METHODS. ML blind deconvolution is an image-processing technique that estimates the original scene from a degraded image. This technique has been used in confocal scanning laser microscopy to remove "out-of-focus" haze in three-dimensional confocal image stacks. ML blind deconvolution requires no prior estimation of the point-spread function (PSF), as opposed to classic linear deconvolution methods. Instead, the algorithm estimates an initial PSF based on the optical setup of the confocal scanning device and optics of the eye and iteratively proceeds to a solution. The improvement in repeatability of height measurements from mean topography images within scan (intrascan) and between scans (interscan) afforded by ML deconvolution was evaluated in a testretest series of HRT images from 40 ocular hypertensive and glaucomatous patients with varying degrees of media opacity.
RESULTS. There was an improvement in intrascan repeatability in 38 out of the 40 mean topography images (median improvement 2.5 µm, inter-quartile range 2.19, P < 0.001), and an improvement in interscan repeatability in 33 of the 40 mean topographies (median improvement, 1.0 µm, interquartile range 3.49, P < 0.001). There was a positive association between the magnitude of the improvement in repeatability and the level of mean pixel height standard deviation (MPHSD), intrascan (P = 0.004) and interscan (P = 0.002).
CONCLUSIONS. ML blind deconvolution algorithm improves the repeatability of topographic height measurements from the HRT. This improvement was greater in patients with poorer quality images.
Confocal scanning laser tomography is a special application of confocal microscopy that regards the ocular fundus as the object. The resolution of the raw 3-D images obtained by the technology is roughly pencil-shaped, with a lateral resolution of
10 µm in diameter and a depth resolution of
300 µm. A topography image is generated by determining the position of peak reflectance in the confocal stack at each pixel. Reflecting layers within the retina (e.g., pigment epithelium, nerve fiber layer, inner limiting membrane) vary in thickness from 200 to 500 µm. Because of the limited depth resolution (300 µm) of the technology, the topography image can be thought of as the position of the mean depth of the interfaces between these layers.12
Campbell and Gubisch13 demonstrated that the reflective properties of the fundus best approximate a diffuse surface, they also showed with Fourier analysis that the optimal pupil diameter is between 2 and 3 mm for imaging applications. If the resolution of optical sectioning of the human retina achieved by CSLT were only limited by diffraction then, according to estimates from theoretical models, the lateral resolution would be between 1.6 and 2.3 µm, and the depth resolution would be between 45 and 48 µm.14 15 These resolutions would lead to topographic height measurements with an axial position precision of
2.3 µm (whereas the current optimal precision is
15 µm).16 However, the resolution achieved in practice is limited by the optics of the human eye, specifically the lens, cornea, and tear film, whereas laser safety standards for the human eye and the confocal pinhole size also inhibit resolution. The lens, cornea, and tear film produce wave aberrations that are unique for each eye. Low-order aberrations can be compensated for by spectacles or lenses placed in front of the instrument optics; however, high-order aberrations remain. Pupil dilation improves resolution up to a critical point. The pupil dilation for optimal lateral resolution has been shown to be different from that which achieves optimal depth resolution.17 Venkateswaran et al.,15 demonstrated how the achievable depth resolution increases with increasing pinhole size.
If we now consider an ideal imaging system, the image acquired would be a scaled version identical to the true object being imaged. Image-restoration algorithms are a collection of image-processing techniques used to estimate the true object of interest given the image obtained and information of the nature of the blurring and noise in the imaging system.18 Image-restoration algorithms can be classified as linear and nonlinear. Nonlinear techniques, based on computationally intensive methods have increasingly gained acceptance, with Jansson et al.19 finding they deliver superior results to linear methods in a broad spectrum of applications. As computational resources are now faster and cheaper, the historic objections against using nonlinear techniques have become inappropriate. Maximum-likelihood (ML) deconvolution is an example of these techniques: this estimates both the true image and the blur from the degraded image using partial information about the imaging system.20 21 Holmes et al.22 were the first to apply the techniques to images obtained in confocal microscopy. In this study, we applied ML blind deconvolution to series of images obtained from a testretest study of HRT topography reproducibility. Our motivation is to investigate whether this technique can improve the repeatability of topographic height measurements and hence increase the utility of the HRT as a technique to assess the ONH.
| Methods |
|---|
|
|
|---|
The 3-D confocal image acquired is postprocessed in an attempt to compensate for random eye movements, such as microsaccades and slow drift, known to occur when a subject is asked to fixate on a target.23 A topography image is then formed by calculating the position of maximum reflectivity at each z-profile (a one-dimensional signal of 32 intensity values parallel to the optical axis). The intensity at each pixel in the resultant topography image represents the surface height of the ONH or surrounding papillary retina. Typically, three topography images are acquired at each visit. The topographies are normally merged to calculate a mean topography. Image registration algorithms within the proprietary HRT software align the topography images for interscan differences in examination positions.
The repeatability of the images obtained was quantified by the mean pixel height standard deviation (MPHSD). This metric is effectively a gauge of the variability of each pixel height measurement across the three topographies used to make up the mean topography and is used in this study to report intrascan repeatability.24 It is calculated from the standard deviations at each pixel across the mean topographic image (i.e., the MPHSD is the mean of 256 x 256 pixel height standard deviations). Previous studies have used MPHSD to evaluate the repeatability of the technology in normal subjects and patients with glaucoma25 26 It has been shown that MPHSD is influenced by lens opacity, age, and degree of astigmatism.11 In this study, MPHSD was also used to report the repeatability of mean topographies, referred throughout as interscan repeatability.
Confocal scanning laser tomography has known limitations. For example, although the optical setup is designed to reject most light from outside the focal plane, it by no means rejects all it and an out-of-focus haze remains.27 28 The resolution in the confocal images is higher in the x and y directions compared with poorer resolution along the optical axis (z-axis). The resolution obtained is also limited by the optics of the eye, aberrations are generated by the cornea and the lens.29 This results in axial smearing: For example, if a spherical point object is being imaged with constant reflectivity properties, the resultant image obtained will appear elongated in the z-axis. Another limitation of the technology is that the detector in the optical setup, is prone to Poisson noise, primarily due to quantum variations in the number of photons recorded.23 This noise obscures real data and randomly creates impossible features such as high-intensity data only one pixel in size.22 A further discussion on the principals and limitation of imaging systems is given by Goodman.30
Deconvolution is an example of an image restoration algorithm that models the imaging system with:
![]() |
ML Deconvolution
The ML estimation is a mathematical optimization strategy designed to produce the best guess of true data that have been corrupted by random noise.33 It is an adaptation of the Richardson-Lucy optimization strategy.34 35 The ML deconvolution approach is known to fail, however, unless strong constraints can be applied to the properties of the PSF. Holmes et al.22 published suitable constraints for data obtained by confocal microscopy, using assumptions about the shape of the hour-glass PSF and optimized for the mathematical nature of the noise present. Previous studies evaluated ML deconvolution in confocal microscopy on simulated and real data.36 37 In this study we applied ML deconvolution, using software developed by Holmes et al., with a commercially available system (AutoDeblur; ver. 9.3.6; AutoQuant Imaging, Inc., now Media Cybernetics, Silver Spring, MD).
The deconvolution software requires input of the optical setup and image medium to allow it to approximate and iteratively constrain the solution of the PSF. The numerical aperture of the lens (0.08), wavelength of the laser beam (680 nm), refractive index of the medium being imaged (assumed to be close to water, 1.333), and spacing of the image obtained is input into the software. For a 10° scan, the x and y spacing are 11.4 µm With the HRT, the depth of scan can be varied from 2 to 4 mm, while always acquiring 32 scans. The z-spacing is set appropriately, depending on the depth of the scan. The z-spacing is set as scan depth divided by 32, and this is directly entered into the deconvolution software.
Clinical Data
The techniques were applied to a testretest data set. The study protocol adhered to the Declaration of Helsinki and had local ethics committee approval; the informed consent of subjects was obtained. All subjects attended the OHT clinic at Moorfields Eye Hospital, London. Strouthidis et al.11 38 originally collected the data to evaluate the testretest variability of the HRT and HRT II. The study design was originally developed to test interoperator and intervisit repeatability. Patients were not excluded on ONH appearance; patients were excluded on the basis of myopia greater than 12 D of spherical power or any history of intraocular surgery. The eye with greater media opacity was selected preferentially. A total of five mean topographies was acquired from each subject, with three single topography images combined to form a mean topography. The images were obtained over a 6-week period in two visits to the clinic. On the first visit, three mean topographies were obtained, and in the follow-up visit a further two were obtained. All subjects had had experience with scanning laser tomography, having been imaged at least three times previously.
Comparison
In this study, we investigated whether ML deconvolution would result in an improvement of intrascan (within-scan) and interscan (between-scan) topographic height measurement repeatability. For this objective, 40 patients image series (HRT Classic) were randomly selected from the testretest dataset.
Measuring Intrascan Repeatability
The mean topography on the first visit was obtained, and the MPHSD was recorded from the HRT software (version 2.01b). Three confocal stacks used to generate the mean topography images were deconvolved. The images were then re-entered into the HRT software and three single topographies were calculated. It was observed that deconvolution had induced an artifact at the border of the single topographies. This is a known phenomenon and Gonzalez et al.39 suggest using an edge-taper function to blur the edge of the image to minimize this effect. As a local spatial filter may affect the results, the artifact at the edges was removed using an image-processing erosion algorithm.32 In this application the erosion algorithm proceeds by labeling pixels with an intensity of 0 as background (or not available), this is the "windowed" area typically seen at the edges of a topography image. The rest of the pixel intensities (1 to 255) are set as foreground. The erosion algorithm proceeds by removing a layer of foreground pixels at the boundary edge between the foreground and background pixels. This makes the windowed area at the edges of the topography image larger. In this application, 5 pixels were eroded from the boundary edge. The mean topography was then calculated, and the MPHSD was recorded for comparison with the preprocessed data.
Measuring Interscan Repeatability
Three mean topographies were randomly selected for each patient. For the purpose of quantifying the repeatability across scans, the mean topography images were input into the HRT database as single topographies. The proprietary image registration algorithms therefore spatially aligned the mean topographies and the computed MPHSD was recorded. The nine confocal images associated with the three mean topography images were deconvolved. Nine single topographies were then generated and the edge artifact was removed as before. Three mean topographies were generated and input into the HRT software as single topographies, and the MPHSD quantifying the interscan repeatability was recorded.
The association between lens opacity and the effect deconvolution had on repeatability was investigated. The lens opacity was measured subjectively with the Lens Opacity Classification System (LOCS) III grading system. Nuclear opalescence (NO), nuclear color (NC), and posterior subcapsular (PS) and cortical (C) scores were recorded.
A paired nonparametric (Wilcoxon) test was performed to determine whether the improvement in repeatability was statistically significant, because the distributions of differences in MPHSD had a positive skew. Spearmans rank correlation test was used to detect whether there was an association between the average MPHSD and the improvement in repeatability before and after deconvolution. The association between improvement and lens opacity was also quantified in this way.
| Results |
|---|
|
|
|---|
|
|
|
| Discussion |
|---|
|
|
|---|
The adaptive optic scanning laser ophthalmoscope is an alternative method to correct wave-front aberrations produced by the optics of the eye,42 43 but this is not yet available in clinical instruments. Wavefront aberrations detected at a wave-front sensor are modeled using Zernike polynomials. Appropriate signals are sent to a deformable mirror in a feedback loop to compensate. By correcting high-order aberrations, the PSF becomes much narrower resulting in more photons passing through the confocal pinhole.15 This technology has achieved lateral and depth resolution improvements an order of magnitude better than the conventional scanning laser ophthalmoscope. However, even after the corrections, measurable wave-front aberrations remain.15 Christou et al.44 have previously shown that deconvolution can remove the residual wavefront aberration after adaptive optics correction. In this study, blind deconvolution was shown to improve repeatability; however, it is likely that the improvement would be greater if knowledge of the wavefront aberrations could be measured and incorporated into the deconvolution algorithm.
In this study, image restoration algorithms were applied to 3-D confocal images before the position of the peak reflectance was estimated, to construct a topography image. The gain achieved may be greater when fitting more complex models to the z-profile. A recent application of the HRT has been in imaging the disease process of macular edema in diabetic retinopathy; here, a more complex model for the z-profile, the "edema index, " is calculated by Gaussian curve fitting. Further work will investigate whether deconvolution has a significant impact in this area.45
Previously, Burk and Rendon46 evaluated the impact of a new subpixel image alignment algorithm on series of HRT images and reported an improvement in intrascan MPHSD. An average improvement from 22.9 ± 82 (SD) µm with the standard alignment procedure to 15.46 ± 6.8 (SD) µm with the new algorithm was reported on 132 eyes. The study did not report on the improvement in interscan repeatability. It should be noted that the improvement achieved here is an incremental improvement over that already achieved by better alignment algorithms.
All forms of cataract result in additional light scattering, reducing the repeatability of measurements made from scanning laser tomography images. Both cataract and glaucoma occur within the same age range and the two conditions frequently coexist. It is recognized that glaucoma surgery is often followed by a speeding up of a preexisting cataract.47 The image restoration techniques presented in this study may be particularly applicable to glaucoma patients and especially in series of data in cases when cataract may develop during the course of follow-up. Strouthidis et al.11 reported an association between MPHSD and lens opacity measured using LOCS grading, and they reported the highest associations using the NO and NC scales. Siik et al.48 reported a statistically significant association between NO, NC and white light scatter. We report a low association between grading of lens opacity and improvement in MPHSD which, however, was stronger for the intrascan, than the interscan, improvement. This finding is perhaps due to the improvement afforded by deconvolution getting swamped by noise induced by misalignment present after applying HRT image registration algorithms. The z-profiles at the rim area are shown after data exploration of the image in Figure 1 . This important area of the image for monitoring glaucoma and detecting damage during follow-up has low light reflectance properties. In cases with high noise, there will be inherent difficulties in accurately determining the position of maximum reflectance, and thence calculating the topographic height (see Fig. 1g ). However, the deconvolved z-profile (Fig. 1h) clearly shows the position of maximum reflection.
The problems in screening a normal elderly population (aged 6589 years) with scanning laser tomography have been highlighted in a recent study that showed that 10% had a MPHSD above 68 µm and a mean MPHSD of 26.8 ±13.3 (SD) µm with the worst 10% removed.10 Strouthidis et al.11 reported in a population with OHT and primary open angle glaucoma, over a mean of five visits, that MPHSD had a mean of 33.5 ± 23.6 (SD) µm and range of 12 to 130 µm. The instrument guidelines categorize MPHSD < 10 as excellent; 10 to 20, very good; 20 to 30, good; 30 to 40, acceptable; 40 to 50, look for ways to improve; >50, low-quality image and suggest not to use as a baseline image. The results highlighted in Figures 2 and 3 suggest that the improvement obtained by deconvolution was greater in these least-repeatable images. For example, the mean improvement in the five patients with an original MPHSD greater than 60 µm was 10.4 µm. Therefore, these techniques may have a real clinical impact in the use of images that are of a poor quality, and it is hoped that the results from this study combined with further work to optimize and test the techniques on further data sets will result in scanning laser tomographys having wider utility in the assessment of the ONH.
| Acknowledgements |
|---|
| Footnotes |
|---|
Supported in part by funding from the Moorfields Eye Hospital Special Trustees, London.
Submitted for publication February 22, 2006; revised May 8 and 31, 2006; accepted August 3, 2006.
Disclosure: A.J. Patterson, AutoQuant, Inc. (F); D.F. Garway-Heath, Heidelberg Engineering (F); D.P. Crabb, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: David P. Crabb, Department of Optometry and Visual Science, City University, Northampton Square, London EC1V 0HB, UK; d.crabb{at}city.ac.uk.