|
|
||||||||
1From the Department of Ophthalmology, Harvard Medical School and the Massachusetts Eye and Ear Infirmary, Boston, Massachusetts; the 2Center for Innovative Visual Rehabilitation, Veterans Administration Hospital, Boston, Massachusetts; the 3Department of Electrical Engineering and Computer Science and the 4Research Laboratory of Electronics, Massachusetts Institute of Technology, Cambridge, Massachusetts; and the 5Nanofabrication Laboratory, Cornell University, Ithaca, New York.
| Abstract |
|---|
|
|
|---|
METHODS. Four hypotheses were tested: (1) epiretinal stimulation can be performed during acute experiments without obviously damaging the retina or degrading vision or the health of the eye; (2) perception can be obtained 50% of the time in blind patients with charge densities below published safety limits; (3) the minimal charge needed to induce perception would be higher in patients with more severe retinal degeneration; and (4) threshold charge would be lower at shorter stimulus durations. Five subjects with severe blindness from retinitis pigmentosa and one with normal vision (who underwent enucleation of the eye because of orbital cancer) were studied. Electrical stimulation of the retina was performed on awake volunteers by placing a single 250-µm diameter handheld needle electrode or a 10-µm thick microfabricated array of iridium oxide electrodes (400-, 100-, or 50-µm diameter) on the retina. Current sources outside the eye delivered charge to the electrodes. Assessment of damage was made by observing the clinical appearance of the eyes, comparing pre- and postoperative visual acuity, obtaining retinal histology in one case, and comparing perceptual thresholds with published safety limits.
RESULTS. No clinically visible damage to the eye or loss of vision occurred. Even at sites removed from stimulation, histology revealed swollen photoreceptor inner and outer segments, which were believed to be nonspecific findings. Percepts could not be reliably elicited with 50-µm diameter electrodes using safe charges in one blind patient. With the two larger electrodes, only the normal-sighted patient had thresholds at charge densities below 0.25 and 1.0 millicoulombs (mC)/cm2 for 400- and 100-µm diameter electrodes, respectively, which is one seemingly reasonable estimate of safety derived from the product of charge per phase and charge density per phase. In blind patients, thresholds always exceeded these levels, although most were close to these limits in patient 6. The range of charge density thresholds with the 400-µm electrode in blind patients was 0.28 to 2.8 mC/cm2. The normal-sighted patient had a threshold of 0.08 mC/cm2 with a 400-µm electrode, roughly one quarter of the lowest threshold in the blind patients. Strengthduration curves obtained in two blind patients revealed the lowest threshold charge at the 0.25- or 1.0-ms stimulus duration.
CONCLUSIONS. Threshold charge densities in severely blind patients were substantially higher than that in a normal-sighted patient. Charge densities in blind patients always exceeded one seemingly reasonable estimate of safe stimulation. The potential adversity of long-term stimulation of the retina by a prosthesis has yet to be determined.
Our primary objective was to transfer techniques for use of ultrathin electrode arrays developed in animal experiments to humans. Our arrays contacted the retina, which provided an opportunity to obtain lower and potentially safer thresholds than previously reported.20 21 We addressed another fundamental questionthe quality of induced perceptsin a study reported in a companion paper in this issue.22
| Methods |
|---|
|
|
|---|
Volunteer Selection
The protocol was approved by the Massachusetts Eye and Ear Infirmary and the Massachusetts Institute of Technology and adhered to the provisions of the Declaration of Helsinki. Participants could not have vision that initially exceeded hand motion perception in the worse eye, which was always the eye studied. After four experiments, the criterion was changed to 20/800 (Table 1) . Volunteers underwent a medical examination and (all but one) psychiatric screening. One volunteer with normal vision whose eye was removed because of orbital cancer was studied.
|
|
|
|
|
|
|
The goal of experiments 1 and 2 was to determine whether percepts could be induced within published charge safety limits. The goal of experiment 3 was to determine whether higher charges would yield more defined percepts; monopolar stimuli (6 or 30 Hz) were generally used. The goal of experiment 4 was to determine whether lower thresholds and better vision could be obtained in a normal-sighted volunteer; monopolar, 20-Hz, 2-ms pulses were used. The primary goal for experiments 5 and 6 was to obtain strengthduration curves; mostly monopolar, 20-Hz stimulation was used.
Assessment of Damage from Surgery or Electrical Stimulation
We sought evidence of ocular injury by comparison of pre- and postoperative visual acuity, intraocular pressure, slit lamp evaluation and funduscopy and by retinal histology in one case. The potential for harm from electrical stimulation was judged by comparison of thresholds to published safety limits. Strengthduration curves obtained in two experiments provided a detailed assessment of threshold charge. Eye examinations were performed by us on the first 2 days after surgery and thereafter by arrangement with local ophthalmologists.
Histology
The retina of the enucleated eye (volunteer 4) was fixed in 10% formalin, divided into eight 2 x 2-mm pieces taken at various eccentricities, dehydrated, and embedded in glycol methacrylate (JB-4; Polysciences, Eppelheim, Germany). Sections (24-µm thick) were stained with 1% neutral red and examined by light microscopy.
| Results |
|---|
|
|
|---|
Evolution of Experimental Technique
Experiment 1 yielded ill-defined percepts. The small electrode size too severely restricted the charge limits, which prompted us to make larger electrodes and a stronger current source. Experiment 2 produced similar results, despite the use of a much higher charge. We suspect that retinal degeneration (rather than electronics) compromised the outcome. Experiment 3 yielded some formed percepts, but thresholds were relatively high. This result motivated us to concentrate on obtaining strengthduration curves in experiments 5 and 6 to assess charge efficiency.
Hypotheses
Hypothesis 1.
Epiretinal stimulation can be performed during acute experiments without obviously damaging the retina or degrading vision or the health of the eye.
Eye discomfort with intraocular pressure of 38 mm Hg developed after the first experiment. Medical therapy achieved normal pressure, which was verified up to day 17. One posterior subcapsular cataract advanced. No visual loss, discomfort, change in the appearance of the retina, or other potentially relevant problems were reported during a mean observation period of 32 months (range: 2836) after surgery. Histology of the eye from the sighted patient showed swelling of photoreceptor inner and outer segments and other nonspecific changes (Fig. 5) .
|
Only single-electrode trials were considered. Reliable thresholds could not be obtained in the first two experiments. In experiment 3, a clear threshold was not obtained with 50-µm diameter electrodes. Accordingly, relevant data were recorded from the 100- and 400-µm diameter electrodes in experiments 3, 5, and 6, especially the strengthduration curves (Figs. 6 7) . Threshold charge generally decreased with shorter pulses, with the lowest charge usually occurring at 0.25 ms. At this duration, charge density was 4.1 and 0.30 millicoulombs (mC)/cm2 for the 100- and 400-µm electrodes, respectively. For blind patients, thresholds with 400-µm electrodes were 0.28 to 2.8 mC/cm2. Safety limits are summarized in the Discussion section.
|
|
The normal-sighted patients charge density threshold was 0.08 mC/cm2 (400-µm diameter electrode, 2 ms), approximately one quarter that of any blind patient (Table 1) . By comparison, threshold charge density at 4 ms was 4 times greater in the patient with 20/800 acuity and 8 and 19 times greater in the patients with 20/1000 and hand-motion vision, respectively. These results required normalization of thresholds to 4 ms, but qualitatively similar results held without normalization (see online Appendix, Part II, Data Analysis).
Hypothesis 4.
Threshold charge is lower at shorter stimulus durations. The strengthduration curves (Fig. 7) support this hypothesis.
| Discussion |
|---|
|
|
|---|
The second hypothesis asserted that perception could be obtained 50% or more of the time in blind patients with charge densities below published safety limits. The outcome depends on the choice of a "safe" limit (see online Appendix).24 25 26 One widely quoted limit, based on the electrode-to-fluid potential remaining below that which generates hydrogen or oxygen (ie, a physical versus a biological limit), is 1 mC/cm2 for cathode-first stimulation with iridium oxide electrodes without DC bias,25 which corresponds to our experiments. (That study also suggested that a positive 0.8-V bias raises the limit to 3.5 mC/cm2.) Strengthduration curves of our blind patients showed thresholds below 1 mC/cm2 for 400-µm diameter electrodes with 4-ms pulses or less, but above 3.5 mC/cm2 for 100-µm diameter electrodes (Fig. 7) .
Stimulation limits for biological safety are less well understood, and we are unaware of experimental data that closely relate to our methods. One model26 based on electrical injury in feline cortex24 27 predicts neural damage when the product of charge density per phase and charge per phase exceeds a certain threshold. Using a threshold of 79 µC2/cm2 per phase, rather than a more conservative limit of 32 µC2/cm2 suggested by Shannon26 (see online Appendix), this model predicts limits of 0.25 and 1.0 mC/cm2 for 400 and 100-µm diameter electrodes, respectively (Fig. 7) . For both electrode sizes, only the normal-sighted patient had thresholds at chargecharge density products below these "safe" values. In blind patients, thresholds always exceeded these values, although most were close to the limits in patient 6.
The third hypothesis asserted that threshold charge increases with more severe blindness. Given the variation in stimulation parameters across experiments (see Evolution of Experimental Technique section and online Appendix), comparison of thresholds required a normalized standard (Table 1) . Compared with the normal volunteer, threshold charge density for 400-µm electrodes at 4 ms was 4 times larger in the patient with 20/800 and 8 and 19 times larger in the patients with 20/1000 and hand-motion vision, respectively. Indeed, thresholds increased with the degree of blindness, which raises concern about whether severely blind patients could safely tolerate prolonged stimulation with an implanted prosthesis.
The fourth hypothesis was affirmed, with the strengthduration curves (Fig. 7) generally revealing the lowest threshold charge at 0.25 ms. This pursuit for the lowest threshold was motivated by the safety potentially gained from reduced electrochemical toxicity.
Comparison to Prior Work
Our methods differ from those of Humayun et al.,20 who first performed intraocular stimulation of human retina, in that we (1) used flexible, microfabricated arrays; (2) paralyzed the eye to achieve closer and more stable alignment between the electrodes and retina; (3) did not illuminate the eye during testing; (4) frequently interspersed control tests; (5) generally performed many more stimulations per subject; (6) obtained strengthduration curves; and (7) quantified the accuracy and reproducibility of responses. Humayun et al. found generally higher thresholds and a greater range of thresholds in blind patients (0.1680 mC/cm2 vs. 0.282.8 mC/cm2 in our study), possibly due to variable separation between their handheld electrodes and the retina. Similarly disparate results were obtained in normal-sighted volunteers (the 0.8 and 4.8 mC/cm2 with a 125-µm diameter wire thresholds of Humayun et al. versus 0.31 and 0.08 mC/cm2 with 100 and 400 µm diameter electrodes in our study).28 Technical factors, such as higher stimulation frequencies, sequential stimulation, or planar (versus slightly rounded) electrodes used in our study, may explain some differences in outcome. Our technique of using a paralyzed eye and our more detailed search for threshold are the most likely explanations for the less variable and lower thresholds in our study. No other data of this type are available on blind patients. By comparison, in two normal patients, Eckmiller et al. reported epiretinal thresholds (verbal communication) of 12 to 95 µA with biphasic 0.1 ms pulses, using five electrodes (500-µm diameter) (Eckmiller RE, et al. IOVS 2002;43:ARVO E-Abstract 2848).
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication August 13, 2002; revised November 18, 2002, and May 21 and June 26, 2003; accepted June 30, 2003.
Disclosure: J.F. Rizzo III (P); J. Wyatt (P); J. Loewenstein, None; S. Kelly, None; D. Shire (P)
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Joseph F. Rizzo, III, Massachusetts Eye and Ear Infirmary, 243 Charles Street, Boston, MA 02114; joseph_rizzo{at}meei.harvard.edu.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Roessler, T. Laube, C. Brockmann, T. Kirschkamp, B. Mazinani, M. Goertz, C. Koch, I. Krisch, B. Sellhaus, H. K. Trieu, et al. Implantation and Explantation of a Wireless Epiretinal Retina Implant Device: Observations during the EPIRET3 Prospective Clinical Trial Invest. Ophthalmol. Vis. Sci., June 1, 2009; 50(6): 3003 - 3008. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Hood, C. E. Lin, M. A. Lazow, K. G. Locke, X. Zhang, and D. G. Birch Thickness of Receptor and Post-receptor Retinal Layers in Patients with Retinitis Pigmentosa Measured with Frequency-Domain Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., May 1, 2009; 50(5): 2328 - 2336. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. I. Fried, A. C. W. Lasker, N. J. Desai, D. K. Eddington, and J. F. Rizzo 3rd Axonal Sodium-Channel Bands Shape the Response to Electric Stimulation in Retinal Ganglion Cells J Neurophysiol, April 1, 2009; 101(4): 1972 - 1987. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Mazzoni, E. Novelli, and E. Strettoi Retinal Ganglion Cells Survive and Maintain Normal Dendritic Morphology in a Mouse Model of Inherited Photoreceptor Degeneration J. Neurosci., December 24, 2008; 28(52): 14282 - 14292. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. de Balthasar, S. Patel, A. Roy, R. Freda, S. Greenwald, A. Horsager, M. Mahadevappa, D. Yanai, M. J. McMahon, M. S. Humayun, et al. Factors Affecting Perceptual Thresholds in Epiretinal Prostheses Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2303 - 2314. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Sekirnjak, P. Hottowy, A. Sher, W. Dabrowski, A. M. Litke, and E. J. Chichilnisky High-Resolution Electrical Stimulation of Primate Retina for Epiretinal Implant Design J. Neurosci., April 23, 2008; 28(17): 4446 - 4456. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. L. Kent, I. V. Glybina, G. W. Abrams, and R. Iezzi Chronic Intravitreous Infusion of Ciliary Neurotrophic Factor Modulates Electrical Retinal Stimulation Thresholds in the RCS Rat Invest. Ophthalmol. Vis. Sci., January 1, 2008; 49(1): 372 - 379. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Scholz Perspectives on: Materials Aspects for Retinal Prostheses Journal of Bioactive and Compatible Polymers, September 1, 2007; 22(5): 539 - 568. [Abstract] [PDF] |
||||
![]() |
P. J. DeMarco Jr, G. L. Yarbrough, C. W. Yee, G. Y. McLean, B. T. Sagdullaev, S. L. Ball, and M. A. McCall Stimulation via a Subretinally Placed Prosthetic Elicits Central Activity and Induces a Trophic Effect on Visual Responses Invest. Ophthalmol. Vis. Sci., February 1, 2007; 48(2): 916 - 926. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Gekeler, A. Messias, M. Ottinger, K. U. Bartz-Schmidt, and E. Zrenner Phosphenes Electrically Evoked with DTL Electrodes: A Study in Patients with Retinitis Pigmentosa, Glaucoma, and Homonymous Visual Field Loss and Normal Subjects Invest. Ophthalmol. Vis. Sci., November 1, 2006; 47(11): 4966 - 4974. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Salzmann, O P Linderholm, J-L Guyomard, M Paques, M Simonutti, M Lecchi, J Sommerhalder, E Dubus, M Pelizzone, D Bertrand, et al. Subretinal electrode implantation in the P23H rat for chronic stimulations Br. J. Ophthalmol., September 1, 2006; 90(9): 1183 - 1187. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Montezuma, J. Loewenstein, C. Scholz, and J. F. Rizzo III Biocompatibility of materials implanted into the subretinal space of yucatan pigs. Invest. Ophthalmol. Vis. Sci., August 1, 2006; 47(8): 3514 - 3522. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Sekirnjak, P. Hottowy, A. Sher, W. Dabrowski, A. M. Litke, and E. J. Chichilnisky Electrical Stimulation of Mammalian Retinal Ganglion Cells With Multielectrode Arrays J Neurophysiol, June 1, 2006; 95(6): 3311 - 3327. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Dagnelie, D. Barnett, M. S. Humayun, and R. W. Thompson Jr Paragraph text reading using a pixelized prosthetic vision simulator: parameter dependence and task learning in free-viewing conditions. Invest. Ophthalmol. Vis. Sci., March 1, 2006; 47(3): 1241 - 1250. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. I. Fried, H. A. Hsueh, and F. S. Werblin A Method for Generating Precise Temporal Patterns of Retinal Spiking Using Prosthetic Stimulation J Neurophysiol, February 1, 2006; 95(2): 970 - 978. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Jensen, O. R. Ziv, and J. F. Rizzo III Thresholds for Activation of Rabbit Retinal Ganglion Cells with Relatively Large, Extracellular Microelectrodes Invest. Ophthalmol. Vis. Sci., April 1, 2005; 46(4): 1486 - 1496. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. I. Loewenstein, S. R. Montezuma, and J. F. Rizzo III Outer Retinal Degeneration: An Electronic Retinal Prosthesis as a Treatment Strategy Arch Ophthalmol, April 1, 2004; 122(4): 587 - 596. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. Rizzo III, J. Wyatt, J. Loewenstein, S. Kelly, and D. Shire Perceptual Efficacy of Electrical Stimulation of Human Retina with a Microelectrode Array during Short-Term Surgical Trials Invest. Ophthalmol. Vis. Sci., December 1, 2003; 44(12): 5362 - 5369. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |