1From the Istituto di Oftalmologia, Università di Catania, Catania, Italy; the 2Institut de Recherche en Ophthalmologie, Sion, Switzerland.
| Abstract |
|---|
|
|
|---|
METHODS. The pulsatile, nonpulsatile, and mean ChBF were measured with laser Doppler flowmetry in 11 healthy volunteers with a mean age of 32 ± 13 (SD) years. The posture of the subjects was changed from standing (90°), to supine (-8°), and back to standing, with a mechanically driven table. During the whole experimental procedure, ChBF and heart rate (HR) were continuously recorded. After 30 seconds in standing position, the subjects were tilted to supine during approximately 30 seconds. They remained in this position for approximately 2 minutes, after which they were tilted back to the standing position (recovery), where they remained for another approximately 2 minutes. Systemic brachial artery blood pressure (BP) was measured in the baseline, supine, and recovery positions. This procedure was repeated to measure the intraocular pressure (IOP) at the different postures.
RESULTS. Mean BP did not change significantly throughout the experimental procedure. As the body was tilted from standing to supine, HR decreased by 16% (P < 0.0004), IOP increased by 29% (P < 0.001), and mean ChBF increased by 11% (P < 0.01). The increase in ChBF was primarily due to an increase in the nonpulsatile component of the blood velocity.
CONCLUSIONS. Based on previously reported experimental data that indicate that the ocular perfusion pressure increases less than predicted by purely hydrostatic considerations when the body is tilted from the standing to the supine position, the observed increase in ChBF suggests a passive response of the choroidal circulation to the posture change.
Body inversion represents a simple and convenient method of changing the ocular perfusion pressure. For that reason, it has been widely used as a provocation test for the ocular circulation. Several studies, in which a variety of techniques were used, have described the effect of body position on the orbital, retrobulbar, optic nerve, and retinal circulations.1 2 3 4 5 6 7 8 9 10 11 For the choroidal circulation, studies have been limited to the assessment of the effect of postural change on the pulsatile component of ChBF, as assessed by the POBF system.12 13 To the best of our knowledge, no information is available on the effect of posture change on the nonpulsatile and mean components of choroidal blood flow in humans. In this work, we used continuous, confocal laser Doppler flowmetry (LDF) to assess the effect of posture change on the three components of subfoveal choroidal blood flow (ChBF).
| Materials and Methods |
|---|
|
|
|---|
Posture Change and Systemic Blood Pressure
An electrically driven tilting table for radiology (DuoDiagnost; Philips, Eindhoven, The Netherlands) was rotated from 90° to -8° and back to 90° to tilt the subjects from the standing to the supine position and back to the standing position (recovery). The motion from the standing to the supine position and vice versa took approximately 30 seconds. Subjects were connected to the table using straps to prevent them from sliding (Fig. 1) . Preliminary trials were first performed to familiarize the subject with the procedure and verify that stable LDF recordings could be obtained during the posture changes.
|
The mean brachial artery pressure was calculated according to the formula3 : BPm= BPdias+ 0.42 · (BPsys - BPdias), where BPdias and BPsys are the diastolic and systolic brachial artery blood pressures, respectively.
Laser Doppler Flowmetry
The principle of ChBF measurements by LDF and the flowmeter used in this work have been described.14 The flowmeter was mounted on a mechanical support fixed on the tilting table and placed in front of the eye. This device delivered a laser beam from a laser diode (
= 785 nm, 90 µW at the cornea) through a confocal optical system. Nominal diameter of the beam at the fundus, when focused, was approximately 12 µm. The light scattered by the red blood cells and surrounding tissue was collected by six fibers, each having a core diameter of 110 µm. These fibers were arranged on a 180-µm diameter circle centered on the incident beam. The collected light was guided to an avalanche photodiode, and the output current was amplified before being analyzed by computer (NeXT; Redwood, CA).15 The LDF photocurrent was fed into a loudspeaker to obtain an audio signal. Its DC, which is proportional to the intensity of the scattered light, was also recorded and displayed.
Subjects were asked to fixate directly on the probing laser beam, which appeared as a weak but clearly visible tiny spot, and to focus it by adjusting the front lens of the flowmeter. Then the operator precisely aligned the device with the eye to obtain maximum loudness of the audio Doppler signal and constant DC during the experiment.
The subfoveal LDF parameters calculated from the analysis of the photocurrent are the velocity, ChBVel (kHz), the volume, ChBVol (arbitrary units [au]) and the flux, ChBF (au), of the red blood cells. These parameters are related to each other through the relationship ChBF = k · ChBVel · ChBVol, where k represents an instrumental constant. Assuming no change in hematocrit during the experiments, the changes in the LDF parameters are proportional to the changes in the actual velocity, volume, and flow of blood in the sampled tissue, respectively. Artifacts caused by rapid eye movements and blinks were easily identifiable in the recordings. They were manually removed, as described in a previous publication.16
By recording the heart pulse using an ear infrared plethysmometer, the average time course of each LDF parameter during the cardiac cycle was obtained by averaging over a period T, taking into account the phase of the heart cycle, as previously described.17 This average time course for ChBVel, which we denote as ChBVelT in Figure 2 , consists mainly of a nonpulsatile and most often of a much smaller pulsatile component. The magnitude of the nonpulsatile component is the minimum of ChBVelT, that of the pulsatile component as the mean of ChBVelT minus this minimum.
|
The baseline of the LDF parameters were defined as the average values of these parameters obtained during the approximately 30 seconds of measurement time in the starting (standing) position. The averaged data obtained during the last minute in supine position were defined as supine LDF parameters and those obtained after 40 seconds of recovery time were averaged to obtain the recovery LDF parameters.
Statistics
Mean changes in the LDF parameters were assessed for significance using the Students paired t-test. P
0.05 was considered significant.
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Tilting the body from standing to supine changed the systemic parameters in accordance with previously reported findings.3 9 Quantitatively, the decrease in BPdias of 7.4% is similar to the 6.6% found by Evans et al.11 and the 4.4% reported by Sayegh and Weigelin.3 Confirming previous findings, we found no change in BPsyst. The 15.8% decrease in heart rate in our study (8269 bpm) is practically identical with that found by James and Smith13 (7059 bpm) but larger than the 10.3% and 7.9% reported by Savin et al.10 (7769 bpm) and Evans et al.11 (7670 bpm), respectively. The differences between our study and the data of others are, most probably, only of statistical nature.
Regarding the IOP, Kothe9 has summarized the changes in IOP occurring between upright and supine obtained by previous investigators. These changes ranged from 0.5 to 4.4 mm Hg. Thus, the average 4-mm Hg increase observed in our subjects lies within this range.
According to Friberg et al.25 this increase in IOP between standing and supine postures appears to be closely related to increased venous pressure in the orbit and possibly to increased choroidal blood volume. This increase in volume may not translated into an increase in ChBVol, because it probably occurs in the large vessels rather than in the choriocapillaristhe region sampled by LDF.
Our results demonstrate a significant increase in the blood volume at recovery. The reason for this increase is not clear. Longer measurements during recovery may help determine the duration of this effect and whether it is due to some spatial redistribution of blood after the decrease in IOP on the return to the standing position.
The tilting of the body from standing to supine increased ChBF by an average of 11%. This increase was mainly due to a statistically significant 8% change in the velocity. This flow increase was entirely reflected in the increase of the nonpulsatile component of the flow. For each LDF parameter, the change in the pulsatile component was not significant, most probably because this component is much smaller than the nonpulsatile component, resulting in a greater variability of its measurement. The reason for this is that the LDF signal originates predominantly from the choriocapillaris,18 where pulsatility is markedly attenuated compared with that in the choroidal arteries.
Pulsatile ocular blood flow, as measured by POBF, decreases from the seated to the supine position.12 13 26 Based on data compiled by Kothe,9 we calculated a mean decrease of pulsatile ocular blood flow of 19% ± 5% from the standing or sitting to the supine posture. We presume, as mentioned before, that the pulsatile component of ChBF in our measurement was too low (11.2%) to reveal such an effect. By comparison, the POBF technique is based on the pulsatility in the large vessels which is markedly higher than in the choriocapillaris. Thus, based on the value of the resistance index (RI; [(peak systolic velocity peak diastolic velocity)/peak systolic velocity]) of 0.68 that was found by Kaiser et al.27 in the short and long posterior ciliary arteries, we estimated that the pulsatile component of the velocity in these vessels is approximately 55% of the mean velocity.
James and Smith13 have assumed that the decrease in pulsatile ocular blood flow is associated with an increase of the nonpulsatile part. Our results confirm this increase in nonpulsatile flow, which is primarily due to an increase in velocity. An increase in velocity was also observed in the ophthalmic artery by Doppler ultrasonography.12
Mean ocular perfusion pressure is defined as OPPm = OABPm - IOP, where OABPm is the mean ophthalmic artery pressure. In this study, OABPm was not measured. Therefore, to assess OPPm in standing and supine postures, we used values of OABPm based on previous studies, involving postural changes, but where ophthalmodynamometric measurements were performed in healthy volunteers under these conditions. From the data of two studies of Sayegh and Weigelin3 28 performed on a total of 254 normal volunteers, we derived a relationship between OABPm and BPm. In upright position OABPm = 0.74 · BPm, and in supine position OABPm= 0.84 · BPm. Applying these values to our BPm and IOP data, we obtain OPPm= 57 mm Hg in the standing and 63.6 mm Hg in the supine positionthat is, an increase in OPPm of 11.6% between these two postures. This value is very similar to the 11% change observed in our study and suggests some passive response of ChBF to the increase in OPPm.
Bill29 has theoretically assessed the expected OPPm in the standing and supine positions in normal subjects. From his analysis, OPPm= BPm -
Pf -
Ph - IOP.
Pf is the loss of pressure (assumed by Bill to be 5 cm H2O, i.e., approximately 4 mm Hg), because of the flow resistance of the vessels between the heart and the eye.
Ph is the static pressure difference of a water column with a height equal to the distance between the heart and the eye (assumed by Bill to be 40 cm, i.e., 29 mm Hg). The BPm and IOP data in Bills study, when converted were 103 and 13 mm Hg, respectively, in upright posture and 96 and 14.8 mm Hg, respectively in the supine position (eyes 10 cm above the level of the heart). These values result in a change in OPPm between upright and supine from 57 to 70 mm Hg (i.e., an increase of 23%).
Applying this analysis to our data and taking into account that in our test, in the supine position, the eyes were approximately 3 cm below the heart (which reduces the OABPm by approximately 4 mm Hg, instead of the 7 mm Hg in Bills analysis), we found that OPPm = 59 mm Hg in the standing and 79 mm Hg in the supine positionthat is, an increase of approximately 34%.
If Bills model were representative of the events occurring in the body, the change in ChBF of only 12% in the face of a 34% increase in OPPm would inevitably lead to the conclusion that an active mechanism is operating to increase choroidal vascular resistance to maintain ChBF in the supine posture close to that in standing position. The ophthalmodynamometric data of Sayegh and Weigelin,3 however, suggest that some compensatory mechanism is already acting between the heart and the eye to buffer most of the increase in the blood pressure induced by the tilting from upright to supine. This could occur in the ophthalmic artery or even at the level of the internal and common carotid arteries.10
In conclusion, our results show that mean subfoveal ChBF increases significantly (by an average of 11%) when the body is tilted from upright to supine position. This increase, which is primarily due to the nonpulsatile component of this flow, results from an increase in blood velocity. Because experimental data indicate that the ocular perfusion pressure increases by a similar percentage, which is less than expected based on purely hydrostatic considerations (Bills model), the response of subfoveal ChBF can be explained by the passive response of the choroidal vascular system to postural change.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication July 18, 2003; revised October 7, 2003; accepted October 14, 2003.
Disclosure: A. Longo, None; M.H. Geiser, None; C.E. Riva, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Martial H. Geiser, Institut de Recherche en Ophthalmologie, Grd Champsec 64, 1950 Sion, Switzerland; martial.geiser{at}hevs.ch.