From The Netherlands Ophthalmic Research Institute, Royal Netherlands Academy of Arts and Sciences, Amsterdam, The Netherlands.
| Abstract |
|---|
|
|
|---|
METHODS. The psychophysical technique of the "direct compensation" method was adapted to make it suitable for routine clinical assessment. In the new approach, called "compensation comparison, " the central test field is subdivided into two half fields: one with and one without counterphase compensation light. The subjects task is a forced-choice comparison between the two half fields, to decide which half flickers more strongly. A theoretical form for the respective psychometric function was defined and experimentally verified in a laboratory experiment involving seven subjects, with and without artificially increased light scattering. The method was applied in a separate multicenter study. Its reliability was additionally tested with a commercial implement (C-Quant; Oculus Optikgeräte, Wetzlar-Dutenhofen, Germany).
RESULTS. A repeated-measures SD of 0.07 log units was achieved, to be compared with differences in the young normal population of 0.4 log units and an increase with healthy aging by 0.5 log units at 80 years and by 1.0 or more log units with (early) cataract or corneal disturbances. Reliability was further found to be high when using the commercial version of the method.
CONCLUSIONS. The compensation comparison method for measuring retinal straylight is suited for clinical use to diagnose patients with complaints caused by large angle light scattering in the eye such as early cataract.
Retinal straylight can be seen as the outer skirt of the pointspread function,10 outside, say, 1°. It causes a veiling luminance over the whole retina that adds to the retinal projection of the visual scene, thereby reducing the contrast of the retinal image. Disability glare, as defined by the Commission International dEclairage,4 corresponds to retinal straylight, which is quantified by means of the concept of equivalent luminance (i.e., the [external] luminance that has the same visual effect as the glare source at some angular distance).4
The first attempts to measure intraocular straylight by means of equivalent luminance involved the comparison of two threshold measurements: one threshold in the presence of a distant glare source and one threshold in the presence of a homogeneous background (equivalent) luminance.4 Van den Berg and IJspeert11 compared the results from various groups, all using this method, and concluded that these results varied considerably. Moreover, the method was not widely used, because it was not easily accessible for clinical application. As easy-to-use alternatives, so-called glare testers were introduced that usually consisted of visual acuity (e.g., ETDRS,12 Ferris-Bailey,13 or Regan14 charts) or contrast sensitivity (e.g., sinusoidal gratings,7 12 14 15 Landolt rings,12 16 or Pelli-Robson charts13 14 17 ) test, with and without a glare source presented at some angular distance in the visual field. Although glare testers were occasionally appraised favorably,14 more often provided unreliable results, demonstrated by their outcomes correlating badly with various validity measures such as outdoor visual acuity in bright sunlight,12 15 a questionnaire assessing perceived visual disability,13 16 or directly measured forward light scatter.14 16 Also, the repeatability and discriminative ability of studied glare tests were found to be inadequate.14 16 A particular example is the omission of the glare measurement results, performed with the Miller-Nadler glare tester, in the final results of the large multicenter PERK study,18 because the glare tester was not sensitive enough to detect small but significant amounts of light scattering,19 which was also mentioned in later studies.14 20 As a result of these issues with glare testers, a standard way of glare measurement was never adopted, and some overview papers discussing glare test problems appeared.21 22 23 24 25
To improve on this situation, Van den Berg5 proposed a new psychophysical method, called the direct compensation method. In short, this method works as follows (Fig. 1) : A bright, ring-shaped, flickering light source is presented at a certain angular distance (
) from a (dark) test field. Because of intraocular scatter, part of the light from the bright straylight source is projected on the retina at the location of the test field, inducing a (weak) flicker in the test field. To determine the exact amount of straylight, variable counterphase compensation light is presented in the test field. By adjustment of the amount of compensation light, the flicker perception in the test field can be extinguished. In this way, there is "direct compensation" for the straylight modulation caused by light scattered from the glare source.
|
In 1990, the direct compensation technique was implemented in a small portable device, called a straylight meter, to accommodate other researchers.30 31 32 This method led to publications, notably by Elliott et al.,6 on a variety of subjects, such as the already mentioned cataract and disability glare test evaluation14 studies. Furthermore, they found increased straylight values after induced corneal edema,8 in contact lens wearers33 as well as in 25% of the subjects 1 year after excimer laser photorefractive keratectomy.34 Advantages of the direct compensation method over alternative methods of assessing wide angle-scatter were mentioned in a paper discussing these methods for use in evaluating visual function in cataract.24 Other researchers used the straylight meter after refractive surgery and found increases in small scatter angles and dilated pupils after radial keratotomy,35 but no increases more than 2 weeks after photorefractive keratectomy,20 36 except in some individuals.37 38 Ocular lubricants were reported to have no adverse effects on the optical quality of the eye.39 The straylight meter again showed increases in patients with retinitis pigmentosa40 41 and also in those with choroideremias.42 Straylight meter readings were found to correlate significantly with clinical grading of lens opacities and lens back scatter,17 with corneal swelling,43 with lens opacity measurements in patients with glaucoma,44 and with posterior capsule opacification.45 Most recently, the direct compensation method was used in a field study investigating the suitability of several glare tests for drivers license applications and was found to be the most promising candidate.16 46
In general, the direct compensation method has given a great boost to the study of retinal straylight. Moreover, it was emphasized in the literature that this technique has much greater sensitivity than do glare tests, for example in patients with corneal edema8 and posterior capsular opacification.45 It was also the gold standard for assessing the validity of glare tests.14 However, outside the laboratory, it was a difficult technique to use.38 45 In a field study16 46 involving 112 subjects drawn from the patients and visitors of the outpatient departments of three clinics, the standard deviations of differences between repeated measurements found in such a field study were 0.15 and 0.18 log units, for two different implementations of the direct compensation method. It appears that the method has some major drawbacks for routine clinical or large-scale use: (1) Judgment of the weak flicker in the test field often appeared to be difficult for untrained subjects. This seemed to be caused by the presence of the strong flicker of the straylight source. (2) Usually, visual tests are based on what subjects actually see. On the contrary, in the direct compensation method, the subjects have to indicate whether the flicker perception has disappeared. The continuous flickering of the straylight source in the periphery made this contraintuitive task even more difficult. (3) The accuracy of the measurement seemed to depend on the adjustment strategy, which could differ considerably between subjects, and on proper explanation of the test. (4) There was no control over an individuals measurement reliability. (5) Subjects had the ability to influence the test outcome. This aspect is particularly important in the field of driver testing.
As a result of these drawbacks, the straylight meter largely remained limited to laboratory use. The instrument could not be used on a large scale, such as clinical diagnosis or occupational health testing. For these applications, the test must be easy to understand, easy and quick to perform, easy to explain, and fraud resistant. Also it should be criterion independent, so that the values have universal validity and results from different locations can be compared.
To overcome these limitations, we proposed a new method to measure retinal straylight, the "compensation comparison" method. In essence, this method presents exactly the same stimuli to the subject as the direct compensation method. Note that in the direct compensation method, the amount of compensation light is varied until the straylight flicker has disappeared. In other words, in the direct compensation method, the subject compares different stimuli sequentially. In contrast, in the compensation comparison method, two stimuli of the direct compensation method are presented to and compared by the subject simultaneously. In this way, the direct compensation method is implemented as a two-alternative forced-choice (2AFC) approach. The characteristics of the psychometric function for this 2AFC method will be reported in this article. This function determines what comparisons would be the best to use. The compensation comparison method has been summarized in abstract form (Van den Berg TJTP, et al. IOVS 2005;46:ARVO E-Abstract 4315) and in a patent.47
The compensation comparison method has been used in a field study involving 2422 subjects (GLARE, see www.glare.be) and in other projects such as a study investigating the wavelength dependence of retinal straylight.48 Some results from the GLARE study, pertinent to the present question, will be used in this report. In this study, several visual tests, including straylight measurements, were performed among a population of drivers in Europe, spread over five age categories. Data were collected in clinics in The Netherlands, Austria, Germany, Spain, and Belgium. Since the study was designed to assess the prevalence of vision impairments in the driving population, the only inclusion criterion was being an active driver. As a result, the measured population consisted of a wide range of subjects, including ages from 20 to 85, visual acuities below 0.5 (logMAR [logarithm of the minimum angle of resolution] 0.3) to more than 1.0 (logMAR 0.0), visual field defects, and other ocular diseases such as glaucoma and cataract. This huge variation in ocular conditions provided an ideal opportunity to evaluate the compensation comparison method in clinical practice.
In the present paper, the principles, design considerations, and advantages of the compensation comparison method with respect to the direct compensation method are discussed, and a model for flicker comparison using this method is proposed and tested in a laboratory experiment. This model comprises a psychometric function designed to describe the (stochastic) characteristics of the responses in a compensation comparison experiment. For simplicity, real error responses (false-positive and false-negative mistakes of the subjects) were not included in the formulas that follow. These values are very low (on the order of 1% or less) in laboratory experiments. Their inclusion is straightforward though, and they were included in the final formulas used for the field study. The reliability of the compensation comparison method was tested with a commercially available embodiment of the method (C-Quant, manufactured by the Germany based firm Oculus Optikgeräte GmbH (Wetzlar-Dutenhofen, Germany).47
| Methods |
|---|
|
|
|---|
To test the compensation comparison method also for conditions of increased scattering, the same subjects were measured with a lightdiffusing filter (Black Pro Mist 2 BPM2; Tiffen Manufacturing, Hauppauge, NY) in front of the tested eye. This filter, among a collection of 23 commercially available lightdiffusing filters, was found to have the best light-scattering characteristics for mimicking (early) cataract or aging effects in the human crystalline lens.49
As mentioned before, the compensation comparison method was evaluated in the European GLARE study. In the course of this study, some improvements were made on the implementation of the test, as will be described at the end of this section.
For stimulus generation, a computer system with either a CRT monitor or combination of digital light processing (DLP) projector and back-projection screen was used. The straylight source was a white light annulus extending from 7° to 14°. Because of the approximate 1/
2 dependence of retinal straylight, this corresponds to a 10° scattering angle.10
To test the reliability of the clinical version of the compensation comparison method (C-Quant; Oculus Optikgeräte), 17 subjects with no experience in the direct compensation and compensation comparison measuring techniques were recruited from a neighboring institute. The average age was 44 years (range, 2881). Except for the oldest subject, all were without ocular defects. Refraction ranged from 7 to +3 D. All measurements were performed monocularly on the subjects preferred eyes, without glasses or contacts. Thoroughly cleaned trial glasses were used when appropriate. All subjects performed six measurements: three without and three with the BPM2 filter in front of the studied eye.
Basics of the Compensation Comparison Method
The test screen layout of the compensation comparisonbased straylight meter is similar to that of the direct compensation method, only the test field is now divided in two halves (Fig. 2) . Compensation light is presented in one of the two test field halves (randomly chosen, referred to as field b in the remainder of the article), whereas no compensation light is present in the other test field half (referred to as field a). As a result, two flickers are perceived, that differ in modulation depth: one results from straylight only (field a), the other is a combination of straylight and compensation light (field b), flickering in counterphase with this straylight. Simplified, the procedure runs as follows (Fig. 3) : during the test, a series of limitedduration stimuli are presented that differ in the amount of compensation light in test field b. After a 2AFC paradigm, the task for the subject is to decide for each stimulus which test field half flickers stronger. The subjects responses are recorded by means of two push buttons, representing the left and right test fields. Using the psychophysical model for this flicker comparison task, which will be described in detail later in the article, a psychometric curve is fitted to the subjects responses, from which both the straylight parameter and a measure for the quality of the measurement can be deduced.
|
|
|
|
The stimuli in the initial phase are equidistant, with a step size of 0.1 log units (except for the first step which is 0.3 log units) and presented in order from high to low straylight (Fig. 4 , increasing numbers). The absolute stimulus values of the initial phase can be placed differently and can be chosen by the operator, but in the GLARE study it was set to adjust for the known population averages as a function of age.10 The example given in Figure 4 is for a 30-year-old subject. For a 70-year-old subject, all initial phase stimuli were shifted upward by 0.3 log units.
In the first stimulus, a very weakly flickering ring is presented (stimulus 1 in Fig. 4 ). Then it is very easy to recognize the test field half with compensation. Subsequently, the intensity of the ring is increased, thereby increasing the difficulty of the flicker comparison task. This relates to the real-life experience of being disturbed more and more by glare sources with higher intensities. In the final phase, the ring flickers at constant intensity, whereas the compensation luminance in field b is varied. The stimuli in the final phase (Fig. 4 , xs) are logarithmically equidistant at 0.05 log units in a fixed interval around the first coarse estimate of the 50% point of the psychometric curve, as based on the data of the initial phase (Fig. 4 , dots). In the final phase, the stimuli are presented in random order, according to the method of constant stimuli.51
The Psychometric Function
As a basis to describe the psychometric function, we started out from the well-known logistic function.52 Comparing two flickering test fields a and b with different modulation depths, the chance probability (P) of choosing one of the test fields as having the stronger flicker was written as (Fig. 6)
![]() | (1) |
![]() | (2) |
|
![]() | (3) |
![]() | (4) |
![]() | (5) |
![]() | (6) |
![]() | (7) |
Once the shape of the psychometric function has been established, estimation of the straylight parameter value s in individual subjects involves shifting of the psychometric function to fit the dataset of that individual. Fitting is achieved by means of the maximum-likelihood procedure, as just outlined. An example of such a fit is given in Figure 4 . In this case, log(s) was found to be 0.85. The straylight value is determined by the horizontal position of the minimum of the curve, where MDb= 0 and Scomp= s. This approach was applied in the European GLARE study involving 2422 subjects in total. In the course of the study, some improvements were made on the implementation of the straylight test: (1) A three-trial instruction phase was added before the real measurement, to familiarize the subject with the flickercomparison task. (2) The subjects responses were displayed to the operator during the measurement, making it possible to interfere in case the response pattern was erratic and start a new measurement after additional explanation. (3) The luminance in the test fields was increased by a factor of 2 in the initial phase, making the measurement easier for older subjects. In total, 1073 subjects were measured with this final version (including these improvements). More detailed reports of this study are in preparation, but some preliminary data will be given herein to test the psychometric function (equation 1) and to illustrate the performance of the test.
| Results |
|---|
|
|
|---|
|
|
The model was further validated by applying it to field measurements of 1073 subjects, performed in the European GLARE study, as described in the previous section. The wide variation in ocular conditions found in this population can be expected to reflect itself in different psychophysical behavior, and therefore in psychometric functions that differ between these 1073 individuals. To analyze this, all measurements were performed twice and divided in nine groups of equal size, sorted on the differences between the two repeated measurements. In each group, equation 1 was then fitted to all data, after normalizing each individual curve for the individual straylight value. Results are given in Figure 8 . The best 67% (top six panels) of the 1073 subjects have a repeated-measurement SD of 0.036, and 89% (all but the last panel) of 0.059 log units, whereas the SD for all measurements is 0.099 log units.
|
|
| Discussion |
|---|
|
|
|---|
This study was provoked, among others, by existing evidence in the literature that there is a clinical need for testing a patients glare sensitivity. As outlined in the introduction, many different glare testers have been proposed, most of which have disappeared from the market. Some studies tried to validate glare testing against straylight as the gold standard, but with questionable results.7 14 16 17 Repeatability was compared between different glare tests and the direct compensation method,14 leading to the conclusion that the direct compensation method performs better. With the improved performance of the compensation comparison method, this will, a fortiori, be the case again. The present compensation comparison technique offers new opportunities to test and validate the performance of glare testers.
To obtain these results, some understanding of the underlying psychometric function was needed. The proposed model describes measured laboratory data well (Fig. 7) for a wide range of straylight values (Table 1) . The log(s) values without BPM2 filter all fall within the normal population range, which has been shown to increase with age.10 From this study, it follows that the relation between straylight parameter s and age can be approximated, in a white population with a 10° scattering angle, by the equation s = 7(1+(age/70)4 ), with an uncertainty of 0.1 log units. In Table 1 , this average age normal population value is given for each subject.
The log(s) values with BPM2 filter show less variation. This is because the total straylight is a combination of the filter (which itself has log(s) = 1.12) and the eye (log(s) values from 0.55 to 1.15). The experimental values for the eyefilter combinations (log(s) values from 1.20 to 1.39) correspond well to values that can be predicted by calculation (log(s) values from 1.22 to 1.44).
Figure 8 shows that the model is capable of accurately describing the psychophysical behavior of a population that varies widely with respect to physical condition of the eye. Subdividing the population according to differences between two repeated measurements reveals different slopes of the psychometric curves of the various subgroups, accounted for in the model by different MDCc values. The model fits fairly well to all subgroups of Figure 8 , except for the subgroup with the largest repeated measurement differences (lower right panel). For some cases in this subgroup, response behavior was so erratic that reliably fitting a psychometric curve and therefore reliably estimating the log(s) value, is not possible. To detect such erratic behavior automatically during measurements, we developed a reliability parameter, as mentioned earlier. This parameter must assume a certain shape of the psychometric function and was based on the analysis in the present paper. After the lowest-quality measurements were filtered out with this parameter, the overall SD of repeated measurements was between 0.06 and 0.1 log units, which is very good, considering the variation in straylight parameter in the (clinical) population.
Figure 9 shows that the compensation comparison method gives highly repeatable results in untrained subjects, over a wide range of straylight values. The measurements with BPM2 filter follow the additive model for eye plus filter very accurately, indicating that the instrument measures absolute straylight values very well. The fitted log(s) value for the filter (1.14) is very well in accordance with the objectively measured log(s) value for this filter of 1.12.49
The compensation comparison method for measuring retinal straylight was designed as an improvement on the direct compensation technique. According to feedback we got from the operators in the clinics who participated in the GLARE study and who also had earlier experience with the direct compensation method, the task is easier and more intuitive (mostly suprathreshold, short stimulus presentations), easier to explain and less dependent on explanation from the operator. The measurement time is fixed and limited, making the test more pleasant for both patient and operator. However, we did not collect systematic statistical data on these subjective assessments. Moreover, the reliability of the compensation comparison method was shown to be very good, and a reliability index was developed, based on the dataset of a tested individual.
Given these advantages, retinal straylight measurement is now possible on a large scale and in the clinical routine. As a result, the compensation comparison method described in this article has been implemented in a commercially available measurement device (the C-Quant; Oculus Optikgeräte). For future development, the model for flicker comparison gives a basis for improving on the measurement performance by studying different measurement strategies, such as adaptive methods.
| Footnotes |
|---|
Disclosure: L. Franssen, None; J.E. Coppens, None; T.J.T.P. van den Berg, (P)
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Luuk Franssen, The Netherlands Ophthalmic Research Institute, Royal Netherlands Academy of Arts and Sciences, Meibergdreef 47, 1105 BA Amsterdam, The Netherlands; l.franssen{at}ioi.knaw.nl
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
I. J. E. van der Meulen, L. A. Engelbrecht, T. C. T. Van Riet, R. Lapid-Gortzak, C. P. Nieuwendaal, M. P. Mourits, and T. J. T. P. van den Berg Contributions of the Capsulorrhexis to Straylight Arch Ophthalmol, October 1, 2009; 127(10): 1290 - 1295. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Anderson, T. Redmond, D. R. McDowell, K. M. M. Breslin, and M. B. Zlatkova The Robustness of Various Forms of Perimetry to Different Levels of Induced Intraocular Stray Light Invest. Ophthalmol. Vis. Sci., August 1, 2009; 50(8): 4022 - 4028. [Abstract] [Full Text] [PDF] |
||||
![]() |
I J E van der Meulen, H Porooshani, and T J T P van den Berg Light-scattering characteristics of explanted opacified Aquasense intraocular lenses Br J Ophthalmol, June 1, 2009; 93(6): 830 - 832. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. V. Patel, K. H. Baratz, D. O. Hodge, L. J. Maguire, and J. W. McLaren The Effect of Corneal Light Scatter on Vision After Descemet Stripping With Endothelial Keratoplasty Arch Ophthalmol, February 1, 2009; 127(2): 153 - 160. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Cervino, R Montes-Mico, and S L Hosking Performance of the compensation comparison method for retinal straylight measurement: effect of patient's age on repeatability Br J Ophthalmol, June 1, 2008; 92(6): 788 - 791. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Franssen, J. Tabernero, J. E. Coppens, and T. J. T. P. van den Berg Pupil Size and Retinal Straylight in the Normal Eye Invest. Ophthalmol. Vis. Sci., May 1, 2007; 48(5): 2375 - 2382. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Franssen, J. E. Coppens, and T. J. T. P. van den Berg Modulation depth threshold in the Compensation Comparison approach J Vis, January 1, 2007; 7(1): 8 - 8. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Costello, S. Johnsen, K. O. Gilliland, C. D. Freel, and W. C. Fowler Predicted Light Scattering from Particles Observed in Human Age-Related Nuclear Cataracts Using Mie Scattering Theory Invest. Ophthalmol. Vis. Sci., January 1, 2007; 48(1): 303 - 312. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |