1From the Eye Research Fund Laboratory, and the 2Departments of Cell and Developmental Biology and 3Physiology, Medical School of the University of Michigan, Ann Arbor, Michigan.
| Abstract |
|---|
|
|
|---|
METHODS. Testing was performed with a photo slit lamp equipped with two 2x teleconverters and a digital camera through a Goldmann-type fundus contact lens that was fabricated for the rodent eye.
RESULTS. Images of the rat and mouse optic discs were obtained that are comparable to those used by ophthalmologists to assess optic neuropathy in glaucoma, a key part of the standard of care and of clinical investigation of this disease. The cup in the optic disc image of these rodents is darker than the neural rim of the disc, rather than lighter, as it is in humans.
CONCLUSIONS. In addition to the application of this imaging method to studies of the effect on optic disc cupping of induced increased intraocular pressure in rats and mice, by detecting and documenting the onset and the course of optic neuropathy, it should be valuable in identifying animal models of glaucoma, in studying neuropathogenic mechanisms, and in assessing the effects of experimental therapies.
The present report describes a method for obtaining images of the optic disc in conscious rats and mice that are comparable to those routinely available to ophthalmologists for detecting the onset and monitoring the course of optic neuropathy in patients with POAG. This is a key part of both the current standard of care and of clinical investigation of this disease.6 By obtaining documenting images of the optic discs over time in rodent models of POAG, changes in the discs appearance that are comparable to those observed clinically in humans can be correlated with the histology and histochemistry of the optic nerve head.
| Materials and Methods |
|---|
|
|
|---|
Goldmann-Type Fundus Lens with Hruby Preset Lens-Type Holder
A special Goldmann-type fundus contact lens (Fig. 1) was fabricated for the rodents: The human fundus contact lens (Haag-Streit) was modified in a 5.7-mm diameter zone centered on its posterior surface that had a 2.5-mm radius of curvature (for the human it is 7.6 mm). A 7-mm diameter flat zone was machined centered on the anterior surface, and a coverslip with a broadband, antireflection coating was cemented to it. A Hruby preset lens holder (Haag-Streit) was modified to support this contact lens.
|
Digital Image Acquisition
This photo slit lamp has a mirror housing that incorporates a front surface mirror that swings into the optical path and closes its flash contact when its shutter release bar is activated. The photo slit lamp settings in this study included: Galilean power changer at 25x, mirror housing diaphragm at 3, slit illuminator spotlight diameter at 3 mm and filter at heat-absorbing, and light source flash at high. Attached to the mirror housing was the objective tube for monophotography, an adapter (model F; Nikon, Melville, NY), two teleconverters (TC-201 2x; Nikon) and the camera.
A digital camera (FinePix S1 Pro; Fujifilm, Elmsford, NY) was used for the imaging. Its picture element is a 23.3 x 15.6-mm charge-coupled device (CCD) with primary color filter with 3.4 megapixels in an interwoven pattern. The camera was tethered to the USB port of a personal computer, a hot shoe-to-PC adapter was mounted to the cameras accessory shoe and connected by a sync cord to the photo slit lamp flash socket. The camera settings were: ISO (equivalent film speed), 320; shutter speed, 1.5 seconds; white balance, auto; and tagged information file format (TIFF) red-green-blue (RGB) and image size 3040 x 2016 pixels. Based on this, the cameras calculated pixel density is 9354 pixels/mm2. A foot switch was wired in parallel to the left button of a computer mouse to activate the camera, swinging its mirror out of the optical path, releasing its shutter, and closing its flash contact.
When satisfactory focus, centering, and light reflex location are obtained, the operator activates the camera with the foot switch and then, within the 1.5-second window, presses the shutter release bar of the photo slit lamp. Because the two flash contacts are in series, the flash fires. The exposure time of the image is effectively the flash duration. The image is saved to the hard drive of the computer.
After the imaging method was developed, additional series of images were obtained in conscious rats and mice with a second digital camera (D1x; Nikon). Its picture element, according to Nikon specifications, is a 23.7 x 15.6 mm, 12-bit RGB charge-coupled device (CCD) with 5.47 megapixels and an effective pixel count of 5.33 megapixels (4024 x 1324 pixels). Based on this, the cameras calculated pixel density is 14,410 pixels/mm2.
Histologic Correlation
The eyes of anesthetized rats and mice were perfusion fixed by injection of 2.5% glutaraldehyde in 0.1 M Sorensens buffer through a needle in the left ventricle. The eyes were then harvested and embedded in paraffin or Epon. Serial longitudinal sections were obtained through the optic nerve heads.
| Results |
|---|
|
|
|---|
|
| Discussion |
|---|
|
|
|---|
Forty years ago clinical POAG research shifted toward cupping of the optic disc, and this focus endures today, because cupping is the crux between elevated IOP and visual field defects, both at the onset and during the course of the neuropathy. However, clinical studies of the optic disc in POAG have the burden of limitations inherent in research in patients and in addition the usual slow progress of this disease. Furthermore, histologic studies of the optic neuropathy of POAG are fraught with limitations of quantity and quality of available human tissue. Reports on these scarce specimens, poignant contributions by patients and their families, are unevenly distributed across the course of the disease, but almost never from the period of greatest interest, its onset. Moreover, the usefulness of these specimens, unavoidably acquired hours after death, declines to the extent research methods increasingly require fresh tissue.
These barriers to research on the optic neuropathy of POAG led to the adaptation of a basic tool of biomedical investigation, the animal model, in experiments in which IOP was elevated in the monkey7 8 by laser damage to the pathway of aqueous humor outflow. A series of productive studies of the optic nerve head in this animal was followed by comparable experiments in rats, after elevated IOP was induced by obstructing the venous outflow from the eye.9 10 11
Fundus photography has been used in the rat,12 and its use in mice has been described.13 Recently, the mouse disc has been imaged with a photo slit lamp through a goniolens.14 Two of the studies of the effect of increased IOP in the rat included optic disc photographs, one using a fundus camera with the cornea flattened by a coverslip,15 and one with a photo slit lamp through a contact lens.10
In a comparative study of the microscope of a photo slit lamp and a fundus camera for in-air resolution and for their ability to record fine vessel detail in normal human optic discs, the instruments performed comparably.16 However, because fundus cameras are telescopic systems, they can bring the optic disc into focus independent of its distance from the camera, and consequently disc image magnification varies with the refractive power of the eye.17 Further, although flattening the rodent cornea with a coverslip overcomes its aberrations, this can introduce optical distortion. Using the rodent Goldmann-type fundus contact lens with the microscope of the photo slit lamp eliminates the effects of both corneal refractive power and corneal aberrations.18 Finally, a mirror in a photo slit lamp instead of a beam splitter maximizes illumination efficiency, both for viewing and imaging. A digital camera provides immediate evaluation of the image for exposure, centering, focus, and the location of light reflexes.
The quality of the rodent optic disc images acquired with the method described improved with practice, but the detail in the images presented of these optic discs, with diameters of approximately 300 µm in rats and 100 µm in mice, probably approaches the current limit of the eye-optical-camera system. Increasing the pixel density of the digital camera did not enhance the quality of the rodent optic disc image, because at the magnification of image acquisition and the image size presented in Figures 2A and 2B , pixel density is not the limiting factor in resolution. Even as advances in digital camera technology increase dynamic range and decrease signal-to-noise ratio, which will extend the contrast range and suppress blooming, achieving significantly more detail in these images is unlikely.
In images of these rodents optic discs obtained with the same system settings, the horizontal diameter of the rat disc was only approximately one quarter larger than in the mouse. However, in the longitudinal histologic sections the size of the optic nerve head at the level of the pigment epithelium in the rat is almost three times larger than in the mouse. Because the effect of the cornea on image magnification was eliminated by the fundus contact lens and the difference in the axial lengths of these species eyes is relatively moderate, the closeness of the disc image size is probably due to a difference in their crystalline lenses. In any event, the surprisingly large optic disc image in the mouse is an additional advantage of this most powerful system for investigating mammalian genetics as a tool toward understanding human POAG.
The shape and orientation of the cup in the rat and mouse optic nerve headnarrow, funnel-shaped, and oblique to the disc surface19 result in a shadowed-cup appearance in the optic disc images: dark against the lighter neural rim of disc tissue. In contrast, the human cup is typically relatively broad and shallow and in images is paler than the surrounding rimcharacteristics that have contributed to the challenge of disc surface contour quantitation.18 Perhaps the different shape of the rodent optic cup will facilitate cup measurements in images of their discs, at least in two dimensions, for detection of cup size differences over time. In assessing rodent cup shape in the histologic material it was noted that it can be affected by the quality of fixation. Nerve fiber swelling can encroach on the cup if fixation is sub-optimal in eyes immersed in fixative.20 Although better fixative access to the optic nerve head can be achieved by opening the freshly enucleated eye, the gold standard is by perfusion fixation, a technique obviously not applicable to human POAG material.
Optic disc imaging in conscious rodents obviates the loss of lens transparency that corneal drying under anesthesia can lead to, especially in the mouse. Although this is reversible, the phenomenon ends the imaging session. More important, repeated sessions in a conscious animal to obtain serial disc images will be performed without the risk of death from multiple anesthesias. This risk is significant in rodents, especially in mice, and would obviously be progressively greater in models to the extent the phenotype is expressed in aging animals.
Three such models are the goal of genetic engineering in a rodent system: elevated IOP without neuropathy, to mimic ocular hypertension in humans; elevated IOP with neuropathy, comparable to classic human POAG; and neuropathy without elevated IOP, which corresponds to human normal-tension glaucoma. Optic disc imaging will be an integral part of identifying each animal model, of deciding when to obtain optic nerve head tissue for study of the fundamental mechanisms of the neuropathy, and for assessing the effects on the neuropathy of experimental therapies.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication January 30, 2002; revised July 2, 2002; accepted July 29, 2002.
Commercial relationships policy: N.
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Bruce E. Cohan, Eye Research Fund Laboratory, 2350 Washtenaw Avenue, Ann Arbor, MI 48104; becohan{at}umich.edu.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Ho, S. Kumar, X. S. Min, Y. L. Kng, M. Y. Loh, S. Gao, and L. Zhuo Molecular Imaging of Retinal Gliosis in Transgenic Mice Induced by Kainic Acid Neurotoxicity Invest. Ophthalmol. Vis. Sci., May 1, 2009; 50(5): 2459 - 2464. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-L. Guyomard, S. G. Rosolen, M. Paques, M.-N. Delyfer, M. Simonutti, Y. Tessier, J. A. Sahel, J.-F. Legargasson, and S. Picaud A Low-Cost and Simple Imaging Technique of the Anterior and Posterior Segments: Eye Fundus, Ciliary Bodies, Iridocorneal Angle Invest. Ophthalmol. Vis. Sci., November 1, 2008; 49(11): 5168 - 5174. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Paques, J.-L. Guyomard, M. Simonutti, M. J. Roux, S. Picaud, J.-F. LeGargasson, and J.-A. Sahel Panretinal, High-Resolution Color Photography of the Mouse Fundus Invest. Ophthalmol. Vis. Sci., June 1, 2007; 48(6): 2769 - 2774. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Shakoor, N. P. Blair, M. Mori, and M. Shahidi Chorioretinal Vascular Oxygen Tension Changes in Response to Light Flicker Invest. Ophthalmol. Vis. Sci., November 1, 2006; 47(11): 4962 - 4965. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Guo, V Tsatourian, V Luong, A G Podolean, D A Jackson, F W Fitzke, and M F Cordeiro En face optical coherence tomography: a new method to analyse structural changes of the optic nerve head in rat glaucoma Br. J. Ophthalmol., September 1, 2005; 89(9): 1210 - 1216. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||