(Investigative Ophthalmology and Visual Science. 2000;41:1229-1238.)
© 2000
by The Association for Research in Vision and Ophthalmology, Inc.
Pupil Perimetry using M-Sequence Stimulation Technique
Helmut Wilhelm,
Jürgen Neitzel,
Barbara Wilhelm,
Stefan Beuel,
Holger Lüdtke,
Ulf Kretschmann and
Eberhart Zrenner
From the Universitäts-Augenklinik, Department of Pathophysiology of Vision and Neuro-ophthalmology, Tübingen, Germany.
 |
Abstract
|
|---|
PURPOSE. M-sequence stimulation technique allows mapping of the retinal function
by multifocal electroretinographic (ERG) recordings. However, the
information provided about visual field is limited to retinal function.
Optic nerve diseases and diseases of the higher visual pathways usually
show normal multifocal ERGs. Using pupillary responses instead of the
electrical retinal responses might enhance the diagnostic possibilities
of this system. The problems of local ERG recordings are very similar
to those encountered in pupil perimetry: Local stimuli have to be dim
to avoid or at least reduce stray-light responses. Dim stimuli, close
to the absolute threshold, elicit only subtle pupillomotor responses.
Therefore, techniques that are able to detect small focal responses are
promising.
METHODS. Pupillography was done by means of an infrared video camera and real
time image processing (50 Hz) using a custom-designed videoboard in a
personal computer (486). Recording conditions: The stimulus was
presented on a monitor (75 Hz) in 26 cm distance from the patients
eyes. It contained 37 hexagons in a 25° visual field. Each element
changed between black (1.6 cd/m2) and white (160
cd/m2) after a binary M-sequence independently from other
elements. Four thousand ninety six different stimulus pictures of
120-msec duration were shown during a single pupillogram recording.
Thirty-seven local pupillograms were calculated in a cross-correlation
of stimulus sequence and the pupil diameter.
RESULTS. The pupillomotor fields in normals showed a shape and sensitivity
distribution as known from conventional pupil perimetry techniques.
Artificial paracentral scotomas (5°) created by masking different
locations could be demonstrated convincingly. Even in patients with
optic nerve lesions it was possible to demonstrate visual field
defects.
CONCLUSIONS. Pupil perimetry using the M-sequence technique is a promising method of
objective perimetry that may find its entrance into clinical
application.
 |
Introduction
|
|---|
Visual field testing is an important method in ophthalmology and
neurology, both for focal diagnosis and follow-up in diseases of the
eyes and the visual pathways. However, perimetry remains a subjective
technique even if performed by automated systems using sophisticated
stimulus presentation and response evaluation algorithms. Objective
perimetry is a strong demand, not only to disprove malingering but also
to examine patients who have difficulty cooperating with conventional
perimetry. Several attempts have been made using the pupil light reflex
as response to a local perimetric stimulus.1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
None of
those techniques has come into widespread use for different reasons.
Pupillomotor threshold using conventional perimetric stimuli is higher
than visual threshold,19
and it varies considerably
between individuals. Additionally, intraindividual variability exists.
In most individuals stimulus intensity has to exceed the visual
threshold considerably to provide a stable and repeatable pupillary
response. On the other hand, brighter stimuli increase stray light and
this limits the maximal level of stimulus brightness, because the local
response is replaced by a stray-light response. In quite a few patients
with visual field defects in conventional perimetry proven by objective
morphologic findings, pupil perimetry fails to demonstrate a scotoma.
If the stimulus is chosen too dim, it may remain below the pupillomotor
threshold; if it is too bright, it will elicit a stray-light response.
It is virtually impossible to find for each individual and each retinal
location a stimulus that elicits a local response without an additional
stray-light effect. Techniques that help to extract local responses
from stray-light "noise" are therefore promising.
These problems encountered in pupil perimetry are similar to those met
in focal electroretinography (ERG). Sutter and Tran have provided a new
solution introducing M-sequence stimulation
technique.20
21
This visual evoked response imaging system
(VERIS) has become a valuable tool in multifocal ERG recordings
providing detailed functional topography in retinal
disorders.22
23
24
It seemed to be logical to use the M-sequence system for pupil
perimetry as well. Instead of feeding an ERG curve to the system, pupil
size in millimeters coded as a voltage (1 V = 1 mm) was used. The
purpose of this article is to describe the methodology and first
applications in normals, artificial scotoma, and patients.
 |
Methods
|
|---|
Stimulation
The stimulus is presented on a black and white monitor (75 Hz) at
a 26 cm distance from the patients eye, and consists of 37 hexagonal
elements within a 25° (radius, see Fig. 1 ). Each element changes its luminance between 1.6
cd/m2 (black) and 160 cd/m2
(white) after a binary M-sequence independently of other elements. In
the center of the monitor a constant visible gray spot serves as a
fixation target. The background of the stimulus-screen is otherwise
black (1.6 cd/m2).

View larger version (31K):
[in this window]
[in a new window]
|
Figure 1. (A) Geometry of the test field with 37 stimulated areas.
(B) Multifocal pupillogram recorded from the right eye of a
normal volunteer. (C) For further analysis the
responses were grouped either into 4 quadrants or into 5 concentric
rings.
|
|
During a single recording 4096 different stimulus pictures were
presented. Each stimulus picture was shown for the duration of nine
monitor frames (120 msec). This does not mean that the complete
stimulus was flickering with such a high frequency. Each single
location of the stimulus changes considerably slower (i.e., each
location remains unchanged for several frames). Our setup led to a
total stimulation time of 8.19 minutes. This net time was divided into
40 segments. Between the segments the subjects were allowed to blink
and move their eyes. If blinks and movements occurred during a segment,
the data were discarded and the segment repeated. Before each segment
the stimulus was presented for 960 msec without measuring the pupil
diameter to allow the subject to find a stable fixation. The average
time for the whole test was 30 minutes. Table 1
compares the setup parameters of ERG and pupil perimetry recording.
Pupillography
The pupil of the stimulated eye was recorded by means of an
infrared-sensitive CCD-camera, while the contralateral eye was
occluded. The video signal was processed by a custom-designed real-time
frame-grabber board occupying the extension slot of a personal
computer. The pupil was tracked horizontally with a frequency of 50 Hz.
Pupil diameter was then converted into proportional voltage changes
(010 V, 1 V for 1 mm). Instead of an ERG signal, this pupil diameter
signal was transferred to a Macintosh Quadra 650 computer, harboring
the VERIS software (provided by Erich Sutter, EDI, San Francisco, CA).
It was stored as a digital signal each 6.67 msec.
Signal Analysis
The pupillary light response was analyzed for 37 stimulated areas
with a fast M-transform algorithm, which generates the first order
kernel, a linear approximation of the pupillary response. Amplitudes
and peak times (i.e., time from stimulus onset to maximum constriction)
were considered for evaluation. It has to be kept in mind that the
latencies calculated by the VERIS software are not pupillographic
latencies because the peak of the pupillary constriction is taken as
the end point of the latency time, not the beginning of the
constriction as usually occurs in pupillography. Therefore it is better
to speak of peak time.
Experiments
Eleven healthy subjects (mean age: 36.8 ± 12.6 years) were
examined by this method to test how the retinal pupillomotor
sensitivity would be displayed by this method. They had given their
informed consent and all experiments were done according to the
Declaration of Helsinki.
Parts of the screen had been covered by black cardboard in one
stimulation. In other subjects one element had been switched off (i.e.,
it remained black during the measurement).
Altogether 6 patients who had given their informed consent were
examined. They had been selected as representative of different types
of visual field defects and according to their ability to perform the
pupil perimetry. In 1 of them the record could not be completed because
of blinks and unstable fixation, and in 2 of them (advanced glaucoma
and advanced retinitis pigmentosa) the resulting pupillograms were
extremely small, hardly distinguishable from noise. Three patients
(optic neuritis, chiasmal lesion, and bilateral occipital lesion) are
demonstrated here.
Conventional perimetry was performed using the Tuebingen Automated
Perimeter (suprathreshold technique, 190 test locations in the 30°
field = program 1). Stimulus size was 10', duration 100 ms,
luminance 1000 cd/m2, and 10 cd/m2
background. The shaded areas in Figures 5 6
7
give the areas in
which the 1000 cd/m2 stimulus was not seen.

View larger version (22K):
[in this window]
[in a new window]
|
Figure 5. Patient with optic neuritis and central scotoma left eye.
(A) Multifocal pupillograms. The central pupillogram is
markedly reduced compared with normals (Fig. 2)
. The scotoma is shown
as a gray shaded area. (B) Averaged
pupillograms from 4 rings as indicated in Figure 1C
. Much lower
sensitivity especially in the center compared with normals (see Figs. 4C
and 4D
).
|
|

View larger version (40K):
[in this window]
[in a new window]
|
Figure 6. Patient with pituitary adenoma and temporal hemifield loss. The scotoma
is shown as a gray shaded area. (A)
Multifocal pupillograms: pupillary response is reduced in comparison to
the nasal field. (B) The same can be demonstrated by
averaged pupillograms from 4 quadrants as indicated in Figure 1C
. The
temporal quadrants that are usually more sensitive than the nasal show
reduced sensitivity.
|
|

View larger version (41K):
[in this window]
[in a new window]
|
Figure 7. In a patient with so-called checkerboard visual field defect after
bilateral occipital lobe infarction, it was not possible to demonstrate
the field defect convincingly. However, the pupillary responses were
relatively low in general. The visual field defect of the right eye is
shown as a gray shaded area. (A) Multifocal
pupillograms. (B) Averaged pupillograms in 4 quadrants.
|
|
 |
Results
|
|---|
The resulting 37 pupillographic traces are shown as a trace array,
each response positioned at its approximate place of origin in the
visual field. Amplitudes and latencies at the trough of the
pupillographic trace were analyzed off-line.
Multifocal pupillograms could be extracted (Fig. 1B)
. The averaged
pupillogram of the whole field showed an amplitude of approximately 1
mm (Fig. 2)
.

View larger version (10K):
[in this window]
[in a new window]
|
Figure 2. Averaged pupillogram of the whole field. Temporal and spatial
parameters of the pupillogram are similar to conventionally recorded
pupillograms; however, the shape of the pupillogram is altered by
averaging many pupillograms with small amplitudes and short peak
times.
|
|
If part of the screen was covered by cardboard, the
pupillographic response was absent or markedly reduced in this area. The same happened, if one hexagon remained black (Fig. 3)
.

View larger version (17K):
[in this window]
[in a new window]
|
Figure 3. (A) When peripheral areas of the visual field are covered by
black cardboard, the pupillary response is absent or reduced for this
area. (B) The same is found when one hexagon remains black
during recording.
|
|
It could be shown that the pupillographic response is highest in the
center and lowest at the edge of the examined field. This difference
was very pronounced. The upper temporal quadrant of the visual field is
the most sensitive quadrant. The temporal half of the visual field is
more sensitive than the nasal half. This is true for both eyes (Fig. 4)
.


View larger version (38K):
[in this window]
[in a new window]
|
Figure 4. (A) Example of one normal subject: averaged pupillograms
from 4 quadrants as indicated in Figure 1C
. (B) Mean results
of 11 normal subjects for visual field quadrants (constriction
amplitude in micrometers per degree squared). The temporal quadrants
are more sensitive than the nasal quadrants. (C) Example of
one normal subject: averaged pupillograms from 4 rings as indicated in
Figure 1C
. (D) Mean results of 11 normal subjects for
different eccentricities (mean of both eyes). The pupillomotor
sensitivity is much higher in the center than in the periphery.
|
|
In an optic neuritis patient with central scotoma the central pupil
response was decreased and the peak time prolonged (Fig. 5) . In a patient with temporal field loss the pupillary defect could also
be shown (Fig. 6)
, but the method failed in a patient with occipital lobe lesion (Fig. 7)
.
 |
Discussion
|
|---|
The linear approximation of the multifocal ERG (first order
kernel) provides a reliable functional topography of the lesion in
diseases of the outer retina (i.e., Stargardts macular dystrophy and
retinitis pigmentosa),25
but the amplitudes were found
normal in patients with optic nerve or higher visual pathway
diseases.26
Recently, Bearse and coworkers27
described a ganglion cell component extracted from the multifocal
ERG. This component is too small to allow an objective perimetry in
high resolution at the present state of development.
Visual evoked cortical potential (VECP) might be used for this purpose.
However, the VECPs do not depend only on the position of the
stimulus in the visual field but also on the spatial relation between
the cortical tissue in which the potentials are generated and the
position of the recording electrodes on the scalp. Barrett et
al.28
described the phenomenon of paradoxical
lateralization of the VECP, that is, during hemifield stimulation
potentials recorded over the ipsilateral occipital cortex were found to
be larger than over the contralateral cortex as expected. With VECP
recordings using the M-sequence stimulation technique cortical
potentials evoked from different positions of the visual field were
recorded simultaneously by Baseler and Sutter.29
But even
in normal volunteers the recordings revealed a "Swiss cheese
pattern" most probably due to cancellation of local responses
generated in neighboring convolutions of the visual cortex. Kilstorner
et al.30
obtained a good correspondence between Humphrey
visual field and multifocal VECP using bipolar electrodes and
evaluating upper and lower fields separately.
Pupillography as another objective method of visual testing deserves a
closer look, and the first attempts have been made by Sutter
himself.31
His results were promising, and therefore we
decided to apply this method.
VERIS pupil perimetry proved to be possible. In normals, we obtained
principally the same qualitative results as Kardon and
Thompson32
with classic pupil perimetry using a Humphrey
Field Analyzer equipped with pupillography. They found the highest
response in the center of the visual field, markedly decreasing in the
periphery, the same that we found using the M-sequence technique. This
has also been visible on Sutters first pupillographic
fields.31
In former experiments using conventional pupil
perimetry the upper temporal quadrant of the visual field showed the
greatest, and the lower nasal quadrant the least, pupillomotor
sensitivity.7
We found the same results concerning the
most sensitive quadrant for both eyes.
When all 37 focal pupillograms were averaged, the resultant summed
pupillogram was very similar to a pupillogram elicited by a single
light stimulus (Fig. 2)
. Amplitude and latency time were in the same
range. However, an averaged pupillogram looks different than a single
flash response: If many small pupillograms are superimposed, the
latency time will be somewhat shorter, the peak time will remain
approximately the same as in a single flash pupillogram, and the
amplitude depends on the number and amplitude of the focal
pupillograms. This explains why the summed pupillogram does not look
exactly like a single flash response.
An experiment was done with one or more hexagons of the stimulus field
covered to demonstrate that the technique is really able to provide
local pupil responses. It has been possible to demonstrate that
nonstimulated areas do not show a local pupil response.
There are important differences between electrophysiological recordings
and pupillography. In ERG recordings sampling rates of more than 1000
Hz are possible,22
whereas in infrared videopupillography
frame rates of 25 or 50 Hz (30 or 60 Hz, respectively) are used.
Because the pupillary movements are comparatively slow and do not
contain high frequency components, this sampling rate is appropriate
for pupillography. On the other hand, even the stimulus modulation has
to be slowed down. The pupil is much slower to follow light responses
than the retina, because its flicker fusion frequency is as low as 3
Hz.33
Therefore, frequency of the local light flux change
must meet the temporal bandwidth of the pupil (Table 1) . This slows
down the procedure, and the examination time would increase to several
hours if we tried to reach the same quality as in ERG recordings. We
managed to obtain reliable results with recording times around 8
minutes per eye. This recording was split into 40 segments,
because artifacts such as blinks or fixation loss would deteriorate the
results. Many segments had to be repeated, therefore the total
examination time was approximately 30 minutes. Changes caused by
fatigue may begin to interfere with the light responses.34
Despite those problems it has been possible to apply the method
in selected patients and to demonstrate their field defects. In two
patients with a central scotoma caused by optic neuritis the foveal
response was markedly reduced corresponding to the field defect. Why is
it not missing completely as the ERG response in macular
lesions?23
We have to keep in mind that conventional
perimetric stimuli are much smaller than pupil perimetric stimuli. In
our cases 10' stimuli were used for conventional perimetry. Dense
scotomas may contain small partially functional islands that may be
stimulated by the large pupillomotor stimuli. We still do not know much
about pupillomotor receptive fields and the contribution of different
channels of the visual system to the pupillary light response.
The VERIS pupil perimetry still depends very much on the subjects
cooperation, therefore it was not always possible to demonstrate field
defects convincingly, mostly because of too much noise. Artifact
management might be a solution, change of the stimulus field to a lower
number of hexagons another. This leads of course to a reduction of the
spatial resolution; however, this might not be a major problem. The
purpose of pupil perimetry as a clinical application is not to replace
conventional perimetry but simply to verify defects (e.g., in suspected
malingering). Usually, it is not the question of small scotomas that
could be feigned but large hemifield or quadrant defects or severe
constriction. Under these circumstances a lower spatial resolution may
be acceptable.
However, the recording of the patient with the occipital lobe lesion
was technically good, but the defect did not show up convincingly, even
the blind upper temporal quadrant showed the best response. One
question still remains to be answered: Does VERIS pupil perimetry
measure the very same visual function as conventional pupil perimetry?
The pupil reacts not only to light but also to changing gratings and
movement.35
In VERIS pupil perimetry light responses and
responses to more complex stimuli may be combined. This might
theoretically open new possibilities, because there are conditions
where light responses and responses to complex stimuli are differently
involved.36
The very central visual field has an
especially high pupillomotor sensitivity, and defects in this area show
clearly. Maybe the stimulation applied here is comparable to the
checkerboard stimulation technique used by Slooter and van
Norren37
who were able to measure visual acuity by means
of pupillography. If the pupil response to M-sequence stimulation was
purely a pattern response, it would not be very promising to try to
find visual field defects. However, our experiments with artificial
scotoma showed that the pupil response is absent in distinct areas if
they are not stimulated. If we cannot detect scotomas in patients this
cannot be explained by the inability of the VERIS system to extract
local responses. Cooperation problems or a dissociation of the visual
and the pupillomotor response may cause failure of the VERIS pupil
perimetry. This has to be differentiated in further studies with
simplified technique and shorter examination times.
Another question related to this problem is whether pupil perimetry
really tests the complete visual pathways. It was previously thought
that the pupillary light reflex depends completely on subcortical
structures (i.e., will not be impaired in retrogeniculate lesions of
the visual system).38
39
However, there are many hints
that the pupillary light reflex may be disturbed even in
retrogeniculate lesions of the visual pathways and even in cortical
lesions.5
11
13
40
41
42
It is not yet clear whether the
pupillary light reflex is really completely absent or reduced in such
cases with retrogeniculate lesions, or whether certain components of
this reflex are spared.36
As an objective method pupil
perimetry in general might have some limitations in patients with
isolated retrogeniculate lesions. Currently, no exact data exist about
the sensitivity of pupil perimetry for lesions of the higher visual
pathways, but there is evidence that visual field defects caused by
such lesions may be demonstrated even pupillographically in most cases.
This is what we would have expected in our patient, too.
Further technical improvements of pupillography will probably lead to a
more widespread use of pupil perimetry. The VERIS pupil perimetry is a
promising new method of objective perimetry that deserves further
study.
 |
Acknowledgements
|
|---|
The authors thank Erich Sutter for his invaluable help during the
setup and first experiments with the M-sequence pupil perimetry.
 |
Footnotes
|
|---|
Submitted for publication May 7, 1999; revised September 28, 1999; accepted October 27, 1999.
Commercial relationships policy: N.
Corresponding author: Helmut Wilhelm, Universitäts-Augenklinik, Department of Pathophysiology of Vision and Neuro-ophthalmology, D 72076 Tübingen, Germany. helmut.wilhelm{at}uni-tuebingen.de
 |
References
|
|---|
-
Enoksson, P, Hallen, B. (1970) Pupillomotoric perimetry in psychogenic defects of the visual fields Acta Ophthalmol Copenh 48,672-676[Medline][Order article via Infotrieve]
-
Alexandridis, E, Krastel, H. (1989) Pupillographische Perimetrie mit dem "Octopus." Fortschr Ophthalmol 86,692-694[Medline][Order article via Infotrieve]
-
Bresky, RH, Charles, S. (1969) Pupil motor perimetry Am J Ophthalmol 68,108-112[Medline][Order article via Infotrieve]
-
Fankhauser, FD, Flammer, J. (1989) Puptrak 1.0a new semiautomated system for pupillometry with the Octopus perimeter: a preliminary report Doc Ophthalmol 73,235-248[Medline][Order article via Infotrieve]
-
Hamann, KU, Hellner, KA, Müller, JA, Zschocke, S. (1979) Videopupillographic and VER investigations in patients with congenital and acquired lesions of the optic radiation Ophthalmologica 178,348-356[Medline][Order article via Infotrieve]
-
Hellner, KA, Jensen, W, MüllerJensen, A. (1978) Fernsehbildanalytische pupillographische Perimetrie bei Hemianopsie Klin Monatsbl Augenheilkd 172,731-735[Medline][Order article via Infotrieve]
-
Kardon, RH, Kirkali, PA, Thompson, HS (1991) Automated pupil perimetry: pupil field mapping in patients and normal subjects Ophthalmology 98,485-495[Medline][Order article via Infotrieve]
-
Krastel, H. (1975) Pupillographische Prüfung der Spektralsensitivität umschriebener Netzhautbezirke Ber Zusammenkunft Dtsch Ophthalmol Ges 73,176-181
-
Roth, HJ, Alexandridis, E. (1975) Objektive Perimetrie mittels IR-reflexionsphotometrischer Registrierung des indirekten Pupillenlichtreflexes Ber Zusammenkunft Dtsch Ophthalmol Ges 73,303-306
-
Trimarchi, F, Baldini, M, Bianchi, PE, Gelmi, C, Franchini, F. (1979) Objective campimetry: new approach to optic pathway lesions Acta Neurochir Suppl Wien 28,340-343[Medline][Order article via Infotrieve]
-
Kardon, RH (1992) Pupil perimetry. Editorial review Curr Opin Ophthalmol 3,565-570[Medline][Order article via Infotrieve]
-
Frydrychowicz, G, Harms, H. (1940) Ergebnisse pupillomotorischer Untersuchungen bei Gesunden und Kranken Verh Dtsch Ophthalmol Ges 53,71-80
-
Alexandridis, E, Krastel, H, Reuther, R. (1979) Pupillenreflexstörungen bei Läsionen der oberen Sehbahn Albrecht Von Graefes Arch Klin Exp Ophthalmol 209,199-208[Medline][Order article via Infotrieve]
-
Jensen, W. (1976) Die Fernsehbildanalyse. Ein Messverfahren zur objektiven Perimetrie Albrecht Von Graefes Arch Klin Exp Ophthalmol 201,183-191[Medline][Order article via Infotrieve]
-
Harms, H, Aulhorn, E, Ksinsik, R. (1973) Die Ergebnisse pupillomotorischer Perimetrie bei Sehhirnverletzten und die Vorstellungen über den Verlauf der Pupillenlichtreflexbahn Dodt, E Schrader, KE eds. Die normale und die gestörte Pupillenbewegung ,72-82 Verlag J. F. Bergmann München.
-
Wilhelm, H. (1996) Pupille und retrogenikuläre Sehbahn Ophthalmologe 93,319-324[Medline][Order article via Infotrieve]
-
Turtschi, S, Bergamin, O, Dubler, B, Schötzau, A, Zulauf, M. (1994) Pupillenperimetrie mit dem OCTOPUS 12-3 Erste Erfahrungen Klin Monatsbl Augenheilkd 204,398-399
-
Barbur, JL, Watson, JDG, Frackowiak, RSJ, Zeki, S. (1993) Conscious visual perception without V1 Brain 116,1293-1302[Abstract/Free Full Text]
-
Alexandridis, E. (1968) Bestimmung der Dunkeladaptationskurve mit Hilfe der Pupillenlichtreflexe Ber Zusammenkunft Dtsch Ophthalmol Ges 68,274-277[Medline][Order article via Infotrieve]
-
Sutter, EE, Tran, D (1992) The field topography of ERG components in man. I. The photopic luminance response Vision Res 32,443-446
-
Sutter, EE (1991) The fast M-transform: a fast computation of cross-correlations with binary m-sequences SIAM J Comput 20,686-694
-
Kretschmann, U, Gendo, K, Seeliger, M, Zrenner, E. (1997) Multifocal ERG recording by the VERIS technique and its clinical applications Wiedeman, P Kohen, L eds. Macular and retinal diseases. Dev Ophthalmol Vol. 29,8-14 Karger Basel.
-
Kretschmann, U, Seeliger, M, Ruether, K, Usui, T, Zrenner, E. (1998) Spatial cone activity distribution in diseases of the posterior pole using a multi-input stimulation technique for mapping of the retinal function in the central visual field by multifocal electroretinography Vision Res 38,3817-3828[Medline][Order article via Infotrieve]
-
Palmowski, AB, Sutter, EE, Bearse, MAJ, Fung, W. (1997) Mapping of retinal function using the multifocal electroretinogram Invest Ophthalmol Vis Sci 38,2586-2590[Abstract/Free Full Text]
-
Kretschmann, U, Seeliger, M, Ruether, K, Usui, T, ApfelstedtSylla, E, Zrenner, E. (1998) Multifocal electroretinography in patients with Stargardts macular dystrophy Br J Ophthalmol 82,267-275[Abstract/Free Full Text]
-
Kretschmann, U, Gendo, K, Wilhelm, H, Schiefer, U, Hettesheimer, H, Zrenner, E. (1998) Objektivierung von Gesichtsfelddefekten mit Hilfe der multifokalen Elektroretinographie Klin Monatsbl Augenheilkd 212,40-49[Medline][Order article via Infotrieve]
-
Bearse, MAj, Sim, D, Sutter, EE, Stamper, R, Leiberman, M. (1996) Application of the multi-focal ERG to glaucoma [ARVO Abstract] Invest Ophthalmol Vis Sci 37(3),S511Abstract nr 2342
-
Barrett, G, Blumhardt, L, Halliday, AM, Halliday, I, Kriss, A. (1976) A paradox in the lateralization of the visual evoked response Nature 261,253-255[Medline][Order article via Infotrieve]
-
Baseler, HA, Sutter, EE (1997) M and P components of the VEP and their visual field distribution Vision Res 37,675-690[Medline][Order article via Infotrieve]
-
Kilstorner, AI, Graham, SL, Grigg, JR, Billson, FA (1998) Multifocal topographic visual evoked potential: improving objective detection of local visual field defects Invest Ophthalmol Vis Sci 39,937-950[Abstract/Free Full Text]
-
Sutter, EE. (1996) The field topography of the pupillary response: rod and cone inputs [ARVO Abstract] Invest Ophthalmol Vis Sci 37(3),S691
-
Kardon, RH, Thompson, S. (1995) Pupil perimetry: methods of threshold determination and comparison with visual responses Mills, RP Wall, M eds. Proceedings of the XI International Perimetric Society Meeting Kugler Amsterdam.
-
Alexandridis, E, Manner, M. (1977) Folgefrequenz der Pupille bei flimmernden Lichtreizen Albrecht Von Graefes Arch Klin Exp Ophthalmol 202,175-180[Medline][Order article via Infotrieve]
-
Lowenstein, O, Feinberg, R, Loewenfeld, IE (1963) Pupillary movements during acute and chronic fatigue Invest Ophthalmol 2,138-157
-
Barbur, JL, Harlow, AJ, Sahraie, A. (1992) Pupil response to stimulus structure, colour and movement Ophthalmic Physiol Opt 12,137-141[Medline][Order article via Infotrieve]
-
Barbur, JL, Harlow, AJ, Weisskrantz, L. (1994) Spatial and temporal response properties of residual vision in a case of hemianopia Phil Trans R Soc Lond B 343,157-166[Abstract/Free Full Text]
-
Slooter, J, van Norren, D. (1980) Visual acuity measured with pupil responses to checkerboard stimuli Invest Ophthalmol Vis Sci 19,105-108[Abstract/Free Full Text]
-
Wilhelm, H, Kardon, RH (1997) The pupillary light reflex pathway (editorial) Neuro-ophthalmology 17,59-62
-
Loewenfeld, IE. (1993) The pupil Anatomy, Physiology and Clinical Applications Wayne State University Press Detroit.
-
Cibis, GW, Campos, EC, Aulhorn, E. (1975) Pupillary hemiakinesia in suprageniculate lesions Arch Ophthalmol 93,1322-1327[Abstract/Free Full Text]
-
Harms, H. (1951) Hemianopische Pupillenstarre Klin Mbl Augenheilk 118,133-147
-
Körner, F, Teuber, HL (1973) Visual defects after missile injuries to the geniculate pathway in man Exp Brain Res 18,88-113[Medline][Order article via Infotrieve]
This article has been cited by other articles:

|
 |

|
 |
 
B. Link, A. Junemann, R. Rix, O. Sembritzki, A. Brenning, M. Korth, and F. K. Horn
Pupillographic Measurements with Pattern Stimulation: The Pupil's Response in Normal Subjects and First Measurements in Glaucoma Patients
Invest. Ophthalmol. Vis. Sci.,
November 1, 2006;
47(11):
4947 - 4955.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M Bach
Latency of the mfVEP to diagnose glaucoma?
Br. J. Ophthalmol.,
September 1, 2006;
90(9):
1076 - 1077.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Stroux, M. Korth, A. Junemann, J. B. Jonas, F. Horn, A. Ziegler, and P. Martus
A Statistical Model for the Evaluation of Sensory Tests in Glaucoma, Depending on Optic Disc Damage
Invest. Ophthalmol. Vis. Sci.,
July 1, 2003;
44(7):
2879 - 2884.
[Abstract]
[Full Text]
[PDF]
|
 |
|