From the Wilmer Ophthalmological Institute, the Johns Hopkins University, School of Medicine, Baltimore, Maryland.
Abstract
PURPOSE. Short-term pattern electrical stimulation of the retina via multielectrode arrays in humans blind from photoreceptor loss has shown that ambulatory vision and limited character recognition is possible. To develop an implantable retinal prosthesis that would provide useful vision, these results need to be sustained over a prolonged period of retinal electrical stimulation. As a first step toward this goal, the biocompatibility and the feasibility of surgically implanting an electrically inactive electrode array onto the retinal surface was tested.
METHODS. A 5 x 5 electrode array (25 platinum discshaped electrodes in a silicone matrix) was implanted onto the retinal surface using retinal tacks in each of the 4 mixed-breed sighted dogs. Color fundus photography, fluorescein angiography, electroretinography, and visual evoked potentials were obtained preoperatively, at 1-week intervals for 2 weeks postoperatively, then at 2-week intervals up to 2 months postoperatively, and thereafter at 1-month intervals. One dog was killed at 2 months after implantation and a second dog after 3 months of implantation. Histologic evaluation of the retinas was performed. The remaining two dogs continue to be followed beyond 6 months after the implantation surgery.
RESULTS. No retinal detachment, infection, or uncontrolled intraocular bleeding occurred in any of the animals. Retinal tacks and the retinal array remained firmly affixed to the retina throughout the follow-up period. Hyperpigmentation of the retinal pigment epithelium was observed only around the site of retinal tack insertion. No fibrous encapsulation of the implant or intraocular inflammation was visible. A- and b-wave amplitudes of the electroretinogram were depressed at the first postoperative week testing but recovered over the ensuing 1 week and were not statistically different from the normal unoperated fellow eye throughout the postoperative period. N1 and P1 wave amplitudes of the visual evoked potentials were not significantly different from the normal fellow eyes at any of the postoperative test intervals. Fluorescein angiography showed that the entire retina including the area under the electrode array remained well perfused. Similarly, histologic evaluation revealed near total preservation of the retina underlying the electrode array.
CONCLUSIONS. Implantation of an electrode array on the epiretinal side (i.e., side closest to the ganglion cell layer) is surgically feasible, with insignificant damage to the underlying retina. The platinum and silicone arrays as well as the metal tacks are biocompatible. With the success of implanting an electrically inactive device onto the retinal surface for prolonged periods, the effects of long-term retinal electrical stimulation are now ready to be tested as the next step toward developing a prototype retinal prosthesis for human use.
Photoreceptor loss from diseases such as retinitis pigmentosa (RP) and age-related macular degeneration is the leading cause of retinal blindness.1 In these disorders, despite a near total loss of photoreceptors, there is relative preservation of the inner retinal neurons.2 3 A number of groups including ours are evaluating the possibility of restoring sight by developing a retinal prosthesis that would electrically stimulate the remaining retinal neurons.4 5 6 7 8 The feasibility of this approach is supported by clinical studies that have shown that controlled electrical stimulation of retinal areas results in localized and retinotopically correct perceptions in patients blind from end-stage RP and age-related macular degeneration.9 Pattern electrical stimulation using a multielectrode array resulted in multiple dots of light, which when viewed together were perceived by the patients as letters and geometric shapes.10 However, these tests used hand-held electrodes and lasted for short periods, usually less than 60 minutes. To develop an implantable device, long-term implantation and electrical stimulation needs to be equally beneficial and safe. Because both mechanical stress and electrical current could be harmful to the retina, the mechanical and electrical biocompatibility issues must initially be studied separately, so as not to confound a negative result. The most straightforward way to do this is to first-test the biocompatibility of the surgery and chronic implantation without electrical stimulation, an approach taken in the development of other neural prostheses.11 Toward achieving this goal, in this study we address (a) techniques for surgical implantation of the electrode array on the retina, (b) biocompatibility of the implanted array, and (c) the mechanical effects of the electrode array on the retina.
Methods
Electrode Array Specifications
The electrode array was composed of 25 platinum disks arranged in
a 5 x 5 square array (Fig. 1)
. A single 25-µm diameter platinum wire was attached to each
disk. The disks and wire were encapsulated in medical grade
silicone, except for the surface of the platinum disks juxtaposed
against the retina, which was not enclosed by the silicone. The exposed
surface of the platinum disks formed an array of planar, stimulating
electrodes in a silicone matrix. The disks were 400 µm in diameter
and mounted on 600 µm centers. The side of the implant that was
placed next to the retina measured 3 x 5 mm and was curved to
match the retina. The implant was less than 1-mm thick. The 25 wires
from the disks formed a cable 600 µm in diameter, extending from the
electrode array. Each wire was individually insulated. The cable was
10-cm long, a sufficient length to allow the cable to exit the eye
through the sclerotomy wound and to be sutured to the sclera, in the
superotemporal quadrant, under the conjunctiva.
|
Color fundus photography, fluorescein angiography, electroretinography (ERG), and visual evoked potentials (VEP) were obtained preoperatively and at postoperative weeks 1 and 2. These tests were then repeated every 2 weeks up to 2 months and then at every 1-month interval.
Fundus Photography and Fluorescein Angiography
The animals were anesthetized and dilated as described above.
Color fundus photography was performed using 100 ASA color film with a
Topcon TRC-50FT retinal camera (Tokyo Optical, Tokyo, Japan). A No. 0.9
Wratten neutral density filter was used to reduce the tapetal reflex
from the flash of the camera. Two- and one-half milliliters of 10%
sodium fluorescein solution was injected intravenously, immediately
before taking the fluorescein pictures. Fluorescein angiography was
done using 400 ASA black-and-white film with a matched pair of DeLori
fluorescein exciter and barrier filters in place.
Data Acquisition System for Electrophysiology
Our electrophysiological setup consisted of a LKC MGS-1 bright
flash stimulator (LKC Technologies, Gaithersburg, MD) and a computer
for stimulus triggering and waveform capture. The light intensities
(given below for each measurement) are set and calibrated at the
factory and conform to International Society for Clinical
Electrophysiology of Vision Standards for the particular measurement.
The evoked potential was amplified (up to x1000) and filtered with a
P15 Grass amplifier (Grass Instruments, Warwick, RI). Additional
amplification of up to x50 was available on the A/D board (model
ATMIO64-F5; National Instruments, Austin, TX) in the desktop computer
(IBMcompatible). The signal was acquired at a 10-kHz sampling rate
(Labview software; National Instruments).
ERG
Animals were anesthetized by inhalation of halothane and oxygen
and kept under stage 3 and degree 2 to 3 of anesthesia (loss of
palpebral and flexor reflexes and eyes in ventral position). The
animals were maintained in a sternally recumbent position, and their
body temperatures monitored. All animals were dark-adapted for 45
minutes before the beginning of the electrophysiological testing.
During the testing, the eyelids of the stimulated eye were kept open
with an eyelid speculum. The cornea was kept lubricated by the
administration of artificial tears. Three electrodes were used: a
corneal recording electrode, reference electrode at the outer canthus,
and ground electrode in the mouth. For the dark-adapted bright flash
ERG, a single full-field flash stimulus (flash intensity of 2.09
candela/m2 [cd/m2]) was
used. The signal was filtered and amplified (0.3300 Hz bandwidth,
x20,000 amplification). A flicker ERG response was measured by
presenting the same bright flash at 31 Hz in the presence of a dim
(25.2 cd/m2) background light. The same
amplification was used, but the filter was narrowed (1100 Hz
bandwidth, x20,000 amplification). Fifty waveforms were averaged for
each recording of the flicker ERG. The averaged 50 waveforms were
digitally processed with Fourier analysis (1024 point fast Fourier
transformation subroutine supplied with MATLAB 5.0; Mathworks, Natick,
MA). Three recordings of the dark-adapted and flicker ERG types were
obtained for each eye. Recordings of the operated eye were compared
with those from the unoperated eye of the same animal at each follow-up
examination. The data were analyzed by measuring the amplitude and
latency of the a-wave and the b-wave. The b-wave amplitude was measured
from the trough of the a-wave. The averaged recordings of unoperated
eye were taken as 100%, and then the averaged recordings of the
operated eye were calculated as a relative percentage of unoperated
eye. Statistical analysis was performed, using a paired Students
t-test. P
0.05 was considered
statistically significant.
Visual Evoked Potentials
Animal anesthesia and positioning were the same as has been
described for the ERG recordings. Two subcutaneous platinum iridium
electrodes were used, one over the visual cortex and the other over the
frontal region. VEPs were recorded as the average of 100 bright flash
stimuli presented at 1 Hz (flash intensity of 2.09
cd/m2, no background illumination, 0.3300 Hz
bandwidth, x20,000 amplification). To exclude the chance that a
far-field ERG was contaminating our VEP responses, we sequentially
moved the recording electrode closer to the eye and showed that the VEP
waveforms quickly diminished as we moved away from the scalp overlying
the visual cortex. Three recordings were done in both of the eyes at
each follow-up. The averaged recordings of the operated eye were
compared with those from the unoperated eye and statistically analyzed
as described for the ERG recordings.
Electrode Impedance Measurement
Impedance of several electrodes was evaluated before implantation
and after explantation in phosphate-buffered saline. The impedance of
the electrode was evaluated using a custom measurement system. The DC
potential of the electrode was held at 300 mV (versus saturated calomel
reference electrode), and a 5 mV sine wave (1 kHz) was used as the
excitation signal. A constant DC bias is used, because platinum
impedance is sensitive to DC potential.12
Light and Electron Microscopy
One animal was killed at 2 months and one at the 3-month
postoperative period. Two animals remain under follow-up which now is
more than 6 months in duration. Eyes of the killed animals were
enucleated and subjected to light and electron microscopic examination.
The unoperated eye served as the control. After hemisecting the eyes,
the posterior halves were immersed overnight in 2% glutaraldehyde and
2% paraformaldehyde in 0.1 molar phosphate buffer at 4°C. The area
under the retinal prosthesis was sectioned and post-fixed for 2 hours
in 2% osmium tetroxide in phosphate buffer, alcohol dehydrated, and
embedded in epoxy resin. Thick sections (1 µg) stained with toluidine
blue were used for light microscopy, and thin sections (6090 nm)
stained with lead citrate and uranyl acetate and were examined with a
JOEL JEM-100 CX2 transmission electron microscope (Hitachi, Tokyo,
Japan).
Results
Ophthalmoscopic Examination and Fluorescein Angiography
Anterior segment and lens remained normal in all dogs despite a
transscleral cable that remained implanted in the subconjunctival
space. No retinal detachment, infection, or excessive intraocular
bleeding occurred in any of the animals. There was minimal vitreous
hemorrhage in dog 1, which cleared within 2 to 3 weeks. Similarly,
there were a few vitreous cells and minimal vitreous haze in dog 3,
which cleared within the first week. A stable array position is
critical to maintaining steady image resolution and stimulation current
levels. The retinal tacks and the electrode array remained firmly
affixed to the retina of all dogs throughout the follow-up period. A
mild retinal fold was observed at one edge of the electrode array
intraoperatively in all dogs. There was no progression of the retinal
fold with time. Around the site where the tack pierced the retinal
pigment epithelium (RPE), hyperpigmentation and hypopigmentation of the
RPE was noted. Hyperpigmentation was ophthalmoscopically visible within
the first week and progressively enlarged, but remained confined to the
tack area and did not involve the entire implanted area in any of the
dogs (Figs. 2
A, 2B). This hyperpigmentation of the RPE blocked the reflex from the
underlying tapetum.
|
|
|
|
(1
array, 8 electrode sites). Postexplant impedance averaged 8.0 k
(2
arrays, 14 electrode sites). The difference was not statistically
significant (P > 0.05).
Light and Electron Microscopy
Dissection of the enucleated eye confirmed that the array was in
close apposition to the retina, as was indicated by the fundus
photography. Light microscopy showed that the photoreceptor layer, the
outer nuclear layer, the inner nuclear layer, and the ganglion cell
layer were all of normal thickness and density under the electrode
array in both dogs that were killed (Figs. 6
A, 6B). The nerve fiber layer appeared undisturbed. RPE, tapetum, and
choriocapillaris appeared normal on light and electron microscopy.
Electron microscopy also showed intact RPE with good intercellular
connections with photoreceptor outer segments (Fig. 6C)
. The area through which the tack was inserted showed loss of tapetum,
RPE, and retinal layers; however, this damage was limited to the tack
insertion site, and no pathology was observed in retinal layers under
the electrode array compared with the normal eye. Last, close
inspection with a high-power magnification surgical microscope revealed
no mechanical degradation of the silicone rubber matrix, the platinum
electrode wires, or disk electrodes of the explanted array. The
structural integrity of the implanted arrays was also confirmed by
measuring electrical impedance of the electrode sites before and after
implantation (see results of electrode impedance testing).
|
Our results show that a multielectrode array can be secured to the inner retinal surface, is mechanically stable, and biologically tolerated well over a 6-month period (i.e., a length of time after which the outcome is unlikely to change at least as far as vitreoretinal surgical procedures are concerned). Although complex vitreoretinal procedures are commonly performed, affixing an implant to the retina is rare. The retinal tissue, with a thickness of 0.3 mm, is delicate and tears easily. The delicate nature of the retina also predisposes it to pressure necrosis from the weight of an implanted epiretinal device. The abundant vascular supply of the retina and the underlying choroid complicates surgical manipulation because of the threat of excessive bleeding. Despite these hurdles, we were able to place the electrode array epiretinally without inducing significant retinal damage. Before tack placement, a complete vitrectomy with posterior hyaloid separation was necessary to reduce postoperative vitreous traction and displacement of the electrode array. The posterior hyaloid separation in dogs was relatively easy to perform provided that careful technique as described in the Methods section was followed. We found the application of the first retinal tack to be critical because if it pulled out not only would this result in the displacement of the electrode array but also result in immediate bleeding and retinal detachment. The second retinal tack placement was easier, and we found the second tack necessary to keep the array affixed in close proximity to the retina. Retinal tacks were placed so as to avoid damaging retinal ganglion cell axons from the areas underlying the electrode sites. We believe that the minor retinal fold, which appeared intraoperatively after applying the tacks, was due to the thickness of silicone matrix of the electrode array, which displaced the retina toward the edge of the implant. Although this effect was minor and did not progress, the fold can be reduced further by making thinner electrode arrays.
The distance from the stimulating electrode to the retina is critical in terms of electrical stimulation threshold and image resolution. An electrode should be positioned as close as possible to the cells that it will stimulate without harming those cells. Such positioning will maximize resolution and minimize the stimulating current. Although no electrical stimulation was performed in these tests, data from other experiments suggest that positioning the electrodes as was done in these experiments will allow safe and effective stimulation of underlying retinal neurons. Specifically, short-term (<60 minutes) electrical stimulation tests in blind humans have used a similar stimulating array positioned over the retina.10 Test subjects, individuals with end-stage outer retinal degeneration, have reported focal visual percepts corresponding to individual electrodes and simple shapes created by multiple electrodes activated simultaneously. The stimulating current density, in a majority of the cases, was within the acceptable limit for long-term stimulation of neural tissue with either platinum or iridium electrodes.12 Based on these results, we believe that surface electrode positioning is capable of creating useful visual percepts. In the current experiments, we sought to show that the array could be maintained in this position over a long period. Evidence that the array position was maintained (i.e., the array is not floating in the vitreous cavity) comes from fundus photography. The retinal fold near one side of the array (caused by the array pressing on the retina, Fig. 3B ) is consistent in appearance throughout the implant period. The array contacted the retina during the surgery to produce the fold. If the array had later moved away from the retina, the fold would have changed in appearance. Finally, during the initial histologic evaluation, the electrode array was observed to be firmly attached to the retina. Great care was necessary when removing the array from its close apposition to the underlying retina to allow the retina and the array to be evaluated by microscopy.
We chose epiretinal positioning for our electrode array because of the ease of surgical implantation. A subretinal location has the potential benefit of obviating the need to affix an implant to the retina by securing the implant between the retina and the RPE. However, placing the implant in the subretinal space will require either a retinotomy or an external approach that would cross the choroid after a partial thickness scleral flap. The retinotomy may not seal especially if the implant is thick and therefore carries the risk of retinal detachment. The thickness of the implant, even if perforated, reduces the choroidal blood flow to the retina and may result in ischemia of the overlying retina.5 6 Last, crossing the choroid to gain entry into the subretinal space can result in excessive bleeding from the highly vascularized choroidal bed.
We selected dogs for electrode array implantation, because of the well-established and readily available National Eye Institute colony of dogs with photoreceptor degeneration.13 14 15 In these dogs, photoreceptors fail to differentiate normally and subsequently degenerate, due to abnormal cyclic nucleotide metabolism caused by the same genetic defect found in some forms of human RP. These dogs form a good model of outer retinal degeneration and can be used subsequently for implantation of an active electrode array to test the electrical stimulation. Sighted mixed-breed dogs were used for the mechanical biocompatibility tests, however. The retinal degeneration dogs are expensive because of the need for prolonged housing to allow the time for them to develop severe photoreceptor degeneration. Moreover, ERGs and VEPs have been used to study toxic retinal reactions, and only sighted dogs provide us with the opportunity to use those tests to quantify the extent of any damage.4 Even the cost of housing sighted dogs for 6 months or more is high, which limited the number of animals used in this study. On the other hand, both electrophysiological and histologic data were consistent among all four dogs in showing little damage due to long-term epiretinal implants. Therefore, we did not see the need to duplicate these results above and beyond shown in here.
Both platinum and silicone are proven biocompatible materials. Silicone has been approved by the FDA for intraocular use (silicon foldable intraocular lens) and has a proven capability of being well tolerated in the cochlea as part of the cochlear implant electrode array.16 17 18 Similarly, platinum iris clip lenses have been successfully used in ophthalmology.19 Platinum metal also has a proven performance record as an electrode material (in cochlear implants) for long-term electrical stimulation because of its high charge injection limits before irreversible reactions occur at the electrode tissue interface.12 Undoubtedly, if the need for smaller closely packed electrode arrays should arise, the current means of hand-fabricating silicone/platinum electrodes will not suffice, and microfabricated arrays will become a necessity.20 21 However, hurdles remain in microfabricating thin film electrode arrays.21 Silicon has been used as a substrate to fabricate densely packed electrode sites, but to attempt to make a silicon electrode array thin enough to be both flexible and lightweight results in silicon that is unacceptably brittle. Polyimide (Kapton) is another substrate for electrode array fabrication, but it is prone to saline leaking as well as a weak material interface, resulting in delamination of the metal at the electrode site.22 The electrodes we used were modified cochlear implant electrodes and have been tested extensively in saline baths with and without passing current and are part of FDA-approved cochlear implants. We did not see any fibrous encapsulation histologically, and visibly the wires to each electrode remained intact. The silicone rubber also showed no discoloration or any structural changes. Similarly, no significant difference was measured between the postexplant impedance and preimplant impedance. Finally, because it is unclear whether a completely wireless method of data and power transmission can be developed for a retinal prosthesis, we elected to evaluate the worst case scenario of testing an electrode array with an attached transscleral cable. Transscleral tubes are tolerated well as part of certain glaucoma procedures.23 We similarly found a transscleral cable to be well tolerated. We believe that as long as the cable is positioned subconjunctivally and the scleral wound is secure, the risk of infection and any other untoward effect remains low. Thus, if needed, an electrode array can be partially powered via such a cable attachment.
Metal alloy tacks were chosen to secure the implants because of their known biocompatibility (FDA-approved for intraocular use).24 The retinal tack worked well to maintain the electrode array position. Tacks have been associated with fibrovascular reaction when placed at the edge of a retinotomy site.24 25 26 However, when placed in uncut retina, this reaction was only limited to the insertion site and did not affect the area below the implant. There were minimal hyperpigmentary changes at the level of the RPE around the tack insertion site, obscuring the underlying tapetal reflex. But this proliferation of RPE was confined around the site of the tack and did not extend much beyond that site. We believe that this limited proliferation is a mild response for the injury caused by the tack. The mild nature of this change is confirmed by the fact that RPE proliferation did not adversely affect the overlying retina as determined by electrophysiology or histologic evaluations. Similar RPE changes are seen after submacular surgery in cases of choroidal neovascular membrane removal in presumed ocular histoplasmosis cases. In these patients the visual acuity often recovers to an excellent level despite these changes, again suggesting that RPE hyperpigmentary changes are not mutually exclusive with good vision.27 We believe that in lieu of the limited proliferation, an intact underlying choriocapillaris, and overlying retina, such an RPE response can be tolerated. The successful use of retinal tacks does not preclude the search for other even less traumatic techniques of fixation. Some of these include cyanoacrylate glue, magnets, using autologous thrombin,28 and even a preshaped electrode that would conform to the curvature of the back of the eye (cantilever design).4 These techniques all have the advantage of not piercing the RPE and choroid and, therefore, minimizing the damage associated with fixation, but each has its own disadvantages. The magnetic attachment would hinder the future use of MRI but has the distinct advantage of making the implant easy to replace. The ease of replacement is also a feature of the cantilever design, but this approach may be limited by the disadvantage of having to build a different device for varying eye curvatures, as might be encountered in different patients. Cyanoacrylate and thrombin glues do not allow easy replacement of the implant and can be associated with toxicity.28
ERG test results are not significantly different in the operated eyes compared with the normal fellow eyes in all dogs. However, a full-field ERG is a mass response of the retina and does not specifically represent the area of retina under the retinal implant. We used two measures to make a better assessment of the functionality of the retinal area to which the array was fixed (area centralis). VEPs and 31-Hz ERG are potentials dominated by the area centralis.29 The ganglion cells of the area centralis project to the visual cortex nearest to the cortical surface, which means activity from the area centralis comprises the bulk of the VEP. Similarly, the 31-Hz ERG is a cone response, and cones are in greater number in the area centralis. VEPs and 31-Hz ERG responses add further support to the full-field ERG and show that the retina underlying the implant likely remains functional to near normal levels. Focal ERG with the incident rays focused on the retina under the electrode array would be yet another method to provide information regarding retinal function from the implanted area.30 However, in lieu of the above electrophysiological results and the histology showing near normal retinal anatomy, we believe that focal ERG testing would not add significantly. Focal ERG may also be limited in our case, because the metal electrodes would block the incident light and may also scatter it beyond what is covered by the background illumination of a typical focal ERG machine.
In conclusion, we designed and fabricated an electrode array and then successfully implanted it epiretinally over a prolonged period. Although, in the future improvements in the material, fixation, and surgical technique will undoubtedly be made, we have a prototype design and method of implantation. These techniques will allow long-term retinal electrical stimulation, the next step toward the development of a retinal prosthesis.
Footnotes
Reprint requests: Mark S. Humayun, Wilmer Ophthalmological Institute, the Johns Hopkins University, School of Medicine, Maumenee 738, 600 N. Wolfe Street, Baltimore, MD 21287.
Supported by grants from the National Eye Institute (1R 29 EY 11888-01), Research to Prevent Blindness, The Whitaker Foundation, and the Ross Clinician Scientist Award.
Submitted for publication July 6, 1998; revised February 18, 1999; accepted April 6, 1999.
Proprietary interest category: N.
References
This article has been cited by other articles:
![]() |
C. Scholz Perspectives on: Materials Aspects for Retinal Prostheses Journal of Bioactive and Compatible Polymers, September 1, 2007; 22(5): 539 - 568. [Abstract] [PDF] |
||||
![]() |
S. R. Montezuma, J. Loewenstein, C. Scholz, and J. F. Rizzo III Biocompatibility of materials implanted into the subretinal space of yucatan pigs. Invest. Ophthalmol. Vis. Sci., August 1, 2006; 47(8): 3514 - 3522. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Walter, Z. F. Kisvarday, M. Gortz, N. Alteheld, G. Rossler, T. Stieglitz, and U. T. Eysel Cortical Activation Via an Implanted Wireless Retinal Prosthesis Invest. Ophthalmol. Vis. Sci., May 1, 2005; 46(5): 1780 - 1785. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Hossain, I. W Seetho, A. C Browning, and W. M Amoaku Artificial means for restoring vision BMJ, January 1, 2005; 330(7481): 30 - 33. [Full Text] [PDF] |
||||
![]() |
J. I. Loewenstein, S. R. Montezuma, and J. F. Rizzo III Outer Retinal Degeneration: An Electronic Retinal Prosthesis as a Treatment Strategy Arch Ophthalmol, April 1, 2004; 122(4): 587 - 596. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kanda, T. Morimoto, T. Fujikado, Y. Tano, Y. Fukuda, and H. Sawai Electrophysiological Studies of the Feasibility of Suprachoroidal-Transretinal Stimulation for Artificial Vision in Normal and RCS Rats Invest. Ophthalmol. Vis. Sci., February 1, 2004; 45(2): 560 - 566. [Abstract] [Full Text] [PDF] |
||||
![]() |