From the Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, Peoples Republic of China.
| Abstract |
|---|
|
|
|---|
METHODS. One hundred thirty-three eyes from 133 healthy subjects with mean spherical equivalent 6.0 ± 4.2 D (range, 13.13 to +3.25 D) were analyzed. Optic disc measurements including disc area, rim area, cup area, cup-to-disc area, and vertical and horizontal ratios were obtained with an optical coherence tomograph (StratusOCT; Carl Zeiss Meditec Inc., Dublin, CA) and a confocal scanning laser ophthalmoscope (Heidelberg Retina Tomograph, HRT 3; Heidelberg Engineering, GmbH, Dossenheim, Germany). The modified axial length method derived from prior published work was used to correct the OCT measurements for ocular magnification. Bland-Altman plots were used to evaluate the agreement for each optic disc parameter. Associations between optic disc area and axial length/spherical equivalent were evaluated by linear regression analysis.
RESULTS. Disc area increased with the axial length/negative spherical equivalent in the HRT and the corrected OCT measurements although opposite directions of associations were found when the OCT measurements were not corrected for magnification. The difference of the corrected OCT and HRT disc area (corrected OCT disc area minus HRT disc area) was correlated with the axial length (r = 0.195, P = 0.025). When the ametropia was limited to 8.0 to +4.0 D, the correlations became insignificant in the HRT. Using the corrected OCT measurements, disc area, rim area, and cup area, cup-to-disc area, and cup-to-disc horizontal and vertical ratios were significantly larger than those measured by the HRT, with a span of 95% limits of agreement at 1.99, 1.33, and 1.86 mm2 for the areas, 0.34, 0.53, and 0.58 for the ratios, respectively.
CONCLUSIONS. While optic disc area generally increased with the axial length and myopic refraction, the HRT measurements demonstrated that optic disc size was largely independent of axial length and refractive error between 8 and +4 D. OCT may overestimate optic disc size in myopic eyes and results in poor agreement between the two instruments.
| Methods |
|---|
|
|
|---|
Optical Coherence Tomography Imaging
Optical coherence tomography was performed with OCT version 3 (StratusOCT, Carl Zeiss Meditec Inc). Fast optic nerve head imaging was performed with six radially linear scans, each with 128 scan points, centered at the optic nerve head. The disc margin is automatically identified in the StratusOCT by detecting the top and inner edges of retinal pigment epithelium on each side of the optic disc. A line is then joined and measured as the disc diameter. The reference plane (the cup offset) was then determined by tracing a line parallel to the disc diameter with an anterior offset of 150 µm. Region above the reference plane is defined as the rim and below as the cup. Data analyzed in each scan are used to compute the composite image measurements. Global measurements including disc area, cup area, rim area (disc area minus cup area), cup-to-disc area ratio, cup-to-disc horizontal ratio, and cup-to-disc vertical ratio were analyzed in this study. All the scans had a signal strength of at least 7.
Although one can input the eyes axial length and refractive correction in the StratusOCT, it has no impact on the analyzed optic disc measurements. The default axial length and refraction in every OCT scan are set to 24.46 mm and 0 D, respectively, for the optic disc measurement. Therefore the actual disc size in an eye with axial length longer than 24.46 mm and/or refraction <0 D would be larger than the printout values and vice versa. The relationship between the measurement of the OCT image and the size of the actual fundus dimension can be expressed as t = p · q · s,15 where t is the actual fundus dimension, s is the measurement on OCT, p is the magnification factor related to the camera of the imaging system, and q is the magnification factor related to the eye. The correction factor q can be determined with the formula q = 0.01306 (x 1.82), where x is the axial length16 ; p is a constant in a telecentric system. In OCT, the magnification factor in an eye with axial length 24.46 mm is 1 (i.e., t = s). Substituting t = s into the formula t = p · q · s, we arrive at p = 1/q and therefore, p can be calculated as 1/[0.01306 (24.46 1.82)] = 3.382. Because t = p · q · s refers to linear magnification, the equation should be modified to t2 = p2 · q2 · s2 for area magnification. Therefore, values obtained from the StratusOCT optic nerve head analysis printout should be corrected by the factor 3.3822 · 0.013062 · (x 1.82),2 where x represents the axial length.
Confocal Scanning Laser Ophthalmoscopy Imaging
Confocal scanning laser ophthalmoscopy was performed with the HRT 3 (Heidelberg Retina Tomograph III [HRT 3]; Heidelberg Engineering, GmbH, Dossenheim, Germany). A three-dimensional topographic image consisting of 384 x 384 x 16 up to 384 x 384 x 64 pixels is constructed from multiple focal planes axially along the optic nerve head. An average of three consecutive scans is obtained and aligned to compose a single mean topography for analysis. Images obtained in this study were reviewed carefully for imaging score and overall quality score. These image-quality checks are generated by the HRT 3 software (Heidelberg Eye Explorer, ver. 1.5.1.0; Heidelberg Engineering, GmbH). Images were selected for analyses only when each of these checks was rated as good or better and with an average pixel height standard deviation of 30 µm or less. Once the contour line is drawn, the software automatically calculates all the optic disc measurements. The reference plane is defined at 50 µm posterior to the mean retinal height between 350° and 356° along the contour line. Area above the reference is defined as the rim and below as the cup.
Visual Field Testing
Standard visual field testing was performed with the static automated white-on-white threshold perimetry (Humphrey Field Analyzer II, Carl Zeiss Meditec, Inc.). A visual field was defined as reliable when fixation losses and false-positive and -negative rates were <25%. A visual field defect was defined as having three or more significant (P < 0.05) nonedge contiguous points with at least one at P < 0.01 on the same side of the horizontal meridian in the pattern deviation plot and classified as outside normal limits in the glaucoma hemifield test. Visual field perimetry was performed to ensure that all the eyes included were normal and demonstrated no glaucomatous damage.
Statistical Analysis
Statistical analyses were performed with commercial software (SPSS ver. 11.0; SPSS Inc, Chicago, IL). The associations between the optic disc measurements and the axial length/refractive error were calculated with linear regression analysis. The HRT and OCT measurements were compared with paired t-test. Bland-Altman17 plots were used to assess the agreements between the HRT and OCT optic disc parameters. P < 0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
Comparison of Optic Disc Parameters between the HRT and OCT
Table 1 compares the disc, cup, and rim areas, and the cup-to-disc area, cup-to-disc vertical, and cup-to-disc horizontal ratios measured by the HRT and OCT (magnification corrected). Fifteen subjects were excluded in this analysis either because the standard 150-µm reference plane in the OCT images was too low to get measurements in shallow optic cups (n = 7) or because the presence of posterior vitreous detachment rendered inaccurate the demarcation of the rim area (n = 8). Disc, cup, and rim areas and cup-to-disc ratios (area, vertical, and horizontal) measured by OCT were significantly larger than that measured by the HRT (all with P < 0.001). The mean difference and the 95% limits of agreement of each of these parameters are shown Table 1 . No association was found between age and any of the optic disc measurements.
|
| Discussion |
|---|
|
|
|---|
A few studies have been published comparing the agreement of optic disc size measured with different imaging technologies.22 23 24 Schuman et al.22 studied 77 eyes of 44 subjects and reported a coefficient of correlation (r) of 0.64 between HRT and StratusOCT. A recent study by Hoffmann et al.23 showed the span of 95% limits of agreement for the HRT 2 and StratusOCT was 1.15 mm2 and r = 0.73 (n = 42 eyes). In the study by Iliev et al.,24 the span of 95% limits of agreement was 1.99 mm2 and the r was 0.5 (n = 49 eyes). It should be noted that in these studies, the OCT measurements were not corrected for magnification. Different limits of agreement and coefficients of correlations could result if correction for magnification had been taken into consideration. In the present study, the mean difference between the HRT and the magnification-corrected OCT disc area was 0.48 mm2 and the span of 95% limits of agreement was 2.04 mm2 (1.5 to 0.54 mm2). Consistent with previous studies, the OCT measured a larger disc area compared with the HRT. Rim area, cup-to-disc area ratio, and vertical and horizontal cup-to-disc ratios were also significantly larger than those measured by the HRT, and this difference is probably related to the larger disc area measured by OCT. These suggest that the determination of disc area has significant influence in the evaluation of all the optic discrelated parameters.
Controversies exist regarding the relationship between optic disc size and refractive error. Jonas6 and Wang et al.7 found that optic disc size was statistically independent, with refractive error within the range of 8 to +4 D. In contrast, the Rotterdam study reported that the disc area increased by 0.033 mm2 linearly for each diopter increase toward myopia, and these increases were not significantly different in eyes with refractive error between 4 and +4 D.8 Of note, in this study, the HRT 3 and the StratusOCT also showed different results. Although correlations between the disc area and axial length-refractive error were found in both instruments with ametropia between 14 and +4 D, correlations were not evident in the HRT, but were in OCT, when the range of ametropia was limited to 8 to +4 D. We believe this discrepancy could be related to the different definitions of disc margin. In the HRT, the disc margin is defined by the outer boundary of Elschnigs ring whereas in OCT, it is the ends of the retinal pigment epithelium (RPE) that are automatically detected. In myopic eyes, peripapillary atrophy is more common.8 It is conceivable that optic discs with peripapillary atrophy, which is an area characterized histologically by Bruchs membrane devoid of RPE and photoreceptors,25 could have been measured larger in size in OCT. It is because the borders of the RPE, which are farther away from the outer edge of Elschnigs ring in eyes with peripapillary atrophy, would be detected as the disc margin. Although the location of the disc margin can be adjusted manually in OCT, determining the disc margin in the cross-sectional optic disc image could be highly subjective. So far, no standard approach has been suggested or validated in this regard. It has been shown that the prevalence of peripapillary atrophy increase with 1.3% for each diopter increase in myopia.8 It is plausible that OCT overestimates optic disc area in eyes with longer axial length and leads to an apparent correlation within the range between 8.0 and +4.0 D. It is supported by the observation that the difference in disc area measured by OCT and HRT was correlated to the axial length (r = 0.195, P = 0.025; the difference in disc area measured by OCT and HRT may well correspond to the area of peripapillary atrophy). The findings in the HRT agree with the notion that optic disc size is largely independent of axial length/refractive error within the range between 8 and +4 D. Cautions should be exercised in the interpretation of OCT disc measurements in eyes with peripapillary atrophy.
The HRT and OCT have theoretical advantages over conventional planimetry. However, none has yet to offer perfect disc measurements. The HRT produces best image quality when the range of ametropia is from 12 to +12 D, and this range limited our selection of myopic subjects. (The maximum negative spherical error in our cohort was 12.5 D [spherical equivalent = 13.1 D].) OCT stands out for its automated algorithm for objective detection of disc margin, and yet, this algorithm crumbles in eyes with peripapillary atrophy in which the borders of the RPE may not correspond to the actual disc margin. In addition, much of the disc information could be missed by the extrapolation of disc measurement from the six linear cross-sectional scans. The limitation in disc measurement is also related to the level of reference plane. The default reference plane in the StratusOCT is determined by tracing a line with an anterior offset of 150 µm parallel to the disc diameter (the line joining the ends of the RPE). Although we found that this reference plane was suitable in detecting glaucomatous change,26 it could be too high in eyes with shallow cups. Although the level of reference plane would not affect the measurement of the disc area, its position has direct impact on the calculation of rim and cup areas. Seven cases were excluded in the analysis of rim area and cup-to-disc ratio for this reason. Finally, we would like to highlight the importance that optic disc measurements should not be directly adopted from the StratusOCT analysis printout. As shown in this study, different conclusions could be drawn if correction for ocular magnification was not taken into consideration.
In summary, optic disc area generally increased with axial length/negative spherical equivalent, although the HRT measurements demonstrated that these associations became insignificant in the range of 8 to +4 D. OCT may overestimate optic disc size in myopic eyes. Most of the disagreements between the HRT and OCT optic disc measurements could be related to the different disc margin definitions of the two instruments. The relationships between disc area and axial length/refractive error should always be interpreted with reference to the range of ametropia and the imaging technology used in the optic disc measurements.
| Footnotes |
|---|
Disclosure: C.K.-s. Leung, Heidelberg Engineering (R); A.C.K. Cheng, None; K.K.L. Chong, None; K.S. Leung, None; S. Mohamed, None; C.S.L. Lau, None; C.Y.L. Cheung, None; G.C.-h. Chu, None; R.Y.K. Lai, None; C.C.P. Pang, None; D.S.C. Lam, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Christopher Kai-shun Leung, Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, 147K Argyle Street, Kowloon, Hong Kong; tlims00{at}hotmail.com.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Tsutsumi, A. Tomidokoro, H. Saito, A. Hashizume, A. Iwase, and M. Araie Confocal Scanning Laser Ophthalmoscopy in High Myopic Eyes in a Population-Based Setting Invest. Ophthalmol. Vis. Sci., November 1, 2009; 50(11): 5281 - 5287. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Lagreze, M. Gaggl, M. Weigel, J. Schulte-Monting, A. Buhler, M. Bach, R. D. Munk, and T. A. Bley Retrobulbar Optic Nerve Diameter Measured by High-Speed Magnetic Resonance Imaging as a Biomarker for Axonal Loss in Glaucomatous Optic Atrophy Invest. Ophthalmol. Vis. Sci., September 1, 2009; 50(9): 4223 - 4228. [Abstract] [Full Text] [PDF] |
||||
![]() |
A M Dubis, J T McAllister, and J Carroll Reconstructing foveal pit morphology from optical coherence tomography imaging Br J Ophthalmol, September 1, 2009; 93(9): 1223 - 1227. [Abstract] [Full Text] [PDF] |
||||
![]() |
C K M Chan, A C O Cheng, C K S Leung, C Y L Cheung, A Y Yung, B Gong, and D S C Lam Quantitative assessment of optic nerve head morphology and retinal nerve fibre layer in non-arteritic anterior ischaemic optic neuropathy with optical coherence tomography and confocal scanning laser ophthalmoloscopy Br J Ophthalmol, June 1, 2009; 93(6): 731 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. G. Strouthidis, H. Yang, B. Fortune, J. C. Downs, and C. F. Burgoyne Detection of Optic Nerve Head Neural Canal Opening within Histomorphometric and Spectral Domain Optical Coherence Tomography Data Sets Invest. Ophthalmol. Vis. Sci., January 1, 2009; 50(1): 214 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Nagai-Kusuhara, M Nakamura, Y Tatsumi, Y Nakanishi, and A Negi Disagreement between Heidelberg Retina Tomograph and optical coherence tomography in assessing optic nerve head configuration of eyes with band atrophy and normal eyes Br J Ophthalmol, October 1, 2008; 92(10): 1382 - 1386. [Abstract] [Full Text] [PDF] |
||||
![]() |
S A Vernon, A P Rotchford, A Negi, S Ryatt, and C Tattersal Peripapillary retinal nerve fibre layer thickness in highly myopic Caucasians as measured by Stratus optical coherence tomography Br J Ophthalmol, August 1, 2008; 92(8): 1076 - 1080. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. He, B. Liu, W. Huang, J. Zhang, Q. Yin, Y. Zheng, D. Wang, and J. Ge Heritability of Optic Disc and Cup Measured by the Heidelberg Retinal Tomography in Chinese: The Guangzhou Twin Eye Study Invest. Ophthalmol. Vis. Sci., April 1, 2008; 49(4): 1350 - 1355. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |