1From the Research, Development, and Innovation Department, Ophthalmologic Institute of Alicante, Alicante, Spain; and 2Department of Optometry and Neuroscience, University of Manchester Institute of Science and Technology (UMIST), Manchester, United Kingdom.
| Abstract |
|---|
|
|
|---|
METHODS. Corneal aberrations were determined in fifteen healthy subjects at 1 second time intervals after a blink, up to a total elapsed time of 15 seconds. Corneal aberrations were obtained from corneal elevation maps measured using a Tomey TMS-2N topographer and custom software. All data were decomposed using Zernike polynomials to yield the root mean square (RMS) wavefront deviations, in micrometers, for two pupil diameters (3 and 7 mm).
RESULTS. Total wavefront aberration decreased slightly with time in the first few seconds after a blink for both pupil diameters, reaching a minimum after approximately 6 seconds. Thereafter aberrations increased steadily, exceeding the immediate postblink level after approximately 10 seconds.
CONCLUSIONS. In normal subjects, the contribution of the anterior cornea to the overall ocular aberration remains reasonably stable over the normal interblink interval (approximately 4 seconds) but rises to levels which could perceptibly degrade retinal image quality under circumstances where the interblink interval is increased to exceed 10 seconds, as may occur during the use of visual display screens or when performing difficult tasks.
In an earlier study,8 an increase was found in total and corneal aberrations 20 seconds postblink compared with immediately after a blink. This was attributed to the increasingly irregular tear film. However, to our knowledge, no time-resolved measurements of optical aberrations of the anterior cornea during the first seconds after a blink have been published to date.
Thus the purpose of the present study was to investigate with good time resolution the pattern of changes of the corneal higher-order aberrations and their impact on the optical quality of the cornea during the first 15 seconds after a blink.
| Methods |
|---|
|
|
|---|
Corneal Aberrations from Corneal Topography
Topographic data were obtained with a Tomey TMS-2N (Tomey Corporation, Nagoya, Japan) instrument. During the initial setting up, measurements in each eye were repeated until a well-focused and aligned image was obtained. Following the procedure used in an earlier study,8 subjects were instructed to keep their eyelids open during the image capture. Topographic images were obtained at 1-second time intervals from 1 to 15 seconds after a blink. Mean pupil diameter during the measurements was 3.2 ± 0.63 mm (photopic conditions, 80 cd/m2). Only the left eye was used for the measurements. To avoid possible longer-term changes to the cornea and/or tear film from successive periods of nonblinking, the experiments on each individual were performed on 3 separate days to yield three sets of measurements. Corneal videokeratographic data were downloaded onto floppy disks in ASCII files, which contained information about corneal elevation, curvature, power, and position of the pupil. The tests were run in a controlled temperature (21 ± 2°C) and humidity (41 ± 3%) at the Instituto Oftalmológico de Alicante (Alicante, Spain).
Data Analysis
The videokeratographic data were fitted with Zernike polynomials up to the sixth order to determine aberration coefficients, from which the wavefront aberration function was reconstructed. The calculation of corneal wavefront aberration was performed using the CT-View 5.0 software (Sarver & Associates, Inc, Merritt Island, FL) for two pupil diameters: 3- and 7-mm. The Zernike coefficients were used to calculate the total monochromatic, anterior corneal aberration, and the aberration contributed by spherical aberration and coma-like aberrations. Corneal wavefront aberrations were calculated relative to the pupil center instead of the vertex normal (videokeratoscope axis) since this is more relevant to visual acuity. The spot diagram corresponding to the corneal aberration was calculated. This showed the results of a geometric ray tracing for an object at infinity for an image plane corresponding to the minimum blur circle. The corneal Point Spread Function (PSF), which is defined as the light distribution in the image of a point source, was also calculated. The shape and width of the PSF depend on the levels of aberrations and the shape of the pupil, and include the effects of diffraction. Both the spot diagram and the PSF reveal how the corneal aberrations, considered alone, affect the final image. The actual retinal image will, of course, additionally depend on the contribution of the posterior cornea and the crystalline lens to the aberrations.
| Results |
|---|
|
|
|---|
|
Figure 2 shows, for a 7-mm pupil, the spot diagrams associated with the anterior corneal aberrations of Figure 1 at different times postblink (1 to 15 seconds). The spot diagram was obtained by geometric ray tracing, using the local slopes of the wave aberration. Only third- and higher-order aberrations were considered. Positive horizontal coordinates indicate temporal cornea for left eyes. The spot diagrams show that the image quality (directly related to the spread of the spots) changes as a function of the time postblink. In the first two seconds after the blink, the spots are relatively scattered but their concentration improves to give a minimal spread after approximately 6 seconds. Thereafter the spot diagram gradually broadens, as the wavefront aberrations increase.
|
Results were reasonably consistent across the three repeated runs on different days for each subject and between individual subjects, although a clear aberration minimum was not obtained on every run. Figure 3 shows the temporal changes in the total amount of higher-order (third to sixth) root-mean-square (RMS) wavefront aberration for a 7-mm pupil diameter, during three runs with the same subject as that illustrated in Figure 1 .
|
|
|
ANOVA one-way tests revealed statistically significant differences in corneal wavefront aberrations versus time for both pupil diameters (3- and 7-mm, P < 0.0001). A Bonferroni test of multiple comparisons showed that values immediately after a blink differed from those obtained at 5 to 8 seconds interval postblink for both pupil diameters (P < 0.01). For a 3-mm pupil diameter, RMS values obtained at 9 and 10 seconds were comparable to that obtained immediately after the blink (P > 0.01). From 11 to 15 seconds postblink, RMS values were significant higher than those immediately after the blink (P < 0.01). For a 7-mm pupil diameter, RMS values obtained in the 9 to 13 seconds range did not differ from that obtained immediately after the blink (P > 0.01). At 14 and 15 seconds postblink RMS values were significantly higher than those immediately after the blink (P < 0.01). For both pupil diameters there was a statistically significant reduction in the corneal wavefront aberration between immediately after the blink and the 5 to 8 seconds range, the minimum value occurring at 6 seconds. After 8 seconds, significance at different times postblink differed, depending on the pupil diameter. In both 3- and 7-mm pupils, corneal wavefront aberration values increased significantly at 15 seconds postblink compared to immediately after the blink. Thus the corneal wavefront aberration pattern revealed that after a blink, there is a reduction in corneal aberrations, a minimum value being reached at around 6 seconds. Aberrations then start to increase, first reaching the initial value (immediately after the blink) and then attaining still higher values at 15 seconds postblink. An alternative approach is to consider whether the data are better fitted by a second-order polynomial than by a linear trend. The squares of the Pearson product moment coefficients for linear fits to the 3- and 7-mm pupil data are R2 = 0.48 and 0.39, respectively, whereas those for corresponding quadratic fits are R2 = 0.88 and 0.87, so that the second-order fits are significantly better.
Spherical Aberration
Figure 5 shows mean values of the RMS spherical aberration (Z04 and Z06), for 3- and 7-mm pupil diameters, as a function of time postblink. Solid lines represent the best second-order polynomial trend equation. For both pupil diameters, there is an increase in spherical aberration after a blink. Over the 15-second observing period, the mean RMS spherical-aberration across all subjects increased from 0.012 µm to 0.044 µm for the 3-mm pupil, and from 0.185 µm to 0.532 µm, for the 7-mm pupil.
Coma-like Aberration
Figure 6 also illustrates mean values of the RMS coma-like aberration (Z3i and Z5i), for 3 and 7-mm pupil diameters, as a function of time postblink. Dashed curves represent the best second-order polynomial fit equations to the RMS coma-like values. There is a reduction immediately after the blink, minimum values being reached at 5 to 7 seconds postblink, followed by a progressive increase up to the measured maximum value at 15 seconds postblink. This behavior differs from that observed for spherical aberration, which showed a progressively increasing trend postblink, with no minimum.
|
| Discussion |
|---|
|
|
|---|
Analysis of the wave aberration in terms of its individual Zernike components suggests that the observed behavior cannot be accounted for by changes in the rotationally-symmetric spherical aberration terms (Z40 and Z60), which tend to increase monotonically with postblink time. This is presumably because the greater rate of evaporation at the center of the palpebral aperture15 causes the tear film to thin more rapidly at the center than peripherally. The resultant progressive change in the shape of the airtear film surface, from a prolate toward an oblate shape, is reflected in a change of spherical aberration toward more negative values (mean Z04 = 0.11 µm immediately after the blink and 0.31 µm 15 seconds postblink).
It seems more reasonable to attribute the temporal minimum in the overall aberration to the changes in coma-like aberrations illustrated in Figure 6 , together with analogous changes in some of the other aberrations which lack rotational symmetry. Such changes presumably result from such factors as the directional characteristics of lid movement, the effects of gravity on the tear movement, the uneven local rates of evaporation associated with the shape of the palpebral aperture, and, perhaps, dynamic change in the contour of the underlying cornea after lid pressure.11
How serious are the effects of these levels of aberration and is it likely that they will cause perceptible effects on vision? With a 3-mm pupil, total monochromatic higher-order RMS aberration is generally
0.06 µm, or about a tenth of a wavelength. This is close to the Maréchal criterion for near diffraction-limited performance (wavefront aberration < fourteenth of a wavelength). Thus, since images are normally also degraded by longitudinal chromatic aberration, it seems unlikely that any perceptible changes in the visual image will normally be noticed when the pupil is small. This agrees with the results of Ridder and Tomlinson,16 who found no systematic change in photopic contrast sensitivity over a time interval 0.1 to 3.2 seconds postblink (blink-induced suppression occurs for the immediate 0.1 second postblink). Tutt et al.3 whose measurements had much poorer time resolution, found no consistent change in contrast sensitivity in the first 10 seconds after a blink, although a noticeable drop occurred after 15 seconds.
The typical interblink interval in normal patients is around 4 to 5 seconds.17 It may, however, be extended to 10 to 20 seconds when "gazing" during difficult tasks,18 higher-speed driving,19 or working with visual displays,20 21 when the increased levels of aberration observed during the present study may well become relevant.
The time course of the aberrational changes observed is likely to be accelerated in dry-eye patients or those with other ocular surface diseases. In a comparison of the overall higher-order ocular aberrations in the eyes of a group of dry-eye patients (mean and SD of tear break-up time 3.8 ± 1.1 seconds) and a group of normals (9.7 ± 3.2 seconds), Montés-Mico et al.22 23 found that third and fourth order aberrations 5 to 10 seconds after a blink were some 2.5x greater for the dry-eye patients. It would clearly be of interest to attempt to correlate wavefront aberration measurements of the present type with the results of other methods for studying the dynamics of the normal and dry-eye tear film, for example, interferometry.24 25
Tear film-induced changes in the aberration contributed by the anterior cornea have obvious relevance to discussions on the extent to which corneal aberrations are balanced by the aberrations of the rest of the eye to yield reduced values of overall aberration.26 27 28 Clearly, exact balancing cannot be achieved if the corneal contribution is variable. While a good degree of compensation of corneal aberration by lenticular aberration has been claimed by some authors, it is difficult to see how any controlled compensatory growth mechanism, analogous to that thought to be involved in emmetropization, could function if one of its inputs was characterized by short-term variability.
Finally the present results suggest, like the data of Buehren et al.,11 that the topography of the airtear film surface can change significantly over the time interval during which videokeratoscope images are typically acquired (up to around 12 seconds). This suggests that, for the most consistent results, efforts should normally be made to complete videokeratography within the time span of the normal interblink interval (approximately 4 to 5 seconds, see also Nemeth et al.12 ).
The aberration introduced by the airtear film interface changes with time. It appears to be minimal approximately 6 seconds after a blink. Under normal photopic conditions, when the pupil diameter is modest and the interblink interval is short (approximately 4 seconds), it appears unlikely that the changes in aberration will produce detectable effects on vision. The increasing amounts of aberration observed after 10 seconds and with larger pupils may, however, lead to perceptible degradation in vision in circumstances, which lead to an increase in interblink interval or for patients with naturally large pupils.
| Footnotes |
|---|
Submitted for publication August 5, 2003; revised December 27, 2003; accepted February 23, 2004.
Disclosure: R. Montés-Micó, None; J.L. Alió, None; G. Muñoz, None; W.N. Charman, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Robert Montés-Micó, Research, Development, and Innovation Department, Instituto Oftalmológico de Alicante, Avda. Denia 111. 03015. Alicante, Spain; roberto.montes{at}uv.es.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. Lopez-Gil, V. Fernandez-Sanchez, R. Legras, R. Montes-Mico, F. Lara, and J. L. Nguyen-Khoa Accommodation-Related Changes in Monochromatic Aberrations of the Human Eye as a Function of Age Invest. Ophthalmol. Vis. Sci., April 1, 2008; 49(4): 1736 - 1743. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Koh, N. Maeda, Y. Hirohara, T. Mihashi, K. Bessho, Y. Hori, T. Inoue, H. Watanabe, T. Fujikado, and Y. Tano Serial Measurements of Higher-Order Aberrations after Blinking in Patients with Dry Eye Invest. Ophthalmol. Vis. Sci., January 1, 2008; 49(1): 133 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Koh, N. Maeda, Y. Hirohara, T. Mihashi, S. Ninomiya, K. Bessho, H. Watanabe, T. Fujikado, and Y. Tano Serial measurements of higher-order aberrations after blinking in normal subjects. Invest. Ophthalmol. Vis. Sci., August 1, 2006; 47(8): 3318 - 3324. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Montes-Mico, J. L. Alio, and W. N. Charman Postblink Changes in the Ocular Modulation Transfer Function Measured by a Double-Pass Method Invest. Ophthalmol. Vis. Sci., December 1, 2005; 46(12): 4468 - 4473. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Montes-Mico, J. L. Alio, and W. N. Charman Dynamic Changes in the Tear Film in Dry Eyes Invest. Ophthalmol. Vis. Sci., May 1, 2005; 46(5): 1615 - 1619. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |