(Investigative Ophthalmology and Visual Science. 2001;42:957-965.)
© 2001
by The Association for Research in Vision and Ophthalmology, Inc.
Comparison of Pupil Perimetry and Visual Perimetry in Normal Eyes: Decibel Sensitivity and Variability
Sungpyo Hong1,
Joanna Narkiewicz2 and
Randy H. Kardon3
1 From the Department of Ophthalmology, Kyungpook National University Hospital, Taegu, Korea;
2 Department of Ophthalmology, St. Adalberts Hospital, Gdansk-Zaspa, Poland; and
3 Department of Ophthalmology and Visual Science, University of Iowa Hospital and Clinic, Iowa City.
 |
Abstract
|
|---|
PURPOSE. To compare the sensitivity and variability of pupil perimetry with
visual perimetry at the same retinal locations in normal subjects.
METHODS. Pupil perimetry was performed on the right and left eyes of 10 normal
subjects using a computerized infrared pupillometer equipped to present
perimetric light stimuli and record pupil light reflexes. Eleven
locations were tested at different intensities along the horizontal
meridian of each eye, and the decibel sensitivity of the pupil light
reflex was compared with the visual threshold at the same location.
RESULTS. The shape and height of the hill of vision (retinal sensitivity) was
very similar between the right and left eyes of each individual using
either pupil perimetry (R2 = 0.69) or
standard threshold perimetry (R2 =
0.62) but was less similar between subjects. Comparisons between pupil
and visual sensitivity revealed a lack of correlation at the same
retinal location in normal eyes (R2 =
0.19).
CONCLUSIONS. The high intereye correlation for either pupil or visual sensitivity
may provide an important tool for detecting focal or asymmetric visual
field damage. Although the basic shape of the sensitivity profile of
pupil and visual responses was similar under the conditions of testing,
the two did not correlate well within each eye among the normal
subjects. This highlights that similarities do exist in the sensitivity
profile of the two pathways, but they do not seem to vary in the same
proportion between normal individuals.
 |
Introduction
|
|---|
Although standard visual field perimetry has become the
accepted clinical tool for evaluating the effect of disease on visual
field threshold, inherent limitations exist in the method. These
include subjectivity of the patients response, learning
effect,1
inability of some patients to maintain central
fixation during the thresholding task,2
3
and higher
variability of the threshold determination in more peripheral
locations4
5
6
7
and in damaged locations of the visual
field.8
9
In an effort to circumvent these limitations and
to supplement the presently available perimetric information, we have
developed a method of "pupil perimetry."10
11
12
Other
investigators have also reported the use of the pupil light reflex to
obtain visual field information.13
14
15
16
17
18
19
20
21
22
23
24
In pupil perimetry, an objective neuronal reflex, the pupil light
reflex, is quantified using a computerized infrared pupillometer
coupled to an automated perimeter. In visual field locations with
reduced sensitivity, the pupil contracts less to a standardized light
stimulus, whereas locations with normal sensitivity elicit a larger
pupil contraction in comparison. This method of perimetry has the
advantage of being objective, requires little patient effort and
attention, and may reflect damage at an earlier stage of
disease.10
In addition, it can also be used to evaluate
and quantify perimetric sensitivity at both threshold and
suprathreshold levels of light stimuli by taking advantage of the
graded pupil response to increasing intensities of light stimuli. This
last advantage may be important in understanding how eye disease
affects the visual system over a range of light intensities and not
just at threshold levels of perception, where almost all current
testing is assessed. Preliminary results in our laboratory have
indicated that optic nerve damage may cause greater deficits in
pupillary response to brighter light stimuli than dimmer light
stimuli.11
The purpose of this investigation is to derive the pupil sensitivity in
units of decibel sensitivity at locations along the horizontal meridian
of the visual field in normal subjects and to compare these values with
the visual threshold (also in units of decibel sensitivity) obtained at
the same locations. This strategy allows a direct comparison between
the pupil and visual sensitivity profile of the hill of vision.
In addition, we also wanted to determine how the pupil and visual
sensitivity profiles vary between the right and left eyes of the same
subject and between normal individuals. This information is important
to understand what may constitute an abnormal pupil sensitivity profile
in patients with focal or asymmetric damage to the visual pathway.
 |
Methods
|
|---|
The study was conducted according to the tenets of the
Declaration of Helsinki. Ten normal volunteers, three men and seven
women, who were between 28 and 44 years of age (mean, 36 years) were
included for this study and examined in our clinic. Normal subjects
tested by pupil perimetry were not receiving any medications known to
influence the pupil light reflex. All the normal subjects had visual
acuity of at least 20/20 in each eye, normal stereo vision, normal
slit-lamp and fundus examination, and normal Humphrey 24-2 SITA
standard perimetry satisfying normal machine criteria for reliability.
Pupillometer
A new computerized binocular infrared video pupillometer (Tom
Cornsweet, Visual Pathways, Inc., Prescott, AZ) was used to record
pupil responses over a range of stimulus light intensities (Fig. 1)
. The infrared pupillometer consisted of a monochrome VGA monitor with
viewing optics used to present light stimuli to the subject, two
identical systems for pupil tracking, one for each eye, and two CCD
cameras for recording the pupil response of each eye simultaneously.
Stimuli were generated on the monochrome VGA monitor located
inside the machine. There was a converging lens system with
polarizing filters between each eye and the internal VGA monitor, so
that the left eye sees only the left side of the monitor and the right
eye only the right side. The internal monitor can be driven in the
Z-axis with a stepper motor toward or away from the patient
to change its optical distance from the eyes over a range of about
±7.00 diopters of refractive error, so that the plane of the stimulus
is at optical infinity for the patient. The infrared pupillometer
portion of the instrument used the bright pupil images in conjunction
with two Philips CCD video cameras (type VC62505T) to record pupil
responses 60 times/sec with 25 µm-resolution of pupil diameter from
an infrared video image of the pupil. The video output of each camera
was processed in a circuit board that measured the horizontal diameter
of the pupil every 1/60th of a second.

View larger version (118K):
[in this window]
[in a new window]
|
Figure 1. The computerized binocular infrared video pupillometer, which provides
computer-controlled light stimulation and records both right and left
pupillary responses simultaneously. Right: the two small
monitors show the bright pupil video image of the right and left
pupils. The stimulus that is presented to the patient inside the
machine (using a monochrome display monitor) is also observed on
the external computer monitor to the right so the
examiner can monitor the experiments progress.
|
|
Perimetric Stimulus
A mesopic background of 3.15 apostilbs was used for this study.
Previous studies in our laboratory have shown that a higher mesopic
background of 31.5 apostilbs (like that used in standard Humphrey
perimetry) causes the pupil to become too small in some individuals,
limiting its response range of movement.10
We stimulated
at a centrally located fixation point and 10 other locations that were
located 3° above the horizontal meridian and separated by 6° (see
inset in Fig. 3B
). Unlike standard threshold perimetry, which uses a
test target size of 0.4° in diameter (Goldmann target III), a larger
4° target was used in this pupil perimetry study. We have recently
found that stimuli of smaller size (e.g., 1.7° or Goldmann size V)
sometimes fail to elicit a large enough pupil response (e.g., >0.3 mm
of contraction) in the nasal visual field in normal subjects.
Therefore, the larger 4° stimulus was used to ensure that we could
obtain adequate pupillary responses over the range of stimulus
intensities used in this study. Stimulus duration was kept at 0.2
seconds (similar to standard threshold perimetry), and a stimulus
interval of 2.5 seconds was used. The units of stimulus intensities
were calculated as equivalent to the units used in Humphrey automated
perimeters so that the visual sensitivity results from patients tested
on the perimeter could be compared directly. Because the Humphrey
perimeter uses a neutral density filter wheel interposed in the light
path to attenuate the light stimulus, a higher value of decibel
attenuation is equivalent to a dimmer light. The following stimulus
light intensities were used (given in units of decibels of attenuation
above background; e.g., 0 dB attenuation = 10,000 apostilbs or
3183 cd/m2 above background): 37 dB (0.64
cd/m2), 30 dB (3.18 cd/m2),
25 dB (10.07 cd/m2), 21 dB (25.28
cd/m2), 17 dB (63.51
cd/m2), and 13 dB (159.53
cd/m2). Subjects were asked to fixate on X-shaped
fixation target in the eye being tested (2° in diameter) during the
test. A background adaptation of 30 seconds was given before the test.
We stimulated perimetric locations in the nasal and temporal fields
alternately for one eye at a time. Testing time for 11 locations at six
different intensities was 2.42 minutes. Subjects were tested by this
strategy with three repetitions for each eye, and the right eye was
tested first. We averaged the three repetitions of pupil response for
the analysis of this study.

View larger version (34K):
[in this window]
[in a new window]
|
Figure 3. Naka-Rushton curve fitting technique was used to establish pupil
sensitivity for each criterion level. (A) Three stimuli were
repeated for each of the 17 intensity steps for one perimetric location
at 3° temporal to the fovea. This was done to establish that the
shape of the stimulusresponse function was well characterized by the
Naka-Rushton curve fit, even at the lower intensities. The
solutions for each of the three Naka-Rushton parameters
(Rmax, n, and
logk) are shown for the curve fit at this location as well
as the high correlation coefficient
(R2). (B) Graph of stimulus
light intensity versus pupil contraction at three different locations
in the visual field for comparison. The inset shows the
graphed test locations in the visual field in relation to the other
locations tested. In this example, an arbitrary criterion level of 0.7
mm is used to calculate the threshold level for each location. At a
given criterion level (y), the intensity (x)
needed to produce that level of pupil contraction is calculated. This
calculation was repeated for a number of different criterion levels.
Therefore, from the stimulusresponse curves, a threshold value can be
determined at each location based on any chosen criterion level. In the
example shown, the vertical arrows indicate the intensity
threshold values for a criterion level of 0.7-mm pupil contraction for
each of the three locations. The foveal location (0, 0; leftmost
curve) shows the most sensitive threshold level.
|
|
Analysis of Pupil Responses
Details of the method that we are currently using to analyze
pupil responses have been published elsewhere.10
We
developed an automated software program that analyzes all pupil
movements to obtain the contraction amplitude, maximum velocity,
maximum acceleration, and latency time of each pupil light reflex (Fig. 2) . On the basis of quantitative studies of the dynamics of random pupil
movements that occur without a light stimulus, we also established a
software routine to help differentiate a small pupil response that was
likely due to a light stimulus from a small random movement that was
noise. From previous analyses of pupil movements in response to small,
focal light stimuli, we found that when the start of a pupil
contraction (time at which the maximum acceleration occurred) fell
within a finite time window (200450 ms after the onset of a
0.2-second duration stimulus) the contraction was a candidate for a
true pupil light reflex. In addition, if either the maximum contraction
velocity or maximum acceleration exceeded 0.1 mm/sec or 0.1
mm/sec2, the pupil contraction was considered a
pupil light reflex. If the pupil movement that occurred during the
specified time window did not equal or exceed the dynamic constraints
imposed, then the pupil response was considered a "no response"
with a contraction amplitude of zero. Using this method in conjunction
with the Naka-Rushton curve-fitting technique, described below,
resulted in the pupil response approaching zero at intensities well
below absolute threshold.

View larger version (21K):
[in this window]
[in a new window]
|
Figure 2. Example of pupil recordings from a pupil perimetry test, as analyzed by
a customized software program. The time scale (in seconds) is at the
top, and the stimulus record is next, followed by the
tracing of right and left pupil diameters. Tracings 2
through 4: the tracing of the degree of anisocoria,
velocity, and acceleration, respectively. The beginning of each pupil
reaction is marked at the time when maximum acceleration takes place.
The peak of the contraction is defined by the point at which velocity
crosses zero.
|
|
The pupil threshold was determined from "stimulusresponse curves"
characterizing the pupil contraction amplitude (average of right and
left pupil contraction for each stimulus) over a 24-decibel range of
stimulus intensities (Fig. 3)
. A sigmoid curve was fit to the data using a Naka-Rushton equation fit
that solves for three parameters
(Rmax, n, and
logk), which define the shape of the
function.25
26
27
The iterative curve fit that maximizes the
correlation coefficient, R2, was
performed in Microsoft Excel (Microsoft Corp., Redmond, WA) using the
Solver function. The Naka-Rushton equation was as follows:
where
y = millimeters of pupil contraction,
x = stimulus intensity in log units attenuation of light,
Rmax = maximum pupil contraction (this parameter is solved for by the equation),
n = slope of function (this parameter is solved for by the equation), and
logk = intensity at which 1/2 Rmax is reached (parameter solved by equation).
Using the curve fit for each location, the decibel sensitivity
was calculated for arbitrarily defined response criteria levels of 0.1,
0.3, 0.5, and 0.7 mm of contraction of the pupil. For example, at a
criterion level of 0.7 mm, the decibel attenuation (x-axis
on graph in Fig. 3
) was calculated that would give a 0.7-mm pupil
contraction based on the curve fit. This in effect, represented the
decibel sensitivity of the pupil light reflex needed to elicit a
criterion response of 0.7 mm. Decibel sensitivities calculated for
lower criterion levels such as 0.1 mm were closer to visual threshold
sensitivity. This strategy was used to provide a method by which the
pupil response would be expressed in units of decibel sensitivity so
that it could be compared with similar units of visual threshold at the
same location. Visual threshold was determined at the same perimetric
locations with standard sized stimuli (0.4° diameter, Goldmann size
III) using a Humphrey automated perimeter (24-2 program, SITA test
strategy, 31.5 apostilb background). In two subjects, a customized
Humphrey perimeter was used to determine visual threshold using a 4°
diameter stimulus on a 3.15 apostilb background, similar to that used
for pupil perimetry.
 |
Results
|
|---|
The stimulusresponse curves were fit for each location tested by
pupil perimetry and showed an excellent fit to the Naka-Rushton
equation (Fig. 3)
. In the one example shown, 17 stimulus intensities
were used to verify that the Naka-Rushton curve fit did reflect the
true shape of the stimulusresponse function for the pupil light
reflex (Fig. 3A)
. In general, more peripheral locations that were less
sensitive resulted in a shift of the stimulusresponse function to the
right (Fig. 3B)
. The mean value of the correlation coefficient,
R2, which is a measure of the goodness
of fit was 0.94 ± 0.05 (11 perimetric locations x 2
eyes x 10 subjects = 220 curve fits). The profile of decibel
sensitivity along the horizontal meridian is shown for the right and
left eye of each of the 10 normal subjects in Figure 4
. In this figure it is apparent that the sensitivity peaks at the foveal
location and decreases toward the periphery. The shape of the profile
varied among the 10 subjects, particularly in terms of the foveal peak.
However, the shape of the profile was quite similar between the right
and left eyes of the same subject. This was also true for the profile
plots of visual threshold for the same subjects (Fig. 5)
. A small decrease in sensitivity at the peripapillary area (15°
temporally and 3° superiorly) was observed with visual threshold
(0.4° diameter stimulus, 31.5 apostilb background) but not with pupil
threshold (4° diameter stimulus, 3.15 apostilb background).

View larger version (23K):
[in this window]
[in a new window]
|
Figure 4. These graphs depict the "hill" of pupil threshold
sensitivity across the horizontal meridian in 10 normal subjects
(each subject is a different graph). Solid line: left
eye pupil sensitivity profiles; broken line: right eye
pupil sensitivity profiles; x-axis (+ degrees, temporal
field; - degrees, nasal field): horizontal location in degrees;
y-axis: the decibel sensitivity for a criterion level of
0.5-mm contraction for all subjects; a higher decibel (dB) sensitivity
on the y-axis represents a more sensitive location. Note
the variability in shape of the island among the 10 subjects
(particularly in the height of the foveal peak), but the similarity in
shape of the island between the two eyes of each subject. Pupil testing
was performed with stimulus size of 4° in diameter and
background = 3.1 apostilbs.
|
|

View larger version (23K):
[in this window]
[in a new window]
|
Figure 5. These graphs of visual threshold are for the same normal
subjects in the same order shown in the graphs of pupil threshold in
Figure 4
. Note that the shape of the hill of vision for visual
threshold is more similar between the 10 subjects compared with their
pupil threshold. As with pupil threshold, there is a high correlation
between the sensitivity profiles between the right and left eyes.
Visual threshold was determined under standard clinical testing
conditions (0.4° stimulus size, background = 31.5 apostilbs).
|
|
Comparisons between the right and left eyes for visual threshold and
pupil sensitivity are shown in the scatter plots in Figure 6 . The effect of different criterion levels (0.1, 0.3, 0.5, 0.7 mm
contraction) on the right and left eye comparisons for pupil
sensitivity is also shown. It was apparent that the intereye
correlations were much higher when criterion levels > 0.1 mm were
used. When higher criterion levels were used, the correlation
coefficient (r2) was similar for pupil
sensitivity and visual threshold (r2
0.6). With higher criterion levels, the decibel pupil sensitivity
also decreased, because a brighter light (less attenuated) was needed
to produce a greater level of pupil contraction. This caused a greater
difference in absolute threshold between the pupil and visual
sensitivity at the higher criterion levels of pupil contraction. This
was seen as an increase in separation between the visual threshold data
points (open triangles) and the pupil threshold data points (closed
circles) in the scatter plots in Figure 6 .

View larger version (53K):
[in this window]
[in a new window]
|
Figure 6. Correlation between right and left eyes for pupil
sensitivity at different criterion levels of pupil contraction (0.1,
A; 0.3, B; 0.5, C; and 0.7-mm,
D; ). , Visual threshold in each graph for the 10
normal subjects. Each symbol represents the threshold for the same
location in the right and left eye for the same subject. The
linear correlation (r2 value) between
pupil threshold of the right and left eyes at the same perimetric
locations was much higher when criterion levels >0.1-mm pupil
contraction were used. Also, with higher criterion levels there was a
greater difference in threshold level between the pupil and visual
sensitivity (e.g., more separation between solid
dots and open triangles in D).
|
|
The effect of criterion level on the shape of the horizontal profile of
pupil sensitivity is shown in Figure 7
, which represents the average threshold for all 10 subjects. For higher
criterion levels, the profile is shifted downward, and the foveal
sensitivity peak became flatter and less conspicuous. It can also be
noticed that the SD between subjects at each location was much higher
using the lower criterion levels. A similar plot of visual threshold
across the horizontal meridian is shown in Figure 8
for the same subjects using either the standard stimulus size (0.4°
diameter on a 31.5 apostilb background) or the same conditions as pupil
perimetry (4° diameter spot size on a 3.15 apostilb background). The
profile shape for visual threshold was very similar to that shown for
pupil sensitivity when standard clinical testing conditions were used,
but the profile shape flattened considerably at the foveal location and
differed from the pupil sensitivity profile when a lower mesopic
background and larger stimulus was used, like that used for pupil
perimetry. The SD among the right eyes of the 10 subjects was much less
for visual threshold than for pupil perimetry (2 SDs =
3
decibels for visual threshold versus 6 decibels for pupil perimetry).

View larger version (35K):
[in this window]
[in a new window]
|
Figure 7. Average profile of pupil threshold + two SDs (between subject
variability) across the horizontal meridian for the right eyes of 10
normal subjects. The tested locations, in degrees from the fovea (+
degrees, temporal field; - degrees, nasal field), are shown along the
x-axis. Each line profile represents the pupil threshold
based on increasing criteria (0.1, 0.3, 0.5, and 0.7 mm of pupil
contraction). At higher criterion levels, it takes a brighter light
(decibel sensitivity is a lower number) to reach that criterion
response, shifting the profile downward. The shape of the profile
remains similar except for the foveal peak, which becomes more blunted
at higher criterion levels. The SD between subjects at each location
was much higher using the lower criterion levels but appears to remain
constant across the different locations.
|
|

View larger version (27K):
[in this window]
[in a new window]
|
Figure 8. This graph shows the average profile (dark solid line)
of visual threshold + 2 SDs (between subject variability) across the
horizontal meridian for the right eyes of the same 10 normal subjects
shown in Figure 7
. The variability of visual threshold (2 SDs) across
the horizontal meridian in 10 normal subjects is less compared with
pupil threshold. Note the similarity in the shape of the profile
compared with pupil threshold (Fig. 7)
. Also shown for comparison
(broken line) is the visual threshold profile for two
subjects tested with the same size stimulus and background as was used
for pupil perimetry testing (4° diameter stimulus, background =
3.15 apostilb). Using the same stimulus parameters as was used for
pupil perimetry caused the visual threshold to be more sensitive
(upward displacement of line), and the profile became flatter at the
foveal location compared with the results from standard clinical
testing conditions and from pupil perimetry.
|
|
Comparisons between the visual sensitivity (standard clinical testing
conditions) and pupil sensitivities at the same locations for all
normal subjects showed very little correlation, regardless of the
criterion level used for pupil threshold (Table 1)
, even though the overall profile shape was similar.
 |
Discussion
|
|---|
In this study, we were able to define the stimulusresponse
function of the pupil light reflex at each perimetric location to
derive the threshold sensitivity at different response criteria of
pupil contraction. This approach enabled us to quantify the pupil light
reflex sensitivity of the retina in decibel units, the same measurement
unit used in clinical practice to quantify visual sensitivity.
Expressing the pupil light reflex sensitivity in decibel units has the
advantage of allowing a direct comparison with visual sensitivity and
its variability in normal and diseased states. Although the variability
in pupil sensitivity between individuals was greater than for visual
sensitivity, the two methods revealed almost the same high correlation
between the right and left eyes when comparing matching locations. As
will be discussed later, this result may provide a powerful means for
detecting focal damage, when it is asymmetric between the two eyes,
either by visual or pupil perimetry. To our knowledge, this intereye
analysis strategy has not been previously reported, and it may be a
useful means of assessing the sensitivity and specificity of pupil and
visual perimetry in different forms of unilateral or asymmetric optic
neuropathy, compared with normal subjects.
Compared with visual perimetry, pupil perimetry has the added appeal of
being an objective reflex, allowing the results to be scrutinized
without having to rely on a patients judgment, which can sometimes be
unreliable. The objectivity of pupil perimetry may provide a means of
validating the results of visual perimetry in cases where the results
are questionable. In addition, central fixation during pupil perimetry
appeared to be less of a problem during the test because the patient is
not required to make judgments (and hence, they do not have a tendency
to look toward peripheral stimuli). In patients with large central
scotomas, the fixation target size can be easily increased or the other
eye can be used for fixation with our binocular instrument. Eye
position monitoring during the test is also possible with the current
instrument and will be used in the future to determine actual stimulus
location on the retina during the test, should a small shift in
fixation occur during the stimulus presentation.
In this study, we chose arbitrary criterion levels of pupil contraction
to calculate decibel sensitivity based on intensity-response functions.
The pupil light reflex is a graded response that increases in
proportion to stimulus light intensity. Characterization of how pupil
contraction increases with increasing light intensity may be an
indirect measure of how neuronal firing rates change with increasing
stimulus light intensity. In this sense, pupil contraction amplitude
may serve as an objective indicator of the behavior of retinal neurons
to incremental changes in light intensity over a 2 to 3 log unit range.
The quantification of pupil sensitivity at increasing criterion levels
of pupil contraction provides a means to study the decibel sensitivity
near threshold levels (lower criterion levels) and at increasing
suprathreshold levels (higher criterion levels). This method will
enable us to determine whether the effect of disease on retinal
sensitivity is more pronounced at higher stimulus levels (low versus
high neuronal firing rates) in patients with different causes of damage
(i.e., compression, ischemia, inflammation). Preliminary results from
our laboratory in patients with unilateral optic neuropathy have shown
that the damage, as revealed by the difference in pupil response
amplitude between the two eyes, is intensity dependent.11
Calculation of the pupil decibel sensitivity using lower criterion
levels brought the pupil threshold nearer to the range of values for
visual threshold. The actual decibel value of pupil sensitivity depends
not only on the criterion level chosen, but also on the properties of
the light stimulus such as size, background, and stimulus duration. In
fact, the steepness of the pupil sensitivity profile from the fovea to
the periphery may vary depending on the size of stimulus used. Smaller
sized stimuli have been shown to cause a steeper fall off of
sensitivity away from the fovea, whereas larger sized stimuli cause a
more flattened sensitivity profile.15
16
28
29
30
31
This
implies that spatial summation properties of the pupil light reflex may
be a major determinant in the sensitivity profile across the retina. We
used a larger sized target for pupil perimetry (4° in diameter) than
for standard visual perimetry (0.4° in diameter) to flatten the pupil
sensitivity profile so it would coincide more to the profile of visual
perimetry. The larger-sized stimuli were also used to produce a robust
pupil contraction, even in the peripheral locations of the 27° radius
field at stimulus intensities within the range where scatter of light
is negligible. The background level of adaptation may also influence
the profile shape of pupil sensitivity and the mechanical range
of pupil size over which the pupil moves. We chose a lower mesopic
background for pupil perimetry compared with visual perimetry (3.1
versus 31.5 apostilbs) to allow the baseline pupil size to stay in a
more advantageous size range that is not as susceptible to mechanical
constraints. Therefore, although the stimulus conditions were different
between visual and pupil perimetry, it accomplished the objective of
producing flattened sensitivity profiles outside of the fovea that were
comparable between visual and pupil perimetry.
Other factors besides retinal sensitivity may affect pupil sensitivity.
These include supranuclear levels of inhibition or excitement
(wakefulness), the innervational status of the pupil sphincter, and
iris mechanics (e.g., pupil size and "stiffness" of iris
tissue).32
These other factors influence the overall
"gain" of the pupil light reflex but would not necessarily affect
one perimetric location differently than a neighboring one. These
factors may help explain the difference in the overall upward or
downward shift in the pupil sensitivity profile across the 30° radius
field observed among normal individuals and may be responsible for a
greater SD in sensitivity between different subjects compared with
visual sensitivity. However, any differences in the shape of
the pupil sensitivity profile across the field that seemed to vary from
one individual to another would have to be explained more by
interindividual variations in the density and distribution of retinal
neurons that provide input into the pupil light reflex. This variation
in shape of the sensitivity profile does not seem to be random, because
there were remarkable similarities in the shape of the profile between
the right and left eyes of the same individual. From this we conclude
that retinal factors may provide a major influence on the shape of the
pupil sensitivity profile across the field. These retinal factors, such
as ganglion cell distribution and number, seem to be more similar
between the two eyes of the same subject than between the two eyes of
different subjects. The topographical similarities of pupil sensitivity
between the right and left eyes of the same normal subject might be
exploited in the application of pupil perimetry (and visual
sensitivity) to clinical problems to detect dissimilarities as an early
sign of retinal or optic nerve disease.
It was also interesting that at lower criterion levels of pupil
response there was lower correlation of decibel sensitivity between the
same retinal location of the right and left eye than at higher
criterion levels (Fig. 6)
. Lower criterion levels also resulted in a
higher interindividual variability in pupil decibel sensitivity (Fig. 7) . This implies that lower levels of neuronal firing, which are
represented by lower criterion levels, are associated with higher
variability of retinal sensitivity. This is reflected by the flatter,
nonlinear shape of the stimulusresponse function at lower stimulus
intensities; at the lower intensity end of the stimulusresponse
function, larger changes in sensitivity are associated with only small
changes in pupil contraction (Fig. 3)
. Therefore, small changes in
pupil contractions at the lower, flat end of the response function are
associated with relatively large sensitivity changes compared with the
steeper, more linear portion of the curve. The relationship between
neuronal firing rate and variability or retinal sensitivity could have
important implications in understanding what testing conditions are
best suited for detecting worsening or improvement in retinal
sensitivity over time in patients with disease. For example, under
standard visual field testing, retinal sensitivity is measured at
threshold (low neuronal firing rates), and this is known to be
associated with very high subject variability over time once damage has
occurred. Pupil perimetry will lend itself to studying the effect of
disease on retinal sensitivity and its variability at different
neuronal firing levels, which are represented by different criterion
levels, and hence different levels of light stimulation.
Comparison of visual and pupil sensitivity in the same decibel units
revealed that these two measures of retinal sensitivity are not highly
correlated across the horizontal retina in normal subjects.
Differences between the visual and pupil sensitivity profiles in the
same eye have been reported previously.15
16
29
30
The
extent of this dissimilarity is influenced to a large extent by the
state of adaptation, stimulus size, and stimulus duration. Differences
in the ganglion cell type, number, and receptive field properties
between the visual and pupillary systems help to explain why the
retinal sensitivity measured by either the visual or pupillary system
may not correlate.33
The differences between the ganglion
cell type and organization in the visual and pupillary systems may
provide a basis for why diseases of the retina and optic nerve may
produce specific patterns of damage in the two systems. From this
standpoint, it will be of interest to compare the perimetric visual and
pupil sensitivity in normal and damaged eyes. A differential effect of
disease on the pupil and visual system in an eye may be of clinical
importance because it may help to reveal the etiology of damage, it may
provide earlier detection of damage, and it may yield a better
assessment for the effectiveness of treatment. It is hoped that future
studies in our laboratory will clarify these issues.
 |
Footnotes
|
|---|
Supported by an unrestricted grant from Research to Prevent Blindness
(New York), a grant from Zeiss-Humphrey Corporation, and a Merit Review
and Career Development Award from the Veterans Administration,
Washington, DC (RHK). RHK is the recipient of a Lew Wasserman Scholar
Award (Research to Prevent Blindness), and SH received a support
visiting scholarship from Kyungpook National University Hospital,
Taegu, Korea.
Submitted for publication October 26, 1999; revised January 31, 2000;
accepted February 10, 2000.
Commercial relationships policy: N.
Corresponding author: Randy H. Kardon, Department of Ophthalmology and
Visual Science, University of Iowa Hospital and Clinic, 200 Hawkins
Drive, Iowa City, IA 52242. randy-kardon{at}uiowa.edu
 |
References
|
|---|
-
Heijl, A, Lindgren, G, Olsson, J. (1989) The effect of perimetric experience in normal subjects Arch Ophthalmol 107,81-86[Abstract]
-
Henson, DB, Bryson, H (1991) Is the variability in glaucomatous field
loss due to poor fixation control Mills, RP Heijl, A eds. Perimetry Update 1990/1991 ,217-220 Kugler Amsterdam/New York.
-
Vingrys, AJ, Demirel, S (1993) The effect of fixational loss on perimetric
thresholds and reliability Mills, RP eds. Perimetry Update 1992/1993 ,521-526 Kugler Amsterdam/New York.
-
Lewis, RA, Johnson, CA, Keltner, JL, Labermeier, PK (1986) Variability of quantitative automated perimetry in normal observers Ophthalmology 93,878-881[Medline][Order article via Infotrieve]
-
Heijl, A, Lindgren, G, Olsson, J. (1987) Normal variability of static perimetric threshold values across the central visual field Arch Ophthalmol 105,1544-1549[Abstract]
-
Young, WO, Stewart, WC, Hunt, H, Crosswell, H. (1990) Static threshold variability in the peripheral visual field in normal subjects Graefes Arch Clin Exp Ophthalmol 228,454-457[Medline][Order article via Infotrieve]
-
Werner, EB, Ganiban, G, Balazsi, G (1991) Effect of test point location on the magnitude of threshold fluctuation in glaucoma patients undergoing automated perimetry Mills, RP Heijl, A eds. Perimetry Update 1990/1991 ,175-181 Kugler Amsterdam/New York.
-
Heijl, A, Lindgren, A, Lindgren, G. (1989) Test-retest variability in glaucomatous visual fields Am J Ophthalmol 108,13013-13015
-
Heijl, A, Lindgren, A, Lindgren, G, Patella, M (1991) Inter-test threshold variability in glaucomaImportance of censored observations and general field status Mills, RP Heijl, A eds. Perimetry Update 1990/1991 ,189-192 Kugler Amsterdam/New York.
-
Kardon, RH, Kirkali, PA, Thompson, HS (19915) Automated pupil perimetry. Pupil field mapping in patients and normal subjects Ophthalmology 98,485-95discussion 495496[Medline][Order article via Infotrieve]
-
Kardon, R. (1995) Pupillary light reflex Curr Opin Ophthalmol 6,20-26[Medline][Order article via Infotrieve]
-
Donahue, SP, Moore, P, Kardon, RH (1997) Automated pupil perimetry in amblyopia: generalized depression in the involved eye Ophthalmology 104,21612-21617
-
Sugita, K, Sugita, Y, Mutsuga, N, Takaoka, Y. (1972) Pupillary reflex perimeter for children and unconscious patients Monogr Hum Genet 6,199[Medline][Order article via Infotrieve]
-
Cibis, GW, Campos, EC, Aulhorn, E. (1975) Pupillary hemiakinesia in suprageniculate lesions Arch Ophthalmol 93,1322-1327[Medline][Order article via Infotrieve]
-
Jensen, W. (1976) A description of a method for objective perimetry Albrecht Von Graefes Arch Klin Exp Ophthalmol 201,183-191[Medline][Order article via Infotrieve]
-
Aoyama, T, Kani, K. (1977) Pupillographic perimetry Doc Ophthalmol Proc Ser 14,207-212
-
Reuther, R, Krastel, H, Alexandridis, E. (1981) Disturbances of the pupil reflex associated with homonymous hemianopic paracentral scotoma Arch Psychiatr Nervenkr 229,259-266[Medline][Order article via Infotrieve]
-
Trimarchi, F, Bianchi, PE, Gelmi, C, Franchini, F, Baldini, M. (1981) Diagnostic value of objective campimetry J Neurol 225,167-173[Medline][Order article via Infotrieve]
-
Alexandridis, E, Krastel, H (1989) Pupillographic perimetry using the "Octopus." Fortschr Ophthalmol 86,692-694[Medline][Order article via Infotrieve]
-
Fankhauser, FD, Flammer, J. (1989) Puptrak1.0a new semiautomated system for pupillometry with the Octopus perimeter: a preliminary report Doc Ophthalmol 73,235-248[Medline][Order article via Infotrieve]
-
Turtschi, S, Bergamin, O, Dubler, B, Schotzau, A, Zulauf, M. (1994) Pupil perimetry with the Octopus 12-3. Initial experiences Klin Monatsbl Augenheilkd 204,398-399[Medline][Order article via Infotrieve]
-
Wilhelm, H. (1996) The pupil and retrogeniculate visual pathway. Overview Ophthalmologe 93,319-324[Medline][Order article via Infotrieve]
-
Bergamin, O, Schoetzau, A, Sugimoto, K, Zulauf, M. (1998) The influence of iris color on the pupillary light reflex Graefes Arch Clin Exp Ophthalmol 236,567-570[Medline][Order article via Infotrieve]
-
Yoshitomi, T, Matsui, T, Tanakadate, A, Ishikawa, S. (1999) Comparison of threshold visual perimetry and objective pupil perimetry in clinical patients J Neuroophthalmol 19,89-99[Medline][Order article via Infotrieve]
-
Massof, RW, Wu, L, Finkelstein, D, Perry, C, Starr, SJ, Johnson, MA (1984) Properties of electroretinographic intensity-response functions in retinitis pigmentosa Doc Ophthalmol 57,279-296[Medline][Order article via Infotrieve]
-
Anastasi, M, Brai, M, Lauricella, M, Geracitano, R. (1993) Methodological aspects of the application of the Naka-Rushton equation to clinical electroretinogram Ophthalmic Res 25,145-156[Medline][Order article via Infotrieve]
-
Evans, LS, Peachey, NS, Marchese, AL (1993) Comparison of three methods of estimating the parameters of the Naka- Rushton equation Doc Ophthalmol 84,19-30[Medline][Order article via Infotrieve]
-
Schweitzer, NMJ (1956) Threshold measurements on the light reflex of the pupil in the dark adapted eye Doc Ophthalmol 10,1-78
-
Bresky, RH, Charles, S. (1969) Pupil motor perimetry Am J Ophthalmol 68,108-112[Medline][Order article via Infotrieve]
-
Hellner, KA, Sautter, H. (1978) Pupil and light sense perimetry with light and dark adaptation Klin Monatsbl Augenheilkd 173,653-657[Medline][Order article via Infotrieve]
-
Okuyama, S, Matsumoto, C, Iwagaki, A, Otsuki, T, Otori, T (1997) Pupil perimetry with the Octopus 12-3 Wall, M Heijl, A eds. Perimetry Update 1996/1997 ,51-58 Kugler Amsterdam/New York.
-
Loewenfeld, IE, Newsome, DA (1971) Iris mechanics. I. Influence of pupil size on dynamics of pupillary movements Am J Ophthalmol 71,347-362[Medline][Order article via Infotrieve]
-
Loewenfeld IE. The light reflex. The Pupil: Anatomy, Physiology,
and Clinical Applications. Detroit, MI: Wayne State University
Press; 1993:142146.
This article has been cited by other articles:

|
 |

|
 |
 
O. Bergamin and R. H. Kardon
Latency of the Pupil Light Reflex: Sample Rate, Stimulus Intensity, and Variation in Normal Subjects
Invest. Ophthalmol. Vis. Sci.,
April 1, 2003;
44(4):
1546 - 1554.
[Abstract]
[Full Text]
[PDF]
|
 |
|