1 From the Department of Ophthalmology, Herlev Hospital, University of Copenhagen, Copenhagen, Denmark; and the 2 Eye Clinic, Steno Diabetes Center, Gentofte, Denmark.
| Abstract |
|---|
|
|
|---|
METHODS. Forty-degree stereo fundus photographs of the macular field and OCT
scans in 84 eyes of 47 patients with diabetes were compared in terms of
both location and area of retinal thickening. Six radiating 6-mm OCT
scans at intervals of 30° were obtained, and the retinal thickness
was compared with the mean retinal thickness ± 2 SD of healthy
control subjects (33 eyes). Subsequently, retinal thickness was mapped
topographically and the subjectively assessed retinal thickening map
was overlaid. The degree of agreement on location was evaluated as
retinal thickening present or absent in nine subfields of the fundus.
Area of retinal thickening was divided into four categories: no retinal
thickening, less than 1 disc area (DA), less than 2 DAs, or less than 3
DAs, expressed as the
statistic.
RESULTS. Exact agreement on location was found in 676 of 756 (84 x 9)
observations (89.4%;
= 0.69; 95% confidence interval [CI]:
0.630.75). Two eyes in which both methods assessed the same amount of
retinal thickening showed no agreement on location. These eyes were
compared only in terms of location. Exact agreement on area was found
in 69 (84.1%) of 82 eyes. Weighted
(
w) = 0.79;
95% CI: 0.760.82).
CONCLUSIONS. The degree of agreement between subjectively and objectively assessed retinal thickening was very good, implying that changes in diabetic macular edema can be accurately and prospectively measured with OCT.
| Introduction |
|---|
|
|
|---|
The objective of this study was to compare subjective evaluation of retinal thickening in stereo fundus photographs with objective assessment of retinal thickening by OCT in patients with diabetes with macular edema and to evaluate the degree of agreement between the two methods.
| Patients and Methods |
|---|
|
|
|---|
CSME, meeting the ETDRS criteria of retinal thickening 500 µm or less from the center of the fovea, was diagnosed in two eyes by fundus examination.4 Two eyes had previously had focal laser photocoagulation for CSME, and both eyes were clinically assessed as still having retinal thickening. Non-CSME (n = 71; defined as retinal thickening present at or within 2 disc diameters [DD] from the fovea, but not meeting the criteria of CSME) or no definite retinal thickening (n = 9) was diagnosed in the remainder of the eyes by fundus examination.4
The control group consisted of 33 eyes in 25 healthy control subjects (15 right eyes, and 18 left eyes): mean age, 48.2 years (range, 2575); female-to-male ratio, 17:8.
Stereo Fundus Photography
ETDRS 7 standard 40° field stereo fundus photographs were
obtained in all patients with a fundus camera (CF60 Uvi; Canon USA,
Lake Success, NY) by a photographer certified for stereo fundus
photography with this camera by the Fundus Photography Reading Center
(University of Wisconsin, Madison, WI). Photographs in which sufficient
stereo effect, clarity, and correct positioning had not been obtained
(because of poor patient cooperation, excessive tearing, poorly dilated
pupils, or unclear media) were excluded.
The stereo color slides were mounted in transparent slide holders in stereo pairs, and retinal thickening was evaluated in ETDRS standard field 2 (centered on the macula) of each photograph set. Retinal thickening was evaluated with a stereo viewer and assessed with respect to area and location of retinal thickening by the first author (CS). These assessments were drawn on transparent maps, one map for each eye. For improving the assessment of the retinal thickening in the fundus photographs, a detailed grid (Fig. 1C) , calibrated for 40° fundus photographs, was used (kindly provided by the Fundus Photography Reading Center).
|
In each eye, six radiating cross-sectional B scans of 6 mm, matching the size of the grid (Fig. 1C) used for assessing retinal thickening by stereo fundus photographs, were obtained, with the center of each scan being the center of the fovea. The procedure used for obtaining OCT scans in this study was standardized by retaking all scans for which correct fixation had not been obtained. An internal fixation spot was available to the patient, and the fixation was monitored on the fundus photograph. The fovea was overrepresented in measuring points with the scanning procedure used in this study. This may to some degree compensate for small deviations from the correct fixation spot, which were not detected immediately by the technician.
The scanning procedure, with the six radiating 6-mm scans, was chosen as the standard procedure for this instrument5 and also for completion within a reasonable amount of time. All scanning sessions were performed by one trained technician.
B scans imaged the retinal thickness by acquiring a sequence of 100 interferometric A scans of the retina. Through the A scans, a retinal profile was automatically processed, limited by the nerve fiber layer at the front and the inner layer of the retinal pigment epithelium (RPE) at the back. Retinal thickness was calculated with the OCT instrument algorithm and entered into a newly developed program (by Palle R. Jensen, Copenhagen, Denmark), which maps the retinal thickness from the six tomograms. To estimate retinal thickness in the wedges between each OCT scan, linear interpolation was performed. In addition, the program allowed for correction of artifacts, because the nerve fiber layer in some cases has too low a signal to be detected by the OCT algorithm.
Comparison of Study Eyes and Healthy Control Eyes
The algorithm used for interpolation of the OCT scans in this
study compared the retinal thickness of the study eyes to a mean value
+2 SD of healthy control eyes. The scans were topographically mapped in
a color code, with red indicating a definitely thickened area (i.e., a
retinal thickness exceeding the mean retinal thickness +2 SD of healthy
control eyes), blue indicating a definitely nonthickened area (i.e., a
retinal thickness below the mean retinal thickness of healthy control
eyes), and green indicating a thickness in between (i.e., a retinal
thickness greater than the mean value of healthy control eyes, but
below the mean +2 SD value, Figs. 1B
). The range of standard
deviations for the 100 A scans of healthy control subjects was 7 to
14 µm.
To ensure that anatomic differences in retinal thickness in the control subjects (e.g., nasal versus temporal, right versus left) would not interfere with the comparison to diabetic eyes, right and left eyes were treated separately.
Comparison of Location of Retinal Thickening
On the comparison of location (i.e., placement of retinal
thickening), the transparent, subjectively assessed retinal thickening
map was compared with the OCT color-coded map, and the area or areas
definitely thickened on OCT were added to the transparent map.
For the purpose of analyzing the degree of agreement on location of retinal thickening, the fundus was divided into nine subfields, as suggested by Hee et al.3 allowing for a differentiated evaluation of agreement on location of retinal thickening in one eye (Fig. 2) .
|
The OCT maps and subjective evaluation of stereo fundus photographs were assessed by the same person with a minimum of 7 days between the two assessments. The subjective assessment of retinal thickening on the fundus photographs took place before assessment of the OCT topographic maps.
Informed Consent
The study adhered to the tenets of the Declaration of Helsinki.
Informed consent was obtained from the subjects after explanation of
the nature and possible consequences of the study. The research was
approved by the institutional human experimentation committee.
Statistics
The
statistic is interpreted as the chance-corrected
proportional agreement between two groups. The weighted
(
w) is the weighted proportional agreement
corrected for chance.6
The usage of the
w is preferable to
for comparison of the
area of retinal thickening, because the degree of disagreement is taken
into account. Thus, in this data set, the categories were ordered, and
disagreements were given weights according to the size of the
discrepancy. However, the comparison of the location of retinal
thickening yielded only two categories. Thus,
w analysis was not required. In addition, a
binomial test, also known as the McNemar test, was performed as a test
for bias in symmetry of the disagreements.
The
statistic has a maximum of 1.00 when agreement is perfect, and
= 0 indicates no agreement better than chance. The
interpretation of
values cannot be absolutely stated; only
guidelines to interpretation can be given. Those given by Landis and
Koch7
suggest that
= 0.61 to 0.80 indicates that
agreement between groups is substantial, and
= 0.81 to 1.00
indicates that agreement is almost perfect. Although the degree of
acceptable agreement depends on circumstances, for most purposes a
or
w < 0.5 indicates poor agreement.
Reproducibility
One hundred sixty-two observations were made (nine per eye in 18 eyes)
and entered in a 2 x 2 table. The
2 statistic gave
a P < 0.0001. A high correlation between the first
and second evaluation by grader 1 (CS) was found, with
r = 0.841 (95% confidence interval [CI]:
0.790.88) and P < 0.0001. There were no
disagreements between the first and the second evaluation of grader 1
on the area of retinal thickening.
Ninety-nine observations were made (nine per eye in 11 eyes) and the
interobserver agreement on location of retinal thickening showed exact
agreement in 90 observations (91%) with
= 0.91 (95% CI:
0.840.96). Only 1 of 11 eyes showed disagreement on area of retinal
thickening from grader 1 to grader 2.
Intraobserver reproducibility obtained in healthy control eyes in the same scanning position, in the same eye, on the same day was 7%, comparable to earlier studies.5
| Results |
|---|
|
|
|---|
statistic was 0.69 (95% CI: 0.630.75), between evaluation of stereo
fundus photographs and OCT, for location of retinal thickening. The McNemar test for symmetry of the data showed that four fields (2, 3, 4, and 5) had significant bias in the symmetry of the observations of discrepancies in favor of the presence of retinal thickening in stereo fundus photographs and absence on OCT. One field (field 9) was also biased in favor of stereo fundus photographs when there were discrepancies, yet not significantly. Two fields (1 and 6) showed no disagreement on discrepancies, and in fields 7 and 8, the discrepancies were biased in favor of OCTs more often showing retinal thickening than stereo fundus photographs, yet not significantly (Table 1) .
|
w = 0.79; 95% CI:
0.760.82). Two eyes in which both methods detected retinal
thickening, but showed no agreement on location were omitted from the
results (Table 2)
.
|
| Discussion |
|---|
|
|
|---|
Comparing an Objective and a Subjective Method
Determining the exact thickness of a thickened area in stereo
fundus photographs is difficult, because it is dependent on the
stereopsis of the observer and the quality of the fundus photographs.
Sometimes, retinal thickening is underestimated in stereo fundus
photographs when compared with binocular ophthalmoscopy.9
Empirically, the fovea is subject to the most thickening in cases of
diffuse edema. This can be observed clinically, histologically, and on
OCT. Farther from the center of the fovea, the thickened areas tend to
be less prominent and much harder to assess for exact thickness. In
cases such as in the present study eyes, focal, and often rather subtle
disease, is usually present. This makes an assessment of the exact
thickness of the thickened areas even more difficult. In more
pronounced edema, a comparison between perceived thickness on stereo
fundus photographs and absolute thickness (even volume), as determined
by OCT, is probably feasible. Although volume measurements could
provide additional information, a direct comparison between the two
methods in terms of both area and location of retinal thickening would
provide the most useful information in cases of more subtle disease
such as those assessed in this study.
One of the advantages of stereo fundus photography is that is has a large lateral resolution over the whole photographic field. OCT does not have this large lateral resolution, unless an immense amount of scans are performed. Thus, by making the comparison on the terms of the subjective stereo evaluation, the approximation of mapping the retina from the six OCT scans seems reasonable. The advantage of the OCT is that it facilitates quantitation of the retinal thickness in an objective way.
The radiating six scans that were performed in this study, with estimated values of retinal thickness in the wedges between scans, covered an area of 28.3 mm2. This area matched the area covered by the grid used for subjective evaluation of stereo fundus photographs. Objective assessments of the retinal thickness with mapping of the posterior pole has also been demonstrated by Gieser et al.10 This technique is different from the OCT technique in optical principle, scanning procedure, and analysis methods of the computer software. Until now, no other studies in this group of patients have been available for comparison to the present study.
Is subjective assessment of retinal thickening in stereo fundus photographs more sensitive than OCT? Table 1 shows that, in this study, stereo fundus photographs tended to be more sensitive than OCT for the detection of retinal thickening. Thus, in five of nine fields, the majority of disagreements were in favor of retinal thickening present on stereo fundus photographs but absent on OCT. Two fields showed no bias on disagreements, and two fields were biased toward OCT being the more sensitive method, yet not significantly. Table 2 shows that 9 of 82 eyes showed a larger area of retinal thickening on stereo fundus photographs than on OCT, whereas only 4 of 82 eyes were found to have a larger area of retinal thickening on OCT than on stereo fundus photographs. In addition, in three eyes, OCT showed no retinal thickening, whereas thickening was present on stereo fundus photographs. Given that the OCT retinal thickness was compared with a mean value of retinal thickness in healthy control subjects +2 SD and not just with the mean retinal thickness, the bias toward fundus photographs showing retinal thickening more often than OCT was expected. Very subtle lesions (i.e., above mean retinal thickness, yet below mean retinal thickness +2 SD in healthy control subjects eyes) simply would not be detectable on the topographical OCT map. However, these subtle changes can be observed on individual OCT retinal profiles.
Previously, OCT has been reported to be more sensitive than slit lamp biomicroscopy for detection of small changes in retinal thickness (i.e., changes of <100 µm greater than the mean retinal thickness in normal subjects).5 Changes of this magnitude are too subtle for the human eye to detect on stereo fundus photographs as well as on binocular slit lamp biomicroscopy, the latter presumably being the better way of assessing retinal thickening.9 The sensitivity of OCT for detection of subtle retinal lesions could be enhanced by generating a topographic map using mean values of retinal thickness in healthy control subjects +1 SD (instead of +2 SD) as a baseline comparison. A longitudinal study of changes in retinal thickening could measure very small changes in retinal thickness, given that the study eye is compared with itself at baseline. However, in this study and for clinical purposes, retinal thickening of less than the mean healthy control value +2 SD was considered not to represent macular edema.
Hard exudates can be an important factor in relation to the detection of retinal thickening by subjective assessment of edema on fundus photographs. ETDRS report 5 describes hard exudates as very likely to be associated with retinal thickening, even if the thickening is not apparent on the fundus photograph.9 The fundus photographs in this study were strictly evaluated with respect to definite retinal thickening. Appreciable macular hard exudates by themselves were not sufficient to qualify the eye as having retinal thickening. Hard exudates increase retinal thickness in the OCT software algorithm. However, the effect of shadowing caused by exudates on the RPE is modest. Typically, a maximum increase of 5% in retinal thickness is found just beneath the hard exudate.
Lens opacification could also present a problem for both methods. Decreased clarity of the photographs was seen with increasing lens opacification, and the quality of OCT scans could also have been impaired. In the present study, three eyes were excluded from the data set due to poor photograph quality. In addition, in one eye with lens opacification, the OCT signal was too impaired, and no retinal profile could be assessed. This eye was also excluded. However, in general, the OCT did not seem to be prone to impairment by cataract (JL Hougaard et al., Department of Ophthalmology, Herlev Hospital, Copenhagen, Denmark, personal communication, November 2000).
OCT is a noninvasive, patient- and operator-friendly technique that has the advantage of quantitating retinal thickness, and it has been shown to be reproducible by several authors. In addition, the results of this study show that OCT assessment of retinal thickening corresponded very well to subjective assessment of retinal thickening by grading of stereo fundus photographs and that the OCT was reliable for determining the location and extent of retinal thickening. These results suggest that changes in diabetic macular edema can be accurately and prospectively measured with the OCT in both clinical trials and clinical practice.
| Acknowledgements |
|---|
| Footnotes |
|---|
Commercial relationships policy: N.
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Charlotte Strøm, Department of Ophthalmology, Herlev Hospital, University of Copenhagen, Herlev Ringvej 75, DK-2730 Herlev, Denmark; charlotte_strom{at}dadlnet.dk.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Bernardes, T. Santos, and J. Cunha-Vaz Increased-Resolution OCT Thickness Mapping of the Human Macula: A Statistically Based Registration Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 2046 - 2052. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. Davis, S. B. Bressler, L. P. Aiello, N. M. Bressler, D. J. Browning, C. J. Flaxel, D. S. Fong, W. J. Foster, A. R. Glassman, M. E. R. Hartnett, et al. Comparison of Time-Domain OCT and Fundus Photographic Assessments of Retinal Thickening in Eyes with Diabetic Macular Edema Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 1745 - 1752. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Virgili, F. Menchini, A. F. Dimastrogiovanni, E. Rapizzi, U. Menchini, F. Bandello, and R. G. Chiodini Optical Coherence Tomography versus Stereoscopic Fundus Photography or Biomicroscopy for Diagnosing Diabetic Macular Edema: A Systematic Review Invest. Ophthalmol. Vis. Sci., November 1, 2007; 48(11): 4963 - 4973. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Sander, D. N. Thornit, L. Colmorn, C. Strom, A. Girach, L. D. Hubbard, H. Lund-Andersen, and M. Larsen Progression of Diabetic Macular Edema: Correlation with Blood Retinal Barrier Permeability, Retinal Thickness, and Retinal Vessel Diameter Invest. Ophthalmol. Vis. Sci., September 1, 2007; 48(9): 3983 - 3987. [Abstract] [Full Text] [PDF] |
||||
![]() |
K A Goatman A reference standard for the measurement of macular oedema. Br. J. Ophthalmol., September 1, 2006; 90(9): 1197 - 1202. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Vujosevic, E. Midena, E. Pilotto, P. P. Radin, L. Chiesa, and F. Cavarzeran Diabetic Macular Edema: Correlation between Microperimetry and Optical Coherence Tomography Findings. Invest. Ophthalmol. Vis. Sci., July 1, 2006; 47(7): 3044 - 3051. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Chen, B. Cense, M. C. Pierce, N. Nassif, B. H. Park, S. H. Yun, B. R. White, B. E. Bouma, G. J. Tearney, and J. F. de Boer Spectral Domain Optical Coherence Tomography: Ultra-high Speed, Ultra-high Resolution Ophthalmic Imaging Arch Ophthalmol, December 1, 2005; 123(12): 1715 - 1720. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Strom and B. Sander Comparison of Objective Retinal Thickness Analysis and Subjective Stereo Fundus Photography in Diabetic Macular Edema Invest. Ophthalmol. Vis. Sci., May 1, 2004; 45(5): 1450 - 1455. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Brown, S. D. Solomon, S. B. Bressler, A. P. Schachat, C. DiBernardo, and N. M. Bressler Detection of Diabetic Foveal Edema: Contact Lens Biomicroscopy Compared With Optical Coherence Tomography Arch Ophthalmol, March 1, 2004; 122(3): 330 - 335. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Guan, C. Hudson, and J. G. Flanagan Comparison of Heidelberg Retina Tomograph II and Retinal Thickness Analyzer in the Assessment of Diabetic Macular Edema Invest. Ophthalmol. Vis. Sci., February 1, 2004; 45(2): 610 - 616. [Abstract] [Full Text] [PDF] |
||||
![]() |
G L Ong, L G Ripley, R S B Newsom, and A G Casswell Assessment of colour vision as a screening test for sight threatening diabetic retinopathy before loss of vision Br. J. Ophthalmol., June 1, 2003; 87(6): 747 - 752. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||