From the Lions Eye Institute, Center for Ophthalmology and Visual Science, University of Western Australia, Nedlands.
| Abstract |
|---|
|
|
|---|
METHODS. A commercial Q-switched Nd:YAG laser was used to produce the fifth harmonic wavelength of 213 nm. Ablation trials were carried out on porcine corneas. Slit ablations of dimensions 0.5 x 2.5 mm were performed using seven values of laser fluence to obtain the most efficient fluence for ablation. The morphology of each ablation was obtained using a computer-automated confocal profiling system. These profiles were then analyzed to determine the ablation depth for the range of fluence values used.
RESULTS. A fluence in the region of 200 mJ/cm2 was found to be the most efficient for ablation. The efficiency in this region was approximately 0.35 mm3/J, and the ablation rate was found to be 0.6 µm/pulse. The ablation threshold was found to occur at a fluence of 50 mJ/cm2. In the region of highest efficiency, the peak varied slightly in the fluence range between 150 and 250 mJ/cm2.
CONCLUSIONS. This study confirms that the corneal ablation properties at 213 nm are comparable with those at the 193-nm excimer laser wavelength. Increased pulse energy was obtained for the fifth harmonic of Nd:YAG lasers at 213 nm through the use of new nonlinear optical crystals to perform the frequency conversion. A solid state laser is feasible to replace the excimer gas laser for performing refractive surgery procedures. For the first time, the increased energy at 213 nm allows large-beam ablations to be performed at this wavelength.
| Introduction |
|---|
|
|
|---|
The Nd:YAG laser has emerged as a possible candidate to replace the excimer gas laser. The fundamental wavelength of this laser is 1064 nm, and three nonlinear optical crystals can be used to produce the fifth harmonic wavelength of 213 nm. As a result, a number of different configurations have been proposed using various combinations of nonlinear crystals. Configurations incorporating the two crystals deuterated cesium dihydrogen arsenate (CD*A) and ß-barium borate (BBO)4 are subject to very low percentage conversion efficiencies.5 6 This has limited the available energy at 213 nm, and, consequently, the laser beam has had to be focused on a small spot to achieve the fluences required for ablation. With the invention of the nonlinear crystal cesium lithium borate (CLBO)7 the overall conversion efficiency of this process has been greatly improved, and the amount of available energy at 213 nm has been increased.8 This increase in energy has enabled the development of a solid state UV laser with the potential to perform ablations using a large beam diameter.
Previous studies using solid state lasers investigating the similarity in ablation properties between the 193-nm excimer wavelength and the 213-nm wavelength of solid state lasers have been constrained to use of low-output pulse energies.5 6 9 10 11 These studies were restricted to using small beam diameters of 0.5 mm or less to achieve a fluence above the ablation threshold.
We present a quantitative analysis of the ablation rate using a solid state laser with increased pulse energy. From the ablation rate data, ablation efficiency is calculated, and a determination of the most efficient fluence for tissue removal is made.
| Materials and Methods |
|---|
|
|
|---|
Ablations were carried out on porcine corneas obtained from an abattoir. All eyes were used within 5 hours of enucleation and stored on ice until required for ablation. Immediately before laser irradiation the epithelium was débrided and the eye vacuum stabilized onto a clamped syringe. The eye was positioned less than 5 mm behind a 0.5 x 2.5-mm slit onto which the collimated laser beam was centered. Therefore, the slit acted as an aperture for the beam rather than using focusing to produce the required beam diameter. The dimensions of the slit were chosen to produce a rectangular ablation profile to simplify the task of measuring the depth. The pulse energy through the 1.25-mm2 slit was measured before and after each ablation using a power meter (model 1825-C; Newport, Santa Ana, CA) with a detector (model 818T-10; Newport). A small fan was used for each ablation to disperse the ablation plume to minimize any absorbing or blocking effect this would have on the impinging laser radiation.12 13
For each ablation trial, the fluence was held constant, and a number of eyes were ablated with a varying number of pulses. This allowed the ablation rate, defined as the etch depth per pulse to be obtained for a wide range of fluence values. This procedure was repeated for fluence values of 80, 120, 150, 180, 210, and 250 mJ/cm2. The number of pulses was varied between 100 and 400. A separate trial was undertaken to determine the fluence value for the threshold of ablation. The method used for this trial was to irradiate eyes with low values of fluence and a large number of pulses until a notable etching effect was observed from a profiling system (described later).
In an effort to eliminate potential depth measurement errors associated with histologic methods,14 15 the ablated eyes were analyzed immediately after ablation. Surface profiles of the ablated corneas across the 0.5-mm cut were measured using a laboratory custom-made, computer-automated confocal profiling system. The design of the profiling system is similar to a confocal microscope and has a depth resolution of 10 µm. Although the resolution is greater than the submicrometer ablation of each pulse, it is suitable for measuring the depth of an ablation resulting from a large number of pulses. Also, a resolution of this magnitude provided a convenient working distance and facilitated surface detection. Calibration of the profiler was achieved using an interferometric method. This involved using the actuator that was used to oscillate the microscope objective to oscillate a plane mirror in one arm of an interferometer. This allowed accurate calibration of the actuator position. Immediately after irradiation, the eye was placed in the profiler while affixed to the syringe to maintain intraocular pressure. The eye and syringe were placed on an x y translation stage (Newport) and positioned in the beam of the profiler so that it was at the latitudinal center of the cut. The microscope objective used to focus the beam onto the surface of the cornea was oscillated with a frequency of 25 Hz. This enabled the depth measurement at each position across the cut to be averaged over many measurements to minimize error. The profile data were then analyzed to obtain the depth of ablation.
The corneal profile data obtained from the confocal microscope were separated into two data series. The first represented the unablated corneal profile. The second series represented the ablated segment of the cornea. Additional data series were then derived and used in the calculation of ablation depth.
Figure 1 shows one such profile. To obtain a meaningful measurement of the ablation depth, the need arose to extrapolate the unablated curvature of the corneal surface. In this way, the depth could be calculated from the surface extrapolation rather than from a straight line across each edge of the cut. This extrapolated curve was determined by applying a second-order polynomial fit to the corneal surface data points. To calculate the depth of ablation, another two polynomials were used. Each of these polynomials was parallel to the fitted curve of the unablated corneal surface. The first was drawn through the deepest point of the ablated region. The second is termed an acquisition line and was heuristically set at 75% of the distance between the surface fit and the deepest point of the ablation. The acquisition line determines which data points from the ablated region are included or acquired for calculation of the cut depth. Included points are therefore those located between the deepest point and the acquisition lines. The magnitude of the depth of each included point was taken as the vertical distance from the corneal surface fit. Those data points that are included are then averaged and an ablation depth obtained. A line displaying the averaged ablation depth is located in the central region of the cut in Figure 1 .
|
| Results |
|---|
|
|
|---|
Figure 2 shows the relationship of ablation rate versus fluence. For the range of fluence values used in the study, this relationship appears to be almost linear. The threshold of ablation was found to occur for a fluence of 50 mJ/cm2. This fluence produced a cut depth of 25 µm after exposure to 1200 pulses corresponding to an ablation rate of 0.02 µm/pulse. The observed cut depth at this fluence is approximately double the resolution of the confocal profiling system. In the fluence range between 150 and 250 mJ/cm2 the rate increases to 0.45 and 0.8 µm/pulse. From these results, the ablation efficiency was calculated for each fluence.
|
|
| Discussion |
|---|
|
|
|---|
An early study by Ren et al.5 using a solid state laser at 213 nm used a larger fluence (1.2 J/cm2) than was used in this study. The ablation rate was obtained by counting the number of pulses required to perforate the cornea. The laser energy used was 0.4 mJ/pulse in a spot size of 0.2-mm diameter. The ablation rate at this fluence was estimated to be 0.3 µm/pulse. This is a lower rate than obtained in this study, even though the fluence value was more than four times the largest fluence we used.
In more recent study Caughey et al.4 estimated the ablation rate in porcine corneas to be 0.4 µm/pulse at a fluence of 200 mJ/cm2. This work was also performed at the 213 nm wavelength with a maximum pulse energy of 1.1 mJ and beam diameters between 0.5 and 1.1 mm. Depth analysis was performed by focusing a microscope on the top and bottom sections of the cut. Our results at this fluence show an agreement with this value to within 50%.
These studies showed that the 213-nm wavelength produces tissue ablation without damage to the surrounding tissue in a fashion similar to the results achieved with the 193-nm excimer laser. The observations in Caughey et al. of surrounding tissue damage were made using histologic methods.
Further work by Shen et al.17 on the ablation rate of human cornea at 213 nm showed a relationship between ablation rate and fluence similar to our results. However, the ablation rates at comparable fluence values used in our study were consistently lower. They also used perforation of a section of tissue of known thickness to obtain the ablation rate.
The first to investigate the most efficient fluence for operation at the 193-nm wavelength were Krueger and Trokel.2 Ablation rates were found by measuring the number of pulses required to achieve corneal perforation for a wide range of fluences. From these data the fluence for highest efficiency was found by calculating the amount of radiant energy that was required to ablate 100 µm of tissue. A refinement to this method was made in later work that quantified ablation efficiency as the volume of tissue removed per unit of input energy.15 The importance of calculating ablation efficiency can be seen in terms of the amount of energy required to remove a given volume of tissue. A less efficient fluence requires more energy than is necessary to remove the same volume of tissue. As a result, a greater proportion of the energy must be spent on processes other than tissue removal. These processes may be manifest in increased heating of the cornea or a larger shock wave on the surface of the cornea.
We were able to characterize the ablation behavior of the solid state laser (Fig. 1) . The variation of the data in the ablated area suggests that the beam used in this study was not uniform. It should be noted, however, that no beam-smoothing techniques in the form of image rotation or spatial filtering were used. We then calculated the ablation efficiency in a range similar to that used in current clinical excimer lasers. The efficiency was obtained by dividing the ablation rate by the fluence and had units of joules per cubic millimeter. The results shown in Figure 3 suggest that the efficiency reaches a maximum for fluence values above 150 mJ/cm2. The work of Krueger et al.2 was performed on bovine corneas, and they found the most efficient fluence for ablation at 193 nm was 200 mJ/cm2. Others have found the most efficient fluence to be within the range of 150 to 400 mJ/cm2.15 The ablation rate at these fluences was approximately 0.4 µm/pulse, which is slightly lower than the value we obtained of 0.48 µm/pulse (at 150 mJ/cm2) although within 20%. The fluence value of 180 mJ/cm2 used with clinical excimer lasers falls within this range.
In addition to the laser parameters used in this study, there is scope to investigate the effect of varying other parameters and the effects this would have on ablation at this new wavelength. These include the effects of pulse duration and repetition rate. Pulse duration has been shown to affect ablation threshold. Using a pulse duration of 25 psec, Hu and Juhasz18 observed an ablation threshold fluence of 3 mJ/cm2 in human corneas. This was achieved using a wavelength of 211 nm and a repetition rate of 1 kHz. This repetition rate represents an increase of two orders of magnitude above the 10-Hz rate used by excimer lasers, and it is therefore difficult to quantify the effect. A limited study of the effect of repetition rate on ablation threshold was conducted for rates of 1, 10, and 25 Hz by Krueger et al.1 using 193 nm. The results showed there was no change in ablation threshold over this range. This is yet to be established at even higher rates in the kilohertz regimen and the longer 213-nm wavelength.
This study confirms that corneal ablation rates at 213 nm are comparable with those achieved with excimer lasers at 193 nm for similar pulse duration and repetition rates. A solid state laser with increased energy at 213 nm capable of performing large-beam ablations may provide an improvement in corneal hydration14 19 through a reduction in the time required to remove a given amount of tissue.
| Footnotes |
|---|
Commercial relationships policy: P.
Corresponding author: Geoffrey T. Dair, Lions Eye Institute, 2 Verdun Street, Nedlands 6009, Perth, Western Australia. E-mail: gdair{at}cyllene.uwa.edu.au