From the Centre for Ophthalmology, University of Tübingen, Germany.
| Abstract |
|---|
|
|
|---|
METHODS. DTL electrodes were used with a four-alternative, forced-choice method to determine psychophysically the strengthduration curves by using the Weiss model of electrical tissue stimulation in 47 subjects: healthy individuals (n = 17; n = 6 under light- and dark-adapted conditions), patients with open-angle glaucoma (n = 9), retinitis pigmentosa (RP; n = 14), amblyopia (n = 3), or homonymous visual field loss (n = 4).
RESULTS. In all subjects, thresholds were safely determined without side effects. Subjects reported a homogenous, central, white, steady phosphene. The rheobase was higher in glaucoma and patients with RP than in healthy subjects (0.1129 ± 0.0314 and 0.6868 ± 0.1054 vs. 0.0383 ± 0.0057 mA; P = 0.05 and <0.00001, respectively), and it increased with age (r = 0.51; P = 0.038). The rheobase was lower in light- than in dark-adapted conditions (difference 0.015 mA; P = 0.016). In patients with RP, no correlation was found between rheobase or chronaxie and retinal thickness (in OCT) or number of years after reading loss, but there was a correlation between rheobase and visual acuity (P = 0.014). In patients with RP or glaucoma, no inhomogeneity of phosphenes was reported, but all patients with homonymous visual field loss reported lateralization of the phosphene into the area of visual loss.
CONCLUSIONS. Use of DTL electrodes to elicit electrical phosphenes is safe, fast, and reliable. It bears significant advantages over corneal electrodes and provides a valuable tool to elucidate ophthalmologic disease processes and to screen candidates for retinal prostheses.
We therefore propose a new system of stimulation with DTL electrodes (named after Dawson, Trick, and Litzkow18 ), which are in common use for clinical registration of electroretinograms. DTL electrodes are easy-to-construct, single-use items and pose minimal discomfort to the patient. We have psychophysically determined thresholds for different pulse lengths to retrieve chronaxie and rheobase using the Weiss model of electrical tissue stimulation in a series of healthy individuals and patients with open-angle glaucoma, retinitis pigmentosa (RP), amblyopia, and homonymous visual field loss.
| Materials and Methods |
|---|
|
|
|---|
/m). A 4-cm piece is clamped on one side by a connector and on the other side by adhesive tape (Fig. 1A) . The latter is then fixed on the skin close to the medial canthus, and the clamp is attached to the cheek of the subject. The thread then lies on the bulbar conjunctiva just below the corneal limbus. The ground was a gold-plated cup electrode (LKC; Gaithersburg, MD) on the ipsilateral temporal skin.
|
Impedance of the electrodes was tested with an electrophysiologic testing unit (ColorBurst and Espion; Diagnosys LLC, Littleton, MA) and never exceeded 8 k
at 25 Hz.
A four-alternative, forced choice-method was used for psychophysical determination of thresholds. At each stimulus duration a train of one, two, three, or four pulses of the same current strength was given and subjects had to name the number of pulses. The correct answer three times out of four was judged as "seen" or suprathreshold. The current amplitude was changed until the subject named the number of pulses correctly fewer than three times. After the threshold was determined, the stimulus duration was changed. The subject was given a prompt when the series of pulses was started. Pulses were delivered approximately every 1.5 seconds starting at 1 ms and then switching between longer and shorter pulse lengths. Subjects were seated in a completely darkened room. Full darkness was necessary to perceive the very subtle phosphenes, which was not possible in a brighter surrounding because of the lack of contrast. The room light was switched on in regular intervals for the light-adapted experiments, to avoid dark-adaptation during the experiment. In a subgroup of normal subjects, dark adaptation was reached after 30 minutes. Subjects were asked about the location, size, color, and temporal properties of the phosphene and to report any missensations such as muscle activation or pain.
Subjects
All experiments were undertaken with the understanding and written consent of each subject, respecting the Code of Ethics of the World Medical Association (Declaration of Helsinki) and in accordance with the European Communities Council Directive of 24 November 1986 (86/609/EEC). The study was approved by the local universitys ethics committee.
Examinations were performed in 47 subjects in the following groups:
Data Evaluation
The empiric Weiss model of electrical tissue stimulation19 was used to perform a nonlinear fitting to determine rheobase (defined as the minimum suprathreshold current amplitude, theoretically at infinite pulse length) and chronaxie (the pulse length at twice the current of the rheobase) of each individual. Before a least-squares fit was performed, a logarithmic transformation had to be performed to adjust the residuals distribution. The Weiss model equation is as follows:
![]() |
e are considered physiologic tissue constants. After analysis of variance (ANOVA) the Dunnett test was used to compare the groups means against the mean of the control group. For comparisons between measurements in dark- and light-adapted conditions in the same individual, a paired t-test was performed. All calculations and statistical analyses were performed with commercial software (JMP IN, version 5.1; SAS Institute, Cary, NC).
A Bland-Altman-analysis20 assessed repeatability in five subjects of the control group, to calculate the mean difference and the coefficient of repeatability (CR; calculated as 1.96 · SD of the differences).
| Results |
|---|
|
|
|---|
Healthy Individuals
All subjects in the control group reported a whitish circular phosphene in the center of the visual field extending approximately 20° in each direction. Brightness usually decreased a little with the second, third, and fourth pulses (if given). Different flash durations were not noticed.
The rheobase was 0.0383 ± 0.0057 mA and chronaxie, 11.27 ± 2.11. The mean of all healthy individuals was fitted using the Weiss model with a high determination coefficient of r2 = 0.889. We found a linear correlation between the standard deviation of the threshold current and the logarithm of stimulus duration (SD = 0.196 + 0.066 · log(t) mA; r2 = 0.740; P < 0.01) and were thus able to plot the 95% confidence intervals for the Weiss model of the means of the group (Fig. 2) .
|
|
|
|
|
|
|
The rheobase in this group was significantly higher than in the control group, but the chronaxie was not significantly different (0.1129 ± 0.0314 mA, P = 0.05; 16.55 ± 4.62 ms, P = 0.783, respectively; Figs. 6B 7 ; Table 1 ). Rheobase levels in three patients are within the 95% confidence interval of the control group: 1 below and 5 above (Fig. 5) .
Patients with Retinitis Pigmentosa.
The subjective description of the phosphenes was not different from that in the control group. All patients reported an evenly distributed phosphene and none of the patients reported any correlation to remaining "islands" of visual function within the visual field (if present). In four patients with RP, no phosphenes were elicitable, despite stimulation with currents approximately 100 times as high as the average current in the control group. These patients were not included in the statistical work-up.
The rheobase in this group was significantly higher than in the control group, whereas the chronaxie was not significantly different (0.6868 ± 0.1054 mA, P < 0.00,001; 13.81 ± 5.17 ms, P = 0.970, respectively; Figs. 6C 7 ; Table 1 ). The fitting for the means of the patients with RP does not run parallel to the fitting of the control groups means. For the long pulses the difference is much greater and far above the upper 95% confidence interval, whereas for the shorter pulses the fitting approximates the fitting of the control group and lies well within the 95% confidence interval (Fig. 6C) . In all patients with RP, rheobase is above the 95% confidence of the control group (Fig. 5) .
There was no correlation between rheobase and the number of years after loss of reading ability or between rheobase and retinal thickness as assessed by OCT. However, there was a statistically significant difference in rheobase between patients with and those without light perception (P = 0.014). The chronaxie did not correlate significantly with any of these values.
Patients with Homonymous Visual Field Loss.
Most striking in this group was that all four subjects unanimously and spontaneously reported that the phosphene was brightest in the area of their visual loss.
The rheobase and chronaxie in this group were not significantly different in comparison with those in the control group (0.02785 ± 0.0106 mA, P = 0.934; 18.59 ± 8.01 ms, P = 0.709, respectively; Fig. 6D 7 ; Table 1 ). The fitting for the means of this group runs parallel to the fitting of the control groups meansthat is, there is no difference between the short and the long pulses in comparison to the control group (Fig. 6D) . All rheobases in this group are outside the 95% confidence interval of the control groupthree of them below and one slightly above (Fig. 5) .
| Discussion |
|---|
|
|
|---|
All healthy subjects reported homogeneous, steady, white flashes in the central visual field extending an estimated 20° in each direction from the center. This corresponds to the area of the highest rod and cone bipolar cell density in the primate retina.25 Although there is ongoing debate on the exact site of electrical stimulation within the retina and some have argued that the photoreceptors are preferentially driven,17 evidence is emerging that photoreceptors and their synapses to bipolar cells are the preferential target because even retinas without any functioning photoreceptors can be stimulated within reasonable thresholds,1 3 4 17 26 27 a finding that is supported by our study where even patients with RP with only slight light perception had near normal thresholds (e.g., patient 2 in Fig. 5 ). Simultaneous stimulation of all wavelength bipolar cells or synapses to photoreceptors would arguably yield a whitish phosphene.
Patients with glaucoma with arcuate superior visual field loss reported the same central homogeneous phosphene as the control group, and no difference was noticed between the affected upper and the unaffected lower hemisphere. Whereas it might be difficult for subjects to describe the exact size of a phosphene, the sparing of the visual field center in our patients (Fig. 1B) could explain the intact phosphene perception in these patients and lack of correlation to the visual field loss. It was expected that patients with absolute glaucoma did not perceive EEPs because ganglion cells which are considered the primarily affected cells in glaucoma28 29 30 represent the most centrally located neurons in the retina, and their loss will inhibit conduction of any signal to the brain. In patients with RP, only central phosphenes were reported, and no correlation with remaining visual islands was found. This is somewhat in contradiction to other studies, which have found phosphenes at thresholds that begin in the periphery and reach the center only at higher amplitudes,10 although homogeneous excitation of the whole retina can be assumed through external stimulation by modeling the volume conductors of the eye.31 Considerable reorganization processes and build-up of nonphysiological synapses in RP retinas32 with misdirection of electrical excitation could be responsible for the perceived flicker phenomena in some patients with RP and seems expectable from the spontaneous observation of these phenomena by some of these patients. In contrast to the homogeneous phosphene in glaucoma and patients with RP, it is very striking that all patients with homonymous visual field loss (quadrantic or hemianopic) did not perceive their visual loss but rather lateralized the phosphene spontaneously and even with a touch of surprise in the area of their loss. It is well known that cortically blind patientsin complete absence of acknowledged awarenessunconsciously localize light stimuli into their impaired hemisphere (known as blindsight33 34 ). It has been argued that extrastriate cortical areas are involved in this unconscious perception,33 and it may be that a comparable phenomenon in our patients led to perception of EEPs in the area of their visual loss by processing of these electrically generated signals in extrastriate areas, with bypassing of the impaired brain region to reach conscious perception.
The values of the fitting curve of diffusely EEPs in patients with amblyopia (Fig. 6A ; Table 1 ) were practically identical with those in the control group, which is in accordance with electrophysiologic findings that show reduced cortical and retinal responses from pattern stimulation and normal responses from full-field stimulation.35
Our finding that thresholds are significantly lower in the light-adapted state is in opposition to previous studies that showed no dependency on retinal adaptation.3 16 The difference was clear, however, in each individual subject and the difference were easily appreciated by determining thresholds directly after pointing an ophthalmic flashlight at the eye and after only 30 seconds of dark adaptation. The speed of this mechanism correlates with the speed of the initial cone sensitivity loss during dark adaptation36 and indicates preferential involvement of the cone or the cone bipolar system in EEP perception.
The rheobase and chronaxie were aptly estimated and used in the statistical work-up of the Weiss model of electrical tissue stimulation19 in all subjects. Using a fitting model to estimate rheobase and chronaxie and evaluate electrical excitability is superior to drawing conclusions from thresholds at discrete pulse durations only: first, because the influence of possible outliers is reduced, which makes evaluation of physiological parameters more robust, and second, because the evaluation of the fitting curves produces more information than does a single threshold alone (e.g., a difference between longer and shorter pulses can be appreciated; Figs. 6B 6C in patients with glaucoma and RP, respectively).
Our rheobase levels are on the order of magnitude of other studies of ocular surface stimulation. Delbeke et al.8 found an average value for the rheobase in 10 healthy individuals of 0.28 mA using periorbital skin electrodes, which is an order of magnitude higher than our average rheobase of 0.038 mA (calculated from the Weiss model; Table 1 ) and can be readily explained by the addition of a serial resistance by the closed eyelid. Dorfman et al.,5 using a Burian-Allen contact lens electrode, found thresholds of between 2 and 3 mA at 1-ms pulse duration in four healthy individuals. This threshold is approximately an order of magnitude higher than our finding of 0.3 mA (Fig. 2B) . Miyake et al.,16 using a corneal contact electrode, found thresholds in healthy individuals at a 5-ms pulse duration of 0.29 mA, which is approximately three times higher than our value of 0.096 mA at 5 ms (Fig. 2B) . It must be considered that Miyake et al.16 and Dorfman et al. 5 used objective measurements of thresholds by registering cortical potentials that certainly yield higher thresholds than subjective estimation. Finally, Morimoto et al.,10 using a Burian-Allen contact lens electrode, found a threshold in healthy individuals of 0.065 mA, which is still twice as high as our rheobase of 0.038 mA. We speculate that the psychophysical determination of thresholds yields lower values than objective tests, such as recording cortical potentials or pupillary responses, such as Morimoto et al. performed. In patients with RP, values of approximately 4 mA at 4 ms are have been recorded (with a Burian-Allen corneal contact electrode), which is 1.5 times higher than our 2.63 mA at 4 ms in these patients (Fig. 6C) . Morimoto et al.10 found a mean threshold of 0.0545 mA in patients with RP, which is somewhat lower than our rheobase of 687 mA. Both studies showed large interindividual variation (SE was 0.411 mA in Morimoto et al. and 0.105 mA in our study). Intraocular stimulation obviously requires lower thresholds, which are reported in the range of 0.1 mA at 16 ms, for epiretinal stimulation in patients with RP37 and are six times lower than our threshold of approximately 0.6 mA with extraocular stimulation (Fig. 6C) . For subretinal stimulation, no thresholds are thus far available in humans, but studies are under way (Zrenner E et al. IOVS 2006;47;ARVO E-Abstract 1538). Ultimately, it must be the goal to establish a method that allows estimation of intraocular thresholds from external stimulation, to select valid patients for retinal prosthesis implantation.
We found no statistically significant difference in the chronaxie across groups in our study. The chronaxie and rheobase are both considered physiological tissue constants.19 38 In simultaneous stimulation of a large number of different retinal cell types, axons, and polarization23 24 the strengthduration curve will inevitably deviate from the original curve of single cell stimulation. The rheobase is in many studies used synonymously to the threshold current amplitude, although chronaxie is less frequently used but has been shown to be the stimulation paradigm to correlate with the least energy.38 The deviating curve through multicell-type stimulation where shorter pulses tend to lie closer to the normal subjects fitting curve (Fig. 6) could be responsible for our finding that no significant differences in chronaxie were found.
Across all groups, the rheobase was significantly higher in our patients with RP than in the control group in analogy to all above mentioned studies. We found no correlation between the clinical parameters retinal thickness (as assessed by OCT) or years after reading loss (Table 2) and the elicitability of EEPs (as expressed by the rheobase). The only parameter that showed a correlation with the rheobase was visual acuity (light perception or no light perception; P = 0.014). However, some eyes without light perception had elicitable EEPs, and some eyes with intact light perception did not perceive them (Fig. 5 , Table 2 ). Morimoto et al.10 found no correlation, whereas Yanai et al.6 found a more distinct correlation than our study. Regarding function testing for a retinal prosthesis, none of these parameters can be used as an isolated predictor of the possibility that phosphenes can be elicited through prostheses. Because valid selection criteria for patients in implant surgery are required, it seems reasonable to use EEPs as a preoperative testing method.
|
In conclusion, the use of standard DTL electrodes appeared safe and reliable throughout our study. In conjunction with a four-alternative, forced-choice method, the fitting algorithms based on the Weiss model allowed a detailed statistical work-up of the chronaxie and rheobase. The obvious advantages of DTL electrodes could help to increase the use of EEPs in research and clinics. Eliciting EEPs by the proposed method seems to be a valid predictor of the excitability of eyes that are to undergo retinal implant surgery, and valuable information can be gained in other ophthalmic diseases, as we have shown by our peculiar and unexpected finding in patients with homonymous visual field loss.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication April 24, 2006; revised July 28, 2006; accepted September 22, 2006.
Disclosure: F. Gekeler, None; A. Messias, None; M. Ottinger, None; K.U. Bartz-Schmidt, None; E. Zrenner, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Florian Gekeler, Centre for Ophthalmology, Schleichstrasse 1216, D-72076 Tübingen, Germany; gekeler{at}uni-tuebingen.de.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. Inomata, K. Tsunoda, G. Hanazono, Y. Kazato, K. Shinoda, M. Yuzawa, M. Tanifuji, and Y. Miyake Distribution of Retinal Responses Evoked by Transscleral Electrical Stimulation Detected by Intrinsic Signal Imaging in Macaque Monkeys Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 2193 - 2200. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||