1From the Department of Ophthalmology, Herlev Hospital, University of Copenhagen, Herlev, Denmark; 2Steno Diabetes Center, Gentofte, Denmark; and 3Retinalyze A/S, Hørsholm, Denmark.
| Abstract |
|---|
|
|
|---|
METHODS. This was a retrospective cross-sectional study of 260 consecutive nonphotocoagulated eyes in 137 diabetic patients attending routine photographic retinopathy screening. Mydriatic 60° fundus photography on 35-mm color transparency film was used, with a single fovea-centered field. Routine grading was based on visual examination of slide-mounted transparencies. Reference grading was performed with specific emphasis on achieving high sensitivity. Computer-assisted automated red lesion detection was performed on digitized transparencies.
RESULTS. When applied in a screening population comprising patients with diabetes with untreated diabetic retinopathy in any eye and patients with diabetes without retinopathy, the automated lesion detection correctly identified 90.1% of patients with retinopathy and 81.3% of patients without retinopathy. A per-eye analysis for methodological purposes demonstrated that the automated lesion detection could be adapted to simulate various visual evaluation strategies. When adapted at high sensitivity, the automated system demonstrated sensitivity at 93.1% and specificity at 71.6%. When adapted at high specificity the automated system demonstrated sensitivity at 76.4% and specificity at 96.6%, closely matching routine visual grading at sensitivity 76.4% and specificity 98.3%.
CONCLUSIONS. Automated detection of untreated diabetic retinopathy in fundus photographs from a screening population of patients with diabetes can be made with adjustable priority settings, emphasizing high-sensitivity identification of diabetic retinopathy or high-specificity identification of absence of retinopathy, covering opposing extremes of visual evaluation strategies demonstrated by human observers.
In the present study we examine the ability of an automated image-analysis system for correct classification of patients as having or not having diabetic retinopathy in a series of untreated eyes with either no retinopathy, mild to moderate nonproliferative diabetic retinopathy, or untreated proliferative diabetic retinopathy. The assumption is that this task is practically meaningful in screening services, because patients who have undergone any treatment by photocoagulation or vitrectomy in either of their eyes should continue to be observed by ophthalmologists, and that these patients will not be considered eligible for automated photographic retinopathy screening. The study involved a consecutive series of fundus photographs from the retinopathy screening service of a large diabetology clinic.
In clinical practice, diabetic retinopathy screening concerns individual patients (pairs of eyes) rather than single eyes, because the decision to refer the patient for ophthalmic examination or treatment is made if one eye has reached or exceeded a threshold level of retinopathy. Consequently, as our primary end point, we evaluated the systems performance in relation to patients, not eyes. The specific objective was to test the ability of automated lesion analysis to identify patients with any diabetic retinopathy in either of their eyes in the subpopulation of diabetic patients who had not previously undergone retinal photocoagulation treatmentthe rationale being that patients who have undergone photocoagulation in either eye should be designated a priori to visual retinopathy grading.
For methodological purposes, a per-eye analysis is relevant when seeking to characterize system performance in relation to specific diseases. Consequently, we used a per-eye analysis to evaluate variation in performance in relation to variation in the lesion-detection sensitivity as controlled by the user-supplied visibility threshold parameter.
The development of the automated red lesion detection applied in this study was based exclusively on fundus photographs selected from other diabetic populations. Images from the screening clinic involved in the present study were not used in the development of the automated lesion-detection system; thus, the results of this study serve as validation of the results presented in a previous study.1
| Materials and Methods |
|---|
|
|
|---|
From the photographic archives, 365 macula-centered photographs obtained in eyes of the 184 patients were retrieved. Thus, three patients were represented by photographs from only one eye each because the fellow eye was blind. Among the remaining 365 eyes, four eyes in four patients had previously been exempted from screening because media opacities precluded adequate visualization of the fundus (two eyes with cataract, one eye with vitreous hemorrhage, and one eye with asteroid hyalosis). Photocoagulation treatment had been performed in both eyes in 43 patients and in one eye in 12 patients. Three eyes in three patients were not included because of decentration of the photographs. One fundus photograph from each of the remaining 260 eyes in 137 patients was included in the study. It follows that 14 of the 137 patients were represented by one eye, because photocoagulation had been applied in the fellow eye (eight patients), the patient had one blind eye (one patient), the photographs of the fellow eye were decentered (two patients), or the eye had previously been exempted from photographic screening because of media opacities (three patients).
The Steno Diabetes Center serves more than 4000 patients with type 1 and more than 1000 patients with type 2 diabetes from the greater Copenhagen area. More than 80% of the patients are systematically screened for diabetic retinopathy by fundus photography under a predefined clinical protocol that was established 12 years before the initiation of the study.
The present study was entirely retrospective and complied with the tenets of the Declaration of Helsinki. The study did not involve patients or biological samples and as such did not require institutional review under Danish law.
Photography
Fundus photography was made after pupil dilation with one or more drops of phenylephrine hydrochloride 2.5% and/or tropicamide 1%, a fundus camera (CF-60UV; Canon Europa NV, Amstelveen, The Netherlands) set at 60° angular field of view, and 35-mm color transparency film (Ektachrome Elite 100; Eastman Kodak, Rochester, NY). A stereoscopic set of photographs was recorded, if possible, from of the macular region of each eye.
Digitization
The fundus photographic 35-mm diapositive slides were digitized (Coolscan LS-2000 film-slide digitizer; Nikon Corp., Tokyo, Japan) at 1350 dpi and 12 bits per color channel. For visual inspection the color and contrast of the digitized fundus images were enhanced according to an automated procedure involving full color stretching separately for each color channel. The full image size was 1947 x 1296 pixels, and the diameter of the circular fundus region defined by the internal mask of the camera was approximately 1530 pixels. The images were displayed on a 21-in. color monitor (Professional Series PT813; ViewSonic Corp., Walnut, CA).
Routine Retinopathy Grading
The slides were originally graded in November and December of 1999 by visual examination of slide-mounted transparency pairs. Ideally, all eyes were represented by stereoscopic pairs of macula-centered fundus photographs. Stereoscopic examination was not possible when no stereo separation was visible or when one of the photographs from the stereo pair was of inferior quality. The grading was made by a retina specialist who assigned each macula-centered photographic field a routine screening grade of diabetic retinopathy, using a multilevel grading system, of which the lowest retinopathy grade was one to five microaneurysms per eye in the absence of other elements of nonproliferative diabetic retinopathy. The eyes were classified as to whether diabetic retinopathy was present or absent in the macula-centered photographic field. The routine screening was conducted before initiation of the study and is representative of the high-volume grading of the institution. The average net time available for review of a set of fundus photographs from both of a patients eyes has previously been found to be 10 minutes (Larsen N, unpublished data, 2001). The routine screening grades, the indication for photocoagulation, and other clinical information were retrospectively obtained from the clinical database.
Reference Retinopathy Grading
A reference grading of all 260 macula-centered slides was made by an experienced ophthalmologist given the task of classifying each eye as having or not having diabetic retinopathy. The photographs were shown in random order of presentation. The reference grader was allowed as much time as needed to review the slides, and the task was repeated after an initial training session including the entire series. The reference grading was made with the aid of magnifiers and a slide projector.
Automated Lesion Detection
A commercial computer software system for automated fundus image analysis (RetinaLyze System; Retinalyze A/S, Hørsholm, Denmark) was used to provide automated red microaneurysm and hemorrhage lesion detection. The systems uses advanced modeling of the gray-level image function of digital images, primarily the green color channel, with custom-developed algorithms that identify dark fundus areas in the color fundus image.1 Only areas that are not contiguous with the branches of the central retinal artery or vein and are not within 1/4 optic nerve head diameter distance of the optic nerve head were considered.
A single, user-supplied parameter, the visibility threshold, controls the lesion-detection sensitivity of the system. The visibility threshold is the cutoff level for identification of a lesion in a three-dimensional matrix representation of the fundus image after complex numerical filtering, including adjustment for uneven illumination, and removal of the optic nerve head and the trunk vessels of the retina. Practical considerations and improvements have motivated the adjustment of the default visibility threshold from 1.2 to 1.0. The former value was used in a previous study.1 The default visibility threshold is the value used by the system when the user has not explicitly specified the threshold. The default visibility threshold represents a preliminary optimization for use in screening of patients with diabetes for diabetic retinopathy, at a high sensitivity priority for detection of patients with retinopathy. No official approval for such use has been obtained at this writing. A scale parameter has also been introduced in the automated lesion-detection algorithm, to specify the resolution of the digital fundus image. This is given as the expected size of the diameter of the optic nerve head measured in pixels (vertical optic nerve head diameter). The default value of the optic nerve head scale parameter of 200 pixels was used throughout the study.
In the first round of analysis, the visibility threshold of the system was used at its default setting. After completion of the first round of analysis, the capacity of the system to adapt to variations in screening priorities was tested by adjusting the visibility threshold to fit the two different visual grading strategies represented in the study. First, the visibility threshold was adjusted to fit the high sensitivity of the reference grading. The objective was to measure the loss of specificity when the automated lesion-detection system was set to achieve per-eye sensitivity above 90%. Second, the visibility threshold was adjusted to fit the sensitivity of the automated lesions detection to that of the routine grading. In other words, the sensitivity of the automated lesion detection was set equal to that of the routine grader, thus enabling a one-dimensional comparison of performance by comparing the respective specificities of the computer and the routine grader.
Computer-Assisted Retinopathy Grading
Until the completion of the study procedures, the routine grader was involved only in routine grading. Finally, the routine grader was given the task of reviewing all 260 fundus photographs included in the study on the computer display. The review was assisted by the automated lesion-detection system, which displayed the fundus image and the outline of each automatically identified lesion as identified at the high-sensitivity setting, as described earlier. This outline was shown as an overlaid image that could be toggled on and off at the discretion of the grader. This part of the study was appended after completion of the primary studies to explore the potential for methodological development of the automated lesion-detection algorithm.
Statistical Analysis
Patients without prior photocoagulation treatment in either eye were classified as having automatically detected diabetic retinopathy if a single red lesion of any type was identified by the automated lesion-detection algorithm in either of the patients eyes. The per-patient performance of the automated lesion-detection algorithm was characterized by the specificity, sensitivity, and accuracythat is, the fraction of the entire population that was correctly classifiedas measured against the high-sensitivity reference grading categorizing patients as having no diabetic retinopathy or any diabetic retinopathy. The receiver operating characteristic (ROC) described the relationship between sensitivity and specificity, with the area under the curve serving as a general measure of performance.2
Eyes without prior photocoagulation treatment were classified as having automatically detected diabetic retinopathy if a single red lesion of any type was identified by the automated lesion-detection algorithm. The per-eye performance of the algorithm was characterized by the specificity, sensitivity, and accuracy measured against the high-sensitivity reference grading categorizing eyes as having no diabetic retinopathy or any diabetic retinopathy.
The routine screening grading was collapsed into a binary classification of no diabetic retinopathy and any diabetic retinopathy. The specificity, sensitivity, and accuracy of the routine grading were measured against the high-sensitivity reference grading.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The United Kingdom Prospective Diabetes Study (UKPDS) identified systemic parameters permitting stratification of patients with identical levels of retinopathy into subgroups with variable progression rates.5 Such information, if accessible in screening clinics, may enable a refined estimation of appropriate follow-up intervals for individual patients in retinopathy screening. This should minimize the concern that lesions in the far periphery, visible only by binocular contact lens biomicroscopic examination, are missed when only photographic methods are used for retinopathy screening. Ultimately, the choice of screening method must be based on an overall assessment of the options available in a given health care environment.
In the present study, we demonstrated the practicability of using automated digital detection of red lesions in photographic diabetic retinopathy screening. The objective was to sort patients who have not undergone photocoagulation treatment for diabetic retinopathy in two groups: those without retinopathy and those with possible or definite retinopathy who must have visual grading of their fundus photographs.
An automated prescreening system that automatically categorizes patients with diabetes into those who should have their photographs visually graded and those whose photographs can reliably be handled automatically should be able to accomplish a series of functions. These include identification of images that are unsuitable for automated lesion detection (i.e., low-quality images caused by refractive media opacity or technical failures in the photographic process). Photographs with such characteristics should be automatically designated to visual inspection. Indeed, a preliminary evaluation of an automated quality-assessment algorithm developed subsequent to the completion of this study has demonstrated a significant potential for increasing the sensitivity of the automated classification by elimination of several of the false-negative gradings in this study, one of which had proliferative diabetic retinopathy (Table 2) . Additional enhancement appears to be possible by incorporating automated detection of bright fundus lesions. Thus, one eye with a false-negative grade (Table 2) had bright lesions only.
An ancillary finding was that the quality of high-volume visual routine grading was enhanced considerably when the attention of the routine grader was directed to potential fundus lesions by the automated lesion-detection system. This part of the study was appended after completion of the primary studies. The results should be considered indicative of a potential for methodological development, but the tentative conclusion should be verified in a larger prospective study.
Clinical ophthalmology emphasizes the detailed assessment of retinopathy at or above treatment thresholds. The task of making a fine distinction between presence and absence of any retinopathy is rarely addressed outside epidemiologic studies. Clinical ophthalmology practice is not traditionally subjected to validation or quality control, and it is therefore difficult to find standards against which new methods can be compared. Even when validation is attempted, there is usually no quality-assurance system in place to ensure the widespread application of certified standards of clinical ophthalmology. It appears that technological progress may enable application of such systems, even in complex clinical fields such as ophthalmology.
Populations that undergo systematic fundus photographic screening have been found to have a favorable and improving visual prognosis.6 Consequently, the task of automatically assisting the separation of patients with retinopathy from patients without retinopathy is a clinically meaningful one, and our results warrant further study of the methods practical applicability in clinical screening practice.
| Footnotes |
|---|
Submitted for publication April 29, 2002; revised August 7, 2002; accepted August 19, 2002.
Commercial relationships policy: E (JG, MG); C (ML, NL, HL-A); P (MG).
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Nicolai Larsen, Department of Ophthalmology, Herlev Hospital, DK-2730 Herlev, Denmark; nicolai.larsen{at}dadlnet.dk.