(Investigative Ophthalmology and Visual Science. 2000;41:3818-3826.)
© 2000
by The Association for Research in Vision and Ophthalmology, Inc.
A Spatial Frequency-Doubling Illusion-Based Pattern Electroretinogram for Glaucoma
Teddy Maddess1,
Andrew Charles James1,
Ivan Goldberg2,
Stephen Wine2 and
Jeffrey Dobinson2
1 From the Centre for Visual Sciences, Research School of Biological Sciences, Australian National University, Canberra; and the
2 Prince of Wales Hospital, Sydney, Australia.
 |
Abstract
|
|---|
PURPOSE. A pattern electroretinogram (PERG) in which stimuli displaying the
frequency-doubling (FD) illusion are presented simultaneously to
multiple parts of the visual field was evaluated for its ability to
diagnose glaucoma. This multiregion FD PERG is referred to in the
current study as the MFP.
METHODS. The nine stimulus regions were temporally modulated at incommensurate
frequencies typically producing an FD percept. Two other spatial scales
of the stimuli were also investigated. The sensitivity and specificity
of MFP were examined using linear and quadratic discriminant methods.
RESULTS. Even with the simpler linear discriminant classification, sensitivities
and specificities of 100% were obtained in eyes with moderate to
severe glaucoma. Of eyes with glaucoma strongly suspected, 67% were
classified as being glaucomatous. Stimulus patterns having differing
spatial scales produced different PERG visual field dependencies.
CONCLUSIONS. The differing results for the 16-fold change in spatial scale may
reflect the accessing of different mechanisms. The MFP method appears
to have significant value for the diagnosis of
glaucoma.
 |
Introduction
|
|---|
Transection of the optic nerve leads to degeneration of the inner
retina and a declining pattern electroretinogram (PERG) in
cats,1
monkeys,2
and humans.3
The PERG has components resulting from current flows in the vicinity of
the inner plexiform (IPL) and ganglion cell layers.4
5
There are several types of retinal ganglion cells6
and
many types of amacrine cells with processes in the
IPL.7
8
9
10
The PERG is thus the sum of currents from a
variety of sources within the proximal retina (for review see Reference
11). Which of these signals, if any, is related to glaucoma and why?
Long-term studies provide evidence that the PERG has predictive value
in glaucoma,12
13
14
but what might be the best stimulus
arrangement for glaucoma diagnosis? Trick15
showed that
increasing the temporal frequency of contrast modulation from 1 to 8 Hz
improved discrimination of normal subjects and patients with primary
open-angle glaucoma (POAG) when using check sizes that have fundamental
spatial frequencies in the range 0.18 to 1.41 cyc/deg (see also
Reference 16). Marx et al.17
found that decreasing spatial
frequency from 3.5 to 0.5 cyc/deg improved sensitivity in experimental
glaucoma for stimulation at 6 Hz. Johnson et al.18
showed
that spatial frequencies of approximately 0.18 cyc/deg (i.e., 2°
checks) modulated at 10 Hz discriminated glaucomatous primate eyes
well. The reader is further referred to an excellent review on the use
of evoked potentials in assessing glaucoma.19
These studies share one factor: The visual stimulus is
presented at the same scale to all parts of the retina, thus
ignoring the 50-fold decline in ganglion cell density between 2° and
10° eccentricity.20
21
There is increasing evidence that
there is diffuse as well as local ganglion cell loss in
glaucoma.22
23
24
25
Anatomical25
and PERG
studies12
13
26
indicate macular as well as peripheral
retinal damage in early glaucoma, and the potential for central retinal
involvement therefore should not be ignored. At the same time,
peripheral scotomas are a diagnostic feature of
glaucoma.27
Ideally, several image regions should be
tested, and the size of these regions and the scale of any textures
contained in them should be tailored to retinal ganglion cell density
and characterization of both diffuse and sectoral losses. Some attempts
have been made in this area, such as use of appropriately scaled dot
patterns28
or separate recordings for different check
sizes.26
29
In multifocal visual evoked potential (VEP)
studies of glaucoma,30
the check sizes used were M-scaled.
The issue of spatial scale is separate from whether there is a
component of the PERG that is suited to glaucoma diagnosis. There is
considerable evidence, both from experimental
glaucoma25
31
and human POAG,32
33
34
35
that
large retinal ganglion cells are disproportionately damaged. In
primates the parasol cells36
project to the magnocellular
layers of the dorsal lateral geniculate nucleus (dLGN)6
37
and so are frequently referred to as M cells. The M pathway contains
(at least) two functional classes that are similar to cat X and Y
cells.6
39
Between 5% and 20% of dLGN M cells respond
with a nonlinear Y-type response39
40
41
42
thus, the term
My cells. The responses of LGN units largely
reflect the characteristics of their retinal ganglion cell
inputs.43
44
45
46
47
Subpopulations of parasol cells that are
either coupled or not coupled to amacrine cells,48
a
feature that discriminates cat X and Y cells,49
may have
some relation to the primate X and Y types. At least three studies
indicate that My-cell receptive
fields39
50
and axons39
40
(based on
conduction velocities) are larger than those of
Mx cells. Therefore, it seems that
My cells may be larger than
Mx cells.
From the perspective of glaucoma diagnosis it is perhaps more important
that My cells appear to have a very low retinal
coverage factor,51
52
53
so that loss of a single cell
perhaps leads to a complete local loss of visual sensitivity at a given
retinal location. These findings, in conjunction with some selective
loss of larger ganglion cells in glaucoma, suggest that the
My-cell subsystem should be accessed to monitor
glaucoma.
How then should the My-cell class be accessed?
One possibility is that the spatial frequency-doubling (FD)
illusion54
55
56
may be related to
My-cell function.52
53
57
58
The FD
illusion is seen when subjects view spatially coarse sinusoidal
gratings with a contrast that is modulated at 15 to 30
Hz54
55
56
or that are presented briefly.59
60
The nonlinearity describing the characteristic response component of Y
cells has the same rectifying form as that governing the FD illusion
(cf. References 61 and 54). Some investigators57
58
62
have shown that when FD is seen, the PERG signal becomes dominated by
components with phase shifts characteristic of the retinal gain control
signal operating on the nonlinear response component of Y
cells.63
This retinal gain control mechanism greatly
increases the nonlinear responses of Y cells to stimuli combining low
spatial frequencies and high temporal
frequencies61
63
64
65
66
67
The nonlinear response component
of Y cells is more amplified by retinal gain control than either the Y
cells own linear response or that of X cells,61
64
68
69
which might explain the appearance of the illusion. Also,
we53
have shown that the spatial density of the units
subserving the FD illusion is very similar to the anatomic expectation
for My cells. The ability of subjects with
glaucoma to see the FD illusion is severely
reduced.52
70
71
72
73
74
75
76
77
78
79
80
Evidence has been provided that FD
stimuli are superior to wide-field flicker in detecting
glaucoma.81
So, as might be predicted if the FD illusion
corresponded to My-cell function, it appears that
the strength of the FD illusion is highly correlated with glaucomatous
damage.
Taking all the evidence into account, we created a visual stimulus that
was correctly scaled with respect to My-cell
density,53
that exploited the FD illusion, and that had
multiple, relatively large stimulus zones.70
In a
companion study82
we compared the MFP results with results
from Humphrey Field Analyser (Humphrey, San Leandro, CA) examinations
and a contrast threshold method using FD stimuli.73
 |
Methods
|
|---|
Visual Stimuli and Data Analysis
Visual stimuli were presented on a model CCID 7551 monitor (Barco,
Kortrijk, Belgium; mean luminance 45 candelas
[cd]/m2). The monitor was calibrated weekly.
Gamma correction was confirmed by nonlinear systems identification
methods.62
Subjects viewed the monitor at 30 cm providing
the stimulus layout illustrated in Figure 1 . A green fixation spot was presented at the screens center. The
screen of the monitor was divided into nine zones or regions (Fig. 1
;
Fig. 3
, inset). Each region contained an achromatic (6500° K)
sinusoidal grating at 95% contrast. In most of the experiments, the
grating orientation in regions 2 through 9 was such that the stripes of
the gratings were normal to lines drawn from the center of the monitor
to the four corners. The grating in central region 1 had vertical
stripes. The scale of the gratings increased with eccentricity
according to cell density.53
Normally, gratings in the
nine regions had the spatial frequencies: region 1, 3/4 cyc/deg;
regions 2 through 5, 1/4 cyc/deg; and regions 6 through 9,
1/8 cyc/deg. In some experiments the spatial frequency of all
zones was scaled up or down by a factor of 4 to make so called coarse
and fine stimulus patterns. The stimulus display was controlled by a
program running on a Vista graphics board (Truevision, Shadeland
Station, IN). Software for data acquisition, online analysis
and data display was written in a commercially available program
(Matlab; The MathWorks, Natick, MA).

View larger version (140K):
[in this window]
[in a new window]
|
Figure 1. Illustration of the MFP stimulus. The spatial scale of the stimulus
present on the monitor screen is demonstrated along with the scale of
the gratings. The displayed spatial frequencies represent the input
spatial frequencies, not the illusory second harmonic observed. For the
so-called coarse and fine stimuli (Fig. 4)
, the spatial frequencies of
all regions were scaled up or down by 4, whereas the region size
remained fixed. The gratings are presented here at different contrasts
to permit the 9 regions to be distinguished but could represent a
single frame of the stimulus sequence, because contrast of the gratings
was modulated asynchronously.
|
|

View larger version (25K):
[in this window]
[in a new window]
|
Figure 3. Average regional MFP response amplitudes by diagnostic group: N,
normal; W, weakly suspect; S, strongly suspect; G, known glaucoma.
Error bars are group SE. (Inset) The diagrammatic
representation of the stimulus indicates the numbering scheme of the
stimulus regions (zones). All eyes tested were right eyes, and visual
field positions and retinal positions therefore corresponded in the
same way for all subjects.
|
|
Data Acquisition and Processing
To permit extraction of responses to the different regions while
recording with one electrode, the contrast of each stimulus region was
sinusoidally modulated in time, each region at a slightly different
temporal frequency (described later). The MFP signals were amplified
125,000 times and sampled in synchrony with the video frames displayed
on the monitor. Electrodes (ERG-Jet; Universo SA, La Chaux-de-Fonds,
Switzerland) were used for the eye contact, and AgCl button
electrodes were used for the indifferent and earth electrodes.
The indifferent electrode was placed on the temple, and the ground on
the cheekbone half way between the midline of the eye and the temple.
Control experiments revealed no contamination by cortical potentials.
The time base for the analogue-to-digital converter (ADC)
sampling (Labmaster MDA, Scientific Solutions, Solon, OH; 16 bit) was
the horizontal line-scan clock (45,473 Hz) supplied by the graphics
board. The video frame rate was 101.50 Hz (noninterlaced), and there
were 448 cycles of the horizontal line-scan clock per frame. A single
stimulus sequence contained 4096 frames of video providing an overall
stimulus duration of 40.4 seconds. Response components were extracted
by the fast Fourier transform (FFT), the run length providing a
temporal frequency resolution
F of 0.0248 Hz.
For the FFT signal extraction process to work, an orthogonal design was
needed. That is to say, it was necessary for all nine stimulus
frequencies (f1,
f2, and
f9) to contain an integral number of
cycles over the 4096 video frames. Because we were interested in the
second harmonics, it was also necessary that no two summed frequencies
(fi +
fj ) should equal any of the nine second
harmonic frequencies. If summed frequencies
fi + fj
appeared in the record, they would represent interaction or light
scattering between the stimulus zones, and these frequencies were
therefore also monitored. The actual stimulus frequencies were the
multiples:
F · (889, 898, 904, 911, 921, 935, 947, and 955). The
resultant second harmonic signals thus ranged from 44.06 to 47.33 Hz
(contrast reversals/sec).
We recorded several repeats of the 40.4-second stimulus sequence. An
FFT of each record was computed, and the resultant complex Fourier
transfer coefficients were averaged. Figure 2
demonstrates graphically the output of the data acquisition program and
the initial analysis. Figure 2A
shows the amplitude spectrum
highlighting the fundamental, second, third, and fourth harmonics.
Figure 2B shows the second harmonics, numbered 1 through 9 and some
other frequencies in the complex plane in an Argand diagram. For this
part of the analysis, we extracted the second harmonics
(fi +
fi =
2fi), the regional interaction
frequencies (fi +
fj, i
j), and all the remaining noise frequencies in the
band f1 to
f9. In the Argand diagram, frequencies
are represented as vectors for which the length from the origin
represents signal amplitude and the orientation represents phase lag.
Note in this case that because the stimulus frequencies were bunched
(range, 1.8 Hz) differences in phase corresponded closely to
differences in conduction delays. In the Argand representation if
response vectors were concatenated from repeated runs and scaled by the
number of repeats (i.e., take the vector average in the complex plane),
then only responses with relatively constant phase would grow in
amplitude. Noise frequencies would have random phase and so would
stagger in a random walk around the origin. The coefficients from the
noise frequencies thus form a bivariate normal distribution that can be
used to measure the significance of the measured harmonics. In Figure 2B
the noise frequency coefficients and regional interaction
frequencies (fi +
fj, i
j) lie inside the circle representing the 95%
significance level. The derivation of that significance level is given
in the next paragraph.
The actual significance of a frequency component is related to its
amplitude, which is the modulus of the Fourier transform coefficient at
that frequency, A(f). A test of
significance for a particular frequency can be performed as follows.
Under a null hypothesis of no signal, the real and imaginary parts
of the Fourier transform coefficients,
Real[A(f)] and
Imag[A(f)], are independent
normal variates with zero mean and some variance
2. The squared modulus of the coefficient
|A(f)|2
is then
2 times a
2
variate on 2 degrees of freedom. An estimate
s2 of
2 is
obtained from frequencies not in the set of second-order stimulus
frequencies, say with n degrees of freedom. An F-test is
then performed on the quotient
(|A(f)|2/2)/(s2/n),
with (2, n) degrees of freedom. For large n the
F-test is closely approximated by a
2 test,
the test statistic being
|A(f)|2/s2,
with 2 degrees of freedom. This method was used to draw the circle
denoting 95% confidence in Figure 2
. Thus, a given frequency component
is significant if, and only if, it lies outside the circle.
As can be seen (Fig. 2B)
, unlike the noise frequencies, the nine signal
components had relatively constant phase and so over four runs had
grown outside the significance level. In practice, the operator viewed
the Argand diagram (Fig. 2B)
and judged whether sufficient repeats had
been obtained to ensure that most or all the signals from the nine
regions exceeded the significance level. Preliminary experiments
determined that all nine responses could be significant in as few as
four repeats (as shown in Fig. 2B
). Therefore, the operator was
instructed to perform at least 4 repeats, but not more than 12, so that
recording time did not exceed 10 minutes. The operator was instructed
to stop after fewer than 12 repeats if all nine regional signals had
reached 95% significance. These averaged responses were used for
subsequent off-line analysis.
Subjects
Right eyes of 77 subjects were tested. We used the 24-2 program of
the Humphrey Field Analyzer (HFA; Humphrey, San Leandro, CA) to obtain
standard perimetric visual field threshold data from the 77 subjects in
the study group. By those and other previously described
criteria73
subjects were classified (initially by IG, and
confirmed by SW and JD) into four groups: normal subjects, persons with
glaucoma weakly suspected (weakly suspect group), persons with glaucoma
strongly suspected (strongly suspect group), and persons with known
glaucoma (glaucoma group). Of the glaucoma group, 12 of 14 met the
strict HFA criteria of Capriolli83
for glaucoma
diagnosis; all met the moderate Capriolli criteria. Strongly suspect
eyes did not meet either of these visual field criteria but had a
highest ever recorded elevated intraocular pressure (IOP), determined
by applanation tonometry, of more than 20 mm Hg and a glaucomatous
change in the optic disc verified by repeated fundus camera photography
of the optic disc over several years (by IG). For a more complete
description of the clinical evaluation of these subjects see Maddess et
al.73
Six of the glaucoma group had a blockage-mechanism form of glaucoma,
and the remainder had POAG. The group was part of a larger study on 330
subjects in which no effect of these differing glaucoma types was found
for visual field thresholds obtained for FD stimuli.73
Weakly suspect eyes had elevated IOP or suspect discs but had no
observed change in the disc over several years. Disc condition was
assessed by fundus photography, and vertical cup-to-disc ratios were
evaluated by color and contour (see Table 1
of Reference 73
). Further
details of the diagnostic criteria are given elsewhere.73
Because of the long-term nature of the parent study,73
the
glaucoma group were medicated, 10 exclusively with ß-blockers and 7
also with miotics. The miotics led to a significant reduction in pupil
diameter of approximately 27% compared with that in normal subjects
(P < 0.005; Table 1
).
The research followed the tenets of the Declaration of Helsinki.
Informed written consent was obtained from the subjects after the
nature and possible consequences of the study were explained to them.
The research was approved by Australian National Universitys Human
Experimentation Ethics Committee under protocol M 881.
Discriminant Analysis
The objective of this analysis was to determine whether the
structure of the data permitted a method that was able to discriminate
normal subjects from those with glaucoma. Note that the differing
covariance of each group suggested the use of quadratic discriminant
analysis (QDA).84
Linear discriminant analysis
(LDA)84
was also conducted for comparison. The Appendix
illustrates the two methods. Although the LDA models are simpler, LDA
was less appropriate than QDA, given the different covariance
structures of the data of normal subjects and those with
glaucoma.84
Sensitivities and specificities were estimated
from receiver operator curves (ROCs), for which in LDA the risk factors
were based on Fishers linear discriminant function , and
the QDA classifier was the likelihood ratio assuming separate variance
matrices.84
Extensive details of LDA, QDA, and the use of
ROCs have been given previously.73
 |
Results
|
|---|
General Results
It was first necessary to determine whether significant zone
amplitudes could be measured. We found that by averaging data from 4 to
12 trials of 40 seconds we could obtain four or more significant MFP
zone amplitudes. The interaction frequencies
(fi +
fj) that would indicate interactions, or
light scattering, between the nine stimulus zones were only
occasionally significant, consistent with the 95% significance test
(Fig. 2
and the Methods section). The average amplitudes for the
different subject groups, normal subjects, and the weakly suspect,
strongly suspect, and glaucoma groups, are shown in Figure 3
. Notice that there were distinct regional differences and that these
were maintained across the diagnostic groups. We also examined regional
response amplitudes in a subgroup of eight normal subjects using visual
stimuli in which the spatial frequencies were scaled up or down by four
times compared with the normal case (Fig. 4)
. The finer stimuli proved the least reliable, with the coarser stimuli
showing a regional amplitude variation similar to that obtained for the
normal test stimuli (cf. Figs. 4A
and 4B
).

View larger version (20K):
[in this window]
[in a new window]
|
Figure 4. Average regional MFP response amplitudes for eight normal subjects for
test stimuli having the spatial frequencies of each of their regions
scaled: (A) Four times lower than the normal case,
(B) the same as for the normal stimuli (Fig. 3)
, and
(C) four times higher than normal. Error bars are group SE.
The horizontal dash-dot line segment in each figure is the
mean 95% confidence limit above which the signals are determined to be
significant.
|
|
Discriminant Analysis
We next examined the sensitivity and specificity of the MFP. We
took two basic approaches to this analysis. The first involved
constructing discriminant models on so-called sorted measures, in which
measures obtained from each stimulus region were sorted according to
their reliability, thus biasing the analysis toward the most
significant differences from normal performance. For the sorted data we
examined the effect of the number of regional signals included in the
analysis. The second approach involved making direct region-to-region
comparisons.
Absolute phase was of little use. This was not unexpected, because at
these frequencies small delays translate into phase shifts exceeding 1
cycle. Instead, we examined relative phase. We used the phase of region
9 as the reference phase for each data set, because it provided the
most reliable signal (Figs. 2A
3
4B)
, being significant at
P < 0.001 in 76% of all subjects and in all normal
subjects at P < 0.05 or better. To construct the
relative phase, we subtracted the phase of region 9 for a given subject
from each of the nine regional phases. In this way, the phase of region
9 for all subjects was brought to 0°, and the phase lags and leads of
the other regional responses relative to that of region 9 were
preserved.
For both the sorted and rotated data sets, we examined amplitude and
relative phase and amplitude relative to the geometric means across
normal subjects. Both decibel and linear amplitude measures were
considered. In a further analysis, we subtracted the (geometric) mean
model for normal subjects from the amplitude data, and in other cases
we divided by the mean model (equivalent to subtracting the decibel
version from data transformed to decibels). This weighting operation
was a conservative operation, because it scaled large, reliable signals
downward and small, less reliable, MFP signals upward. For all the
cases we examined linear (LDA) and quadratic (QDA) discriminant
analysis models (see the Methods section and Appendix) in which the
models were constructed from the data from our normal and glaucoma
groups. These discriminant functions were then used to classify the
subjects in the weakly and strongly suspect groups. Further details are
supplied later for specific cases.
One of the simplest measures, regional amplitudes divided by the
geometric mean for normal subjects, so-called scaled amplitudes, was
the most reliable diagnostically, providing a sensitivity of 92.9% at
91.7% specificity for the linear discriminant model, and sensitivities
and specificities of 100% for the quadratic case. Note that means and
covariances for the normal and glaucoma groups were used to form the
discriminant functions () and then the vectors,
xi, of nine scaled amplitudes from each
subject were used as the inputs for classification
(x0 in ). Sorted
amplitude differences from normal performance gave poor performance at
71.4% sensitivity for 75% specificity (simultaneously highest values,
85.7% specificity at 83.3% sensitivity for the quadratic case).
Relative phase alone was as good as sorted amplitudes at 71.4%
sensitivity for 75% specificity for LDA. There was no general pattern
of relative phase changes with respect to visual field location.
Including both amplitude and relative phase gave 100% sensitivity and
specificity even in the LDA case. This was true for both scaled and
unscaled amplitudes. In these models the vectors,
xi, of nine amplitudes and nine relative
phases from each subject were used as the inputs for classification
(x0 in ). The scaled
amplitude LDA model diagnosed 52.7% of weakly suspect eyes and 66.7%
of strongly suspect eyes as glaucomatous. Performance for the QDA model
was 50.0% and 60.0% for weakly and strongly suspect eyes,
respectively.
 |
Discussion
|
|---|
The high sensitivities and specificities (100%) reported for the
more complex quadratic discriminant models require further
verification; however, even the simpler linear discriminant model using
amplitude and relative phase performed at 100%. Of our 14 subjects
with glaucoma 12 did not meet the strict HFA criteria of
Capriolli83
for glaucoma diagnosis although all met the
moderate criteria. Thus, it could be argued that we were only matching
HFA performance. Nevertheless, 67% of eyes that were strongly suspect
had glaucoma diagnosed by the same analysis (see also Reference 82).
Combined measures considering both the N95 and P50 component of the
conventional PERG seem to provide sensitivity and specificity in the
region of 95%13
for confirmed glaucoma. Relative phase
appeared to add some extra information for discrimination of our data.
Similarly, small effects have been reported for absolute phase in
conventional PERGs.85
86
Inspection of Table 1
shows that the mean defect (MD) and pattern
standard deviations (PSDs) of our normal subjects were slightly high.
Although this may at first seem at odds with our high reported
sensitivities and specificities for the MFP method, it may also be
related to earlier findings that HFA MD is not well correlated with
glaucomatous damage.73
In any case, improved diagnosis in
our normal subjects would only improve our sensitivities and
specificities.
The large MFP regions used in this study mean the detected loss was
relatively diffusely distributed. This apparent diffuse loss is in
accord with a study in which 330 subjects performed an FD-based
contrast threshold test up to seven times over 2 years.73
In that study the thresholds were obtained with large (FD) patterns
similar to the MFP stimuli used in the this study, and the tests
provided median sensitivities of 90.8% at specificities of 94%. These
two studies indicate that tests based on the FD illusion can pick up
diffuse early glaucomatous loss, as in the strongly suspect group in
this study.82
Such a result is consistent with
anatomical22
23
24
25
87
and conventional PERG
studies,12
13
26
indicating diffuse cell loss in glaucoma.
An interesting feature of the MFP data was the differing responses by
the different visual field quadrants over a 16-fold range of spatial
frequencies (Fig. 4)
. Visual field dependencies similar to those found
for our usual stimulus condition (Figs. 3 4B)
have been reported
before in which the four quadrants were tested sequentially with an
8.33-Hz, 0.5-cyc/deg pattern.88
A study examining the
density of the units subserving the FD illusion has shown similar
differences53
that were found to be related to known
anatomic retinal ganglion cell densities. Thus, it seems with these
spatially coarse fast stimuli, we may be accessing a different retinal
pathway. A study comparing the diagnostic utility for glaucoma of a
range of low spatial frequencies, modulated at 27 Hz and presented in
the periphery, showed that the usual spatial frequencies used in this
study are approximately optimal for glaucoma diagnosis.81
Virtues of Using the My Pathway for the Early Detection
of Retinal Damage
Examining the My-cell system may provide
greater sensitivity to glaucomatous damage than examining other more
populous systems.52
53
Improved sensitivity to damage
would arise from two sources: cell size and retinal coverage factor.
Early suggestions that in humans larger retinal ganglion cells were
more susceptible to glaucomatous damage22
32
33
34
have been
supported by many studies31
32
33
34
indicating that primate
retinal ganglion cell loss in glaucoma is proportional to ganglion cell
size. The size dependence extends even to the fovea.25
At
least three studies indicate that My cells are
larger than Mx cells.39
40
42
As mentioned, perhaps the primary reason for
My-cell sensitivity to glaucomatous damage may
reside in retinal coverage factors.51
52
53
Anatomical89
90
and electrophysiological44
estimates of the ganglion cell coverage factors (or number of receptive
fields/image point) range between 2 and 7 for M cells (cf.
24 for P cells44
; for review see Reference 91). Evidence
from studies on humans53
and macaques39
40
41
42
indicates that the My cells make up 20% or less
of the M pathway, and as a result their coverage factor could be less
than 1. It follows that the performance of the My
system would be particularly susceptible to any diffuse cell loss
across the retina, because there would be little redundancy to hide the
loss. Such low coverage factors would lead to spatial aliasing effects,
which have been reported.51
53
Problems
Perhaps the major problem with the present method is that in a few
eyes we could not obtain significant amplitudes from all nine regions,
even after 12 repeats. Some nonsignificant amplitudes can be expected
in glaucoma; however, at 40 seconds per repeat this would make the
method uncompetitive with frequency doubling technology (FDT; Zeiss
Humphrey Systems, Dublin, CA) perimetry.71
72
74
ERG recording itself has its problems19
and advances in
VEP recording methods for glaucoma30
may make that avenue
more attractive.
 |
Summary
|
|---|
PERG amplitudes measured for gratings scaled over a 16-fold range
showed quite different visual field dependencies. These differences
could reflect different retinal mechanisms being accessed by the
differently scaled stimuli. PERG signal components associated with the
FD illusion57
58
62
appear to be highly selective for
glaucomatous damage. The good sensitivity obtained for our strongly
suspect group (and not for the weakly suspect group) indicates that
information about early glaucoma is obtained. The low spatial
frequencies required to produce the FD illusion are a positive
advantage from the clinical standpoint, because little demodulation of
contrast can be expected even by ±5 D of defocus.
 |
Appendix 1
|
|---|
Suppose there are m observations of the multivariate
random variable X' = (X1,
X2 ...
Xp) from population
1, and n observations of the same
format from population
2. From these data sets
the correct classification is assumed to be known. The technique of
discriminant analysis derives a classification rule for any new data
point, which seeks to minimize the cost of a misclassification.
Let us assume that:
- The two populations have probability density functions
f1(x) and
f2(x) respectively.
- The probabilities of a new data point coming from populations
1 or
2 is
p1 and
p2 = 1 -
p1 respectively (the prior
probabilities).
- The cost of misclassifying a point that is really from population
1 as being from the other population is
C1, and the cost of misclassifying a
point that is really from population
2 as
being from the other population is C2.
Note that the probability density function
f1(x) is the conditional
probability density for observing the value x, given that
the point came from population
1. The joint
probability density of observing the value x and that the
point came from
1 is thus
f1(x)p1,
and the marginal probability of observing the value x
irrespective of the class is f(x) =
f1(x)p1
+
f2(x)p2.
Bayes rule can be used to invert the conditional probabilities: The
probability that an observed value x came from population
1 is
p(
1|x) =
f1(x)p1/f(x),
and the probability that it came from population
2 is
p(
2|x) =
f2(x)p2/f(x).
For a given new observation x, the expected cost if we
assign to class 2 is the probability that it was really from population
1 multiplied by that cost:
C1f1(x)p1/f(x).
The expected cost if we assign to class 1 is likewise
C2f2(x)p2/f(x).
The optimal rule is now clear: We should assign to class 1 if that
expected cost is lessthat is, if
Rearranging, this is true when
or, taking the natural logarithms, a monotonic increasing
function,
The further assumption may be made that each of the two
populations follows a multivariate normal distribution, for which we
estimate the population mean vectors by the sample means
u1 and u2
(p x 1), and we estimate the population
covariance matrices by the sample covariance matrices,
S1 and
S2 (p x
p).
If the covariances are assumed to be equal, they are both estimated by
the pooled covariance, Spooled =
[(m - 1)S1 +
(n -
1)S2]/(m +
n - 2). The multivariate normal probability density
for population
1 is then estimated as
and likewise for population
2. The
condition for assigning to class 1 using the logarithm of the ratio is
and can be rearranged as
This condition can be written in the form
 | (A1) |
where c is a scalar and b is a row vector of
coefficients. This is sometimes referred to as Fischers linear
discriminant or classification rule.
If the covariance matrices are not equal, i.e.,
S1
S2, then the resultant rule is to
classify an observation x0 as being from
population 1 (
1) if the following quadratic
condition holds.
 | (A2) |
where
Geometrically, the linear discriminant function defines a decision
boundary, or separatrix, that is a (possibly diagonal)
straight line across the plane in the case of a bivariate observation
and in general is a hyperplane in the multivariate case. The quadratic
discriminant rule defines a decision boundary that is (possibly
rotated) a parabola in the bivariate case and is a paraboloidal surface
in the general multivariate case (for examples see73
).
 |
Acknowledgements
|
|---|
The authors thank the reviewers for advisory assistance.
 |
Footnotes
|
|---|
Submitted for publication June 10, 1999; revised January 26 and June 19, 2000; accepted July 11, 2000.
Commercial relationships policy: P (TM, ACJ); N (all others).
Corresponding author: Teddy Maddess, Centre for Visual Sciences, Research School of Biological Sciences, Australian National University, Canberra ACT 0200, Australia. ted.maddess{at}anu.edu.au
 |
References
|
|---|
-
Holländer, H, Bisti, S, Maffei, L, Hebel, R. (1984) Electroretinogram responses and retrograde changes of retinal morphology after intracranial optic nerve section A quantitative analysis in the cat Exp Brain Res 55,483-493[Medline][Order article via Infotrieve]
-
Maffei, L, Fiorentini, A. (1981) Electroretinographic responses to alternating gratings before and after section of the optic nerve Science 211,953-954[Abstract/Free Full Text]
-
Harrison, JM, OConnor, PS, Young, SL, Kincaid, M, Bentley, R. (1987) The pattern ERG in man following surgical resection of the optic nerve Invest Ophthalmol Vis Sci 28,492-499[Abstract/Free Full Text]
-
Sieving, PA, Steinberg, RH (1987) Proximal retinal contribution to the intraretinal 8-Hz pattern ERG of cat J Neurophysiol 57,104-120[Abstract/Free Full Text]
-
Baker, CL, Hess, RR, Olsen, BT, Zrenner, E. (1988) Current source density analysis of linear and nonlinear components of the primate electroretinogram J Physiol 407,155-176[Abstract/Free Full Text]
-
Shapley, RM, Perry, VH (1986) Cat and monkey retinal ganglion cells and their visual functional roles Trends Neurosci 9,229-235
-
Kolb, H, Nelson, R, Mariani, A. (1981) Amacrine cells, bipolar cells and ganglion cells of the cat retina: a Golgi study Vision Res 21,1081-1114[Medline][Order article via Infotrieve]
-
Kolb, H, Linberg, KA, Fisher, SK (1992) Neurons of the human retina J Comp Neurol 318,147-187[Medline][Order article via Infotrieve]
-
Mariani, AP (1990) Amacrine cells of the monkey retina J Comp Neurol 301,382-400[Medline][Order article via Infotrieve]
-
Vaney, DI (1990) The mosaic of amacrine cells in the mammalian retina Prog Retinal Res 9,49-100
-
Zrenner, E. (1990) The physiological basis of the pattern electroretinogram Prog Retinal Res 9,427-464
-
Falsini, B, Colotto, A, Porciatti, V, Porrelo, G. (1992) Follow-up study with pattern ERG in ocular hypertension and glaucoma patients under timolol maleate treatment Clin Vision Sci 7,341-347
-
ODonaghue, E, Arden, GB, OSullivan, F, et al (1992) The pattern electroretinogram and ocular hypertension Br J Ophthalmol 76,387-394[Abstract/Free Full Text]
-
Pfeiffer, N, Tillmon, B, Bach, M. (1993) Predictive value of the pattern electroretinogram in high-risk ocular hypertension Invest Ophthalmol Vis Sci 34,1710-1715[Abstract/Free Full Text]
-
Trick, GL (1985) Retinal potentials in patients with primary open-angle glaucoma: physiological evidence for temporal frequency tuning effects Invest Ophthalmol Vis Sci 26,1750-1758[Abstract]
-
Towle, VL, Moskowitz, A, Sokol, S, Schwartz, B. (1983) The visual evoked potential in glaucoma and ocular hypertension: effects of check size, field size, and stimulation rate Invest Ophthalmol Vis Sci 24,175-183[Abstract/Free Full Text]
-
Marx, MS, Podos, SM, BodisWollner, I, Lee, PY, Wang, RF, Severin, C. (1988) Signs of early damage in glaucomatous monkey eyes, low spatial frequency losses in the pattern ERG and VEP Exp Eye Res 46,173-184[Medline][Order article via Infotrieve]
-
Johnson, MA, Drum, BA, Quigley, HA, Sanchez, RM, Dunkelberger, GR (1989) Pattern-evoked potentials and optic nerve fiber loss in monocular laser-induced glaucoma Invest Ophthalmol Vis Sci 30,897-907[Abstract/Free Full Text]
-
Graham, SL, Klistorner, A. (1998) Electrophysiology: a review of signal origins and applications to investigating glaucoma Aust N Z J Ophthalmol 26,71-85[Medline][Order article via Infotrieve]
-
Wässle, H, Grünert, U, Röhrenbeck, J, Boycott, BB (1989) Cortical magnification factor and the retinal ganglion cell density of the primate retina Nature 30,643-646
-
Wässle, H, Grünert, U, Röhrenbeck, J, Boycott, BB (1990) Retinal ganglion cell density and cortical magnification factor in the primate Vision Res 30,1897-1911[Medline][Order article via Infotrieve]
-
Quigley, HA, Addicks, EM, Green, RW (1982) Optic nerve damage in human glaucoma, III: quantitative correlation of nerve fibre loss and visual field defect in glaucoma, ischaemic neuropathy, papilledema and toxic neuropathy Arch Ophthalmol 100,135-146[Abstract]
-
Airaksinen, PJ, Lukowski, R, Drance, SM, Price, M. (1986) Colour vision and retinal nerve fiber layer in early glaucoma Am J Ophthalmol 101,208-213[Medline][Order article via Infotrieve]
-
Schulzer, M, Mikelberg, FS, Drance, SM (1987) A study of the value of the central and peripheral isoptres in assessing visual field progression in the presence of paracentral scotoma measurements Br J Ophthalmol 71,422-427[Abstract/Free Full Text]
-
Glovinsky, Y, Quigley, HA, Pease, ME (1993) Foveal ganglion cell loss is size dependent in experimental glaucoma Invest Ophthalmol Vis Sci 34,395-400[Abstract/Free Full Text]
-
Bach, M, Pfeiffer, N, BirknerBinder, D. (1992) Pattern electroretinogram reflects diffuse retinal damage in early glaucoma Clin Vis Sci 7,335-340
-
Capriolli, J. (1991) Automated perimetry in glaucoma Am J Ophthalmol 111,235-239[Medline][Order article via Infotrieve]
-
Zrenner, E, Nelson, R. (1988) Spatial characteristics of scotopic threshold responses in the corneally recorded electroretinogram elicited by multispot patterns Clin Vision Sci 3,29-44
-
Pfeiffer, N, BirknerBinder, D, Bach, M. (1991) Das Muster-ERG bei okulärer Hypertension und Glaukom: Einfluss von Karogrösse, Kontrast und retinaler Exzentriztät Fortschr Ophthalmol 88,815-818[Medline][Order article via Infotrieve]
-
Klistorner, A, Graham, S, Grigg, J, Billson, F. (1998) Multifocal topographic visual evoked potential: improving objective detection of local visual field defects Invest Ophthalmol Vis Sci 39,937-950[Abstract/Free Full Text]
-
Glovinsky, Y, Quigley, HA, Dunkelberger, GR (1991) Retinal ganglion cell loss is size dependent in experimental glaucoma Invest Ophthalmol Vis Sci 32,484-491[Abstract/Free Full Text]
-
Quigley, H, Sanchez, R, Dunkelberger, GR, LHernault, NL, Baginski, TA (1987) Chronic glaucoma selectively damages large optic nerve fibres Invest Ophthalmol Vis Sci 28,913-920[Abstract]
-
Quigley, HA, Dunkelburger, GR, Green, WR (1988) Chronic human glaucoma causing selectively greater loss of large optic nerve fibres Ophthalmology 95,357-367[Medline][Order article via Infotrieve]
-
Quigley, HA, Dunkelburger, GR, Green, WR (1989) Studies of retinal ganglion cell atrophy correlated with automated perimetry in human eyes with glaucoma Am J Ophthalmol 107,453-464[Medline][Order article via Infotrieve]
-
Quigley, H, KerriganBaumrind, L, Pease, M, Kerrigan, D, Mitchel, R. (1999) The number of retinal ganglion cells in glaucoma eyes compared with threshold visual field data in the same eyes [ARVO Abstract] Invest Ophthalmol Vis Sci 40(4),S582Abstract nr 3059.
-
Polyak, SL (1941) The Retina University of Chicago Press Chicago.
-
Grünert, U, Greferath, U, Boycott, BB, Wässle, H. (1993) Parasol (Pa) ganglion-cells of the primate fovea: immunocytochemical staining with antibodies against GABA-a receptors Vision Res 33,1-14[Medline][Order article via Infotrieve]
-
EnrothCugell, C, Robson, JG (1966) The contrast sensitivity of retinal ganglion cells of the cat J Physiol 187,517-522
-
Kaplan, E, Shapley, RM (1982) X and y cells in the lateral geniculate nucleus of macaque monkeys J Physiol 330,125-143[Abstract/Free Full Text]
-
Marrocco, RT, McClurkin, JW, Young, RA (1982) Spatial summation and conduction latency classification of cells of the lateral geniculate nucleus of macaques J Neurosci 2,1275-1291[Abstract]
-
Derrington, AM, Lennie, P. (1984) Spatial and temporal contrast sensitivities of neurones in lateral geniculate nucleus of macaque J Physiol 357,219-240[Abstract/Free Full Text]
-
Blakemore, C, VitalDuran, F. (1986) Organization and post-natal development of the monkeys lateral geniculate nucleus J Physiol 380,453-491[Abstract/Free Full Text]
-
Kaplan, E, Shapley, RM (1986) The primate retina contains two types of ganglion cells, with high and low contrast sensitivity Proc Natl Acad Sci USA 83,2755-2557[Abstract/Free Full Text]
-
Crook, JM, LangeMaleki, B, Lee, BB, Valberg, A. (1988) Visual resolution of macaque retinal ganglion cells J Physiol 396,205-224[Abstract/Free Full Text]
-
Michael, CR (1988) Retinal afferent arborization patterns, dendritic field orientations, and the segregation of function in the lateral geniculate nucleus of the monkey Proc Natl Acad Sci USA 85,4914-4917[Abstract/Free Full Text]
-
Purpura, K, Kaplan, E, Shapley, RM (1988) Background light and the contrast gain of primate P and M retinal ganglion cells Proc Natl Acad Sci USA 85,4534-4537[Abstract/Free Full Text]
-
Purpura, K, Trachina, D, Kaplan, E, Shapley, RM (1990) Light adaptation in the primate retina: analysis of changes in the gain and dynamics of monkey retinal ganglion cells Vis Neurosci 4,75-93[Medline][Order article via Infotrieve]
-
Dacey, DM, Brace, S. (1992) A coupled network for parasol but not midget ganglion cells in the primate retina Vis Neurosci 9,279-290[Medline][Order article via Infotrieve]
-
Vaney, DI (1991) Many diverse types of retinal neurons show tracer coupling when injected with biocytin or neurobiotin (letter) Neurosci 125,187-190
-
Blakemore, C, VitalDurand, F. (1981) Distribution of x- and y-cells in the monkeys lateral geniculate nucleus J Physiol 320,17P
-
Maddess, T, Henry, GH. (1990) Density of nonlinear visual units and glaucoma [ARVO Abstract] Invest Ophthalmol Vis Sci 31(4),S231Abstract nr 1134
-
Maddess, T, Henry, GH (1992) Nonlinear visual responses and visual deficits in ocular hypertensive and glaucoma subjects Clin Vis Sci 7,371-383
-
Maddess, T, James, AC, Hemmi, J. (1998) Evidence for spatial aliasing effects in the Y-like cells of the magnocellular visual pathway Vision Res 38,1843-1859[Medline][Order article via Infotrieve]
-
Kelly, DH (1981) Nonlinear visual responses to flickering sinusoidal gratings J Opt Soc Am 71,1051-1055[Medline][Order article via Infotrieve]
-
Kelly, DH (1966) Frequency doubling in visual responses J Opt Soc Am 56,1628-1633
-
Tyler, CW (1974) Observations on spatial-frequency doubling Perception 3,81-86[Medline][Order article via Infotrieve]
-
Maddess, T, Bedford, S, James, AC, Rose, KA (1997) A multiple frequency, multiple region pattern electroretinogram investigation of nonlinear retinal signals Aust N Z J Ophthalmol 25,94-97[Medline][Order article via Infotrieve]
-
Bedford, S, Maddess, T, Rose, KA, James, AC (1997) Correlations between observability of the spatial frequency doubled illusion, a multi-region PERG Aust N Z J Ophthalmol 25,91-93
-
Maddess, T, Kulikowski, JJ (1999) Apparent finess of stationary compound gratings Vision Res 39,3404-3416[Medline][Order article via Infotrieve]
-
Kulikowski, JJ (1975) Apparent fineness of briefly presented gratings: balance between movement and pattern channels Vision Res 15,673-680[Medline][Order article via Infotrieve]
-
Victor, JD (1988) The dynamics of the cat retinal Y cell subunit J Physiol 405,289-320[Abstract/Free Full Text]
-
James, AC, Maddess, T, Rouhan, K, Bedford, S, Snowball, M. (1995) Evidence for My-cell involvement in the spatial frequency doubled illusion as revealed by a multiple region PERG for Glaucoma J Opt Soc Am 1,314-317
-
Shapley, RM, Victor, JD (1980) The effect of contrast on the non-linear response of the Y-cell J Physiol 302,535-547[Abstract/Free Full Text]
-
Victor, JD, Shapley, RM (1979) The nonlinear pathway of Y ganglion cells in the cat retina J Gen Physiol 74,671-689[Abstract/Free Full Text]
-
Shapley, RM, Victor, JD (1981) How the contrast gain control modifies the frequency responses of cat retinal ganglion cells J Physiol 318,161-179[Abstract/Free Full Text]
-
Benardete, EA, Kaplan, E, Knight, BW (1992) Contrast gain control in the primate retina: P cells are not X-like, some M cells are Vis Neurosci 8,483-486[Medline][Order article via Infotrieve]
-
Benardete, EA, Kaplan, E. (1999) The dynamics of primate retinal ganglion cells Vis Neurosci 16,344-368
-
Victor, JD (1987) The dynamics of the cat retinal X cell centre J Physiol 386,219-246[Abstract/Free Full Text]
-
Victor, JD, Shapley, RM (1979) Receptive field mechanisms of cat X and Y retinal ganglion cells J Gen Physiol 74,257-298
-
James AC, Maddess T. Glaucoma testing using non-linear systems identification techniques, 1996; US Patent No. 5,539,482.
-
Maddess T. Method and apparatus for use in diagnosis of glaucoma. US patent 5,065,767; 1991.
-
Maddess TL. Early detection of glaucoma. US Patent 5,912,723; 1995.
-
Maddess, T, Goldberg, I, Dobinson, J, Wine, S, Welch, AH, James, A. (1999) Testing for glaucoma with the frequency doubled illusion Vision Res 39,4258-4273[Medline][Order article via Infotrieve]
-
Johnson, CA, Samuels, S. (1997) Screening for glaucomatous visual field loss with frequency doubling perimetry Invest Ophthalmol Vis Sci 38,413-425[Abstract/Free Full Text]
-
Sponsel, W, Argano, S, Trigo, Y, Mensah, J. (1998) Clinical classification of glaucomatous visual field loss by frequency doubling perimetry Am J Ophthalmol 125,830-836[Medline][Order article via Infotrieve]
-
Quigley, H. (1998) Identification of glaucoma-related visual field abnormality with the screening protocol of frequency doubling technology Am J Ophthalmol 125,819-829[Medline][Order article via Infotrieve]
-
Blumenthal, EZ, Sample, PA, Zangwill, L, Lee, AC, Kono, Y, Weinreb, RN (2000) Comparison of long-term variability for standard and short-wavelength automated perimetry in stable glaucoma patients Am J Ophthalmol 129,309-313[Medline][Order article via Infotrieve]
-
Burnstein, Y, Ellish, NJ, Magbalon, M, Higginbottom, EJ (2000) Comparison of frequency doubling perimetry with Humphrey visual field analysis in a glaucoma practice Am J Ophthalmol 129,328-333[Medline][Order article via Infotrieve]
-
Cello, KE, NelsonQuigg, JM, Johnson, CA (2000) Frequency doubling perimetry for detection of glaucomatous field loss Am J Ophthalmol 129,314-322[Medline][Order article via Infotrieve]
-
Patel, SC, Friedmann, DS, Varadkar, P, Robin, AL (2000) Algorithm for interpreting the results of frequency doubling perimetry Am J Ophthalmol 129,323-327[Medline][Order article via Infotrieve]
-
Maddess, T, Severt, WL (1999) Testing for glaucoma with the frequency doubling illusion in the whole, macular and eccentric visual fields Aust N Z J Ophthalmol 27,194-196[Medline][Order article via Infotrieve]
-
Maddess, T, James, AC, Goldberg, I, Wine, S, Dobinson, J. (2000) Comparing a parallel PERG, automated perimetry, and frequency-doubling thresholds Invest Ophthalmol Vis Sci 41,3827-3832[Abstract/Free Full Text]
-
Capriolli, J. (1990) Automated perimetry in glaucoma Walsh, TJ eds. Visual Fields Examination and Interpretation ,71-106 American Academy of Ophthalmology San Francisco.
-
Johnson, RA, Wichern, DW (1992) Applied Multivariate Statistical Analysis 3rd ed. PrenticeHall Englewood Cliffs, NJ.
-
Korth, M, Horn, F, Storck, B, Jost, J. (1989) The pattern-evoked electroretinogram (PERG): age related alterations and changes in glaucoma Graefes Arch Clin Exp Ophthalmol 227,123-130[Medline][Order article via Infotrieve]
-
Falsini, B, Colotto, A, Porciatti, V, Buzzontelli, L, Coppe, A, De Luca, LA (1991) Macular flicker- and pattern-ERGs are differently affected in ocular hypertension and glaucoma Clin Vision Sci 6,423-429
-
Harwerth, RS, CarterDawson, L, Shen, F, Smith, EL, III, Crawford, MJL (1999) Ganglion cell losses underlying visual field defects from experimental glaucoma Invest Ophthalmol Vis Sci 40,2242-2250[Abstract/Free Full Text]
-
Horn, F, Mardin, C, Korth, M, Martus, P. (1996) Quadrant pattern ERG with SLO stimulation in normals and glaucoma patients Graefes Arch Clin Exp Ophthalmol 234(suppl 1),S174-S179
-
Perry, VH, Cowey, A. (1985) The ganglion cell and cone distributions in the monkeys retina: implications for central magnification factors Vision Res 25,1795-1810[Medline][Order article via Infotrieve]
-
Grünert, U, Martin, PR (1991) Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity J Neurosci 11,2742-2758[Abstract]
-
Lee, BB (1996) Receptive field structure in the primate retina Vision Res 36,631-644[Medline][Order article via Infotrieve]
This article has been cited by other articles:

|
 |

|
 |
 
T. Maddess, A. C. James, I. Goldberg, S. Wine, and J. Dobinson
Comparing a Parallel PERG, Automated Perimetry, and Frequency-Doubling Thresholds
Invest. Ophthalmol. Vis. Sci.,
November 1, 2000;
41(12):
3827 - 3832.
[Abstract]
[Full Text]
|
 |
|