METHODS. This was a prospective comparative study of 102 eyes of 102 patients with chronic glaucoma or ocular hypertension. Intraocular pressure (IOP) measurements by the Goldmann tonometer (GT) were compared with self-measured readings with the PPT. Patients evaluated the ease of home use of the PPT. The last 15 patients were asked to stop their glaucoma medications, and the ability of the PPT to detect an elevated IOP during self-tonometry was studied.
RESULTS. The mean ± SD difference between PPT and GT readings was 0.24 ± 1.57 mm Hg. Comparing the PPT with the GT, 86% of the readings were within ± 2.0 mm Hg, and 91% were within ± 3.0 mm Hg. Spearmans correlation coefficient was 0.91. A Bland-Altman plot showed that the 95% limits of agreement between the two methods lay between 2.90 and 3.38 mm Hg. Within-subject coefficients of variation for the GT and the PPT were 4.4% and 7.3%, respectively. In detecting an elevated IOP of more than 21 mm Hg, the sensitivity and the specificity of the PPT were 92.3% and 98.6% respectively. The mean satisfaction score for home use of the PPT was 87.4 ± 16.3 (maximum 100).
CONCLUSIONS. With proper training and technique, self-tonometry with the PPT appears to be accurate up to at least 25 mm Hg and is reproducible. The PPT was sensitive and specific in detecting an elevated IOP of more than 21 mm Hg. As patients were expected to seek ophthalmic care before the self-measured IOP reaches 25 mm Hg, the PPT may have a value for self-monitoring. Patients rated the PPT as satisfactory for home use. Because the PPT is portable and relatively inexpensive and requires no topical anesthesia or direct contact with the eyeball, it may have potential as an instrument for home self-tonometry.
No ideal home IOP monitoring device is currently available. Such a device should be accurate and reliable, simple to use, safe, acceptable to the patient, and inexpensive. Several instruments have been proposed,2 4 5 6 7 but they are too expensive, too complicated, or not safe enough to be operated by patients.
The pressure phosphene tonometer (PPT) is a new device using the entoptic phenomenon of pressure phosphene to evaluate IOP.8 The PPT is applied through the eyelid, and no direct eyeball contact or topical anesthesia is required. Previous data have shown that the PPT, when used by professionals on patients, can produce accurate IOP measurements comparable to the Goldmann tonometer (GT), the gold standard of IOP measurement.8 There are, however, no data on using PP self-tonometry in patients with glaucoma. The intent of this study was to determine the accuracy and reproducibility of PP self-tonometry compared with the GT in patients with glaucoma or ocular hypertension. We then studied whether the PPT could detect an elevated IOP during self-tonometry.
| Methods |
|---|
|
|
|---|
In the first study visit, patients underwent training for the use of the PPT (Fig. 1 ; Proview Eye Pressure Monitor; Bausch & Lomb Inc., Tampa, FL), as follows: The PPT was first reset to the lowest pressure reading of 8 mm Hg on the graticule. The patient was then instructed to look down and out and to place the PPT on the superonasal portion of the eyeball through the upper eyelid. The patient was taught to make the applicator surface lie tangential to the eyeball; in other words, the direction of application was perpendicular to that part of the eyelid and the eyeball. Afterward, the application pressure was increased gradually. The end point was taken as the moment when the patient clearly perceived a well-formed phosphene with a central dark spot, with or without a glowing outer rim (Fig. 2) . The patient was instructed to remove the tonometer from the eyelid as soon as the end point was reached. The IOP could then be read from the graticule on the dial. The graticule has a fiduciary indicator that does not move from the highest reading until it is reset. All training was given by a single trainer (EYYC) for better standardization of technique. The training of each patient lasted approximately 1 hour. The characteristics of the phosphene perceived were recorded. Afterward, the patient was requested to practice with the PPT for approximately 1 hour to ensure he or she had acquired the proper technique with adequate skill and safety.
|
|
The patient was then asked to measure his or her IOP in each eye at home by the PPT three times a day (morning, afternoon, and evening) for 1 week and to record the readings in a logbook. This served as a further self-practice in using the PPT, as well as a test for its ease of use at home.
In the second study visit at week 1, we similarly measured IOP three times by GT, and three PPT measurements were taken by patients. The two tonometries were again performed in random order. All measurements were again taken in the afternoon between 1 and 5 PM. The median of three readings was calculated. The patient was then requested to rate the ease of use of the PPT in home self-tonometry and the ease to perceive the phosphene. A simple rating system was adopted to determine the ease of use of the PPT at home and of perceiving the phosphene. The rating scale was a continuous scoring system from 0 to 100, with the following definitions, according to the patients experiences: 0, very great difficulty; 25, great difficulty; 50, some difficulty; 75, easy; and 100, very easy to use the PPT at home and perceive the phosphene.
The last 15 patients were further asked to stop their glaucoma medications for 10 days, after taking the baseline IOP. During the 10 days, they were requested to perform home self-tonometry with the PPT at 2 PM every day. They were asked to come back immediately for advice if their home tonometry readings were 30 mm Hg or above. A further visit on day 10 was scheduled for repeating GT and PPT measurements at 2 PM. The patients resumed all glaucoma medications on exit.
Statistical Analysis
All data were analyzed on computer (Statistical Package for the Social Sciences for Windows, ver. 11.0; SPSS Science, Inc., Chicago, IL). A Bland-Altman plot12 of the difference between PPT and GT readings against the average of the two was drawn to assess the agreement between the two methods and the presence of systemic bias. The medians of the PPT and GT readings at the initial and week-1 visit were compared using the Wilcoxon signed rank test or the Kruskal-Wallis one-way analysis of variance for subgroups. The mean differences and Spearmans correlation were computed for median PPT versus GT readings. The within-subject reproducibility for PPT and GT measurements was examined by calculating the within-subject coefficient of variation (CV) using the root mean square method as previously reported.13 P < 0.05 was defined as statistically significant. Pair-wise data exclusion was used for handling any missing data.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
Fifteen patients were asked to stop their glaucoma medications. At the initial visit, the mean ± SD. IOP readings were 17.9 ± 3.1 mm Hg by the GT and 17.6 ± 3.4 mm Hg by the PPT. After partial cessation of medications, at day 10 the mean ± SD IOP was 23.7 ± 4.5 mm Hg by the GT and 23.0 ± 4.5 mm Hg by the PPT (Table 2) . Wilcoxon signed rank test between readings at the initial and day-10 visits showed a statistically significant difference with P = 0.001. This is true for both GT and PPT measurements. There was no case in which IOP was elevated to more than 30 mm Hg. The correlation coefficients between GT and PPT were 0.93 (initial visit) and 0.98 (day 10 visit) in this group. Using the IOP measured by GT as the gold standard, we calculated the sensitivity and specificity of the PPT in diagnosing increased IOP (>21 mm Hg). Based on data at the week-1 visit, in 24 eyes results were true positive, in 2 eyes were false negative, in 70 eyes were true negative, and 1 eye are false positive. These gave a sensitivity of 92.3% and a specificity of 98.6%.
Three (2.9%) of the 102 subjects failed to perceive a pressure phosphene (group 1). Only one of the three subjects had visual field defects, with a mean defect of 6.73 dB (P < 0.5). For the remaining two subjects, their best-corrected Snellen visual acuities were fair at 0.6 and 0.2, respectively. Seventy-six subjects (74.5%) could see the black spot plus the glowing outer rim clearly (group 2). There were 23 (22.5%) subjects who could see only the black spot without the glowing outer rim (group 3). There was no statistically significant difference in age between groups 2 and 3. We compared the GT IOP readings of the three groups. The mean ± SD IOP by the GT for groups 1 (3 eyes), 2 (76 eyes), and 3 (23 eyes) were 17.4 ± 3.8, 18.5 ± 4.0, and 18.7 ± 4.9 mm Hg, respectively. A Kruskal-Wallis one-way analysis of variance showed that there was no statistically significant difference among the IOP readings in the three groups, with respect to GT (between groups P = 0.870) or the PPT (between groups P = 0.823). We also compared the performance of groups 2 and 3 using PPT and GT measurements: the Spearman correlation coefficient between PPT and GT was 0.91 for group 2 and 0.88 for group 3.
With the rating system as described previously, the mean ± SD score for the ease of use of the PPT for home self-tonometry was 87.4 ± 16.3. The mean ± SD score for the ease of perceiving pressure phosphene was 78.0 ± 20.4.
There were no complications or ocular injuries reported from use of the instrument in all our subjects.
| Discussion |
|---|
|
|
|---|
The reproducibility of both the GT and the PPT is considered to be good, as the within-subject CV for GT was 4.4% and that for the PPT was 7.3%. The latter is comparable to that of NCT (7.2%).16
We studied how accurate the PPT was in detecting an increasing IOP during home self-tonometry, by asking patients to stop their glaucoma medications for 10 days. During this period, they measured their IOP with the PPT at home. There was a statistically significant increase in mean IOPs comparing the initial (17.9 mm Hg by GT and 17.6 mm Hg by PPT) versus day-10 visit (23.7 mm Hg by GT and 23.0 mm Hg by PPT; P = 0.001; Wilcoxon signed rank test) for both GT and PPT readings. The Spearman correlation of GT versus PPT was excellent (0.930.98). Thus, our data suggest that patients can accurately measure an elevation of their IOPs in self-tonometry by the PPT. We also found the PPT to be sensitive (92.3%) and specific (98.6%) in detecting an IOP level of more than 21 mm Hg. The results were comparable to a study on NCT in which the sensitivity was 100%, whereas specificity was 88% in detecting an increase in IOP of more than 21 mm Hg.17
We also evaluated the ease of use of the PPT at home. Using a simple scoring system, patients rated the PPT to be easy to use (mean scores, 87.4 ±16.3) and the phosphene easy to perceive (mean scores, 78.0 ± 20.4). Our data suggest that the PPT is acceptable to patients for home self-tonometry.
From the Bland-Altman plot (Fig. 6) , it appears that there was a tendency to underestimate IOPs for GT readings more than 25 mm Hg. In the initial calibration of the PPT, linearity was good for the first 30% of the compression range of the instrument spring, and this corresponded to IOP ranges of 9 to 21 mm Hg.8 We thought that this might not necessarily lessen the value of the PPT as a tool for self-monitoring for three reasons. First, the role of the PPT is to detect an elevation of IOP above a target pressure, which in most, if not all patients with glaucoma, a value in the low to mid teens. Second, therefore, patients would have been expected to seek ophthalmic care before the self-measured IOP reached 25 mm Hg. Third, the sensitivity of the PPT in detecting an elevated IOP of more than 21 mm Hg was 92.3% and the specificity was 98.6%, both within a good range. Therefore, with proper expectation and patient education, the PPT may have a value for self-monitoring.
In recent years, it has become evident that central corneal thickness (CCT) is important in interpreting Goldmann applanation tonometry readings. A thick cornea gives rise to an artificially elevated GT reading. The optimal correction formula is yet undefined, though it was suggested that a correction of 1 mm Hg per 40-mm deviation from 525 mm can be made.18 Even if we take into account this effect, the mean central corneal thickness in our subjects (560.3 mm) deviated for only 35 mm, which translates into only 0.88 mm Hg of deviation in GT readings.
The PPT works on the principle of eyeball deformation, causing a tangential stretching of the retina and eliciting the sensation of light (phosphene). The PPT probe has the same application area as the Goldmann applanation tonometer prism.8 The application of force over a given area can be related through the Imbert-Ficks Law. In our experience, there were a few important potential sources of error that one should avoid during self-tonometry: First, the direction of self-application of the tonometer should be perpendicular to the superonasal eyelid covering the eyeball. The calibration of the PPT is only good when the force of application is along the axis of the spring and allows the probe surface to be applied tangentially to the eyelid covering the eyeball. Second, patients should not make the end point too early, before they truly appreciated the phosphene (underestimating true IOP), or too late, when the phosphene is already well formed and enlarging on excessive pressures (overestimating true IOP). We emphasize that for an accurate reading, the desired endpoint should be the very first moment that the patient clearly perceives a phosphene. From our experience, some patients may need to be reminded of this intermittently to avoid those errors.
Although the majority of subjects (74.5%, group 2) saw the glowing rim of pressure phosphene, a significant proportion (22.5%, group 3) did not. The reason for this phenomenon is uncertain. Our analysis did not reveal any statistically significant difference in age and IOP measured by the GT and the PPT between these two groups. The performance in using the PPT was also comparable in both groups.
There were no ocular injuries or complications detected arising from the use of the PPT in our study. Theoretically, the PPT is noninvasive because it requires no direct contact with the eyeball. The duration of application of the PPT onto the eyelid is in terms of seconds. Patients should be educated about the proper technique for using the PPT, and should be warned to withdraw the instrument as soon as a phosphene is perceived and that a further push is neither required nor desirable, to avoid any unnecessary self-inflicted damage. We recognize the need to collect further data on the safety of the PPT.
We excluded patients with previous ocular surgery, because we do not yet know how a possible change of scleral rigidity after operations might affect the measurements. We recognize that patients may have been excluded who had more advanced diseases and who may have been more likely to have trabeculectomy or other ocular surgery performed. Our recruited subjects with glaucoma appeared to have less advanced disease, as 51.5% of them have not yet had glaucomatous field loss, with overall mean Snellen best corrected visual acuity at 0.73. Further study to investigate the performance of the PPT in patients with advanced disease appears warranted.
The patients recruited into this study performed self-tonometry well with the PPT. They were relatively young, with a mean age of 54.3 ± 12.6 years. We anticipated that patients with problems in coordination, (e.g., arthritis of the upper limb, significant hand tremor), which may be more common in the elderly, would have difficulty with this self-tonometry. Further study on a wider age range of patients appears warranted.
Future study on the diurnal curves obtained by the PPT compared with other instruments will better validate its usefulness for home self-tonometry. In this study we have not addressed the value of home tonometry in the management of glaucoma. We envisage that the true value of self-monitoring of IOP with the PPT can only be determined with a randomized controlled trial to see whether patients with glaucoma using the PPT for home tonometry will have better long-term outcomes than those who do not.
In conclusion, to our knowledge, this is the first study of self-tonometry with the PPT in patients with glaucoma. Our results suggest that self-tonometry with the PPT by patients with glaucoma or ocular hypertension, with proper training and technique, is accurate and reproducible up to at least 25 mm Hg. The PPT is a sensitive and specific method for detecting an elevation of IOP to more than 21 mm Hg. Because patients are expected to seek ophthalmic care before the self-measured IOP reaches 25 mm Hg, the PPT will be of value for self-monitoring. Patients rated the PPT to be satisfactory and easily comprehensible for home use. The PPT is relatively inexpensive and portable and requires no or topical anesthesia or direct contact with the eyeball. It may have potential as an instrument for home self-tonometry.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication February 5, 2004; revised April 1, 2004; accepted April 14, 2004.
Disclosure: D.S.C. Lam, Bausch & Lomb (F); D.Y.L. Leung, None; T.Y.H. Chiu, None; D.S.P. Fan, None; E.Y.Y. Cheung, None; T.-Y. Wong, None; J.S.M. Lai, None; C.C.Y. Tham, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Dennis S. C. Lam, Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong; 3/F, Hong Kong Eye Hospital, 147K Argyle Street, Kowloon, Hong Kong; dennislam{at}cuhk.edu.hk.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A Cervino Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure. Br. J. Ophthalmol., December 1, 2006; 90(12): 1444 - 1446. [Full Text] [PDF] |
||||
![]() |
J. M. Martinez-de-la-Casa, J. Garcia-Feijoo, A. Castillo, and J. Garcia-Sanchez Reproducibility and Clinical Evaluation of Rebound Tonometry Invest. Ophthalmol. Vis. Sci., December 1, 2005; 46(12): 4578 - 4580. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |