1From the Kresge Eye Institute, Wayne State University, Detroit, Michigan; 2Division of Ophthalmology, The Childrens Hospital of Philadelphia, Philadelphia, Pennsylvania; 3Department of Ophthalmology, Childrens University Hospital, Dublin, Ireland; 4Department of Ophthalmology, University of Pennsylvania, Philadelphia, Pennsylvania; and 5Ciba Vision Corporation, Duluth, Georgia.
| Abstract |
|---|
|
|
|---|
METHODS. By applying signal processing techniques to existing PCI tracings of human ocular axial length measurements, a signal modeling algorithm was developed and validated to determine the position and variability of a postretinal peak that, by analogy to animal studies, likely corresponds to the choroidal/scleral interface. The algorithm then was applied to diurnal axial eye length datasets.
RESULTS. The postretinal peak was identified in 28% of subjects in the development and validation datasets, with mean subfoveal choroidal thicknesses of 307 and 293 µm, respectively. Twenty-eight of 40 diurnal PCI datasets had at least two time points with identifiable postretinal peaks, yielding a mean choroidal thickness of 426 µm and a mean high-low difference in choroidal thickness of 59.5 ± 24.2 µm (range, 25.9–103 µm). The diurnal choroidal thickness fluctuation was larger than twice the SE of measurement (24.5 µm) in 16 of these 28 datasets. Axial length and choroidal thickness tended to fluctuate in antiphase.
CONCLUSIONS. Signal processing techniques provide choroidal thickness estimates in many, but not all, PCI datasets of axial eye measurements. Based on eyes with identifiable postretinal peaks at more than one time in a day, choroidal thickness varied over the day. Because of the established role of the choroid in retinal function and its possible role in regulating eye growth, further development and refinement of clinical methods to measure its thickness are warranted.
High-frequency ultrasound can identify choroidal thickness in animal eyes5 6 7 with reasonable accuracy. However, because of the large size of the human eye, energy dispersion precludes safe application of high-frequency ultrasound to the human posterior segment. In recent years, posterior segment imaging of the human eye has been revolutionized by optical coherence tomography (OCT). To date, OCT has not proven effective for assessing choroidal structure because of signal attenuation posterior to the retinal pigment epithelium (RPE),8 9 though future modifications of OCT methodology may provide some information about choroidal structure (Unterhuber A, et al. IOVS 2006;47:ARVO E-Abstract 3507).
Partial coherence interferometry (PCI) offers the possibility of providing a detailed quantitative measurement tool for posterior ocular structures with simultaneous axial length measurement. Axial PCI measurements provide data analogous to those obtained with A-scan ultrasound but with much greater precision; they have been clinically applied for biometric calculations for cataract surgery (e.g., IOLMaster; Carl Zeiss Meditec, Oberkochen, Germany). In chicks and humans, the PCI reflectance signal from the posterior eye walls consists of multiple peaks10 11 12 (Fig. 1) . In chick, the likely anatomic sources of these peaks were identified using sequential dissection of the outer layers of an eye, and it was found that the inner scleral surface generates a reliable and identifiable PCI signal.12 Using both the robust peak from the RPE/Bruch membrane interface and the signal from the inner scleral surface, PCI thus enables precise measurements of choroidal thickness and daily choroidal thickness fluctuations in chick.13 14
|
Compared with that of chicks, which have less ocular pigmentation than humans, the P4 PCI signal of the human choroid/scleral region is attenuated and broadened, often with multiple peaks. The presence and morphology of the P4 signal varies widely between human subjects and between measurements within subjects. Because of a low signal-to-noise ratio of the attenuated PCI reflectance signal deep to the RPE, P4 is most evident in signal-averaged data (Fig. 2) .
|
| Methods |
|---|
|
|
|---|
Procedures
Subjects from each study group underwent PCI axial eye length measurements using our previously described instrument and protocol.15 16 During each measurement session, the subjects head position and corneal reflex were monitored and adjusted so that the central corneal reflex and alignment beam were coaxial. Subjects were encouraged to maintain fixation on the alignment beam of the PCI for the 0.8 second required for each measurement, and 80 PCI axial length tracings were acquired over a period of 5 to 7 minutes. If the subject changed fixation from the alignment beam, lateral readjustment of the subjects eye was necessary to obtain an axial length measurement. This, together with continuous monitoring of the corneal reflex, ensured that the subject was fixating on the alignment beam. The first two datasets used in this study contained one measurement session on a single eye of each subject15 (Greenberg KP, et al. IOVS 2003;44: ARVO E-Abstract 3612). The third dataset contained bilateral diurnal axial length measurements on each subject, obtained six times over the course of 1 day at 3-hour intervals between 7 AM and 10 PM and another set of diurnal axial length measurements obtained on another day in the same time frame.16
Algorithm for Determination of P4
Using a variation of a previously reported algorithm,15 we developed an algorithm in MatLab (version 6; The Math Works, Natick, MA) to determine the position and variability of the choroidal/scleral peak (P4) in PCI axial length tracings. Briefly, the algorithm analyzed the raw data consisting of up to 80 PCI tracings from a given subject. Each waveform was examined to determine whether it contained eye movements (characterized by large amplitude signal transients), whether reflections were in a consistent location, and whether there was a high signal-to-noise ratio. Based on these criteria, the algorithm accepted tracings that contained potential peaks and excluded tracings containing only noise or artifact. The accepted tracings with potential peaks (i.e., n tracings) were signal averaged and filtered to generate one posterior eye wall waveform that was curve-fit with three Gaussian curves (Fig. 3) . The resultant curve-fit waveform was then analyzed for the presence of a P4 signal. In summary, a P4 was present if the P4 Gaussian peak was above the level of the noise and was within 550 µm from P3 (i.e., the RPE). If a subjects curve-fit waveform contained a P4 signal, the variability of the P4 determination was estimated from the original dataset using a bootstrap technique.17 Novel bootstrap datasets were generated by random sampling with replacement n times from the original accepted dataset of n tracings.17 The bootstrap datasets were signal averaged and filtered to generate one bootstrap waveform that was curve-fit and analyzed for the presence of a P4 signal. The process was repeated until 200 bootstrap P4s were extracted, from which the variability was estimated. A more detailed description of the algorithm is available in the Appendix, online at http://www.iovs.org/cgi/content/full/50/1/5/DC1.
|
In contrast, for subjects in the diurnal dataset, there were 12 measurement sessions for each eye (six on each day). The additional measurement sessions in essence increased the signal-to-noise ratio of the P4 signal and the number of subjects whose data had an identifiable P4. Figure 4 illustrates how the availability of multiple measurement sessions at different time points for a single eye improved the identification of P4. It provided face validity to the approach used for the diurnal dataset, whereby the composite data from each eye were examined by one of the authors (JSB), and the algorithm was applied to each time point when there was visible evidence of a P4.
|
| Results |
|---|
|
|
|---|
|
Application of the Algorithm to Detect Diurnal Fluctuations in the P4-P3 Distance
To determine whether the P4-P3 distance fluctuated during the day, the algorithm was applied to the datasets from both eyes of 10 subjects who had undergone multiple axial PCI measurements on two different days. These axial PCI data had been accumulated as part of a study to examine the relation of IOP and axial length.16 In the analysis of these 40 diurnal datasets, a P4 waveform was identified in the data of at least two measurement times in 28 datasets (Table 2) . Seven subjects contained a P4 waveform in each eye on each day of testing, and the remaining three contained no P4 waveform in either eye on either testing day.
|
There was no consistent relation between the diurnal fluctuations in choroidal thickness and axial length in these data. Representative waveforms illustrate the inconsistency of comparing the diurnal fluctuations in the two parameters, in which choroidal thickness and axial length were normalized to their respective daily mean values (Fig. 5) . Figure 5A shows a group of two datasets with a statistically significant change in axial length and in which the qualitative change in choroidal thickness was in approximate antiphase with the change in axial length. Figure 5B shows a dataset in which the axial length had no statistically significant diurnal variation and the choroidal thickness also showed little diurnal change. In 5 of the 18 datasets in which there was a significant change in the axial length, the diurnal changes in choroidal thickness were in approximate antiphase with the axial length fluctuations at all but one time point (examples are provided in Fig. 5C ). In a few datasets, the relation between axial length and choroidal thickness suggested that they were unrelated (Figs. 5D 5E) . Figure 5D shows one dataset in which there was a statistically significant change in axial length and the choroidal thickness was essentially unchanged. Figure 5E shows two datasets that had no statistically significant change in axial length and in which the mean change during the day in choroidal thickness was more than twice the SEmeasurement.
|
| Discussion |
|---|
|
|
|---|
Although techniques to estimate choroidal thickness require further development to increase the proportion of subjects with recordable choroidal signals, our methods of analyzing PCI axial data demonstrate a structure deep to the Bruch/RPE interface in 28% to 70% of test subjects. When our algorithm was applied to a single measurement session taken at one time point, only 28% of tested subjects across all ages showed evidence of P4 (Table 1) . When multiple measurement sessions at different times over a day were analyzed, we used a subjective judgment of the waveforms along with the P4 algorithm to aid in identifying P4 (Fig. 4) . Repeat measurements significantly increased the number of identifiable P4 signals by taking advantage of averaging of the inherently low signal-to-noise ratio of the PCI choroidal/scleral signal. Of the 10 subjects in the diurnal study reported here, seven (70%) showed evidence of a P4 signal during at least 2 of 6 measurement sessions for each eye.
Despite the increased percentage of subjects showing evidence of a P4 with more data in the diurnal study, three subjects still did not have evidence of a P4 in either eye. In the diurnal study the presence of a P4 was subject dependent, not eye dependent. Either subjects had a P4 signal in both eyes or the P4 signal was not detected in either eye. The most likely reason for the absence of a P4 signal was decreased signal intensity from absorption because of some anatomic characteristic unique to the subjects, such as thicker choroids, increased choroidal pigmentation, or some property of Bruch membrane. To the extent that race may be a surrogate for choroidal pigmentation, we found no correlation between the presence and absence of a P4 signal based on race in any of the study cohorts.
Choroidal thickness is likely a dynamic parameter influenced by oscillations in IOP or the cardiac cycle. IOP undergoes relatively rapid oscillation associated with the cardiac cycle, and this may translate to a rapid change in choroidal thickness over seconds. However, given that we averaged up to 80 measurements to obtain a single estimate of choroidal thickness, any transient variation on the order of seconds in the choroidal thickness would be expected to average out and probably not influence the mean measurements obtained at each time point.
Other investigators and we19 20 have previously shown that humans undergo diurnal fluctuation in axial length, as previously demonstrated in chicks,5 6 13 21 rabbits,22 and marmosets.7 The physiologic basis for this variation and its implications are not understood. Results in chicks, rabbits, and humans indicate that the diurnal fluctuations in ocular dimensions are not solely passive expansions and contractions of the globe in response to IOP fluctuations.5 6 16 19 22 Axial diurnal fluctuations in humans have also been shown to have no correlation to intraday variations in corneal thickness (Glassman RD, et al. IOVS 2006;47:ARVO E-Abstract 1798).
The choroids of chicks and marmosets, measured with high-frequency ultrasound, undergo diurnal fluctuations in thickness that are in antiphase with axial length fluctuations.6 7 13 In contrast, the results of the present study do not establish a consistent phase relationship between diurnal fluctuations of the choroidal thickness and axial eye length, perhaps because of inherent variability in the patterns of humans in contrast to those of laboratory animals.
Comparison with Other Methods of Choroidal Measurement
A major problem in developing methods to measure human choroidal thickness in vivo is the lack of good comparison data from similar techniques. Overall, the mean P4-P3 distance for every measurement session from all subjects described herein with a P4 was 326 µm (range, 204–490). With the increased number of measurement sessions used in the diurnal study, we were able to estimate choroidal thickness in an increased proportion of subjects (70% compared with 28%). The increased number of measurement sessions in the diurnal study also resulted in a greater mean choroidal thickness (426 µm) compared with the one measurement session (307 and 293 µm), whereas the equipment and the measurement technique were otherwise unaltered, possibly because the increased number of measurements allowed the detection of smaller signals in the noise of the waveform. Given that the power of the PCI signal would be expected to decrease as it proceeded through the eye and was scattered at density interfaces, it is logical that reflections further back in the eye would have smaller signal-to-noise ratios and therefore would be identifiable only by increased averaging of data. Thus, we speculate that the improved sensitivity in detecting P4 in the diurnal study was associated with the ability to detect signals that had penetrated further through the choroid. Whether P4 represents the histologic choroidal/scleral junction or some other prominent vascular or other structure in the outer choroid cannot be established from these data. Assuming that P4 represents the same structure for a given eye when detected, this limitation in the method does not invalidate the conclusions of diurnal variation. Improvements in detecting P4 signals beyond the present methods may lead to further increases in estimated mean value for choroidal thickness. Part of the difference in mean choroidal thickness may also be an artifact of the time of day in the studies with a single session.
Historically, accurate histologic measurements of choroidal thickness have been difficult to obtain, at least in part because of postmortem drainage of blood and loss of tone from the choroidal blood vessels.23 Thickness estimates of 220 µm were reported in 1912 for the posterior choroid based on in vitro injection of choroidal vascular channels,24 a technique still considered likely to underestimate the true choroidal thickness.25 This thickness estimate, however, continues to be cited in contemporary texts.26 An in vivo technique using radiofrequency signals resulted in thicker choroidal measurements. Choroidal thickness was calculated with the use of extrapolated ocular radiofrequency velocities; measurements taken in the macula ranged from 420 to 450 µm and had a reported error of 20 µm.25 We are unaware of the application of any other methods other than our technique and radiofrequency signal analysis to estimate choroidal thickness in the living human eye.
Current Limitations and Future Directions
Although the PCI instrument used in this study provides highly precise measurements of the retina from the inner limiting membrane to Bruch membrane at the fovea, there are limitations for its use to assess in vivo choroidal thickness in humans. Specifically, the intrasubject variability in the choroidal/scleral P4 peak may limit its generalizability. On the other hand, the intrasubject variability may be a clue to individual differences in function that are related to differences in structure. The relatively small and demographically limited diurnal study population also limits the generalizability of these results. However, the diurnal cohort does provide the first evidence that diurnal variation in choroidal thickness may occur in humans and serves as a rationale for continuing efforts to analyze choroidal thickness.
In addition, the PCI used in this study makes use of a light source with a longer wavelength than that of the only commercially available axial PCI device (IOLMaster; Carl Zeiss Meditec). This difference may enhance the ability of our device to penetrate the RPE and choroid because the absorption coefficient of melanin declines with increasing wavelength.27 The PCI used in this study provides higher spatial resolution than the commercially available axial PCI device (IOLMaster; Carl Zeiss Meditec) because of its superluminescent diode and shorter coherence length (30 vs. 120 µm). The resultant output eliminates the secondary artifactual peaks that are generated by the multimode laser diode of the commercially available axial PCI device (IOLMaster; Carl Zeiss Meditec). These observations suggest that using still longer wavelengths may facilitate choroidal measurements. Recently, an OCT system using a longer wavelength at 1040 nm was demonstrated to be capable of producing images of the superficial choroid of approximately 200 µm (Unterhuber A, et al. IOVS 2006;47:ARVO E-Abstract 3507). Extending the wavelength for axial PCI measurements to use the local minimum in water infrared absorption at 1050 nm while minimizing melanin absorption may improve the signal-to-noise ratio of the P4 signal. Multispectral measurements may further enhance the signal quality from the choroid. Correlation of the PCI measurement with multispectral high-resolution OCT imaging may also give insight into the exact structures measured by the P4 signal. These technical issues merit further evaluation because a device that is better able to penetrate the heavily pigmented human choroid and thus provide a stronger reflectance signal from the choroidal/scleral interface would facilitate exploring the complex interaction between choroidal thickness, emmetropization, eye growth, and other disease processes.
| Summary and Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
Submitted for publication January 22, 2008; revised July 26, 2008; accepted October 27, 2008.
Disclosure: J.S. Brown, None; D.I. Flitcroft, None; G. Ying, None; E.L. Francis, None; G.F. Schmid, None; G.E. Quinn, None; R.A. Stone, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Jamin S. Brown, Department of Ophthalmology, Kresge Eye Institute, 4717 St. Antoine Street, Detroit, MI 48236; jsbrown{at}med.wayne.edu.