1From the Department of Ophthalmology, Caritas Medical Centre, Hong Kong, Peoples Republic of China; and the 2Departments of Ophthalmology and Visual Sciences and 3Physiology, Faculty of Medicine, The Chinese University of Hong Kong, Hong Kong, Peoples Republic of China.
| Abstract |
|---|
|
|
|---|
METHODS. This was a cross-sectional study consisting of 41 normal subjects and 30 with early and 40 with advanced glaucoma. RNFL thickness and ONH parameters were measured with reference planes positioned at 95, 150, and 205 µm above the level of retinal pigment epithelium (RPE). Discriminating power for early glaucoma detection and correlation with visual field MD for each parameter at different levels of reference plane were compared by using the analyses of area under the receiver operating characteristic curves (AUCs) and linear regression, respectively.
RESULTS. All ONH measurements were significantly different between normal and glaucoma groups, irrespective of the level of reference plane. In normal eyes, changing the reference plane position resulted in significant differences in ONH measurements. Among all the parameters examined, integrated rim volume and RNFL thickness measured at 150 µm above the RPE showed the largest AUC (0.966) for early glaucoma detection, and the strongest correlation with visual function (r = 0.793), respectively.
CONCLUSIONS. OCT analysis of the ONH and RNFL is useful for early glaucoma detection. Among the three reference planes examined in this study, measurements analyzed at 150 µm above the RPE demonstrated the best performance for glaucoma detection and correlation with visual function. Compared with ONH measurements, RNFL thickness may be a better indicator, reflecting retinal ganglion cell function and monitoring disease progression.
In contrast, most published studies on ONH topology analysis are oriented around HRT, less attention has been focused on assessing the capability of analyzing the optic disc changes in glaucoma using OCT. Theoretically, OCT is more advantageous over HRT in offering a higher axial resolution, automated outlining of optic disc margin, and a consistent and stable reference plane for delineation of the neuroretinal rim boundary. The determination of the position of the reference plane is the most essential element in ONH topographic analysis, because accurate calculations of ONH parameters depend heavily on an appropriate choosing and marking of a reference plane. OCT determines the level of the reference plane in relation to the fixed anatomic landmarksthe cutoff ends of the retinal pigment epithelium (RPE) around the optic cup, and thus allows a consistent and reproducible estimation of cupdisc boundary and neuroretinal rim area. By default, the reference plane is located 150 µm above the level of RPE. However, this is only an arbitrary location, and no published investigation has been performed to assess the effect of shifting the position of the reference plane in the analysis of ONH parameters and at what level the reference plane should be positioned to yield the maximal performance in glaucoma detection or monitoring. The purpose of this study was to use the OCT to assess the impact of different reference planes on evaluation of glaucoma by varying the positions of the anterior offset from the RPE. We also compared the RNFL thickness and topographic ONH parameters in terms of their performances for detection of early glaucoma and their functional correlations with visual field mean sensitivities.
| Materials and Methods |
|---|
|
|
|---|
1.2 million. The study was conducted in accordance with the ethical standards stated in the 1964 Declaration of Helsinki and approved by Hong Kong Hospital Authority Kowloon West Cluster Clinical Research Ethics Committee, with informed consent obtained. The inclusion criteria were best corrected visual acuity not worse than 20/40, spherical refractive error within the range of 6.00 D to + 3.00 D, and reliable Humphrey visual field test results (Carl Zeiss Meditec), defined as having fixation loss <20% and false-positive and false-negative errors <25%. Subjects were excluded if they had a history of any retinal disease, surgery or laser procedures, optic disc anomalies such as coloboma or optic disc drusen, or any kind of neurologic diseases that might cause visual field defects. Automated visual field tests were tested in all subjects (Humphrey Field Analyzer II, Carl Zeiss Meditec) using the central 30-2 threshold program. Normal subjects were individuals with no visual field defect, no structural optic disc abnormalities, and no history of intraocular pressure >21 mm Hg. Glaucoma patients were identified entirely based on the presence of visual field defects. A field defect was defined as having three or more significant (P < 0.05) nonedge-contiguous points with at least one at the P < 0.01 level on the same side of the horizontal meridian in the pattern deviation plot and classified outside normal limits in the glaucoma hemifield test. Any detected field defect was confirmed on at least one other attempt to have been considered abnormal. Subjects were then assigned to the early group if the severity of visual field defect, as reflected by the overall sensitivity index MD, was at or better than 7.0 dB and to the advanced group if the MD was worse than 7.0 dB. Among all 70 subjects with glaucoma, 25 had low-pressure glaucoma, 10 had primary angle-closure glaucoma, 33 had primary open-angle glaucoma, and 2 had uveitic glaucoma. In this study, the patient with open-angle glaucoma with no history of documented intraocular pressure >21 mm Hg was defined as having low-pressure glaucoma, whereas the patient with intraocular pressure >21 mm Hg with no identifiable secondary causes was classified as having primary open-angle glaucoma. Glaucoma with a closed angle on gonioscopy with no evidence of secondary angle closure detected in the first visit was defined as primary angle-closure glaucoma.
OCT Measurements
The third-generation OCT (Stratus OCT or OCT 3; Carl Zeiss Meditec) was used in the study. The optical principles and applications of OCT have been described by Huang et al.9 ONH analysis was performed with the fast optic nerve scan protocol. Six radially linear scans centered over the ONH were analyzed cross-sectionally as shown in Figure 1 . The algorithm detected and showed the location of the top and inner edges of RPE on each side of the optic disc where a line was joined and signified as the disc diameter (Fig. 1 , green line). The reference plane (the cup offset) was then determined by tracing a line parallel to the disc diameter with an anterior offset of 95, 150, or 205 µm (Fig. 1 , white line). The neuroretinal rim area in each cross-sectional scan was estimated by the area bounded using the reference plane as the posterior border and the lines extending perpendicularly from the ends of the disc diameter as the lateral boundaries (Fig. 1 , red area). The nerve width at the disc on each side was a straight line from each disc reference point to the nearest point on the anterior surface (Fig. 1 , yellow line). Data analyzed in each scan were then incorporated and formed the composite image measurements including integrated rim volume (calculated by integration of vertical cross-sectional rim area), integrated rim width (calculated by integration of average nerve width at the disc), disc area, cup area, rim area (disc area cup area), cup-to-disc area ratio, cup-to-disc horizontal ratio, cup-to-disc vertical ratio, and cup volume. Retinal NFL thickness was measured by averaging the results of three sequential circular scans (512 scan points) with diameter 3.4 mm centered at the ONH. The NFL thickness was determined by the difference in distance between the vitreoretinal interface and a posterior boundary based on a predefined reflectivity signal level. A good-quality scan was one with a signal-to-noise ratio of >35, 100% accepted A-scans and good delineation of the anatomic boundaries. Subjects were not included in the study if the quality of the OCT image was suboptimal.
|
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
We believe OCT may provide a more reliable ONH analysis than HRT because of its consistent location of the reference plane and automated determination of the ONH margin based on a fixed anatomic landmark, the RPE. By default, the reference plane in OCT is determined by drawing a line 150 µm above the RPE. This is in contrast to the HRT in which the reference plane is set below the variable retinal surface. For the determination of the ONH margin, instead of manually tracing the disc boundary as in the HRT, OCT is able to position the disc margin automatically at locations where the RPE ends. Keeping both the reference plane and ONH margin (or the contour line) at constant positions, subsequent ONH measurements can therefore be analyzed with low variability. The reproducibility of OCT-measured ONH parameters was recently demonstrated by Paunescu et al.16
Accurate determination of the level of the reference plane is the most essential aspect of ONH analysis, because it has a direct impact on all the subsequent analyses of ONH parameters. In the Stratus OCT (Carl Zeiss Meditec), the default reference plane divides the rimcup boundary at an arbitrary level, 150 µm above the RPE. To identify the optimal reference plane for evaluation of ONH parameters, analyses were performed and compared at three different reference planes located at 95, 150, and 205 µm above the RPE. Significant differences in ONH measurements were observed when the reference plane offset was changed (Table 2) . In this study, relatively high sensitivities and specificities for detection of early glaucoma were evident in most of the ONH parameters examined (Table 5) . Integrated rim volume, cup-to-disc area ratio, and vertical cup-to-disc ratio analyzed with the default reference plane attained the largest AUCs (AUC = 0.966, 0.960, and 0.962, respectively) among all the ONH measurements. These findings were consistent with an earlier report by Schuman et al.17 Among five individual ONH parameters examined by OCT3 (disc area, cup-to-disc area ratio, cup area, cup volume, and rim volume), they found cup-to-disc area ratio and rim volume had the largest AUCs when comparing the glaucoma group with the normal and suspected glaucoma groups (AUCs = 0.79 and 0.77, respectively). In terms of correlation with visual function, integrated rim width, rim area, and cup-to-disc area ratio had the highest correlation (all with r > 0.700). The correlations were weakened in rim area and cup-to-disc area ratio when the level of reference plane was changed. The coefficients of correlation remained the same for integrated rim width, as its calculation was independent of the levels of reference plane. Our results showed that the default reference plane had the optimal performance in terms of glaucoma detection and correlation with visual function.
RNFL thickness, in contrast to ONH parameters, is less controversial, and numerous reports have arrived at the consistent conclusion that it is a useful surrogate marker for assessment of structural damage in glaucoma.5 6 7 8 Depending on the selection of subjects and the severity of visual field defect at the time of inclusion, the diagnostic sensitivity of RNFL thickness could achieve a level as high as >90% at a specificity of 90%. Yet, it is still uncertain whether RNFL thickness or ONH measurements is a better clinical indicator for the assessment of glaucomatous damage. In the present study, both RNFL thickness and integrated rim volume demonstrated comparable performance for detection of early glaucoma in each of the reference planes tested. Because all the optic nerve fibers finally converge toward the ONH, one would expect a corresponding change in neuroretinal rim area/volume when there is reduction in RNFL thickness. (This was reflected by the high correlation between RNFL thickness and the integrated rim width/volume/rim area as shown in Table 6 .) This concurred with the findings that both RNFL thickness and most of the ONH measurements attained similar performance in diagnostic sensitivity for glaucoma. However, a stronger correlation with visual function was evident in RNFL thickness compared with all the ONH measurements (Table 6) . Theoretically, RNFL is a more direct measure of ganglion cell function in contrast to neuroretinal rim volume, which comprises other supportive connective tissue structures and glial cells besides the RNFL. Moreover, the characteristic cupping of ONH in glaucoma involves not only loss of the axons but also change of supporting tissue integrity in the lamina cribrosa. Therefore, structural thinning of the RNFL is more closely related to the real-time functional loss of ganglion cells and then to the visual field defect. It is interesting to note that although integrated rim volume demonstrates the best performance in detection of early glaucoma, it is the rim area that shows the best correlation with visual function among all the ONH measurements. The observed difference may also be explained by the fact that whereas integrated rim volume is a three-dimensional measurement of the bulk of neuroretinal rim consisting of both nerve fiber bundles and their supporting tissues, rim area is a two-dimensional measurement, calculated by subtracting the cup area from the disc area and therefore reflecting more contribution from the nerve fiber bundles in proportion to supporting tissues. This is in agreement with the finding that rim area and integrated rim width had the best correlation with RNFL thickness compared with all ONH measurements. Collectively, although both RNFL and ONH measurements are useful for diagnosing glaucoma, RNFL thickness may be a better surrogate marker for monitoring progression of glaucomatous damage compared with ONH measurements. A prospective study would be useful to address this question fully.
We did not find any difference in ONH parameters among primary open-angle, primary angle-closure, and normal-tension glaucoma, although significant differences were evident in all ONH measurements, except the optic disc area, between normal and the three glaucoma subgroups. Controversy still exists regarding the relationship between optic disc size and glaucoma. Although some studies have reported larger optic disc size in glaucoma when using the HRT or OCT,17 18 others did not find differences in optic disc size between normal subjects and those with glaucoma.11 19 In the Blue Mountains Eye Study with 3654 subjects examined, larger optic disc size was found in patients with glaucoma, and a significant difference in mean disc diameter was found only when comparing the normal and high-pressure glaucoma groups, but not the low-pressure group.20 In agreement with the Blue Mountains Eye Study, the Reykjavik Eye Study (1040 study subjects) also reported larger disc size in glaucoma patients.21 Using OCT in the present study, our results echoed the findings in these two epidemiology studies. Whether it has any role as an indicator of susceptibility for primary open-angle glaucoma has yet to be determined.
Study Limitations
In the present study, the AUCs observed in most of the parameters examined were relatively high. The AUC represents the probability that a random pair of normal and abnormal parameters will be correctly ranked as to their disease state. Therefore, the performance of the ROC curve of any particular testing parameter is dependent on the diagnostic criteria of the disease state. In terms of visual field changes, glaucoma can be classified as mild (MD better than 6 dB), moderate (MD between 6 and 12 dB), or severe (MD worse than 12 dB), according to the classification by Hodapp et al.22 A higher AUC is derived when one compares a normal group to a glaucoma group with more severe mean visual field defects. This difference was exemplified in a recent study by Reus and Lemij.23 Using confocal scanning laser polarimetry, they found the parameter nerve fiber indicator could achieve an AUC of 0.96 when comparing normal and mild glaucoma subjects (average visual field MD, 2.01 dB). The AUC further increased to 0.99 when they compared the normal group with the moderate glaucoma group (average visual field MD, 5.44 dB). With the confocal scanning laser ophthalmoscope, Ahn and Kee24 reported that with the parameter cup-to-disc ratio, 79.3% of subjects who had mild visual field loss (defined as MD better than 6 dB) were identified as having glaucoma. Whereas in the group of moderate visual field loss (defined as MD between 6 and 12 dB), the ratio of classifying them as glaucomatous reached 100%. In the present study, 7 dB was selected as the cut off to divide the glaucoma subjects into early and advanced groups. If the Hodapp et al.22 grading scale were applied, the early glaucoma group would include individuals with mild to moderate glaucomatous visual field defect. We believe the high AUCs observed in most of the parameters examined can be explained by the relatively low visual field MD in our selected sample. Because the primary objective of the present study was to assess the impact of different reference planes on evaluation of glaucoma and to compare the diagnostic performance of RNFL thickness and topographic ONH parameters in the selected samples, emphases were on the comparison of the relative values of AUCs rather than their absolute figures. AUCs obtained in individual parameters should only be extrapolated to other population samples with caution.
Nevertheless, even without the aid of any imaging devices, clinical assessment of the ONH has been found to achieve high values of AUC. In the study by Greaney et al.,6 they found an AUC of 0.93 for detection of glaucoma, using qualitative assessment of ONH stereophotographs (average visual field MD of the glaucoma group, 3.9 dB). Uchida et al.25 also found an AUC of 0.93 for detecting subjects with glaucoma by using qualitative optic disc evaluation (average visual field MD of the glaucoma group, 4.8 dB). It is hoped that by quantifying the cup-to-disc measurements with imaging devices, the diagnostic performance for glaucoma detection can be enhanced.
Because the calculation of the ONH parameters was estimated based on six linear scans cutting cross-sectionally over the optic disc, disc areas lying between the scan lines were not subjected to the analysis. As a result, the sensitivity in detecting a highly localized optic disc abnormality may be reduced. Second, with the current version of the analysis software, it is not yet possible to get individual clock-hour ONH measurements. Future enhancement of the analysis protocol by increasing the number of linear scans and incorporation with sectorial measurements in the analysis software may increase the measurements accuracy and their diagnostic performance in glaucoma detection.
| Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
Disclosure: C.K. Leung, None; W. Chan, None; Y. Hui, None; W. Yung, None; J. Woo, None; M. Tsang, None; K.R. Tse, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Christopher Kai-shun Leung, Department of Ophthalmology, Caritas Medical Centre, 111 Wing Hong Street, Sham Shui Po, Hong Kong, Peoples Republic of China; tlims00{at}hotmail.com.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. G. Strouthidis, H. Yang, B. Fortune, J. C. Downs, and C. F. Burgoyne Detection of Optic Nerve Head Neural Canal Opening within Histomorphometric and Spectral Domain Optical Coherence Tomography Data Sets Invest. Ophthalmol. Vis. Sci., January 1, 2009; 50(1): 214 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K.-s. Leung, A. C. K. Cheng, K. K. L. Chong, K. S. Leung, S. Mohamed, C. S. L. Lau, C. Y. L. Cheung, G. C.-h. Chu, R. Y. K. Lai, C. C. P. Pang, et al. Optic Disc Measurements in Myopia with Optical Coherence Tomography and Confocal Scanning Laser Ophthalmoscopy Invest. Ophthalmol. Vis. Sci., July 1, 2007; 48(7): 3178 - 3183. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Ajtony, Z. Balla, S. Somoskeoy, and B. Kovacs Relationship between Visual Field Sensitivity and Retinal Nerve Fiber Layer Thickness as Measured by Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., January 1, 2007; 48(1): 258 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Medeiros, L. M. Zangwill, C. Bowd, P. A. Sample, and R. N. Weinreb Influence of disease severity and optic disc size on the diagnostic performance of imaging instruments in glaucoma. Invest. Ophthalmol. Vis. Sci., March 1, 2006; 47(3): 1008 - 1015. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K.-s. Leung, K. K.-L. Chong, W.-m. Chan, C. K.-F. Yiu, M.-y. Tso, J. Woo, M.-K. Tsang, K.-k. Tse, and W.-h. Yung Comparative Study of Retinal Nerve Fiber Layer Measurement by StratusOCT and GDx VCC, II: Structure/Function Regression Analysis in Glaucoma Invest. Ophthalmol. Vis. Sci., October 1, 2005; 46(10): 3702 - 3711. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |