1From the Department of Cell Biology, Neurobiology, and Anatomy, Medical College of Wisconsin, Milwaukee, Wisconsin; 2The Jackson Laboratory, 3The Howard Hughes Medical Institute, Bar Harbor, Maine.
| Abstract |
|---|
|
|
|---|
METHODS. Servo-null electrophysiology was used to measure IOP in anesthetized adult zebrafish. A pulled-glass microelectrode was used to penetrate the cornea into the anterior chamber, and pressure transduction was recorded. IOP, histology, and visual behavioral assays were used to assess the brass mutant.
RESULTS. Significant differences in IOP were detected between genetically distinct zebrafish strains. IOP averages were highest in the long fin (LF) strain (20.5 ± 1.2 mm Hg) and lowest in the Oregon AB (AB) strain (10.8 ± 0.3 mm Hg). In brass mutants, which show a reduced melanin phenotype, IOPs were elevated and mild iris hypoplasia was noted. No defects were observed within the retina or in visual function.
CONCLUSIONS. In the present study, an accurate method was developed for measuring IOP in adult zebrafish. Both the ability to measure IOP in the genetically powerful zebrafish model and the conservation of elevated IOP and anterior segment anomalies between albino mammals and hypopigmented fish are important. They strongly support the use of zebrafish for studying the complex genetics of elevated IOP and glaucoma.
To lay the groundwork for using zebrafish to study the genetics of elevated IOP and its relationship to other glaucoma phenotypes and to extend our previous work with mice,9 10 11 we sought to establish and validate a procedure for measuring IOP in zebrafish. The small size that makes zebrafish advantageous for genetic studies, makes physiological experiments challenging. For example, the anterior segment, in which IOP is measured, is on average 1.4 mm in diameter and has an average maximum depth of 0.25 mm at the peripheral angle. The small surface area and relative thickness of the zebrafish cornea (4258 µm at the region adjacent to the angle) preclude the use of instruments such the handheld tonometer (TonoPen; Mentor, Norville, OH). To measure IOPs in the small eyes of zebrafish, we used servo-null electrophysiology to quantify the hydrostatic pressure in the anterior chamber. With this technique, pressures of aqueous solutions are measured by maintaining in dynamic equilibrium the conductivity at the tip of a sensing electrode. This is accomplished by integrating external positive and negative air pressure sources with a servo-null amplifier, piezoelectric valve, and pressure transducer (Fig. 1) . Servo-null electrophysiology is well established for measuring pressures in rat kidney tubules, small-diameter blood vessels, and the chick embryonic heart.12 13 More recently, a servo-null method has been used to measure IOP in the mouse eye.14 Here, we report a reliable servo-null approach for measuring IOP in zebrafish. Zebrafish studies provide an important complement to those in other species. They are expected to allow more rapid and less expensive genetic studies of IOP and associated glaucoma phenotypes than are possible, even using mice.
|
| Materials and Methods |
|---|
|
|
|---|
Experimental Design
Adult zebrafish, 10 to 14 months old and 24 to 34 mm in length, were used for measuring IOP. Eye diameters measured from 1.8 to 2.1 mm. Fish were anesthetized in 400 mL of tricaine (0.16 mg/mL ethyl 3-aminobenzoate methanesulfonate) for approximately 2 minutes. Each fish was anesthetized one at a time until they were unconscious as judged by lack of response to physical stimuli. IOP was measured as soon as the fish failed to respond to physical stimuli. If necessary, anesthesia was maintained during IOP measurement by applying tricaine directly to the gills with a Pasteur pipette.
When anesthesia was obtained, the fish was placed on top of a sterile sponge that rested in a dish of tricaine. The sponge acted to keep the protective mucosal coating of the fish moist and to maintain anesthesia. The AgCl reference electrode was then placed underneath the zebrafish. A drop of tricaine was placed on the eye, and a micromanipulator was used to advance the recording electrode into the drop of tricaine. Once the recording electrode was positioned above the eye, the null meter on the servo-null device was zeroed by setting the resistance at the tip of the electrode to 8 to 10 k
. The micromanipulator was then used to advance the recording electrode through the cornea overlying the pupil and into the anterior chamber. The position of the electrode was visualized under a dissecting microscope, and care was taken to ensure that the shank did not cause corneal deformation and that the electrode tip was not in contact with the lens.
Once the position of the electrode was verified in the anterior chamber, the junction between the electrode and cornea was further sealed with a drop of cyanoacrylate adhesive (Elmers Products, Inc., Columbus, OH). This rapidly formed a hardened barrier around the junction of the electrode and corneal surface. The IOP was then recorded at a rate of 240 Hz for a period greater than 30 seconds. The needle was then quickly removed from the eye and the recording electrode was returned to the tricaine bath to ensure that the null meter remained zeroed. The polymerized adhesive was then removed from the eye with a pair of number 5 student forceps, and the fish was transferred and revived in a holding tank of fresh fish water.
Instrumentation
Electrodes.
Recording electrodes were made with borosilicate glass capillaries (TW100 to 4, 1.0 outer diameter [OD]/0.75 inner diameter [ID]; World Precision Instruments, Sarasota, FL) pulled with a pipette puller (Flaming Brown P-85; Sutter Instruments, Novato, CA). The electrodes were designed to come to a fine point very quickly to create a short, strong shank that could cleanly penetrate the zebrafish cornea. The tips of the electrodes were broken by hand with forceps to an approximate ID of 10 µM. Electrodes were back filled with 3 M KCl, by using 35-gauge microfill needles (World Precision Instruments). These electrodes displayed a resistance of 0.8 to 0.1 M
when introduced into the tricaine bath solution.
Servo-null Micropressure System.
The recording electrode was connected to a commercially available servo-null system (Model 900A, Micropressure System; World Precision Instruments). This system consists of a recording electrode, a reference electrode, a pressure transducer, and a servo-null amplifier. The servo-null system constantly measures the resistance to electrical flow at the tip of the KCl-filled glass electrode. When the recording electrode is inserted into the anterior chamber, hydrostatic forces within the eye push aqueous humor into the electrode tip. As the aqueous humor/KCl interface moves within the electrode, the servo-null device measures the change in resistance and responds via the pressure transducer and piezoelectric vacuum/pressure pump. This device creates a counter pressure that is equal to the hydrostatic pressure within the anterior chamber. The analog output from the servo-null device was converted to a digital signal with a data acquisition device (DI-700; Dataq Instruments, Akron, OH). Pressures were constantly displayed on a computer monitor. All data were recorded and analyzed with waveform recording software (WinDaq; Dataq Instruments).
Criteria for Inclusion of Data Points
The following conditions were necessary for a trial to be considered successful and included in the final data set of IOPs:
0.05. Approximately 67% of IOP measurement trials were successful based on the above criteria. The most common reason that a trail was unsuccessful was an inadequate seal between the cornea and the recording pipette. However, application of the criteria ensured that only recordings in which rapid and tight junctional seals had been achieved were included for analysis. Of the trials presented herein for establishing mean IOPs in different strains, >80% required less than 1 minute to establish a stable IOP.
Histologic Analysis and Morphometrics
Eyes were removed from adult fish after terminal anesthesia and fixed overnight at 4°C in 2.5% gluteraldehyde/1% paraformaldehyde in phosphate-buffered sucrose (pH 7.4). The next morning, eyes were dehydrated and infiltrated with Epon/Araldite. Transverse sections 1 to 2 µm thick were heat mounted on glass slides and stained with 1% methylene blue/1% azure blue II in 1% borax.
For ocular morphometric measurements, eyes from five wild-type fish and five mutant fish were obtained to cut central ocular sections. To assess iris thickness, measurements were taken from the three most central, sagittal sections, as judged by maximum lens diameter and proximity to the optic nerve. These were averaged and constituted one data point. Iris thickness was measured halfway from the tip at the pupil to the base, where the iris forms the angle with cornea. Retinal thickness was measured from the edge of the outer nuclear layer to the edge of the inner limiting membrane. Cells within the retinal ganglion cell (RGC) layer were quantified by counting nuclei within the RGC layer from five central retinal fields of view (500 µm in width). A 40x objective was used for all morphometric analysis.
Visual Behavioral Assays
Optokinetic Reflex Assay.
The optokinetic reflex (OKR) assay was performed as previously described.15 Briefly, 7-day-old fish were placed in a small glass beaker containing 4% methyl cellulose to partially immobilize the larvae. The beaker was then placed within a rotating drum which contained alternating black and white vertical stripes on the inside (15° wide). For each larva, the drum was rotated in both directions for 15 seconds each, and the number of saccades during a 1-minute trial was scored.
Startle-Response Assay.
The startle-response (SR) assay was performed as previously described.16 Briefly, adult fish were placed in transparent rectangular containers and allowed to acclimatize for 30 minutes before testing. When fish faced the observer, black paper was passed in front of their field of view, a stimulus that normally elicits an escape response. The number of times the fish displayed a rapid-escape response during 10 startles was scored. The time required for testing varied from 3 to 8 minutes per fish. All tests were performed in 200-lux broad-spectrum illumination.
| Results |
|---|
|
|
|---|
|
We next investigated whether inserting and sealing a recording electrode into the zebrafish anterior chamber resulted in damage to structures within the eye. After test IOP recordings, three fish were allowed to recover for 2 weeks. During this time, the fish were visually inspected for signs of trauma or infection. These fish were then terminally anesthetized and both control and electrode-penetrated eyes were removed for histologic analysis. Slight corneal scaring was detected in only one of the electrode-penetrated eyes. No damage was detected to other components of the eyes, and there were no signs of inflammation (data not shown).
Twofold Difference in IOP between Distinct Inbred Zebrafish Strains
After validation of the procedure for measuring IOP in zebrafish, we tested whether differences could be detected between genetically different strains of zebrafish. In the mouse, various inbred strains have been shown to have significantly different IOPs.9 10 A representative IOP recording from zebrafish is shown in Figure 2C . The following strains, Oregon AB (AB), Long Fin (LF), Ekkwill (EK), Tübingen (TU), and Tübingen Long Fin (TL) were analyzed for average IOP (Fig. 3) . Overall, we measured IOPs of 14 fish per family from the five strains (Fig. 4) . These strains are approximately 70% percent different across the genome based on a panel of microsatellite repeat sequences. This degree of nucleotide polymorphism has been invaluable for positional cloning projects to identify the genetic basis of phenotype variability.3 4 For the strains tested, average IOPs varied from 10.8 to 20.6 mm Hg (Table 1) . Statistical analysis indicated that pair-wise comparisons between most strains showed significant differences in IOPs (P < 0.0001, Students t-test; Table 1 ). This twofold difference is very similar to that reported between distinct mouse strains and to the range of IOPs in the normal human population.10 18
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Very Similar IOP Distributions in Mammals and Zebrafish
An approximately twofold range of IOP exists in different mammalian species. In mice, IOP between strains ranges from approximately 10 to 20 mm Hg, with different strains having statistically significant differences in mean IOP.10 The mean IOP considering mice of all inbred strains included in a large study was 14.5 mm Hg.10 Similarly, within a nonselected sample of 5941 normal people, IOP varied from 10.5 to 20.7 mm Hg (within the ±2-SD range), and the mean IOP was 15.65 mm Hg.28 Subsequent clinical studies have shown similar results,18 and human IOP differs between isolated populations.2 Herein, we report zebrafish IOP and show that in this nonmammalian vertebrate, both the range of IOP and mean IOP are remarkably similar to mammalian values. IOP ranges from 10.8 to 20.5 mm Hg between distinct fish strains. The strain with the lowest IOP was AB (10.8 ± 0.3 mm Hg), whereas the strain with the highest average IOP was LF (20.5 ± 1.2 mm Hg). The mean IOP across all reported zebrafish strains is 16.5 mm Hg. The striking similarity in normal IOP distributions between zebrafish and mammals supports the use of zebrafish for genetic analyses of IOP and other glaucoma phenotypes.
Effect of Pigmentation Mutations on IOP in Mammals and Fish
Analysis of IOP in brass mutants, which are hypopigmented, suggests further physiological conservation between zebrafish and mammals. We have shown that albino mice of strain C57BL/6J have higher IOPs than otherwise genetically identicalbut pigmented C57BL/6J mice housed in the same environment.10 Similarly, we show here that the average IOP of hypopigmented brass mutants (16.9 mm Hg) is higher than the mean value of non-mutant, strain matched fish (11.0 mHg). The observation of an intermediate phenotype in brass heterozygotes for IOP, but not for pigmentation, is interesting. This result suggests that the brass gene product may function in a dose-dependent fashion for IOP, whereas pigmentation is normal above a certain threshold of brass function. The eventual molecular identification of the brass gene is important for a better understanding of these results.
Because of their elevated IOP, brass mutants provide an important resource for future zebrafish studies for the study of glaucoma. The ability to measure IOP in zebrafish is the first step toward exploiting the zebrafish system for glaucoma research. Another important step is the generation of zebrafish models with pathologically elevated IOP and RGC death. Sensitized screens involve the use of mutant strains predisposed to a particular phenotype in order to identify other genes that can interact with the initial mutation to either exacerbate or suppress the phenotype. The brass mutation significantly elevates IOP, but does not cause RGC death. Thus, brass is a useful sensitizing mutation for use in mutagenesis screens. By crossing the brass mutation with mutagenized lines, mutations that interact with brass to cause pathologically elevated IOP and/or glaucomatous optic nerve degeneration may be identified.
Effect of Pigmentation Mutations on Ocular Development in Mammals and Fish
In mice, mutations in the tyrosinase gene cause albinism and elevated IOP.10 We recently demonstrated that the tyrosinase gene participates in anterior segment development in mice.19 We found that in mice, tyrosinase mutations modify the severity of anterior segment abnormalities caused by mutations in the mouse orthologues of two human genes (CYP1B1 and FOXC1). Both of these loci contribute to glaucoma in humans. In human albinos, the prevalence of anterior segment dysgenesis is increased.25 26 In addition to tyrosinase-deficient individuals, anterior segment dysgenesis also occurs in tyrosinase-positive individuals who often have reduced pigmentation of the eyes, skin, and hair. This suggests that additional, yet to be identified pigmentation genes can affect glaucoma phenotypes. In the current study, the hypopigmented zebrafish with the brass mutation had iris hypoplasia, consistent with a role for the pigmentation pathway in modulating anterior segment formation in fish. Similar to other species, the production of melanin in zebrafish is regulated by the tyrosinase pathway. Although the brass mutation has not been identified, linkage analysis indicates that the affected gene is not tyrosinase itself.23 It will be interesting to identify the genetic lesion in brass mutants and to test whether the brass mutation can modify other zebrafish anterior segment dysgenic phenotypes. The importance of pigmentation status for normal ocular development provides further evidence for conservation of molecular pathways between mammals and zebrafish.
In conclusion, by developing a technique to measure IOP in zebrafish, we have laid the foundation for studying the complex genetics of elevated IOP and glaucoma in this genetically attractive organism. We anticipate that zebrafish studies will complement studies in various species. The coupling of forward and reverse genetic approaches in zebrafish and mice provides a powerful and synergistic approach for unraveling the complexities and relationships of glaucoma phenotypes.
| Footnotes |
|---|
Submitted for publication May 19, 2004; revised August 15, 2004; accepted September 4, 2004.
Disclosure: B.A. Link, None; M.P. Gray, None; R.S. Smith, None; S.W.M. John, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.