(Investigative Ophthalmology and Visual Science. 2004;45:610-616.)
© 2004 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.03-0313

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (16)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Guan, K.
Right arrow Articles by Flanagan, J. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Guan, K.
Right arrow Articles by Flanagan, J. G.

Comparison of Heidelberg Retina Tomograph II and Retinal Thickness Analyzer in the Assessment of Diabetic Macular Edema

Kit Guan,1,2 Chris Hudson,1,2,3 and John G. Flanagan1,2,3

1From the Multi-Disciplinary Laboratory for the Research of Sight Threatening Diabetic Retinopathy and the 2Department of Ophthalmology, University of Toronto, Toronto, Ontario, Canada; and the 3School of Optometry, University of Waterloo, Waterloo, Ontario, Canada.


    Abstract
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 References
 
PURPOSE. To compare the within-session variability of the Macular Edema Module of the Heidelberg Retinal Tomograph II (HRT II; Heidelberg Engineering, Heidelberg, Germany) and the Retinal Thickness Analyzer (RTA, Talia Technology Ltd., Neve-Ilan, Israel) in patients with diabetes and nondiabetic subjects and to determine the agreement of both instruments to clinical observation.

METHODS. Seventeen patients with diabetic macular edema (DME) and 17 nondiabetic subjects were examined. Three scans of the posterior pole were acquired for each volunteer with both the HRT II and the RTA. The edema index and foveal average thickness were derived for a 600-µm radius circle centered on the fovea for the HRT II and RTA scans. The coefficient of variation (COV) was calculated. Clinical examination of detectable edema was performed using stereo fundus biomicroscopy and the level of agreement between each instrument and clinical observation was determined using a zonal analysis.

RESULTS. Individual COVs for the HRT II and RTA ranged from 2.3% to 24.6% (median 8.3%) and 2.1% to 46.7% (median 6.4%), respectively, in diabetic subjects and 2.0% to 37.5% (median 6.0%) and 2.3% to 14.7% (median 8.5%), respectively, in nondiabetic subjects. Clinical assessment identified edema in a total of 47 sectors in patients with DME. In comparison to clinical assessment, the HRT II gave a sensitivity of 92% and a specificity of 68% and the RTA gave a sensitivity of 57% and a specificity of 71%.

CONCLUSIONS. Both instruments have good within-session repeatability. The HRT II showed better agreement with clinical assessment than the RTA. The agreement between instruments was poor.


Diabetic macular edema (DME) is one of the leading causes of visual impairment among patients with diabetes.1 2 It results from fluid accumulation in the retina due to the breakdown of the blood–retinal barrier.3 4 The clinical assessment of DME uses contact lens stereo fundus biomicroscopy, which relies on the subjective recognition of thickening of the transparent retina by a clinician. It is problematic to differentiate early retinal thickening from normal variation in retinal thickness.5 Substantial differences exist between experienced retinal specialists in defining the extent and location of retinal thickening in a given patient,6 which illustrates the difficulty in assessing the location and extent of retinal edema.

Several objective techniques for the assessment of retinal edema have been developed including computed tomography, automated slit biomicroscopy and optical coherence tomography (OCT). The Heidelberg Retina Tomograph (HRT; Heidelberg Engineering, Heidelberg, Germany) is a confocal scanning laser tomograph that sequentially acquires two-dimensional, lateral (i.e., x, y plane) section images along the optical (z) axis. The distribution of reflected light intensity along the optical axis for a given pixel is described as the z-profile or confocal intensity profile. A new technique determines the signal width (at half peak height) and peak reflectance intensity of the z-profile. Studies have demonstrated a broadening of the z-profile signal width (SW) and a decrease in peak reflectance intensity (IN) in areas of edema.7 Normalization of the reflectance values reduces the variation in intensity between successive scans: Edema indexi = SWi/INi.

An edema index can be derived for each pixel, which is sensitive to edematous changes of the retina.7 A resultant map of these edema indices gives a measure of the location and extent of retinal edema (Fig. 1) .7 It should be noted that the edema index is not a measure of retinal thickness but of the optical effect of edema within the retina. The edema index methodology has been validated in diabetic retinopathy but not in other disease states. Change of the edema index has been shown to correlate with change of visual function, including logarithm of the minimum angle of resolution (logMAR) visual acuity, conventional automated static perimetry and short-wavelength automated perimetry, in patients undergoing grid laser treatment for clinically significant macular edema (CSME; Flanagan JG, Hudson C, manuscript submitted).7 8 9



View larger version (50K):
[in this window]
[in a new window]
 
FIGURE 1. Left: The z-profile and signal width at half maximum reflected intensity generated from optical cross sections of the HRT. In areas of edema, the peak reflectance intensity was reduced and there was a broadening of the signal width. Right: Edema map of a patient with DME after normalizing reflectance intensity. Brighter pixels represent areas of higher edema index.

 
The Retinal Thickness Analyzer (RTA, Talia Technology Ltd., Neve-Ilan, Israel) works on the principle of slit lamp fundus biomicroscopy. It sequentially scans vertical slits across an area of the retina to generate rapidly a topographic map of retinal thickness.10 11 The reflections are captured by a black-and-white charge-coupled device (CCD) camera and stored for analysis. A bi-Lorenztian curve fitting is performed to delineate the internal limiting membrane (ILM) and retinal pigment epithelium (RPE) separation with an axial resolution of approximately 50 µm (Fig. 2) .10 12 The instrument has been used extensively in the evaluation of retinal diseases including DME.13 14 15 16 17 18 The RTA has been reported to have a high level of reproducibility.12 19



View larger version (34K):
[in this window]
[in a new window]
 
FIGURE 2. Left: Bi-Lorenztian curve fitting that delineates the ILM and RPE separation of captured slits using the RTA. Right: Generated topographic map of the posterior pole. Red: thickened retina; blue: thinner retina.

 
Of the techniques used to assess retinal thickness, OCT is perhaps the most established method. As described by Huang et al.20 OCT uses low-coherence light from a superluminescent diode coupled with a Michelson interferometer to measure optical reflections from the retina. The OCT processes 100 to 512 longitudinal A-scans to generate a cross-sectional slice of the imaged area, depending on the instrument version.21 By using six radial slices, it can create a topographic map of the posterior pole.22 The instrument in its current form provides axial resolution on the order of 10 to 17 µm.20 The OCT has been used in the assessment of a variety of diseases, such as macular holes, retinal edema, and nerve fiber layer thickness.23 24 25

A newer version of the Heidelberg Retina Tomograph, the HRT II (Heidelberg Engineering), samples 384 x 384 pixels across a 15° x 15° field of view. The area scanned is approximately 4.5 x 4.5 mm on the retina and corresponds to a spatial resolution of 12 µm per pixel. In the central 500-µm radius circle, approximately 7100 points are sampled. The RTA samples 32 x 32 pixels across a 6 x 6-mm area. Its spatial resolution is 187 µm per pixel. In the central 500 µm radius circle, 32 points are sampled. The OCT 2000 (software version A6.1; Humphrey Systems Division, Carl Zeiss Meditec, Dublin, CA) employs six radial scans to cover a circular area with a 3000-µm radius.26 It samples 600 points in total, with 100 of these in the central 500-µm radius circle that results in a spatial resolution of approximately 89 µm per pixel. However, the spatial resolution of the radial-spoke methodology declines with increasing eccentricity from the center of the scan to an average of 474 µm per pixel in the periphery. This level of spatial resolution may miss lesions that are a third of a disc diameter or smaller if it falls between the scan lines, such that subtle lesions outside of the central fovea could be missed.27 This variable, and relatively poor, level of spatial resolution precluded our use of the OCT in the mapping of early DME (despite the obviously superior axial resolution of the OCT). The OCT 3000 recently released by Humphrey Systems (division of Carl Zeiss Meditec) increases scan resolution to 512 A-scans along a scan line and has a reported spatial resolution from 5 to 60 µm.

Both the HRT II and RTA should be examined in a clinical setting to determine their relative diagnostic test performance. An objective and repeatable method of assessing early DME would be of great benefit in monitoring the progression and treatment of the disease. The sensitivity and specificity of each instrument in the detection of DME should be established and examined with respect to the current gold standard method of stereo fundus biomicroscopy if either instrument is to be accepted for wide spread clinical use.


    Materials and Methods
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 References
 
Sample
Seventeen patients (mean age, 60 ± 7 years) with DME diagnosed in at least one eye and 17 nondiabetic subjects (mean age, 49 ± 8 years) were recruited from patients and staff of the Toronto Western Hospital (Table 1) . For the patients with DME, the average duration of diabetes was 14 ± 4 years. One eye was randomly assigned to the study in all nondiabetic subjects and in those patients with DME in both eyes. The study was approved by the Research Ethics Board of the University Health Network (University of Toronto, Canada) and followed the tenets set out in the Declaration of Helsinki for research involving human subjects. Informed consent was obtained from all participants after the nature and possible consequences of the study were explained.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Characteristics of Patients with Diabetes

 
Procedures
Each study eye underwent refraction, logMAR visual acuity (Early Treatment of Diabetic Retinopathy Study charts at 96% contrast), pupillary dilation (using 1% tropicamide), assessment of lens opacity using the Lens Opacity Classification System III (LOCS III) and stereo fundus biomicroscopy. In addition, assessment with the HRT II and RTA were performed in a randomized fashion.

Clinical Assessment
The clinical gold standard of stereo fundus biomicroscopy was used throughout the study. A standardized method was established to permit the quantitative comparison between clinical assessment and both imaging modalities. A 20° x 20° retinal image centered on the fovea of the study eye was acquired in each volunteer, with the HRT. A clear acetate sheet was placed over the printed retinal image, and the retinal capillary pattern of each volunteer was carefully traced. The resultant vascular trace provided no other cues that might suggest the presence of retinal edema, such as hard exudates and microaneurysms. This methodology was previously used by Hudson et al.6 to accurately document the area and extent of retinal edema. The location and extent of DME was documented on the vascular trace by one of six retinal specialists (WCL, RGD, MM, PTH, JL, and JC) while performing stereo fundus biomicroscopy. All assessors were masked to the results of the imaging modalities. Each vascular trace defining the location and extent of DME was divided into five sectors (central 600-µm radius circle plus four peripheral quadrants, Fig. 3 ). The original HRT scan was not used in the subsequent analysis.



View larger version (41K):
[in this window]
[in a new window]
 
FIGURE 3. Left: 20° x 20° vascular map of the left eye of a patient with DME in which the location and extent of retinal thickening was clinically documented by stereo fundus biomicroscopy. Center: 15° x 15° edema map generated by the HRT II. Areas of brighter pixels represent higher edema index values. Right: 20° x 20° topographic map generated by the RTA. Red: areas of thickened retina; blue: areas of thinner retina. The two instrument scans have been scaled to match the vascular map. In this patient, both instruments agreed closely with clinical assessment, although both suggested a greater extent of retinal thickening.

 
HRT II: Macular Edema Module
The HRT II (Heidelberg Engineering) sequentially acquires 16 two-dimensional (that is, x, y plane) confocal section images per millimeter along the optical axis (that is, z-axis). The edema index analysis developed by Flanagan and Hudson has been incorporated within the HRT II as the Macular Edema Module (MEM). Each HRT II scan consists of 3 sequential 15° x 15° tomographic images of the retina that are averaged to display a mean topography image. The mean topography image that is subsequently analyzed has a resolution of 384 x 384 pixels. The average edema index was derived for a 600-µm radius circle centered on the fovea using the MEM software (ver. 1.0.0.4). This circle size was chosen to correspond to the same area of the Foveal Average Thickness (FAV) of the RTA. Each of the three sets of scans was divided into five sectors identical with that used for clinical assessment and areas of increased edema index were visually noted by two expert MEM assessors (JGF and CH). Assessors were masked from the subject’s clinical status. At least two of the three scans needed to manifest evident edema for any one sector to be assigned as edematous.

Retinal Thickness Analyzer
The Retinal Thickness Analyzer (RTA; Talia Technology Ltd., Neve-Ilan, Israel) sequentially scans 16 vertical slits across a 3 x 3-mm area. Retinal thickness is calculated for 16 points along each slit, giving a single scan resolution of 256 points (16 x 16) per scan. Five overlapping scans of the posterior pole (central, superior temporal, superior nasal, inferior temporal and inferior nasal) were performed and combined to give a total scan area of approximately 6 x 6 mm (total scan resolution of 1024 points, 32 x 32 pixels). Three sets of the five standard scans were acquired. Analysis of the topographic maps was performed using RTA software version 4.075 (Talia Technology Ltd.). The foveal average thickness (FAV) of a 600-µm radius circle was calculated for each scan. Each of the three sets of scans was divided into five sectors identical with clinical assessment, and areas of increased retinal thickness were visually noted by two expert RTA assessors (KG and CH). Assessors were masked from the clinical status of each volunteer. At least two of the three scans needed to manifest evident edema for any one sector to be assigned as edematous.

Analysis
Differences between groups were compared using a two-tailed Student’s t-test with {alpha} = 0.05. The coefficient of variation (COV = SD/mean) was calculated for the central 600-µm radius circle of both instruments. The level of agreement to clinical assessment and between each instrument was determined using a zonal analysis of the five predefined areas (Fig. 3) . The zonal analysis adjusted for the fact that the area assessed using stereo fundus biomicroscopy and using the RTA was 20° x 20° compared with the area assessed using the HRT II scan which was 15° x 15°. Combining the results of patients with DME and normal subjects, the sensitivity [true positives/(true positives plus false negatives)] and specificity [true negatives/(true negatives plus false positives)] of each instrument with respect to clinical assessment were determined.


    Results
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 References
 
There was a significant difference in logMAR visual acuity between patients with diabetes and nondiabetic subjects (0.13 ± 0.15 vs. -0.13 ± 0.09, respectively, P < 0.001). The assessment of lens clarity using LOCS III for both groups are presented in Table 2 . Patients with diabetes had significantly greater nuclear opalescence, nuclear color, and cortical cataract (P = 0.015, P = 0.049, and P = 0.001, respectively).


View this table:
[in this window]
[in a new window]
 
TABLE 2. LOCS III Assessment of Lens Opacity in Patients with DME and in Nondiabetic Subjects

 
The group mean edema index value for the central 600-µm circle of the HRT II was significantly higher in patients with DME than in nondiabetic subjects (2.02 vs. 1.07, P < 0.001). The median COV for the central 600-µm circle of the HRT II was 8.3% (range, 2.3%–24.6%) in patients with DME and 6.0% (range, 2.0%–37.5%) in nondiabetic subjects. The group mean FAV of the RTA was significantly higher in patients with DME than in nondiabetic subjects (237.5 µm vs. 165.5 µm, P = 0.005). The median COV for the FAV of the RTA was 6.4% (range, 2.1%–46.7%) in patients with DME and 8.5% (range, 2.3%–14.7%) in nondiabetic subjects. The values for both instruments are summarized in Figure 4 .



View larger version (16K):
[in this window]
[in a new window]
 
FIGURE 4. Box plots of edema index values (arbitrary units) for HRT II (left) and foveal average thickness in micrometers for RTA (right) in patients with DME and nondiabetic subjects.

 
Clinical assessment of the patients with DME identified 47 sectors having retinal thickening and 38 sectors without thickening (total number of sectors = 85). The HRT II agreed with clinical assessment for 45 of 47 sectors with thickening, and the RTA agreed with 27 of 47 sectors. An additional 31 sectors were identified as being edematous by HRT II, and 20 sectors were identified as being thickened by RTA, that were not identified by clinical assessment. In nondiabetic subjects, 77 of a possible 85 sectors were correctly identified as having no edema using the HRT II and 69 of 85 sectors by the RTA. A summary of the results are presented in Table 3 . Overall, the sensitivity and specificity of the HRT II was 92% and 68%, respectively, and for the RTA was 57% and 71%, respectively, when compared with clinical assessment. With respect to the HRT II that identified 76 sectors with retinal edema (45 of which agreed with clinical assessment), the RTA agreed in 44 sectors in patients with DME. In nondiabetic subjects, the RTA agreed with HRT II in 61 of a possible 77 sectors.


View this table:
[in this window]
[in a new window]
 
TABLE 3. Sectors of Agreement between HRT II and RTA Compared with Clinical Assessment and Each Other in Patients with DME and Nondiabetic Subjects

 

    Discussion
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 References
 
Both the HRT II and RTA are able to identify DME objectively. Each instrument uses distinctly different methodologies to assess the presence of edema and its effect on retinal thickening. The HRT II analyses the optical effect of change in the profile of reflectance intensity as a function of depth attributable to retinal thickening and the concomitant reduction in retinal reflectivity most likely attributable to changes in refractive index with retinal edema. The RTA determines retinal thickness by identifying the optical separation of the ILM and RPE.

The HRT II had a higher sensitivity in detecting DME than did the RTA (92% versus 57%, respectively). The specificity was approximately the same (HRT II 68% versus RTA 71%) across the two instruments. In comparison, the OCT has a reported sensitivity and specificity in detecting CSME of 89% and 96%, respectively, within the central fovea.28 However, it has been shown that due to its poorer spatial resolution outside the central fovea, the OCT tends to miss subtle edema beyond 500 µm of the fovea.27 It should be noted that with the additional resolution of the OCT 3000, there is increased likelihood of detecting extrafoveal edema. Future studies are needed to assess the ability of the OCT 3000 in the mapping of early edema and the comparison of its diagnostic test performance to clinical assessment and other imaging modalities.

There are two possible explanations for the high sensitivity and moderate specificity of the HRT II. The first would be that its diagnostic test performance is limited by the definition of the gold standard (i.e., the additional edematous areas represent early edema undetected by the gold standard). The second explanation would be that the HRT II is identifying false positives. It is not possible in this cross-sectional study to determine which of these two possibilities determine the diagnostic test performance of the HRT II. However, it is unlikely that the HRT II would have such a high sensitivity along with the moderate specificity, if it were identifying so many false positives. This is further supported by the observation that most of the false positives were from the diabetic subjects—that is, on the whole, the technique does not suggest that there is edema within the normal group (8/85 possible sectors). Prospective studies are ongoing that will enable us to investigate this issue.

Concerning the RTA, which has poor sensitivity and moderate specificity, its diagnostic test performance bears little similarity to the clinical gold standard. The poor sensitivity of the RTA may be explained by the dependence of the instrument on clear media and a minimum 4-mm dilated pupil, not always easy to achieve in patients with diabetes. The curve fitting algorithms used by the RTA can erroneously identify areas of apparent edema that actually represent artifact from blurred or distorted images (Fig. 5) .26 Neubauer et al.29 used both the OCT and the RTA to assess foveal thickness in subjects with DME and in nondiabetic subjects. They found that the RTA, in some cases, showed high retinal thicknesses in patients without clinically detectable retinal thickening when compared with OCT. This was attributed to slit images with blurred edges, and it resulted in a high interpatient COV of 33%. Our results with the RTA are further supported by the poor correlations found in two other studies comparing it with OCT26 and clinical assessment.18



View larger version (59K):
[in this window]
[in a new window]
 
FIGURE 5. Example of the RTA curve-fitting algorithm giving erroneously high localized retinal thickness values, probably due to media irregularities in a nondiabetic subject.

 
The level of agreement between the HRT II and RTA is poor, especially in patients with DME. This is to be expected considering the instruments’ respective diagnostic test performances and given that they measure different aspects of DME.

This study used six retinal specialists to determine clinically evident edema, the standard by which we judged the diagnostic test performance of the instruments. Consequently, the definition of the location and extent of retinal edema is influenced by the relative discriminative ability of each individual. However, the use of multiple retinal specialists overcomes the undoubted bias that would be introduced by a single assessor.

The median COV of approximately 7% for both instruments in the central 600-µm radius circle centered on the fovea was good. For the HRT II, this compares favorably with previously published literature that found the COV for the edema index to be approximately 14% to 22% in normal subjects using a prototype system.7 9 Investigators in prior studies have reported the COV for the FAV of the RTA to be approximately 5% to 6% for normal subjects and patients with diabetes, which is comparable to our study.16 17 In comparison, the OCT has an average COV of 6% in normal subjects and 8% in diabetics.22 Our study found a large range of individual COVs for DME subjects from 2.3% to 24.6% for the HRT II and 2.1% to 46.7% for the RTA, reflecting intersubject variation. This intersubject variation was also shown with OCT with a range of COVs in individual subjects reported from 3% to 30% in the central macula.22 30 31 Localized differences in the variability of retinal thickness measurements, particularly in areas of changing topography, have been noted previously for the HRT II, RTA, and OCT.30 32 33 At the time of this writing, new software versions for the RTA (ver. 4.10) and the HRT II (ver. 1.2.0) were being released. It is expected that, like the OCT, newer algorithms will continue to improve the variability of measurements gained with these instruments.

There was a significant age difference between patients with diabetes and nondiabetic subjects (mean age, 60 ± 7 years vs. 49 ± 8 years, respectively), but the comparison of the two instruments with clinical assessment was consistent across both groups. This age difference may cause differences in tear film quality and media clarity between the study groups, affecting the quality of the scans taken by both instruments. The LOCS III assessment of lens clarity is clearly poorer in patients with DME. Patients with established DME tend to have poorer glycemic control,4 34 35 36 which has been associated with reduced lens clarity.37 38

In summary, both instruments had low variability of results within 600 µm of the fovea. The HRT II showed better agreement with clinical assessment than the RTA. The agreement between instruments was poor.


    Acknowledgements
 
The authors thank the Mustafa Rawji and Tien Wong of the Multi-Disciplinary Laboratory for Research of Sight-Threatening Diabetic Retinopathy for their support and advice and ophthalmologists Wai-Ching Lam, Robert G. Devenyi, Mark Mandelcorn, Patricia T. Harvey, Jeffrey Lim, and Jordan Cheskes of the Toronto Western Hospital for the assessment of DME.


    Footnotes
 
Presented in part at the 62nd annual meeting of the American Diabetes Association, San Francisco, California, 2002.

Supported by Canadian Institutes of Health Research Operating Grant MOP-43844 (CH, JGF), a New Investigator Award (CH), a Premier’s Research Excellence Award (CH), and a Vision Science Research Program Fellowship, University of Toronto (KG).

Submitted for publication March 26, 2003; revised July 17 and September 3, 2003; accepted September 4, 2003.

Disclosure: K. Guan, None; C. Hudson, None; J.G. Flanagan, Heidelberg Engineering (C)

The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.

Corresponding author: Chris Hudson, Department of Ophthalmology, University of Toronto, Toronto, Ontario M5T 2S8, Canada; chudson{at}uwaterloo.ca.


    References
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 References
 

  1. Klein R, Klein BE, Moss SE, Cruickshanks KJ. The Wisconsin Epidemiologic Study of Diabetic Retinopathy: XVII. The 14-year incidence and progression of diabetic retinopathy and associated risk factors in type 1 diabetes. Ophthalmology. 1998;105:1801–1815.[CrossRef][ISI][Medline][Order article via Infotrieve]
  2. Klein R, Moss SE, Klein BE, Davis MD, DeMets DL. The Wisconsin epidemiologic study of diabetic retinopathy.XI. The incidence of macular edema Ophthalmology. 1989;96:1501–1510.
  3. Porta M, Bandello F. Diabetic retinopathy: a clinical update. Diabetologia. 2002;
  4. Bresnick GH. Diabetic macular edema: a review. Ophthalmology. 1986;93:989–997.[ISI][Medline][Order article via Infotrieve]
  5. Kinyoun J, Barton F, Fisher M, Hubbard L, Aiello L, Ferris F., III. Detection of diabetic macular edema: ophthalmoscopy versus photography—Early Treatment Diabetic Retinopathy Study Report Number 5.The ETDRS Research Group. Ophthalmology. 1989;96:746–750.[ISI][Medline][Order article via Infotrieve]
  6. Hudson C, Flanagan JG, McLeod D. A clinical vision science perspective of the management of diabetic macular edema. Focus Diabetic Retinopathy. 2001;8:4–9.
  7. Hudson C, Flanagan JG, Turner GS, McLeod D. Scanning laser tomography Z profile signal width as an objective index of macular retinal thickening. Br J Ophthalmol. 1998;82:121–130.[Abstract/Free Full Text]
  8. Hudson C, Flanagan JG, Turner GS, Chen HC, Young LB, McLeod D. Correlation of a scanning laser derived oedema index and visual function following grid laser treatment for diabetic macular oedema. Br J Ophthalmol. 2003;87:455–461.[Abstract/Free Full Text]
  9. Hudson C, Flanagan JG, Turner GS, Chen HC, Young LB, McLeod D. Scanning laser derived edema index topographic maps: correlation with visual function assessment in patients undergoing laser photocoagulation for clinically significant diabetic macular edema. Wall M Wild JM eds. Perimetry Update 1998–99, Proceedings of the XIIIth International Perimetric Society Meeting. 1999;503–510. Kugler Publications The Hague, The Netherlands.
  10. Zeimer R, Shahidi M, Mori M, Zou S, Asrani S. A new method for rapid mapping of the retinal thickness at the posterior pole. Invest Ophthalmol Vis Sci. 1996;37:1994–2001.[Abstract/Free Full Text]
  11. Gieser JP, Rusin MM, Mori M, Blair NP, Shahidi M. Clinical assessment of the macula by retinal topography and thickness mapping. Am J Ophthalmol. 1997;124:648–660.[ISI][Medline][Order article via Infotrieve]
  12. Shahidi M, Zeimer RC, Mori M. Topography of the retinal thickness in normal subjects. Ophthalmology. 1990;97:1120–1124.[ISI][Medline][Order article via Infotrieve]
  13. Shahidi M, Ogura Y, Blair NP, Rusin MM, Zeimer R. Retinal thickness analysis for quantitative assessment of diabetic macular edema. Arch Ophthalmol. 1991;109:1115–1119.[Abstract]
  14. Shahidi M, Ogura Y, Blair NP, Zeimer R. Retinal thickness change after focal laser treatment of diabetic macular oedema. Br J Ophthalmol. 1994;78:827–830.[Abstract/Free Full Text]
  15. Asrani S, Zeimer R, Goldberg MF, Zou S. Application of rapid scanning retinal thickness analysis in retinal diseases. Ophthalmology. 1997;104:1145–1151.[ISI][Medline][Order article via Infotrieve]
  16. Weinberger D, Axer-Siegel R, Landau D, Yassur Y. Retinal thickness variation in the diabetic patient measured by the retinal thickness analyser. Br J Ophthalmol. 1998;82:1003–1006.[Abstract/Free Full Text]
  17. Oshima Y, Emi K, Yamanishi S, Motokura M. Quantitative assessment of macular thickness in normal subjects and patients with diabetic retinopathy by scanning retinal thickness analyser. Br J Ophthalmol. 1999;83:54–61.[Abstract/Free Full Text]
  18. Nishiwaki H, Shahidi M, Vitale S, et al. Relation between retinal thickening and clinically visible fundus pathologies in mild nonproliferative diabetic retinopathy. Ophthalmic Surg Lasers. 2002;33:127–134.[ISI][Medline][Order article via Infotrieve]
  19. Zeimer RC, Mori MT, Khoobehi B. Feasibility test of a new method to measure retinal thickness noninvasively. Invest Ophthalmol Vis Sci. 1989;30:2099–2105.[Abstract/Free Full Text]
  20. Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science. 1991;254:1178–1181.[Abstract/Free Full Text]
  21. Hee MR, Izatt JA, Swanson EA, et al. Optical coherence tomography of the human retina. Arch Ophthalmol. 1995;113:325–332.[Abstract]
  22. Hee MR, Puliafito CA, Duker JS, et al. Topography of diabetic macular edema with optical coherence tomography. Ophthalmology. 1998;105:360–370.[CrossRef][ISI][Medline][Order article via Infotrieve]
  23. Coker JG, Duker JS. Macular disease and optical coherence tomography. Curr Opin Ophthalmol. 1996;7:33–38.[CrossRef][Medline][Order article via Infotrieve]
  24. Schaudig UH, Glaefke C, Scholz F, Richard G. Optical coherence tomography for retinal thickness measurement in diabetic patients without clinically significant macular edema. Ophthalmic Surg Lasers. 2000;31:182–186.[ISI][Medline][Order article via Infotrieve]
  25. Mistlberger A, Liebmann JM, Greenfield DS, et al. Heidelberg retina tomography and optical coherence tomography in normal, ocular-hypertensive, and glaucomatous eyes. Ophthalmology. 1999;106:2027–2032.[CrossRef][ISI][Medline][Order article via Infotrieve]
  26. Polito A, Shah SM, Haller JA, et al. Comparison between retinal thickness analyzer and optical coherence tomography for assessment of foveal thickness in eyes with macular disease. Am J Ophthalmol. 2002;134:240–251.[CrossRef][ISI][Medline][Order article via Infotrieve]
  27. Strom C, Sander B, Larsen N, Larsen M, Lund-Andersen H. Diabetic macular edema assessed with optical coherence tomography and stereo fundus photography. Invest Ophthalmol Vis Sci. 2002;43:241–245.[Abstract/Free Full Text]
  28. Goebel W, Kretzchmar-Gross T. Retinal thickness in diabetic retinopathy: a study using optical coherence tomography (OCT). Retina. 2002;22:759–767.[CrossRef][ISI][Medline][Order article via Infotrieve]
  29. Neubauer AS, Priglinger S, Ullrich S, et al. Comparison of foveal thickness measured with the retinal thickness analyzer and optical coherence tomography. Retina. 2001;21:596–601.[CrossRef][ISI][Medline][Order article via Infotrieve]
  30. Baumann M, Gentile RC, Liebmann JM, Ritch R. Reproducibility of retinal thickness measurements in normal eyes using optical coherence tomography. Ophthalmic Surg Lasers. 1998;29:280–285.[ISI][Medline][Order article via Infotrieve]
  31. Massin P, Vicaut E, Haouchine B, Erginay A, Paques M, Gaudric A. Reproducibility of retinal mapping using optical coherence tomography. Arch Ophthalmol. 2001;119:1135–1142.[Abstract/Free Full Text]
  32. Gilmore ED, Hudson C. Eccentricity and measurement variability and repeatability with the Retinal Thickness Analyzer. Br J Ophthalmol. .In press
  33. Chauhan BC, LeBlanc RP, McCormick TA, Rogers JB. Test-retest variability of topographic measurements with confocal scanning laser tomography in patients with glaucoma and control subjects. Am J Ophthalmol. 1994;118:9–15.[ISI][Medline][Order article via Infotrieve]
  34. Klein R, Klein BE, Moss SE, Davis MD, DeMets DL. The Wisconsin epidemiologic study of diabetic retinopathy.IV. Diabetic macular edema. Ophthalmology. 1984;91:1464–1474.[ISI][Medline][Order article via Infotrieve]
  35. Stratton IM, Kohner EM, Aldington SJ, et al. UKPDS 50: risk factors for incidence and progression of retinopathy in Type II diabetes over 6 years from diagnosis. Diabetologia. 2001;44:156–163.[CrossRef][ISI][Medline][Order article via Infotrieve]
  36. Service FJ, O’Brien PC. The relation of glycaemia to the risk of development and progression of retinopathy in the Diabetic Control and Complications Trial. Diabetologia. 2001;44:1215–1220.[CrossRef][ISI][Medline][Order article via Infotrieve]
  37. Kato S, Shiokawa A, Fukushima H, et al. Glycemic control and lens transparency in patients with type 1 diabetes mellitus. Am J Ophthalmol. 2001;131:301–304.[CrossRef][ISI][Medline][Order article via Infotrieve]
  38. Bron AJ, Sparrow J, Brown NA, Harding JJ, Blakytny R. The lens in diabetes. Eye. 1993;7:260–275.



This article has been cited by other articles:


Home page
IOVSHome page
E. D. Gilmore, C. Hudson, R. K. Nrusimhadevara, R. Ridout, P. T. Harvey, M. Mandelcorn, W.-C. Lam, and R. G. Devenyi
Retinal Arteriolar Hemodynamic Response to a Combined Isocapnic Hyperoxia and Glucose Provocation in Early Sight-Threatening Diabetic Retinopathy
Invest. Ophthalmol. Vis. Sci., February 1, 2008; 49(2): 699 - 705.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
E. D. Gilmore, C. Hudson, R. K. Nrusimhadevara, P. T. Harvey, M. Mandelcorn, W. C. Lam, and R. G. Devenyi
Retinal Arteriolar Diameter, Blood Velocity, and Blood Flow Response to an Isocapnic Hyperoxic Provocation in Early Sight-Threatening Diabetic Retinopathy
Invest. Ophthalmol. Vis. Sci., April 1, 2007; 48(4): 1744 - 1750.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
A. J. Patterson, D. F. Garway-Heath, and D. P. Crabb
Improving the repeatability of topographic height measurements in confocal scanning laser imaging using maximum-likelihood deconvolution.
Invest. Ophthalmol. Vis. Sci., October 1, 2006; 47(10): 4415 - 4421.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Ophthalmol.Home page
K A Goatman
A reference standard for the measurement of macular oedema.
Br. J. Ophthalmol., September 1, 2006; 90(9): 1197 - 1202.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
S.-W. Teng, H.-Y. Tan, J.-L. Peng, H.-H. Lin, K. H. Kim, W. Lo, Y. Sun, W.-C. Lin, S.-J. Lin, S.-H. Jee, et al.
Multiphoton autofluorescence and second-harmonic generation imaging of the ex vivo porcine eye.
Invest. Ophthalmol. Vis. Sci., March 1, 2006; 47(3): 1216 - 1224.
[Abstract] [Full Text] [PDF]


Home page
Arch OphthalmolHome page
M. Kisilevsky, C. Hudson, J. G. Flanagan, R. K. Nrusimhadevara, K. Guan, T. Wong, M. Mandelcorn, W.-C. Lam, and R. G. Devenyi
Agreement of the Heidelberg Retina Tomograph II Macula Edema Module With Fundus Biomicroscopy in Diabetic Maculopathy.
Arch Ophthalmol, March 1, 2006; 124(3): 337 - 342.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
H. Zou, X. Zhang, X. Xu, and S. Yu
Quantitative In Vivo Retinal Thickness Measurement in Chinese Healthy Subjects with Retinal Thickness Analyzer
Invest. Ophthalmol. Vis. Sci., January 1, 2006; 47(1): 341 - 347.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
A. J. Patterson, D. F. Garway-Heath, N. G. Strouthidis, and D. P. Crabb
A New Statistical Approach for Quantifying Change in Series of Retinal and Optic Nerve Head Topography Images
Invest. Ophthalmol. Vis. Sci., May 1, 2005; 46(5): 1659 - 1667.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (16)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Guan, K.
Right arrow Articles by Flanagan, J. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Guan, K.
Right arrow Articles by Flanagan, J. G.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS