From the Contact Lens and Visual Optics Laboratory, School of Optometry, Queensland University of Technology, Brisbane, Queensland, Australia.
| Abstract |
|---|
|
|
|---|
METHODS. Corneal topography data were acquired for 100 young adult subjects by a method that allows central and peripheral maps to be combined to produce one large, extended corneal topography map. This computer-based method involves matching the common topographical features in the overlapping maps. Corneal height, axial radius of curvature, and axial power data were analyzed. The corneal height data were also fit with Zernike polynomials.
RESULTS. Conic fitting to the corneal height data revealed the average apical radius (Ro) was 7.77 ± 0.2-mm and asphericity (Q) was 0.19 ± 0.1 for a 6-mm corneal diameter. The conic fit parameters were both found to change significantly for increasing corneal diameters. For a 10-mm corneal diameter, Ro was 7.72 ± 0.2 mm and Q was 0.36 ± 0.1. A slight but significant meridional variation was found in Q, with the steepest principal corneal meridian found to flatten at a slightly greater rate than the flattest meridian. The RMS fit error for the conic section was found to increase markedly for larger corneal diameters. Higher-order polynomial fits were needed to fit the peripheral corneal data adequately. Analysis of the axial power data revealed highly significant changes occurring in the corneal best-fit spherocylinder with increasing distance from the corneal center. The peripheral cornea was found to become significantly flatter and to decrease slightly in its toricity. Individual subjects exhibited a range of different patterns of central and peripheral corneal topography. Several of the higher order corneal surface Zernike coefficients were found to change significantly with increasing corneal diameter.
CONCLUSIONS. Highly significant changes occur in the shape of the cornea in the periphery. On average, the peripheral cornea becomes significantly flatter and slightly less astigmatic than the central cornea. A conic section is a poor estimator of the peripheral cornea.
A simple and commonly used method to describe corneal shape quantitatively is as a conic section, defined by the apical radius (Ro) and the asphericity parameter (Q), which describes the type of conicoid that best fits the corneal shape (i.e., the degree to which the surface departs from a sphere) where Q = 0 describes a sphere, Q > 0 describes an oblate or steepening ellipse, and 1 < Q < 0 describes a prolate, or flattening, ellipse.
Several studies have derived the average Ro and Q for a normal adult population.11 12 13 14 15 These studies have produced average Ro ranging from 7.68 to 7.85 and average Q ranging from 0.33 to 0.18 (i.e., nearly all subjects exhibit corneas with a prolate elliptical shape). Although the average results across these studies have been relatively consistent, most investigators note that large variations in corneal shape exist between normal subjects. Recently, more complex quantitative descriptors of shape such as Fourier series analysis16 and Zernike polynomials17 18 have been used to describe the shape of the cornea.
Most studies of corneal shape have investigated the central 6 mm of the cornea. Although data from the central cornea are obviously the most important for vision, this represents only approximately one fourth of the corneas total surface area. Information from the peripheral cornea is particularly important in the design and fitting of contact lenses. The area of the cornea measured by videokeratoscopes is much larger than that of the traditional keratometer. For videokeratoscopes based on the Placido disc principle, corneal coverage is limited by the fact that the instrument is based on specular reflection from the corneal surface and further limited by obscuration of the ring image by the subjects nose, brow, and eyelashes.19 Corneal coverage with Placido-based videokeratoscopes is improved by using a small measurement cone, as obscuration by the nose and brow is avoided. Videokeratoscopes based on other principles (such as slit scanning and raster stereography techniques) have the potential to measure larger corneal areas than Placido-based systems,20 although they are still limited by interference from the eyelids and eyelashes. However, these other techniques have generally not proved to be as precise or accurate in their measurements as small cone Placido-based systems.21 22
There have been relatively few studies investigating the shape of the peripheral cornea, outside the central 6-mm diameter. Mandell23 used offset fixation points in a videokeratoscope to measure peripheral corneal topography in 11 subjects and found that for peripheral measurements, an ellipse was a poor descriptor of cornea shape, due to increased flattening of the corneal surface in the periphery. An average tangential radius of 11.29 ±1.82 mm was found at 4.5 mm from corneal center for the 11 subjects tested. Reddy et al.3 classified subjects corneal astigmatism based on central and peripheral corneal topography data.
We have recently developed a technique that allows measurement of the peripheral cornea and the subsequent combination of central and peripheral corneal topography data to provide a total corneal topography map.24 Using this technique, we measured the total corneal topography of the right eyes of 100 subjects. The purpose of this study was to provide normative information regarding the shape of the total cornea for a large population of young, healthy, adult subjects with a range of normal refractive errors.
| Methods |
|---|
|
|
|---|
Of the 100 subjects participating, eight subjects were excluded from subsequent analyses due to poor correlation between central and peripheral corneal topography maps (the criteria for subject exclusion due to poor map correlation is described later). This left 92 subjects in the final data analysis. The subjects ages ranged from 18 to 35 years, with a mean age of 24 years. Of the 92 subjects, 54 were women. The subjects exhibited a range of refractive error, with the group mean best sphere refractive error being 1.1 ± 1.8 D (range, +0.63 D to 8.13 D) and group mean astigmatic refractive error being 0.32 ± 0.6 D (range, 0 D to 2.75 D). So that we could rule out any significant anterior eye or tear film abnormalities, each subject underwent a preliminary slit lamp examination.
All corneal topography measurements were taken with a videokeratoscope (model E300 videokeratoscope; Medmont Pty. Ltd., Victoria, Australia). The E300 is based on the Placido disc principle and has been shown to have a high degree of accuracy and precision for measuring inanimate test objects.21 Tang et al.21 found the E300 to exhibit a mean height error of 2 µm for measuring spherical and aspheric test surfaces. This instrument has also exhibited highly repeatable results for measurements on corneas in vivo.22 Cho et al.22 found that two repeated measurements were required with the instrument to ensure a precision in corneal elevation data of 2 µm. The videokeratoscope has a sophisticated range-finding device that determines the distance from the corneal apex to the instruments camera and automatically captures the videokeratoscopic image only when good focus and alignment of the eye are attained. To reduce the effects of any diurnal variation in corneal topography, all measurements were taken in the morning.25 As prior visual tasks may also affect corneal topography, subjects were also asked to refrain from performing significant close work immediately before testing.25 26
The procedure used for measuring the total corneal topography has been described in detail.24 The corneal topography of one eye is measured while the subject views an external-fixation target positioned 1.5 m away from the videokeratoscope with the fellow eye (through the use of a mirror). The external-fixation target has a central target and six peripheral targets in different peripheral angles of gaze (at 0°, 60°, 120°, 180°, 240°, and 300° and approximately 30 cm from the central target). Three videokeratoscope images are captured with the subject fixating on the instruments internal central fixation target. The subject is then instructed to fixate on the external target with the fellow eye, and videokeratoscope images are captured with the subject looking at each of the six peripheral fixation targets in turn. A total of four videokeratoscope images are captured for each of the six peripheral directions of gaze (i.e., 3 central images and 24 peripheral images are captured).
The corneal height data for each of the videokeratoscope images is exported from the instrument and is analyzed with custom-written software24 that examines the correlation between central and peripheral corneal topography data. The purpose of this process is to find the point in the peripheral map corresponding to the center of the central topography map (the vertex normal). Essentially, it locates the point on the peripheral map (defined by the radial distance, azimuthal angle, and degree of cyclorotation from the center of the peripheral map) where the sum of squares of differences in the overlapping portions of the central and peripheral corneal topography maps is minimized.
The central topography map is correlated with each of the four peripheral maps in one direction of peripheral gaze, and the peripheral map that gives the best correlation (i.e., the smallest sum of squares difference) with the central map is chosen as the best peripheral map for that direction of gaze. This procedure is repeated for each of the six directions of peripheral gaze. The one central map and six peripheral maps are then chosen that give the overall smallest sum of squares of differences between central and peripheral corneal topography data.
After this map correlation process, the central and peripheral maps are combined. The point of best correlation in each peripheral map is rotated to make this point the common vertex normal, and the peripheral maps are then combined with the central map. In the combined map, the central 6 mm of the central map is preserved, and the peripheral data are added past this point. The combined map is output in the same format as the standard Medmont height map except that the combined map has 46 rings of data in 300 semimeridians (the standard Medmont height map has 32 rings of data). Combined axial radius of curvature maps were also calculated based on the combined corneal height maps. Figure 1 shows an example of central and peripheral corneal axial radius of curvature maps and a combined data map for one subject.
|
If central and peripheral maps correlate well, then the combined map exhibits a smooth transition between central data (the central 6-mm diameter) and peripheral data (outside of the central 6-mm diameter). A feature of combined maps with poor correlation is a large change at the junction between the central and peripheral corneal data.
To determine what would be an unacceptable difference in axial curvature at the junction between the central and peripheral data, every semimeridian of axial curvature data from every subjects standard central map was analyzed and a second-order polynomial function fit to a small region of data just inside 3 mm from the corneal center for each map. The difference between the actual axial radius and the axial radius predicted by the polynomial fit was then calculated for the point just outside 3 mm from center for each semimeridian. The average and SD of the difference between actual and predicted axial radius was found to be 0.04 ± 0.07 mm with an outer 95% confidence interval of 0.18 mm.
The combined central and peripheral axial radius of curvature maps were analyzed in a similar fashion. A polynomial function was fit to the final five central map data points just inside 3 mm from corneal center, and the predicted data were then compared to the first two peripheral map data points at 3.13 and 3.26 mm from corneal center. Any maps showing a difference between actual and predicted data of greater than ±0.2 mm in more than 40 semimeridians at the junction between central and peripheral data was considered to have a poor correlation (one poorly correlated peripheral map would be expected to result in an abrupt boundary change in 50 semimeridians). Of the 100 subjects who had corneal topography measurements taken, 8 were excluded due to poor correlation between central and peripheral topography data.
It is possible that changes in extraocular muscle tension (as occurs in our protocol for capturing the peripheral topography data) may cause changes in corneal topography.27 28 29 A control experiment was performed to investigate whether the eye movements (and subsequent changes in extraocular muscle tension) used in our protocol to capture the peripheral corneal topography maps had an effect on the peripheral corneal topography data. Topography maps of the temporal corneal periphery were taken for two subjects, first by moving the videokeratoscope camera (i.e., no eye movements used) and second by changing fixation (i.e., eye movements used). The peripheral corneal topography data from the two conditions was then compared. No significant difference was found between the peripheral corneal data with or without eye movement. We concluded from this, that the alteration of extraocular muscle tension associated with the change in fixation in our peripheral map capturing protocol (approximately 11°) was not enough to cause significant changes in the peripheral corneal topography data.
The center of standard corneal topography maps is located at a point where the optical axis of the videokeratoscope is perpendicular to the cornea, known as the vertex normal.30 The position of the vertex normal relative to the geometric center of the cornea is known to differ from person to person.31 Therefore to have each subjects corneal topography map centered to a common reference point, we rotated each combined map to the corneal geometric center. The best central videokeratoscope image (as determined through the map correlation process) for each subject was analyzed to determine the position of the corneal geometric center using customized computer software that locates the corneal limbus in the videokeratoscope image.32
Analysis
We calculated the average size of the topography maps in the vertical and horizontal dimensions for both standard (central fixation) and combined corneal topography, by finding the shortest complete semimeridian of data within five meridians (6°) of the horizontal (temporal and nasal) and vertical (superior and inferior) directions. The shortest semimeridians in the temporal and nasal direction were combined, and the shortest semimeridians in the superior and inferior direction were also combined. This provided an estimate of the minimum horizontal and vertical map diameters for both the standard topography and the combined topography maps. Individual subject data were combined to calculate the group mean map size for each technique.
Each subjects corneal height data were averaged across all semimeridians, and the average Ro and Q were calculated based on Bakers equation for conic sections: y2 = 2Rox px2, where y is the distance from corneal center and x is the corneal height.33 A linear in parameters least squares fitting was performed to calculate the Ro and p for the averaged semimeridian data. The asphericity parameter Q is related to p by the equation Q = p 1. Ro and Q were calculated for 6-, 8-, 9-, and 10-mm corneal diameters. For this average across all semimeridians, only subjects with at least 200 semimeridians of complete data were included in the analysis. Any missing data (even in the combined maps) tend to be in the superior semimeridians due to interruptions from the brow and eyelashes; therefore, the larger-diameter analyses will be slightly biased toward the horizontal regions. We would expect this bias to have only a slight effect and only on the 9- and 10-mm diameter data. The root mean square (RMS) fit error was also calculated for each corneal diameter. A repeated-measures analysis of variance (ANOVA) with one within-subject factor (corneal diameter) was used to investigate whether Ro and Q changed significantly for the different corneal diameters tested.
The conic fitting was also performed for the steepest and flattest meridians of corneal data for each subject. The steepest and flattest meridians were calculated based on the best fit spherocylinder to the corneal axial power for an 8-mm corneal diameter. The meridian of data corresponding with the corneal cylinder axis and the two adjacent meridians on either side were averaged to provide the corneal data along the flattest corneal meridian. The same procedure was performed for the steepest meridian data. The best fitting Ro and Q were then calculated for each meridian for each subject. To investigate meridional variations in Ro and Q, repeated-measures ANOVA was performed with two within-subject factors (corneal diameter and corneal meridian).
Franklin et al.24 showed that for larger corneal diameters, a polynomial function fit corneal height data better than a conic fit. They found that a fourth-order polynomial was needed to fit 7-mm diameter data and that for a 10.7-mm diameter a ninth-order polynomial fit was necessary. The average corneal height data were also therefore fit with a polynomial function of the form x = Ay + By2 + Cy3 + Dy4... and so on (where x is the corneal height and y is the distance from corneal center). For each subject, the third- through to the ninth-order polynomial functions were all fit to the corneal height data for 6-, 8-, 9-, and 10-mm diameters. The RMS fit error was also calculated for each of the polynomial orders and for each corneal diameter.
For each subject, axial power maps were calculated based on the combined corneal height data. The axial power data for each subject were analyzed to calculate the best fit corneal spherocylinder using the method of Maloney et al.34 We found these fitting routines to be highly sensitive to any missing data points, therefore only subjects with 300 complete semimeridians of axial power data to the edge of the outer diameter were included in each of the analyses. The best fit spherocylinder data for each subject was converted into the power vectors M (best sphere), J0 (astigmatism 90°/180°) and J45 (astigmatism 45°/135°),35 to allow the group mean and SD to be calculated. The best fit corneal spherocylinder was calculated for corneal diameters of 6, 7, 8, and 9 mm. To investigate for significant changes in the corneal spherocylinder with increasing diameter, a repeated-measures ANOVA was used with one within-subject factor (corneal diameter).
Changes in the corneal spherocylinder with increasing corneal diameter give some indication of variations in the peripheral cornea. However, the large-diameter corneal spherocylinder fits contain data from both the peripheral and the central cornea. Therefore some changes in the peripheral cornea may be somewhat masked by data from the central cornea. To overcome this, and to analyze further the peripheral cornea, we broke the cornea down into concentric annuli of data. This analysis has the advantage that the data in each annulus are mutually exclusive (i.e., the inner annulus provides information about the central cornea only and the outer annulus provides information about the peripheral cornea only). The best fit spherocylinder was then calculated for each annulus of corneal axial power data and again broken down into the power vectors M, J0, and J45. For this analysis, only subjects with complete semimeridians of axial power data to 8-mm diameter were included.
We performed this analysis for annuli of 2-mm width and for those of 0.5-mm width. The 2-mm width analysis provides a measure of the central cornea from the inner annulus (from 0- to 4-mm diameter) and a measure of the peripheral cornea from the outer annulus (from 4- to 8-mm diameter). From this analysis we were able to classify broadly each subjects central cornea as either type 1 (spherical central cornea with astigmatism < 0.75 D) or type 2 (astigmatic central cornea with astigmatism >0.75 D). The peripheral cornea was classified as type a (peripheral astigmatism stable, changing <0.25 D from the central cornea), type b (increasing peripheral astigmatism of
0.25 D), or type c (decreasing peripheral astigmatism of
0.25 D). Previous investigators have classified corneas according to central and peripheral astigmatism in a similar fashion.3 4
The 0.5-mm analysis provides a higher resolution investigation of the changes in the corneal spherocylinder with increasing distance from corneal center. Figure 2 illustrates the corneal spherocylinder annulus analysis for one subject. To investigate changes in the annulus corneal spherocylinder with distance from corneal center, a repeated-measures ANOVA was used with one within-subject factor (corneal annulus diameter) and one between-subjects factor (central corneal type).
|
| Results |
|---|
|
|
|---|
The average conic fit parameters (Ro and Q), polynomial function, and RMS fit errors for the average corneal height data across all meridians are presented in Table 1 . The average Ro for a 6-mm corneal diameter was found to be 7.77 ± 0.2 mm and the average Q value was 0.19 ± 0.1. For a 10-mm corneal diameter, the average Ro was 7.72 ± 0.2 and Q was 0.36 ± 0.2. This indicates an increase in the rate of corneal flattening for the peripheral cornea. Repeated-measures ANOVA revealed that both Ro and Q changed significantly with increasing corneal diameter (P < 0.0001 for both Ro and Q). The RMS fit error was found to increase dramatically for larger corneal diameters. For the conic fitting of the corneal data, the RMS fit error increased from a mean of 0.79 ± 0.4 µm for the 6-mm diameter to 21.18 ± 11.1 µm for the 10-mm corneal diameter. These fit errors highlight the inadequacy of the conic section to describe the peripheral cornea. To reduce the RMS fit error for larger corneal diameters, polynomial fitting to the data was required. For increasing corneal diameters, progressively higher-order fits were needed to fit the data reasonably (i.e., to reduce the RMS fit error). For a 6-mm corneal diameter, a fourth-order polynomial had an average RMS fit error of 0.03 ± 0.01 µm. For the 10-mm corneal diameter, a ninth-order polynomial fit gave an average RMS fit error of 0.28 ± 0.4 µm.
|
|
|
|
|
The group mean best-fit corneal spherocylinder data for the 0.5-mm annulus analysis are presented in Table 4 and emphasize the changes occurring in the peripheral cornea. The average best sphere M, J0, and J45 all decreased in magnitude with increasing annulus diameter. This manifests itself as a general flattening, a slight change in corneal cylinder axis and a slight reduction in corneal astigmatic power in the more peripheral cornea. Repeated-measures ANOVA revealed the changes occurring in M, J0, and J45 with increasing annulus diameter to be highly significant (P < 0.0001 for M and J45 and P < 0.001 for J0). Based on the classification according to central corneal type there were 42 subjects with a spherical central cornea (<0.75 D astigmatism) and 36 subjects with an astigmatic central cornea (>0.75 D astigmatism). A significant interaction was found to occur between corneal annulus diameter and central corneal type for the change in J0 (P = 0.0002). This indicates that the reduction occurring in J0 in the peripheral cornea was greater in subjects with greater central corneal astigmatism. Figure 5 shows the group mean corneal cylinder power and axis as a function of distance from corneal center for the 0.5-mm annulus analysis (data for both the central spherical and central astigmatic corneas are shown). It can be seen in this figure that the change in astigmatic power in the peripheral cornea was much greater in the subjects with astigmatic central corneas (astigmatic eyes showed an average reduction in cylinder from center to periphery of 0.43 D, whereas subjects with spherical central corneas showed only a 0.09-D reduction). The average change in cylinder axis in the peripheral cornea is similar between the two groups with both the astigmatic and spherical central cornea groups showing a slight anticlockwise shift in cylinder axis (right eye) with increasing distance from corneal center.
|
|
|
Zernike polynomials up to and including the sixth radial order were fit to the corneal height data for 6-, 8-, and 9-mm diameters. Repeated measures ANOVA revealed several of the higher order Zernike coefficients to exhibit significant change with increasing corneal diameter. The Zernike polynomial coefficients Z31 (P = 0.001), Z31 (P = 0.002), Z42 (P = 0.009), Z40 (P < 0.0001), Z44 (P = 0.006), and Z60 (P < 0.0001) all showed highly significant change (P < 0.01) with increasing corneal diameter. Figure 7 shows the group mean third- and fourth-order corneal surface Zernike polynomial coefficient values (and Zernike term Z60, as this was the only fifth- or sixth-order term exhibiting highly significant change) for the 6-, 8-, and 9-mm corneal diameters. It is evident in Figure 7 that the fourth-order term Z40 was the higher-order coefficient of the largest magnitude and exhibited the largest change with increasing corneal diameter.
|
| Discussion |
|---|
|
|
|---|
|
We found a small but significant meridional variation in the asphericity parameter Q, indicating that the steeper principal corneal meridian flattens at a slightly faster rate than the flattest meridian. Kiely et al.11 also investigated meridional variation in the asphericity of the cornea and could find no specific trend for Q to differ from one meridian to the other. Other investigators have noted some small meridional variations to exist in corneal asphericity.13 15 The difference that we found between the two principal corneal meridians was small in magnitude, but was highly statistically significant.
We have shown that the conic fit parameters Ro and Q are highly dependent on the diameter of the cornea that is measured. With increasing corneal diameter Ro reduces and Q becomes more negative (indicating an increased rate of flattening in the peripheral cornea). With increasing corneal diameter, the RMS fit error for the conic section also increases markedly. This indicates that although it is convenient and simple to understand, the conic section is a poor estimator of the peripheral cornea. To accurately specify the contour of the peripheral cornea, more complex fitting is required. We found that the use of a ninth-order polynomial function provided an excellent fit to the average corneal contour over a 10-mm diameter, producing an RMS fit error 75 times smaller than that given by the simple conic fitting.
The changes occurring in the contour of the cornea in the periphery will have little influence on foveal vision. Peripheral corneal changes, however, play a role in off-axis aberrations and vision. It is therefore possible that the changes occurring in the peripheral cornea are a part of the balancing of the peripheral refraction or aberrations. It is more likely though that the marked flattening of the cornea in the periphery occurs to produce a smooth transition from the cornea to the flatter scleral surface. The slight meridional variation in corneal asphericity (i.e., the steeper meridians flattening more rapidly leading to a slight reduction in peripheral corneal toricity) may occur to minimize curvature change at the corneoscleral limbus, if we assume that the sclera is not toric. The corneal collagen orientation has been found to change in the limbal zone, becoming circumferentially oriented in the peripheral cornea.40 41 This marked change in corneal stromal collagen orientation is the possible anatomic reason for the flattening and reduction in astigmatism found in the peripheral cornea.
We investigated the average corneal spherocylinder based on the corneal axial power data and found that most of our subjects exhibited some degree of with-the-rule (WTR) astigmatism (i.e., the steepest corneal meridian was oriented near vertical). This is consistent with many previous studies of corneal astigmatism in groups of young, healthy, adult subjects.11 38 42 43 44 The best-fit corneal spherocylinder was also found to change significantly with increasing distance from corneal center. On average, the best fit sphere becomes flatter and the amount of corneal astigmatism reduces slightly in the peripheral cornea. This reduction in toricity of the peripheral cornea is consistent with the meridional variation found in the asphericity parameter Q. That is, as the steeper central meridian flattens at a faster rate in the periphery, then the degree of corneal astigmatism decreases in the corneal periphery.
Whereas, on average, the corneal astigmatism was found to decrease slightly in the periphery, several individual patterns of central and peripheral astigmatism were noted. The majority of subjects (59%) exhibited stable peripheral astigmatism or decreasing peripheral astigmatism (30%), with only 10% exhibiting an increase in corneal astigmatism in the corneal periphery. Guillon et al.12 noted that most of their subjects displayed similar central and peripheral levels of astigmatism, but also noted some individual variations, with some subjects exhibiting a reduction or an increase in peripheral astigmatism. Reddy et al.3 also classified their subjects corneal topography based on central and peripheral corneal astigmatism. In contrast to our study, they found the most common form of astigmatism to be an increased or irregular astigmatism in the periphery, with stable or reducing corneal astigmatism found to be less common. This difference may be due to the subject selection procedures. In our study, we sought to examine the average corneal topography of normal healthy subjects, whereas the study by Reddy et al. involved subjects who were fitted with soft toric contact lenses. Thus, their population would be expected to have a much larger proportion of subjects with high degrees of corneal astigmatism (in fact, only 6% of their subjects had spherical central corneas). Their method for calculating peripheral astigmatism was also different to ours and was based on four corneal topography data points 3.5 mm from the center of the cornea.
The use of Zernike polynomials provides a mathematical description of the entire corneal surface (which does not require averaging across meridians of data as occurs for the conic and polynomial fitting to the average corneal contour). As we were interested in the shape of the cornea, we fit Zernike polynomials to the corneal surface shape, as opposed to deriving the corneal wavefront error (or corneal aberrations) from the corneal shape (which will give information on the corneas optical effects). Particularly for the large corneal diameters that we have analyzed, it is not practical or realistic to consider the optical effects of the peripheral cornea. Figure 7 shows that the Zernike coefficients that made the greatest contribution to the higher order terms for all corneal diameters were the third-order Zernike terms (Z33, Z31, Z31) and the fourth order term Z40. The two terms displaying the most significant change with change in corneal diameter were the terms Z40 and Z60 (these corneal surface Zernike terms are analogous to the spherical aberration corneal wavefront error terms). The highly significant changes in these terms are due to the significant flattening occurring in the corneal periphery. Studies into corneal aberrations in normal healthy corneas have also found that the third-order terms and fourth-order spherical aberration term Z40 to be the dominant higher-order aberration terms.45 46 47 Increasing the corneal diameter over which the corneal aberrations are measured has also been found to cause a general increase in the higher order corneal aberrations, particularly for the spherical aberration and coma terms.48 49 50
The peripheral cornea does not contribute to foveal vision, but its shape is of great importance from an anatomic and mechanical point of view. It is of particular significance in the design and fitting of contact lenses. To characterize the shape of the normal central and peripheral cornea, we specifically excluded those subjects who had corneal disease or had undergone corneal surgery. Future research using these techniques to measure the peripheral cornea in subjects undergoing corneal refractive surgery or orthokeratology may improve our understanding of the changes to the peripheral cornea accompanying these procedures. Investigating the peripheral cornea in subjects with keratoconus or other corneal ectatic disorders may also provide further insight into the etiology of these corneal diseases.
In summary, we found the method of combining central and peripheral corneal topography data to provide a much larger area of corneal coverage than traditional videokeratoscopy. This method has allowed us to present a detailed analysis of the topography of the normal cornea to a large diameter. On average, the cornea flattens significantly and becomes slightly less astigmatic in the periphery. However, several different "patterns" of central and peripheral astigmatism were found in individual subjects. A conic section was found to be a poor estimator of the corneal shape for the peripheral cornea. Higher-order polynomial fits are necessary to describe the corneal contour adequately in the periphery.
| Acknowledgements |
|---|
| Footnotes |
|---|
Disclosure: S.A. Read, None; M.J. Collins, None; L.G. Carney, None; R.J. Franklin, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Scott A. Read, Contact Lens and Visual Optics Laboratory, School of Optometry, Queensland University of Technology, Room B547, O Block, Victoria Park Road, Kelvin Grove 4059, Brisbane, Queensland, Australia; sa.read{at}qut.edu.au.