of signal intensity in the outer band therefore demonstrates that it was not RPE cells that were responsible for the generation of signal but their subcellular components such as melanin and lipofuscin. It follows that the innermost aspect of the outer band is likely to correspond to the apical region of the RPE layer, because it is within this 3-µm layer that melanin granules are concentrated. This was confirmed in our in vitro study by the observation that the location of the inner border of the outer band remained constant and close to the RPE, even with progressive ablation.
Another significant optical effect of melanin seen in these studies was an attenuation of signal from deeper tissues. The most notable examples were seen with bone spicules when the signal from deeper tissues was almost absent. Further, the apparent "unveiling" of signal from tissues deep to focal regions of RPE hypopigmentation is also consistent with a downstream signal-attenuating role for melanin. The candidate optical phenomena responsible for the duality of high signal generation and downstream signal attenuation are absorption, reflection, and scatter. Absorption is unlikely to account for the downstream signal attenuation, because only 3% of incident light at a wavelength of 830 nm is absorbed by RPE melanin.25 Similarly, the degree of attenuation seen deep to melanin cannot be caused by reflection of light by melanin, because a maximum of only 4% of light is reflected by polished interfaces in optical instruments.26 Scatter is the only optical phenomenon that can be responsible for both observations, and melanin is already known to be a strong scatterer of light.27 The role of scatter in the attenuation of downstream signal is likely to have two components, each resulting in a significant reduction in the light returned to the detector. The first is the deviation of light away from deeper tissues so that less is incident on them. Second, any light that reaches deeper tissues and is returned toward the detector may be further scattered by the melanin and therefore deviated away from the detector.
The clinical significance of this phenomenon is that diseases affecting tissues deep to the RPE may not be directly demonstrable with OCT. Changes relating to choroidal neovascularization have been described elsewhere and are limited to a description of the thickness and irregularity of the outer band.28 Changes of this nature were observed in our study of laser lesions and, although the effect of RPE pigmentation was noted in the previous study, the possible influence of this on their findings was not discussed.29
The influence of tissues on the signal from deeper tissues is not exclusive to melanin and was seen with the progressive ablation of tapetal retina. In this case, the thickness of the outer band increased linearly with progressive ablation of overlying retinal tissue, presumably because of a loss of attenuation from those tissues. That this was not significant with nontapetal or human retina until after ablation of the RPE may be accounted for by the attenuation effect of their RPE melanin, in contrast with the nonpigmented RPE of tapetal retina.
In contrast with the discreteness of the inner border of the inner band and both borders of the outer band, the deep border of the inner band was more diffuse and less easy to identify. One explanation of this phenomenon is that there may be less difference in the optical properties of adjacent layers of tissue. Alternatively, this may relate to the presence of significant noise or a trailing edge effect of a high signal, giving a greater apparent thickness to each band. This would also have the effect of masking narrow intervening bands of low signal. Indeed, the band of low signal corresponding to the GCL seen in Toth et al.14 was not identified in our studies. This is likely to be a consequence of our use of extramacular tissue, which by definition has a GCL of two or fewer nuclei in thickness (1020 µm), compared with five or more at the macula (6080 µm).17 Even in the absence of noise, the resolution of OCT is not sufficient to demonstrate such a layer outside the macula, with the result that the inner band in our study corresponded to at least the RNFL, GCL, and IPL in all cases.
This phenomenon may also have accounted for some of the findings of Schuman et al.7 19 In their studies, it was assumed that the inner band surrounding the optic disc corresponded to the retinal nerve fiber layer. They found that this band was thickest in the superior quadrant and thinnest nasally. The thickness of the inner band temporally was found to be between 45% and 69% of that of the inner band superiorly. This differs markedly from the findings of Varma et al.30 in a study of cadaveric normal human eyes. Although the ranking of RNFL thickness by quadrant was the same, the temporal RNFL was 78% of the thickness of the superior RNFL. This difference may have been caused either by an overestimate of the thickness of the RNFL superiorly or by an underestimate of its thickness temporally in the OCT study. It is known that the GCL in all quadrants around the disc is one nucleus thick, except in the temporal quadrant, where it is three or more nuclei thick.13 Thus, it is likely that the inner band in the superior, inferior, and nasal quadrants incorporated the GCL and IPL signal bands and that measurements of its thickness consequently exceeded true RNFL thickness. The inner band temporal to the disc may, however, have approximated to the true RNFL thickness, because it may have been resolved from the high-signal IPL band by virtue of a sufficiently thick intervening low-signal GCL band. Because the early loss of ganglion cells and RNFL in glaucoma occurs in the central macula,30 there may be corresponding OCT changes. It has already been demonstrated that total retinal thickness may be reduced in glaucoma31 and this may be seen on OCT. In addition, the innermost band may also change in thickness with a pattern dependent on the differential rate of loss, if any, of RNFL and GCL thickness. If the RNFL is lost faster, a thinning of the innermost band may be seen initially. However, if the GCL is lost preferentially, a loss of resolution between the RNFL and IPL high-signal bands may follow, resulting in an anomalous increase in the inner band thickness.
In our studies we found the reproducibility of inner band thickness measurements to be 15% between observers and 30% between images of a single subject taken on a single occasion. Schuman et al.19 found the reproducibility of mean RNFL thickness measurements using OCT to be lower between visits than on a given visit. In our study, the most significant factor affecting reproducibility was the variation in polarization settings. This could also have been the reason for the findings of Schuman et al., because the polarization setting could not be recorded and therefore would have varied between visits.
In the present study, the change in the thickness of the inner band with progressive ablation had two phases. The first was linear, where the thickness of the inner band was reduced by the same amount as the ablation step height. After this, the inner band maintained an almost constant thickness and migrated downward into the retina until it became indistinguishable from the outer band. The first phase was consistent with direct relation of OCT signal to specific tissues, because removal of tissue resulted in removal of high signal. The second phase, however, was not consistent with a tissue-specific origin of OCT signal. It may be explained by the abrupt change in refractive index and scattering properties at the primary optical interface of the tissue in the path of the OCT beam. The submersion of samples in glutaraldehyde, which has a refractive index close to that of vitreous, reproduced the optical conditions present in vivo. The primary interface component must therefore also exist in the images of retina in vivo, but its contribution to the intensity and measured thickness of the inner band is uncertain, because its interaction with signal from inner retinal tissues is also unknown. It is likely to be of greatest significance when the inner band is thin. These results would certainly explain the observation that an inner band can often be seen at the fovea and clivus in vivo, when there is a thick GCL but no RNFL or IPL.
Because the two most consistent generators of high OCT signal have been shown to be the surface of the retina and the RPE, the precise and accurate measurement of total retinal thickness is readily achieved with OCT.
Overall, although OCT bands may be, in part, related to specific retinal layers, this is only a consequence of their individual and combined optical properties, many of which are poorly understood. Currently, there are also many other limitations to the resolution of OCT, including light source wavelength and bandwidth, image processing, noise, and the influence of overlying tissues. Finally, the role of the operator in altering the polarization has a major influence on measurements acquired. All these factors preclude the direct comparison of measurements of retinal tissue components collated by different centers.
Acknowledgements
The authors thank Timothy J. ffytche, Ann Patmore, Anne Weston, and Christopher Stephenson for their help throughout this study.
Footnotes
Supported by a grant from the Lady Anne Allerton Fund and forms part of a contribution toward a thesis for a Doctor of Medicine degree (DSC) at the University of London.
Submitted for publication October 16, 1998; revised March 3, 1999; accepted April 12, 1999.
Proprietary interest category: N.
Corresponding author: Devinder Singh Chauhan, Department of Ophthalmology, Rayne Institute, United Medical and Dental Schools of Guys and St. Thomas, St. Thomas Hospital, Lambeth Palace Road, London SE1 7EH, UK.
References
This article has been cited by other articles:
![]() |
U. E. K. Wolf-Schnurrbusch, L. Ceklic, C. K. Brinkmann, M. E. Iliev, M. Frey, S. P. Rothenbuehler, V. Enzmann, and S. Wolf Macular Thickness Measurements in Healthy Eyes Using Six Different Optical Coherence Tomography Instruments Invest. Ophthalmol. Vis. Sci., July 1, 2009; 50(7): 3432 - 3437. [Abstract] [Full Text] [PDF] |
||||
![]() |
M M K Muqit, J C B Gray, G R Marcellino, D B Henson, L B Young, S J Charles, G S Turner, and P E Stanga Fundus autofluorescence and Fourier-domain optical coherence tomography imaging of 10 and 20 millisecond Pascal retinal photocoagulation treatment Br. J. Ophthalmol., April 1, 2009; 93(4): 518 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
A M Tammewar, D-U Bartsch, I Kozak, R Rosen, I A Falkenstein, P Garcia, and W R Freeman Imaging vitreomacular interface abnormalities in the coronal plane by simultaneous combined scanning laser and optical coherence tomography Br. J. Ophthalmol., March 1, 2009; 93(3): 366 - 372. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Abbott, N. A. McBrien, U. Grunert, and M. J. Pianta Relationship of the Optical Coherence Tomography Signal to Underlying Retinal Histology in the Tree Shrew (Tupaia belangeri) Invest. Ophthalmol. Vis. Sci., January 1, 2009; 50(1): 414 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Kelty, J. F. Payne, R. H. Trivedi, J. Kelty, E. M. Bowie, and B. M. Burger Macular Thickness Assessment in Healthy Eyes Based on Ethnicity Using Stratus OCT Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2668 - 2672. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. DeLeon Ortega, L. M. Sakata, B. Kakati, G. McGwin Jr, B. E. Monheit, S. N. Arthur, and C. A. Girkin Effect of Glaucomatous Damage on Repeatability of Confocal Scanning Laser Ophthalmoscope, Scanning Laser Polarimetry, and Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., March 1, 2007; 48(3): 1156 - 1163. [Abstract] [Full Text] [PDF] |
||||
![]() |
K A Goatman A reference standard for the measurement of macular oedema. Br. J. Ophthalmol., September 1, 2006; 90(9): 1197 - 1202. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. DeLeon-Ortega, S. N. Arthur, G. McGwin Jr, A. Xie, B. E. Monheit, and C. A. Girkin Discrimination between Glaucomatous and Nonglaucomatous Eyes Using Quantitative Imaging Devices and Subjective Optic Nerve Head Assessment. Invest. Ophthalmol. Vis. Sci., August 1, 2006; 47(8): 3374 - 3380. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Ugarte, A A Hussain, and J Marshall An experimental study of the elastic properties of the human Bruch's membrane-choroid complex: relevance to ageing Br. J. Ophthalmol., May 1, 2006; 90(5): 621 - 626. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Van de Moere, S S Sandhu, and S J Talks Correlation of optical coherence tomography and fundus fluorescein angiography following photodynamic therapy for choroidal neovascular membranes. Br. J. Ophthalmol., March 1, 2006; 90(3): 304 - 306. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Polito, M. Del Borrello, M. Isola, N. Zemella, and F. Bandello Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography Arch Ophthalmol, October 1, 2005; 123(10): 1330 - 1337. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Bandello, A Polito, M Del Borrello, N Zemella, and M Isola "Light" versus "classic" laser treatment for clinically significant diabetic macular oedema Br. J. Ophthalmol., July 1, 2005; 89(7): 864 - 870. [Abstract] [Full Text] [PDF] |
||||
![]() |
B Sander, M Larsen, L Thrane, J L Hougaard, and T M Jorgensen Enhanced optical coherence tomography imaging by multiple scan averaging Br. J. Ophthalmol., February 1, 2005; 89(2): 207 - 212. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Costa, D. Calucci, M. Skaf, J. A. Cardillo, J. C. Castro, L. A. Melo Jr, M. C. Martins, and P. K. Kaiser Optical Coherence Tomography 3: Automatic Delineation of the Outer Neural Retinal Boundary and Its Influence on Retinal Thickness Measurements Invest. Ophthalmol. Vis. Sci., July 1, 2004; 45(7): 2399 - 2406. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hillenkamp, A. A. Hussain, T. L. Jackson, J. R. Cunningham, and J. Marshall The Influence of Path Length and Matrix Components on Ageing Characteristics of Transport between the Choroid and the Outer Retina Invest. Ophthalmol. Vis. Sci., May 1, 2004; 45(5): 1493 - 1498. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. N. Zacks and M. W. Johnson Transretinal Pigment Migration: An Optical Coherence Tomographic Study Arch Ophthalmol, March 1, 2004; 122(3): 406 - 408. [Full Text] [PDF] |
||||
![]() |
A. Pallikaris, D. R. Williams, and H. Hofer The Reflectance of Single Cones in the Living Human Eye Invest. Ophthalmol. Vis. Sci., October 1, 2003; 44(10): 4580 - 4592. [Abstract] [Full Text] [PDF] |
||||
![]() |
D.-U. G. Bartsch, A. Aurora, N. Rodanant, L. Cheng, and W. R. Freeman Volumetric Analysis of Macular Edema by Scanning Laser Tomography in Immune Recovery Uveitis Arch Ophthalmol, September 1, 2003; 121(9): 1246 - 1251. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Varma, S. Bazzaz, and M. Lai Optical Tomography-Measured Retinal Nerve Fiber Layer Thickness in Normal Latinos Invest. Ophthalmol. Vis. Sci., August 1, 2003; 44(8): 3369 - 3373. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Hougaard, L. Kessel, B. Sander, K. O. Kyvik, T. I. A. Sorensen, and M. Larsen Evaluation of Heredity as a Determinant of Retinal Nerve Fiber Layer Thickness as Measured by Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., July 1, 2003; 44(7): 3011 - 3016. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Gloesmann, B. Hermann, C. Schubert, H. Sattmann, P. K. Ahnelt, and W. Drexler Histologic Correlation of Pig Retina Radial Stratification with Ultrahigh-Resolution Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., April 1, 2003; 44(4): 1696 - 1703. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sanchez-Tocino, A. Alvarez-Vidal, M. J. Maldonado, J. Moreno-Montanes, and A. Garcia-Layana Retinal Thickness Study with Optical Coherence Tomography in Patients with Diabetes Invest. Ophthalmol. Vis. Sci., May 1, 2002; 43(5): 1588 - 1594. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Muscat, S. Parks, E. Kemp, and D. Keating Repeatability and Reproducibility of Macular Thickness Measurements with the Humphrey OCT System Invest. Ophthalmol. Vis. Sci., February 1, 2002; 43(2): 490 - 495. [Abstract] [Full Text] [PDF] |
||||
![]() |
V Tanner, D S Chauhan, T L Jackson, and T H Williamson Optical coherence tomography of the vitreoretinal interface in macular hole formation Br. J. Ophthalmol., September 1, 2001; 85(9): 1092 - 1097. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Massin, E. Vicaut, B. Haouchine, A. Erginay, M. Paques, and A. Gaudric Reproducibility of Retinal Mapping Using Optical Coherence Tomography Arch Ophthalmol, August 1, 2001; 119(8): 1135 - 1142. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L Jones, N. J L Sheen, R. V North, and J. E Morgan The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer Br. J. Ophthalmol., June 1, 2001; 85(6): 673 - 677. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Hettinger, M. d. l. P. Mattozzi, W. R. Myers, M. E. Williams, A. Reeves, R. L. Parsons, R. C. Haskell, D. C. Petersen, R. Wang, and J. I. Medford Optical Coherence Microscopy. A Technology for Rapid, in Vivo, Non-Destructive Visualization of Plants and Plant Cells Plant Physiology, May 1, 2000; 123(1): 3 - 16. [Abstract] [Full Text] |
||||
![]() |
D. S. Chauhan, R. J. Antcliff, P. A. Rai, T. H. Williamson, and J. Marshall Papillofoveal Traction in Macular Hole Formation: The Role of Optical Coherence Tomography Arch Ophthalmol, January 1, 2000; 118(1): 32 - 38. [Abstract] [Full Text] [PDF] |
||||
Read all eLetters
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |