1From the Department of Ophthalmology and the 2Institute of Medical Statistics and Biometry, University of Dresden, Dresden, Germany.
| Abstract |
|---|
|
|
|---|
METHODS. In a prospective clinical trial 75 eyes of 75 patients undergoing phacoemulsification were examined. Before phacoemulsification, the anterior chamber was cannulated at the temporal corneal limbus. In a closed system the IOP was directly set to 15, 20, or 35 mm Hg with a manometric water column. IOP measurements taken by DCT were compared to intracameral measurements with a precision reference pressure sensor.
RESULTS. Measurements from 60 patients were suitable for statistical analysis. At IOP of 15 mm Hg, the mean difference between IOP measured by DCT and intracameral IOP was –0.02 ± 1.32 mm Hg; at 20 mm Hg it was –0.2 ± 1.44 mm Hg and at 35 mm Hg, –0.84 ± 1.90 mm Hg. The concordance coefficient according to Lin was 0.9763, showing good agreement between DCT- and intracamerally measured IOP. There was a statistically significant correlation between the difference in IOP measured by DCT minus intracameral IOP and CCT (P = 0.0291, R2 = 0.00012). All other parameters had no statistically significant effect on the difference between DCT and intracameral IOP (corneal curvature, P = 0.6094, R2 = 0.00367; age, P = 0.9198, R2 = 0.000003; astigmatism, P = 0.1564, R2 = 0.08497; and axial length, P = 0.9484, R2 = 0.00008).
CONCLUSIONS. Measurements with the DCT showed good concordance with intracameral IOP. CCT exerted a statistically significant but clinically irrelevant effect on measurements with the DCT.
Factors that may affect the GAT readings, are either caused by the measurement procedure itself (e.g., thin or thick rings) or by properties of the eye (e.g., corneal thickness, corneal curvature, astigmatism, and axial length).1 Particularly, the effect of corneal thickness on GAT has been discussed extensively.2 3 4 5 A correction factor has been suggested of from 0.19 to 1 mm Hg per 10-µm deviation from the average corneal thickness. However, in most of the studies, correlations in a large cohort of subjects were calculated without knowing the real IOP. To overcome this problem, intracameral measurements seem to be the most appropriate approach to assessing the effect of properties of the eye on IOP measurements by comparing measured IOP with the real (intracameral) IOP, as suggested by Ehlers.6 7 In a recent study, our group showed a correction factor for GAT of 1 mm Hg per 25-µm difference from a central corneal thickness (CCT) of 550 µm in intracameral measurements, showing that corneal thickness has a major impact on GAT measurements.5 However, in a given patient, it is still unclear whether a correction of the measured IOP according to correction tables is the appropriate way to determine the real IOP.
A dependency of IOP readings on CCT seems to be unfavorable, as a thin corneal thickness was considered to be a major risk factor for conversion to glaucoma in the OHTS study8 9 and for progression in patients with glaucoma.10 It is still unclear whether this finding is caused by the corneal-thickness dependency of GAT or is an independent risk factor.
Therefore, it seems desirable to have a tonometer that is, ideally, independent of corneal thickness and other ocular parameters. Recently, the Pascal Dynamic Contour Tonometer (DCT) was introduced.11 The idea behind this instrument is that the tonometer tip has a special shape allowing a direct IOP measurement by minimizing corneal stress. The purpose of the present study was to examine in an intracameral in vivo study whether measurements with the DCT are affected by CCT, corneal curvature, and axial length and to answer the question of whether IOP measurements by DCT represent the real IOP.
| Methods |
|---|
|
|
|---|
Exclusion criteria were a history of previous ocular trauma or surgery, any corneal disease or irregularity, or proliferative diabetic retinopathy.
All patients signed an informed consent before entering the study. The study was performed in accordance with the Declaration of Helsinki and was approved by the institutional ethics committee of the Medical Department of the University of Dresden.
Before entering the study, all patients underwent a complete ophthalmic examination. In addition, keratometry was performed with a keratometer (Zeiss Bombe; Carl Zeiss Meditec, Inc., Dublin, CA), axial length was measured by A-scan ultrasonography (Sonomed 2500; Technomed, Maastricht, The Netherlands), and CCT was measured with an ultrasonic pachymeter (IOPac; Heidelberg Engineering, Heidelberg, Germany).
After surgery, the patients received peribulbar anesthesia (2.5 mL bupivacaine 0.75%, 2.5 mL mepivacaine 2%, and hylase 150 IU in 1 mL), and 0.5% proxymetacaine eye drops were used to anesthetize the cornea. After the peribulbar injection a Vörösmarthy oculopressor was used with a pressure of 30 mm Hg for 10 minutes. This device applies pressure via an inflatable balloon to the outside of the eye through the closed eyelid to soften the eye and improve the effectiveness of the local anesthetic.12
IOP measurements were performed before actual surgery. All measurements and the subsequent cataract surgery were performed by one examiner (AGB). Before the IOP measurements began, the experimental tubing system was adjusted to the height of the eye according to a laser water scale (Fig. 1) .
|
|
|
The instrument provides an internal quality assessment of the measurement in five levels: from 1 representing the best quality to 5 representing the worst. The manufacturer recommends that only measurements equal to or better than quality level 3 be used. Therefore, only measurements of a quality of 3 or better were used for the statistical evaluation, and measurements of quality levels 4 and 5 were excluded.
After completion of the study, each measurement curve was reevaluated by a masked observer. If the IOP was determined in a stable part of the measurement curve and if the measurements were not disturbed by artifacts, the measurements were used for the analysis. All other measurements were excluded.
Before starting with the 60 patients who were included in the study, we performed 10 measurements to avoid having learning effects bias the study. During these training measurements, it was apparent that if the positioning of the cannula caused high tension on the cornea, it led to large differences between measured IOP and reference IOP. For example, in one case the DCT initially showed an IOP of 9 mm Hg lower than reference. Looking at the site of cannulation, it was apparent that the cannula was under tension, as it was bent upward. After reducing the tension by changing the position of the cannula the difference between DCT and reference was reduced to 0.5 mm Hg. Therefore, we used great caution to insert the cannula into the anterior chamber with the least resulting tension for the measurements. As a consequence, if the cannula showed high tension on the cornea the measurements were aborted and not used for analysis.
For statistics, a linear model of analysis of covariance (mixed procedure) was used. In the model, all 60 patients were viewed as randomized effects. As an assumption, identical correlations between the intraindividual repeated measurements were used (compound symmetry).
For comparison of the DCT and reference measurements, the analysis evaluates the agreement of two methods with different measurement error variance. The model includes IOP levels (15, 20, and 35 mm Hg); periods 1, 2, and 3; and methods (DCT, reference) as experimental factors and the dual and triple interactions of the experimental factors. To assess the grade of concordance of both quantitative measurements, the concordance correlation coefficient according to Lin was calculated. In addition, Bland-Altman plots were used. The reproducibility was assessed by calculating intraclass coefficients.
For the evaluation of the effect of different factors on the IOP difference between DCT and reference, the model additionally used CCT, corneal curvature, age, astigmatism, and axial length as quantitative covariables.
The analysis was performed with commercial software (SAS, ver. 9, using PROC MIXED; SAS Institute Inc., Cary, NC).
| Results |
|---|
|
|
|---|
Mean CCT was 562 ± 36 µm (mean ± SD) ranging from 464 to 626 µm. Corneal curvature was 43.7 ± 1.7 D (40.6–50.9 D), astigmatism 1.0 ± 0.9 D (0–5.25 D), and axial length 23.05 ± 1.33 mm (21.27–27.81 mm), respectively. The descriptive statistics of the IOP measurements are summarized in Table 1 .
|
|
|
|
|
The difference between DCT and reference IOP increased with increasing IOP level (P < 0.001). Comparison of the single IOP levels showed that the differences were significantly higher at 20 than at 15 mm Hg (P = 0.0465), at 35 mm Hg than at 15 (P < 0001), and at 35 than at 20 mm Hg (P < 0.001), respectively.
In the multivariate regression analysis, CCT had a statistically significant effect on the difference between DCT and reference IOP (P = 0.0291). An increase in CCT of 92 µm led to an IOP increase of 1 mm Hg. All other parameters had no statistically significant effect on the difference between DCT and reference IOP (corneal curvature, P = 0.6094; age, P = 0.9198; astigmatism, P = 0.1564; and axial length, P = 0.9484). However, in the multivariate model the R2-values were very low for all parameters including CCT with R2 = 0.00012. The other R2-values were 0.00367 for corneal curvature, 0.000003 for age, 0.08497 for astigmatism, and 0.00008 for axial length. The univariate linear regression comparing CCT with the difference between DCT and reference IOP is shown in Figure 7 .
|
| Discussion |
|---|
|
|
|---|
The reason for the reduced concordance at the highest IOP level is not clear. One possible explanation is that the experimental setup was more susceptible to errors at higher IOP levels than at lower levels. This hypothesis is supported by the fact that the difference between DCT and real IOP increased with increasing IOP level and the scattering of measurements increased toward higher IOPs.
A similar trend was seen with regard to the reproducibility. Although the overall reproducibility was good, the reproducibility decreased toward higher IOP levels. In the literature a good reproducibility of the DCT measurements is supported by many authors. The reproducibility of the DCT readings seems to be even better than GAT readings.13 14 15 16 17 18 19 20
The idea in the present study was to compare IOP measured intracamerally with that measured by DCT. Differences between the two can mainly be caused by four sources of error: First, properties of the eyes affecting the measurements; second, the experimental setup; third, the measurement procedure itself; and fourth, the possibility that the instrument does not measure IOP correctly. As the same type of pressure sensor was used to measure both the intracameral and the transcorneal IOP, the fourth possibility seems to be irrelevant in our setup.
Effect of Eye Properties on the Measurements
Corneal curvature, age, astigmatism, and axial length showed no statistically significant effect on the difference between intracameral IOP and DCT measurement. However, CCT had a significant effect on the difference of intracameral IOP and DCT reading. A CCT increase of 92 µm led to an increase in the DCT reading of 1 mm Hg. However, compared with data from a previous study comparing intracameral IOP with GAT, in which a 25-µm change in CCT leads to an error of 1 mm Hg,5 which is almost fourfold, the effect of CCT on DCT seems to be very small. In addition, the correlation of CCT and DCT was very weak with R2 = 0.00012 and P = 0.0291 and therefore seems to be clinically irrelevant. This finding is in agreement with studies comparing DCT and GAT measurements showing that DCT readings seem to be less (or not) affected by CCT than are GAT measurements.16 18 19 20 21 22 23
Critical Evaluation of the Experimental Setup
The tonometer used in the study is not identical with the commercially available Pascal DCT. As the measurements were performed with the subjects supine, a hand-held DCT was used, which is a DCT tip mounted on the body of a Perkins tonometer, allowing measurements independent of body position. As exactly the same measurement tip is used in the hand-held device and the commercially available Pascal DCT, it seems unlikely that IOP readings differ significantly between both instruments. This is supported by a study performed by Roberts et al. (IOVS 2007;48:ARVO E-Abstract 1254), comparing both devices and showing good agreement between both instruments, suggesting that the results of the present study are also applicable to the commercially available Pascal DCT.
In the present study, both the hand-held DCT and intracameral IOP measurements were obtained with identical DCT tips. Therefore, the same pressure gauge in the DCT tip was used for both measurements. If the pressure gauge delivered incorrect pressure measurements, it could lead to a systematic error. However, the calibrations of the pressure sensors of both instruments were verified against a water column before the experiments. In addition, a verification of calibration was performed automatically during each measurement, when the intracameral pressure sensor was connected to the water column (Fig. 3 , phase 2). Therefore, this source of error can be excluded.
As the measurements were performed before cataract surgery, the eyes were anesthetized by peribulbar injection and softened by application of pressure from outside with an oculopression device. Both procedures could affect the aqueous dynamics and therefore be a potential source of error. However, as the IOP was set to certain IOP levels manometrically, and reference as well as DCT measurements were taken simultaneously, it seems unlikely that this had a significant effect on the results.
As pointed out in the Material and Methods section, the initial measurements before the study began showed possible measurement errors, if tension of the cannula was present. Therefore, caution was exercised to minimize the tension of the cannula on the cornea. However, it cannot be fully excluded that at least to some extent such tension influenced the results. However, despite this possible problem, the concordance between measurements was good. If this effect was present during the measurements, one would hypothesize that the agreement between intracameral and measured IOP may have been even better than reported in this article. Therefore, we have initiated a study in which the astigmatism induced by the cannulation is measured to evaluate how this might affect measurements.
It is also possible that the tension of the cornea is more pronounced at higher IOPs, as according to Laplaces law the tension in the cornea increases with increasing IOP. As we described in our empiric data, increased tension resulted in a falsely low measurement of IOP. This false reading may be at least partly an explanation of why the measured IOPs were significantly lower compared with the intracameral IOP at higher IOPs.
In contrast to other studies, we chose to compare the measured values with the values obtained with the reference IOP instead of comparing them with the adjusted IOP level (15, 20, and 35 mm Hg). The reason is that the IOP is not steady, but fluctuates due to the heart action. This fluctuation is measured by the instrument as ocular pulse amplitude, indicating the difference between the diastolic and systolic IOP. Measurements were performed in a closed system. The DCT considers the diastolic IOP only. However, if the system were closed during systole, the diastolic IOP measured by DCT would be lower by approximately the amount of the pulse amplitude. Another advantage of using the intracameral measurement is the possibility of comparing it continuously with the measured IOP (Fig. 3) . Therefore, using the measured IOP reduces possible sources of error.
Of interest, at the IOP level of 35 mm Hg, the difference between adjusted IOP level and IOP measurements was higher than at the other two IOP levels. The reason for this finding is that although the stopcocks were closed during the measurements, the system was not totally closed. Trabecular meshwork and uveoscleral outflow remained open and allowed outflow with a consecutive IOP decrease. Therefore, it seems to be even more important at higher IOP levels not to use the adjusted but the measured intracameral IOP as a reference. Although the model used is not an ideal system, it comes as close as possible to a closed system, and at lower IOPs, the effect seems to be negligible.
Comparison of DCT with GAT
GAT has been the gold standard of tonometry for more than 50 years. Over the years, it has been a helpful tool in the diagnosis and treatment of glaucoma. However, there are still some problems involved in GAT measurements. Because of its measurement principle, it does not measure IOP, but the force that is needed to applanate a circle with a diameter of 3.06 mm. Under certain conditions, the force is directly proportional to the IOP and is valid for most patients; however, for many others, it is not.
That this is really a relevant issue is reflected in reports in which IOP measured with GAT was high due to measurement error, whereas the intracameral IOP was normal.24 25 In addition, patients with normal-pressure glaucoma tend to have a thinner CCT, and those with OHT tend to have a thicker CCT compared with normal subjects, suggesting that at least some of these patients are misclassified because of a measurement error caused by the CCT deviation.26 Because of the dependency of GAT measurements on CCT, the use of nomograms to correct measurements has been suggested. However, adjusting IOP measurements according to a nomogram seems to be problematic, as it is possible that the corrected IOP will be even more different from the true IOP than in the initial measurement.27
Therefore, it would be much more desirable to have a tonometer that measures independent of or is at least less dependent on the properties of the eye (e.g., CCT). As shown in our study, the DCT provides measurements with a high accuracy and reproducibility. The concordance between true IOP and measured IOP seems to be higher compared with the GAT measurements, as CCT and other parameters had no clinically relevant effect, and therefore no correction of the IOP measurements was necessary. According to our data, the DCT measurements came close to the true IOP and seem to be at least a good alternative to GAT, especially if CCT needs to be considered.
In summary, measurements with the Pascal DCT show good accuracy and reproducibility. The measurements are not dependent on astigmatism, corneal curvature, and axial length. There was only a small statistically significant effect of CCT on DCT readings, which seems to be clinically irrelevant. Therefore, the DCT measurements come close to true IOP.
| Footnotes |
|---|
Disclosure: A.G. Boehm, None; A. Weber, None; L.E. Pillunat, None; R. Koch, None; E. Spoerl, None