1 From the Department of Ophthalmology, University of Freiburg, Germany; and 2 the Institute of Applied Physics, University of Berne, Switzerland.
| Abstract |
|---|
|
|
|---|
METHODS. Free running (
= 250 µsec) Er:YAG laser pulses were
transmitted through a zirconium fluoride (ZrF4) fiber
guarded by quartz rod (d = 1000 µm). Laser pulses
were applied to the retinal surface of enucleated pig eyes. Eyes were
mounted in a specially designed rotating sample holder. The fiber probe
was elevated 1.0 ± 0.3 mm above the retinal surface with
perfluorodecaline serving as transmitting medium. The laser energy was
applied in a circular pattern with a radius of 3.0 mm. Radiant
exposures were set to 1, 3, 5, and 10 J/cm2.
RESULTS. Tissue ablation linearly increased with radiant exposure from 3.2 ± 3.7 µm at 1 J/cm2 up to 40.9 ± 12.9 µm at 10 J/cm2. Thermal tissue changes extended 70 ± 10 µm vertically into the retina and 25 ± 5 µm horizontally. Distortion of outer photoreceptor segments was noticed when the retina was exposed to radiant exposures of 3 J/cm2 or higher.
CONCLUSIONS. The Er:YAG laser in combination with perfluorodecaline produced precise ablation of the pig retina, which suggests the feasibility of this technique for safe ablation of epiretinal membranes.
| Introduction |
|---|
|
|
|---|
Currently, the common method of treatment is an attempt to remove these membranes from the retinal surface by engaging the edge and peeling it off with a microforceps.5 6 7 8 However, this technique frequently does not result in complete removal because the membrane strongly adheres to the retina. A retinectomy of the area surrounding the membrane is performed to release the tractional forces on the retina. When a retinectomy is done, a retinal opening is produced that exposes the vitreous cavity to pigment epithelial cells, which are a major cause of proliferative retinopathy.3 4 6 9 It would be desirable to be able to ablate membranes without exerting traction on the retina.
From 1994 to 1996, Brazitikos, DAmico, and associates performed extensive research on the application of the Er:YAG laser in retinal and vitreous surgery. In 1995, Brazitikos et al. published an outstanding paper on Er:YAG laser surgery of the vitreous and the retina and reported on the possible use of perfluorochemicals with the Er:YAG laser. They concluded that the Er:YAG laser is an effective tool for the complete transection of vitreous membranes as well as the creation of retinectomies in detached retina. In epiretinal membrane surgery some problems have been reported.10 11 12 13
However, the Er:YAG laser has been shown to produce precise tissue transection and ablation.14 15 16 17 18 19 The main problem in using infrared-laser (IR-laser) radiation in an aqueous environment such as the vitreous is the formation of fast expanding and subsequent collapsing water vapor bubbles at a submerged fiber tip and thereby the generation of strong pressure transients.20 21 22
The intention of the present study therefore was to systematically evaluate the effects of Er:YAG laser radiation on the pig retina using a perfluorodecaline/retina interphase.
| Materials and Methods |
|---|
|
|
|---|
|
All experiments were performed with a fluid/retina interphase. The eye cavity was filled with perfluorodecaline (DK-Line, Opsia Laboratoires) fluid, routinely used as an intraoperative nonpermanent vitreous substitute. It also has a low absorption coefficient at 2.94 µm (µa = 0.5 cm1) allowing the transmission of Er:YAG laser radiation.
Laser Experiments
The eye holder, with the eye cup, was centered on a rotating
aluminum plate that was attached to the axle of an electric motor
rotating at a speed of 1.88 rpm (Fig. 2)
. The laser delivery system was securely fastened to a metal arm above
the eye, while the tip of the probe was immersed in perfluorodecaline.
|
The macroscopic visible changes on the retina were recorded on videotape. In addition, fast flash video-shadow photography (illumination time t = 20 nsec) was used in a separate experiment (Fig. 3) to document the laser tissue interaction at the time of the laser impact on the retina. For that reason, the retina was cut into small flat pieces, necessary for taking images at the perfluorodecaline-retina interface.
|
Measurements of Distance between Fiber Tip and Retina
An identical setup environment was used in five different pig eyes
to determine variations in distance from the tip of the laser fiber to
the surface of the retina along the predetermined path of laser spots.
For these measurements the laser handpiece was replaced by a standard
ophthalmic A-scan biometry probe connected to a Technomed digital
A-scan A5500 ultrasound unit. The ultrasound probe was immersed into
the perfluorodecaline at a distance of 3 mm from the revolving axis of
the eye holder and approximately 1 mm above the retinal surface. A
total of 16 measurements were taken along the circular path per eye.
Sample Preparation
After laser irradiation, perfluorodecaline was removed and the
posterior eye segment submerged in a combination of 4% buffered
formaldehyde and 4% buffered glutaraldehyde (Karnowskys fixative) for
15 to 20 minutes to achieve rapid and proper fixation of the retina
followed by storage in 4% buffered formaldehyde for a minimum of 48
hours.
If artificial retinal detachment occurred, the specimen was embedded after prefixation in an agar gel using a 2.5% agar solution. To cover the eye on all sides, the segment was placed in a small metal container, and liquid agar was poured into the eye until the entire container was filled. After cooling, the agar hardened to a mechanically solid gel, and the specimen was placed in 5% buffered formaldehyde for a minimum of 48 hours.
For microscopic examination, a disc containing the optic nerve head, the posterior pole, and the circular path of laser applications was cut from the eye cup using a 13-mm trephine. The disc was then sectioned along the vertical diameter and stained with Masson trichrome.
Measurement of Ablation Depth
Ten to 12 light microscopic sections were made from each tissue
sample and examined under a routine light microscope (Olympus BH2).
Laser-induced changes were identified using low-power fields
(x4x10). Measurements of the horizontal extension of the tissue
ablation were performed with a measuring ocular (Olympus
WHK
x10/20 L) using a low-power field (x10). Measurement of the vertical
ablation depth was carried out using a drawing attachment (Olympus
BH2-DA). This allowed for the simultaneous observation of the
histologic section and a drawing tablet located next to the microscope.
An outline of characteristic retinal layers (internal limiting
membrane, ILM; outer nuclear layer, ONL) could be sketched on paper
(Fig. 4) . After calibration, the distance between the ILM and the ONL was
determined using a millimeter scale. For each slide,
sketches of normal retina and ablated retina were drawn, and 17
measurements 1-cm apart were taken for normal and ablated retinas. The
mean distances and the difference in the measurements between normal
and treated retina were calculated and mathematically transformed into
the retinal ablation depth. For each treatment group a total of 33
laser spots and corresponding normal retinal areas were measured and
compared.
|
| Results |
|---|
|
|
|---|
|
Accuracy of Light Microscopic Measurements
The mean distance between the ILM and the ONL was calculated to be
4.1829 ± 0.0192 mm, corresponding to a retinal distance of
103.2 ± 0.47 µm. The SEM was determined to be 0.212 µm.
Laser Experiments
The surface of the retina that underwent laser exposure changed
transparency to a whitish-gray discoloration. This effect became more
visible with higher energies. Using a high magnification (x6x7), it
was noticed that the surface of whitened areas appeared rough and
irregular. Small membranous flakes protruded from the surface of the
retina, whereas some areas exhibited a relatively smooth whitish
surface. Almost simultaneously with the laser pulse, small bubbles
visible to the naked eye emerged from the retina, ascending to the
surface of perfluorodecaline. Adhesion of such bubbles to the tip of
the fiber was sometimes followed by a more intense laser burst. During
our experiments the perfluorodecaline did not exhibit noticeable
clouding or an accumulation of bubbles at the airperfluorodecaline
interphase. Visibility of the retinal surface was unimpaired at all
times.
The fast flash photography revealed the creation of a vapor bubble at the surface of the retina (Fig. 6) . Occasionally, a vapor bubble was noticed within the perfluorodecaline liquid projecting from the tip of the fiber toward the retinal surface. The bubbles were relatively small and their expansion and collapse slow. No pressure transients were measurable even at radiant exposures of 10 J/cm2. In case perfluorodecaline was replaced by water, pressure transients between 45 ± 5 bars (1 J/cm2) and 132 ± 20 bars (10 J/cm2) visualizing the collapse of the water vapor channel were recorded (Fig. 7) .
|
|
|
|
| Discussion |
|---|
|
|
|---|
The introduction of new laser instruments seems to be an appealing approach to improve vitreoretinal surgery and promises to overcome several problems that are related to conventional instruments. Lasers operating in the infrared part of the spectrum have several advantages over ultraviolet lasers used for retinal ablation in the past.24 25 IR-lasers are solid state, rendering them more compact and less costly to operate than excimer lasers.26 In addition, they do not inherit the potential risk of mutagenic effects compared with ultraviolet wavelengths.
The wavelength of an Er:YAG laser emitting at 2.94 µm corresponds to
a maximum absorption peak of water that was shown to be around 10,000
cm1. Such high absorption translates into an
optical penetration depth in water of 1 µm, making it impossible to
transmit Er:YAG radiation through the cornea and the vitreous. Because
of the high absorption in water, a fiber delivery system is mandatory
to irradiate retinal tissue. Fiber delivery systems suitable for
infrared radiation are sapphire and zirconium fluoride. Both materials
can be built into a flexible fiber. A fiber delivery system using a
flexible zirconium fluoride fiber guarded by a low OH quartz at the
distal tip combines an acceptable rate of transmission (
70% through
a fiber 2 m in length).
Laser Experiments
We performed Er:YAG laser experiments on pig retinas to examine
the efficacy and accuracy of partial retinal surface ablation using a
flexible ZrF4 fiber, and perfluorodecaline as
transmitting medium. The retina was selected as a model tissue to
examine potential applications of the Er:YAG laser for vitreoretinal
surgery. Ablation of the inner layer of the retina provided a useful
model for the ablation of thin epiretinal membranes.
Perfluorodecaline was selected as a transmitting fluid for two major reasons. First, many cases of vitreoretinal surgery, in particular complicated cases, are performed using perfluorodecaline as a temporary vitreous substitute. In these cases perfluorodecaline is used to facilitate surgical maneuvers on a mobile detached retina. Second, perfluorocarbon compounds such as perfluorodecaline exhibit an acceptable transmission rate for infrared radiation (absorption coefficient µa [2,94 µm] = 0.5 cm1) and can therefore be used as a transmitting medium for Er:YAG radiation.27 This is especially important because the Er:YAG laser radiation causes a violently expanding and subsequently collapsing water vapor bubble at the distal fiber tip in an aqueous environment.20 The formation and the collapse of the vapor bubble have been shown to be associated with the emission of strong pressure transients (Fig. 7) . However, in water this vapor bubble is essential for a successful ablation of tissue because it establishes a delivery path for the highly absorbed infrared radiation through the water.28
To avoid any vapor bubbleinduced damage to the retina, the aqueous environment was replaced by perfluorodecaline. In fact, the flash photography revealed only a small vapor bubble on the surface of the retina, and pressure recordings within perfluorodecaline showed amplitudes of less than 100 mbar at a distance of 2 mm away from the fiber tip. The creation and the collapse of these small bubbles do not correspond to the bubbles created in water.27 Instead, they are characterized by the disruption and vaporization of tissue and expulsion from the surface of the retina. The creation and collapse of these bubbles is much slower than those seen in an aqueous environment, supporting the fact that no pressure transients were measurable.
Despite these advantages, several disadvantages of perfluorodecaline have been reported in the literature,13 including damage to the underlying retina in an attempt to dissect epiretinal membranes and rapid opacification of the perfluorodecaline liquid due to circulating ablation debris and accumulation of bubbles. Our experiments did not exhibit any noticeable clouding of perfluorodecaline. To the contrary, visibility of the retina was unimpaired at all times, and the transmission of perfluorodecaline used even after multiple extensive ablations decreased only to a small extent. Although bubble formation was noted with laser bursts at the surface of the retina, accumulation did not occur in our experiments.
Sample Preparation
Fixation of laser-exposed posterior pig eye segments in
Karnowskys fixative proved to have several advantages over the
routinely used formaldehyde fixative. Benefits were a faster fixation
of the retina and favorable staining characteristics when used in
combination with Masson trichrome stain. The rapid fixation of the
retinal tissue was able to reduce artificial retinal detachment. In
addition, the internal vitreoretinal border region with the ILM and the
radial fibers of the Mueller cells were much more clearly visible with
the Masson trichrome stain after fixation in Karnowskys fixative
compared with formaldehyde fixation.
Agar gel provided some advantages in the processing of retinal tissue where retinal detachment occurred. Besides the fact that it is readily available and inexpensive, agar gel behaves identically to real tissue during routine tissue processing but does not stain with most widely used staining techniques (i.e., hematoxylin and eosin, periodic acidSchiff, and Masson trichrome). It is therefore possible to process the entire tissue-agar sample in the same way one would process the tissue sample itself.
Light Microscopy and Ablation Depth
In light microscopy sections, multiple artifacts might have been
produced during histologic processing, mounting, and staining of the
tissue samples. Not only were there differences in thickness of
laser-exposed areas but also normal retina displayed great differences
in thickness of different layers.
However, the radial fibers of the Mueller cells and the ILM were clearly visible. Thus, miniature changes in the appearance of the ILM could be easily demonstrated. One scale unit of the measuring eyepiece represented a distance of 2.5 µm with the high-power magnification (x40).
Laser ablation of retinal tissue increased with the radiant exposure in a linear fashion. The graphical presentation of our data shows a relatively wide range of tissue ablation for each individual radiant exposure. This could lead to the conclusion of an unprecise ablation of the Er:YAG laser. However, several factors contributed to the wide range of ablation depths.
First, a major impact is related to the distance between the fiber tip and the retina. Although we were able to demonstrate that the variation of the distance could be maintained between 1.0 ± 0.3 mm, there is still a variation of the radiant exposure of up to ±13%.
Second, miniature vitreous remnants remaining on the retinal surface block infrared wavelengths and reduce the measurable effect on the retina.
An interesting finding that was encountered when histologic sections were studied was that outer photoreceptor segments were distorted and diverted in one direction after irradiation with radiant exposures exceeding 3 J/cm2. This was commonly observed at the peripheral region of laser-induced areas of retinal ablation. Possible explanations for this phenomenon could be a temperature-induced coagulation or a pressure-induced mechanical distortion of the outer segments. However, a temperature effect seems rather unlikely because photoreceptor outer segments in the center of laser spots do not exhibit such changes. The fact that pressure transients could not be detected within perfluorodecaline does not exclude the possibility of a more static pressure force of rather low amplitude originating from the bubble expansion. Whether this phenomenon is associated with permanent damage of the photoreceptor function is the subject of further in vivo studies.
| Conclusion |
|---|
|
|
|---|
Our findings suggest that a radiant exposure between 1 and 3 J/cm2 in combination with a pulse duration of 250 µsec and a proberetinal distance of 1 mm are suitable for precise ablation of retinal tissue.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication April 26, 1999; revised September 15, 1999; accepted September 23, 1999.
Commercial relationships policy: Cc5.
Corresponding author: Peter Janknecht, Department of Ophthalmology, University of Freiburg, Kilianstrasse 5, D-79106 Freiburg, Germany. janknecht{at}aug.ukl.uni-freiburg.de